HT INSTRUMENTS I-V600 Professional I-V Curve Tracer Instruction Manual
- July 29, 2024
- HT INSTRUMENTS
Table of Contents
I-V600 Professional I-V Curve Tracer
“`html
Specifications
-
Model: I-V600
-
Professional I-V curve tracer up to 1500V, 40ADC
-
Compliant with IEC/EN60891, IECEN60904-1-2, and IEC/EN62446
guidelines -
Tests Monofacial and Bifacial PV modules/strings
-
Equipped with a professional internal BMS (Battery Management
System)
Product Usage Instructions
1. Setup
Ensure the I-V600 is connected to a stable power source before
use. Verify that the device is calibrated and ready for
testing.
2. Testing Procedure
a. Connect the PV module/string to the I-V600 following the
manufacturer’s guidelines.
b. Select the appropriate test mode on the device for Monofacial
or Bifacial modules.
c. Initiate the test to measure Voc, Isc, and other performance
parameters.
3. Battery Management
Utilize the internal BMS for extended battery life. Allow the
system to recover energy from module discharges and input voltages
after tests.
FAQ
Q: How do I know if the I-V600 is ready for testing?
A: The device will display a ready status indicator when it is
calibrated and prepared for testing.
Q: Can I test both Monofacial and Bifacial PV modules with the
I-V600?
A: Yes, the I-V600 is designed to test the performance of both
Monofacial and Bifacial PV modules/strings.
“`
I-V600
Rel. 1.07 17/01/24
Professional I-V curve tracer up to 1500V, 40ADC
Pag 1 of 4
The I-V600 model is an I-V Curve and functional test verification (Voc, Isc)
instrument compliant with IEC/EN60891, IECEN60904-1-2 and IEC/EN62446
guidelines. I-V600 tests the performance and functionality of Monofacial and
Bifacial PV modules/strings.
I-V CURVE TRACER (PERFORMANCE/ACCEPTANCE TEST) I-V600 verifies the performance
of PV strings in compliance with IEC/EN60891 guideline by tracking the I-V
curve on installations up to 1500VDC and 40ADC. Through solar irradiation and
temperature measurements of the PV modules (in wireless combination with the
SOLAR03 remote unit), I-V600 extrapolates the @STC curves (Standard Test
Condition: 1000W/m2, 25°C, AM 1.5) comparing them with the ratings provided by
the module manufacturer. The large internal database stores up to 1000
different manufacturers and up to 1000 modules associated with each
manufacturer directly, easily programmable by touch-screen display.
FUNCTIONAL TEST (IVCK) I-V600 verifies the functionality of PV strings in
accordance with IEC/EN62446 guideline by measuring, with or without solar
radiation, the open circuit voltage (Voc) and the short circuit current (Isc)
in operating conditions (@OPC) up to 1500VDC and 40ADC. By measuring solar
radiation and temperature of the PV modules (in wireless combination with the
SOLAR03 remote unit), I-V600 extrapolates the values @ STC (Standard Test
Condition: 1000W/m2, 25°C, AM 1.5) and compares them with the ratings provided
by the module manufacturer.
I-V600 EXTENDS THE CHARGE OF THE INTERNAL BATTERIES To increase the autonomy
of the batteries and allow them to be recharged, I-V600 is equipped with a
professional internal BMS (Battery Management System) algorithm which
automatically recovers energy from the discharge of the module capacities at
the end of an I-V test and from the voltage present on the inputs. A valid aid
in case you need to carry out many tests in rapid succession.
HT ITALIA SRL Via della Boaria, 40 48018 Faenza (RA) Italy T +39 0546 621002 | F +39 0546 621144 M vendite@ht-instruments.com | ht-instruments.com
WHERE WE ARE
I-V600
Rel. 1.07 17/01/24
Professional I-V curve tracer up to 1500V, 40ADC
Pag 2 of 4
1. ELECTRICAL SPECIFICATIONS
Accuracy calculated as ±[%reading + (number dgts*resolution)] at 23°C ± 5°C,
<80%RH
DMM Multimeter function DC Voltage
Range [V]
Resolution [V]
3 ÷ 1500
1
Accuracy ± (1.0%reading + 2dgt)
I-V CURVE TEST
DC Voltage @ OPC
Range [V]
Resolution [V]
Accuracy (*)
15.0 ÷ 1500.0
0.1
±(0.2%Voc)
(*) In compliance with IEC/EN60904-1; The measurement starts if VDC > 15V and module capacitance <30µF
DC Current @ OPC
Range [A]
Resolution [A]
0.20 ÷ 40.00
0.01
(*) In compliance with IEC/EN60904-1; Iscmin = 0.2A and module capacitance <30µF
Accuracy (*) ±(0.2%isc)
DC Power @ OPC (VDC > 30V)
Range [W] 50 ÷ 9999 10.00k ÷ 59.99k
VDC Voltage 30V and module capacitance <30µF
Resolution [W] 1
0.01k
Accuracy ±(1.0%reading+6dgt)
DC Voltage @ STC
Range [V] 3.0 ÷ 1500.0
Resolution [V] 0.1
DC Current @ STC
Range [A] 0.20 ÷ 40.00
Resolution [A] 0.01
DC Power @ STC (referred @ 1 module)
Range [W]
Resolution [W]
50 ÷ 9999
1
Accuracy ±(4.0%reading+2dgt)
Accuracy ±(4.0%reading+2dgt)
Accuracy ±(4.0%reading+2dgt)
HT ITALIA SRL Via della Boaria, 40 48018 Faenza (RA) Italy T +39 0546 621002 | F +39 0546 621144 M vendite@ht-instruments.com | ht-instruments.com
WHERE WE ARE
I-V600
Rel. 1.07 17/01/24
Professional I-V curve tracer up to 1500V, 40ADC
Pag 3 of 4
FUNCTIONAL TEST (IVCK)
DC Voltage @ OPC
Range [V]
Resolution [V]
Accuracy (*)
15.0 ÷ 1500.0
0.1
±(0.2%Voc)
(*) In compliance with IEC/EN60904-1; The measurement starts if VDC > 15V and module capacitance <30µF
DC Current @ OPC
Range [A]
Resolution [A]
0.20 ÷ 40.00
0.01
(*) In compliance with IEC/EN60904-1; Iscmin = 0.2A and module capacitance <30µF
DC Voltage @ STC
Range [V] 3.0 ÷ 1500.0
Resolution [V] 0.1
Accuracy (*) ±(0.2%isc)
Accuracy ±(4.0%reading+2dgts)
DC Current @ STC
Range [A] 0.20 ÷ 40.00
Resolution [A] 0.01
Accuracy ±(4.0%reading+2dgts)
HT ITALIA SRL Via della Boaria, 40 48018 Faenza (RA) Italy T +39 0546 621002 | F +39 0546 621144 M vendite@ht-instruments.com | ht-instruments.com
WHERE WE ARE
I-V600
Rel. 1.07 17/01/24
Professional I-V curve tracer up to 1500V, 40ADC
Pag 4 of 4
2. GENERAL SPECIFICATIONS
DISPLAY AND MEMORY Characteristics: Type of memory: Module database: Storable data:
Color TFT, capacitive touch screen, 7″, 800x480pxl Memory card, max 32GB (not expandable) ca. 63,000 saved modules 9999 test IVCK or I-V curve
POWER SUPPLY: Internal power supply: External power supply: Battery charging
algorithm: Battery charging system (BMS): Consumption: Low battery indication:
Charging time: Battery life (@ 0°C ÷ 40°C):
Auto Power OFF:
8×1.5V alkaline battery type LR6, AA or
8×1.2V rechargeable battery NiMH type LR6, AA
100-440VAC/15VDC, 50/60Hz
CAT IV 300V (use only HT adapter)
via inputs P1, C1, P2, C2
energy recovered from I-V curve measurements
8W
” ” symbol shown on the display
approx. 4 hours
8 hours in the following conditions: Battery capacity: 2000mAh PV string
voltage: 800V Work cycles: 80 measurements/hour Instrument connected to the
modules for 30s/measurement Instrument disconnected for 15s/measurement
1 ÷ 10min selectable (disabling)
OUTPUT INTERFACE PC interface: Interface with SOLAR03:
USB-C and WiFi Bluetooth connection (up to 100m in free space)
MECHANICAL CHARACTERISTICS Dimensions (L x W x H): Weight (included batteries): Mechanical protection:
336 x 300 x 132mm (13 x 12 x 5in) 5.5kg (11lv) IP40 (open case), IP67 (closed case)
ENVIRONMENTAL CONDITIONS OF USE
Reference temperature:
23°C ± 5°C (73°F ± 41°F)
Operating temperature:
-10°C ÷ 50°C (14°F ÷ 122°F)
Operating humidity:
<80%RH
Storage temperature:
-20°C ÷ 60°C (-4°F ÷ 140°F)
Storage humidity:
<80%RH
Max. height of use:
2000m (6562ft)
REFERENCE GUIDELINES Safety: EMC: Safety measurement accessories: I-V Test: IVCK Test: Insulation: Pollution degree: Radio: Measurement category:
IEC/EN61010-1, IEC/EN61010-2-030, IEC/EN61326-1 IEC/EN61010-031 IEC/EN60891, IECEN60904-1-2 IEC/EN62446, IECEN60904-1-2 double insulation 2 ETSI EN300328, ETSIEN301489-1, ETSIEN301489-17 CAT III 1500VDC, max 1500VDC between inputs
This instrument complies with the requirements of the European Low Voltage
Directive 2014/35/EU (LVD), the Directive 2014/30/EU (EMC) and the RED
regulation 2014/53/EU
This instrument complies with the requirements of the European Directive
2011/65/EU (RoHS) and the European Directive 2012/19/EU (WEEE)
HT ITALIA SRL Via della Boaria, 40 48018 Faenza (RA) Italy T +39 0546 621002 | F +39 0546 621144 M vendite@ht-instruments.com | ht-instruments.com
WHERE WE ARE
Diensten van EURO-INDEX
EURO-INDEX is fabrikant, importeur en distributeur van diverse A-merken op het gebied van test- en meetinstrumenten. Daarnaast leveren wij een groot aantal diensten om het gebruik van deze instrumenten in uw bedrijfsvoering te optimaliseren. Dit omvat uiteraard onderhoud, reparatie en kalibratie van de instrumenten, maar ook kennisdeling via EURO-INDEX Academy en verhuur van instrumenten.
Geautoriseerd Service Centrum EURO-INDEX b.v. is van alle vertegenwoordigde merken een Geautoriseerd Service Centrum. Dit betekent dat uw instrumenten worden behandeld door technici die zijn opgeleid door de fabrikant en beschikken over de juiste gereedschappen en software. Er worden uitsluitend originele onderdelen toegepast en de garantie van uw instrument, evenals de certificering (ATEX, EN50379, etc.) blijven intact.
Kalibratielaboratorium
Het laboratorium in Nederland beschikt over een RvA accreditatie
naar EN-ISO/IEC 17025. Deze accreditatie geldt voor grootheden,
zoals gespecificeerd in de scope bij accreditatienummer K105.
RvA kalibratiecertificaten zijn internationaal geaccepteerd en is gelijkwaardig aan BELAC.
Meer info
Mobiele Service
Naast de vaste kalibratielaboratoria in Zaventem en Capelle aan den
IJssel beschikken wij ook over een laboratorium op wielen met de
naam “Mobiele Service”. Dit biedt vertrouwde service en kwaliteit, bij u voor de deur!
Meer info
KWS®
KWS® is een uniek servicesysteem voor uw meetinstrumenten
met periodiek onderhoud en kalibratie tegen vaste, lage kosten.
Via een gratis webportal (mijnkws.be) heeft u altijd en overal beschikking over uw kalibratiecertificaten.
Meer info
Verhuur van meetinstrumenten
Uitgebreid assortiment
Nauwkeurigheid aantoonbaar door actueel kalibratiecertificaat
Deskundig advies Complete levering inclusief accessoires
Meer info
EURO-INDEX Academy Trainingen, seminars en workshops Demonstratie- en instructievideo’s Application notes
Meer info
Servicebalie
Onderhoud, reparatie en kalibratie
Trainingen en seminars
Mobiele Service
Wijzigingen voorbehouden EURO-INDEX® VL 23001
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>