MEAN WELL GSM60U 60W AC-DC Reliable Green Slim Wall Mounted Medical Adaptor Owner’s Manual
- June 12, 2024
- MEAN WELL
Table of Contents
- MEAN WELL GSM60U 60W AC-DC Reliable Green Slim Wall Mounted Medical
- Product Information
- Product Usage Instructions
- Features
- Applications
- GTIN CODE
- Description
- Model Encoding
- SPECIFICATION
- Derating Curve
- Mechanical Specification
- DC output plug
- Installation Manual
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
MEAN WELL GSM60U 60W AC-DC Reliable Green Slim Wall Mounted Medical
Adaptor
Product Information
Model Number| Order Number| DC Voltage| Rated Current| Output Power| Input
Voltage Range| Working Temperature| Working Humidity| Safety Standards
---|---|---|---|---|---|---|---|---
GSM60U05-P1J| GSM60U05| 5V| 6A| 30W| 80 ~ 264VAC, 113 ~ 370VDC| -30°C ~ +70°C|
20% ~ 90% RH non-condensing| ANSI/ AAMI ES60601-1/ES60601-1-11(3.1 version),
CAN/CSA-C22 NO.
60601-1 3rd edition approved
GSM60U07-P1J| GSM60U07| 7.5V| 6A| 30W| 80 ~ 264VAC, 113 ~ 370VDC| -30°C ~
+70°C| 20% ~ 90% RH non-condensing| ANSI/ AAMI ES60601-1/ES60601-1-11(3.1
version), CAN/CSA-C22 NO.
60601-1 3rd edition approved
GSM60U09-P1J| GSM60U09| 9V| 5.5A| 30W| 80 ~ 264VAC, 113 ~ 370VDC| -30°C ~
+70°C| 20% ~ 90% RH non-condensing| ANSI/ AAMI ES60601-1/ES60601-1-11(3.1
version), CAN/CSA-C22 NO.
60601-1 3rd edition approved
GSM60U12-P1J| GSM60U12| 12V| 4.5A| 30W| 80 ~ 264VAC, 113 ~ 370VDC| -30°C ~
+70°C| 20% ~ 90% RH non-condensing| ANSI/ AAMI ES60601-1/ES60601-1-11(3.1
version), CAN/CSA-C22 NO.
60601-1 3rd edition approved
GSM60U15-P1J| GSM60U15| 15V| 4A| 30W| 80 ~ 264VAC, 113 ~ 370VDC| -30°C ~
+70°C| 20% ~ 90% RH non-condensing| ANSI/ AAMI ES60601-1/ES60601-1-11(3.1
version), CAN/CSA-C22 NO.
60601-1 3rd edition approved
GSM60U18-P1J| GSM60U18| 18V| 3.33A| 30W| 80 ~ 264VAC, 113 ~ 370VDC| -30°C ~
+70°C| 20% ~ 90% RH non-condensing| ANSI/ AAMI ES60601-1/ES60601-1-11(3.1
version), CAN/CSA-C22 NO.
60601-1 3rd edition approved
GSM60U24-P1J| GSM60U24| 24V| 2.5A| 30W| 80 ~ 264VAC, 113 ~ 370VDC| -30°C ~
+70°C| 20% ~ 90% RH non-condensing| ANSI/ AAMI ES60601-1/ES60601-1-11(3.1
version), CAN/CSA-C22 NO.
60601-1 3rd edition approved
GSM60U48-P1J| GSM60U48| 48V| 1.25A| 30W| 80 ~ 264VAC, 113 ~ 370VDC| -30°C ~
+70°C| 20% ~ 90% RH non-condensing| ANSI/ AAMI ES60601-1/ES60601-1-11(3.1
version), CAN/CSA-C22 NO.
60601-1 3rd edition approved
Product Usage Instructions
- Ensure that the input voltage range is within the specified range of 80 ~ 264VAC or 113 ~ 370VDC.
- Choose the appropriate model number and order number based on the desired DC voltage and rated current.
- Connect the AC input using a compatible plug and cable. Refer to page 4~5 for more information on plug and cable types.
- Connect the DC output plug according to the pin assignment. The standard plug is P1J with a center positive polarity. The DC plug can be changed through customization or by using a quick adapter accessory (sold separately).
- Ensure that the working temperature is within the specified range of -30°C to +70°C.
- Check the working humidity, which should be between 20% and 90% RH non-condensing.
- Make sure to comply with the safety standards mentioned, including ANSI/ AAMI ES60601-1/ES60601-1-11(3.1 version) and CAN/CSA-C22 NO. 60601-1 3rd edition.
Features
• Ultra slim, compact size
• Medical safety approved (2 × MOPP) according to
ANSI/AAMI ES60601-1/ES60601-1-11
• Extremely Low leakage current
• 2 pole USA plug, Class II power unit
• No load power consumption <0.15W
• Energy efficiency Level VI ( except 5~9V for Level V)
-30~+70°C wide range working temperature
•Protections: Short circuit / Overload / Over voltage
Fully enclosed plastic case
• Various DC plug quick adapter accessory available (Plug kit sold sperately,
please refer to : https://www.meanwell.com/upload/pdf/DC plug.pdf)
• 3 years warranty
Applications
• Blood glucose meter
• Blood pressure meter
• Nebulizer
•Inhaler
• Portable medical device
• Sleep apnea devices
GTIN CODE
MW Search: hitps://www.meanwell.com/serviceGTIN.aspx
Description
GSM60U is a 60W ultra slim wall-mounted style single-output green medical
adaptor series, which is
compact and convenient for carry. GSM60U is a class II power unit (no FG)
equipped with the standard
2-Pin U.S. plug, accepting the input range from 80VAC to 264VAC. The whole
series provides different
models with output voltages ranging between 5DC and 48VDC that it can satisfy
the demands for various
types of medical electronic devices. The circuitry design meets the
international medical standards (2*MOPP),
having an ultra low leakage current (<100 A), fitting the medical devices in
direct electrical contact with
the patients.
With the efficiency up to 90% and the extremely low no-load power consumption
below 0.15W, GSM6OU is
compliant with the latest U.S. energy regulation EISA 2007/Do(Level VI). The
supreme feature allows
the adaptor to save the energy when it is under either the operating mode or
the standby mode. The entire
series is approved for international safety regulations; moreover, it adopts
the 94 V-0 flame retardant
plastic case that it can effectively prevent users from electric hazards.
Model Encoding
SPECIFICATION
ORDER NO.| GSM60U05- P1J| GSM60U07- P1J| GSM60U09-
P1J| GSM60U12- P1J| GSM60U15- P1J| GSM60U18-
P1J| GSM60U24- P1J| GSM60U48- P1J
---|---|---|---|---|---|---|---|---
OUTPUT
| SAFETY MODEL NO.| GSM60U05| GSM60U07| GSM60U09| GSM60U12| GSM60U15|
GSM60U18| GSM60U24| GSM60U48
DC VOLTAGE Note.2| 5V| 7.5V| 9V| 12V| 15V| 18V| 24V| 48V
RATED CURRENT| 6A| 6A| 5.5A| 4.5A| 4A| 3.33A| 2.5A| 1.25A
CURRENT RANGE| 0 ~ 6A| 0 ~ 6A| 0 ~ 5.5A| 0 ~ 4.5A| 0 ~ 4A| 0 ~ 3.33A| 0
~ 2.5A| 0 ~ 1.25A
RATED POWER (max.)| 30W| 45W| 50W| 54W| 60W| 60W| 60W| 60W
RIPPLE & NOISE (max.) **Note.3| 100mVp-p| 100mVp-p| 100mVp-p|
100mVp-p| 120mVp-p| 120mVp-p| 120mVp-p| 150mVp-p
VOLTAGE TOLERANCE Note.4| ±5.0%| ±5.0%| ±5.0%| ±5.0%| ±5.0%| ±3.0%|
±3.0%| ±3.0%
LINE REGULATION Note.5| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%|
±1.0%| ±1.0%
LOAD REGULATION Note.6| ±5.0%| ±5.0%| ±5.0%| ±3.0%| ±3.0%| ±2.0%|
±2.0%| ±2.0%
SETUP, RISE, HOLD UP TIME**| 1000ms, 100ms, 10ms/230VAC 1000ms,
100ms, 10ms/115VAC at full load
INPUT
| VOLTAGE RANGE Note.7| 80 ~ 264VAC 113 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY (Typ.)| 80%| 85%| 87%| 88%| 88%| 88%| 88%| 90%
AC CURRENT| 1.5A / 115VAC 0.8A / 230VAC
INRUSH CURRENT (max.)| Cold start 50A / 115VAC 100A / 230VAC
LEAKAGE CURRENT(max.)| Touch current < 100μA/264VAC
PROTECTION
| ****
OVERLOAD
| 105 ~ 160% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is
removed
OVER VOLTAGE
| 110 ~ 140% rated output voltage
Protection type : Clamp by zener diode, output short
ENVIRONMENT
| WORKING TEMP.| -30 ~ +70℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -30 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03% / ℃ (0 ~ 40℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X,
Y, Z axes
SAFETY & EMC
(Note. 8)
| SAFETY STANDARDS| ANSI/ AAMI ES60601-1/ES60601-1-11(3.1 version),
CAN/CSA-C22 NO. 60601-1 3rd edition approved
ISOLATION LEVEL| Pimary – Secondary: 2 x MOPP
WITHSTAND VOLTAGE| I/P-O/P:5656VDC
ISOLATION RESISTANCE| I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION
| Parameter| Standard| Test Level / Note
Conducted emission| FCC PART 18| Class B
Radiated emission| FCC PART 18| Class B
OTHERS
| LIFE| 3 years : 100% load 40℃, 8hours / day
MTBF| 400Khrs min. MIL-HDBK-217F(25℃)
DIMENSION| 75.53247.5mm (LWH)
PACKING| 200g ; 60pcs / 13kg / CARTON
CONNECTOR
| PLUG| See page 4~5 ; other types available by customer requested
CABLE| See page 4~5 ; another type available by customer requested
NOTE
- All parameters are specified at 230VAC input, rated load, 25°C 70% RH ambient.
- DC voltage: The output voltage set at point measure by plug terminal & 50% load.
- Ripple & noise are measured at 20MHz by using a 12″ twisted pair terminated with a 0.1f & 47f capacitor.
- Tolerance: includes set up tolerance, line regulation, load regulation.
- Line regulation is measured from low line to high line at rated load.
- Load regulation is measured from 10% to 100% rated load.
- Derating may be needed under low input voltage. Please check the derating curve for more details.
- The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the EMC directives. For guidance on how to perform these MC tests, please refer to “EMI testing of component power supplies.” (as available on http://www.meanwell.com)
≥ Product Liability Disclaimer : For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx
Derating Curve
Static Characteristics
Mechanical Specification
DC output plug
Standard plug: P1J
DC plug changeable through:
(1) Customization of the standard part with an optional DC plug according to
the table (MOQ applicable)
(2) Quick adapter accessory (sold separately without MOQ)
Please refer to below table and online selection guide:
https://www.meanwell.com/upload/pdf/DC_plug.odi
Example quick adapter accessory:
Optional DC plug:
(Available in customized cable or quick adapter)
Installation Manual
Please refer to : http://www.meanwell.com/manual.html
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>