MEAN WELL MFM-10 Series 10W High Reliable Green Medical On Board Type Instruction Manual
- June 12, 2024
- MEAN WELL
Table of Contents
MFM-10 Series 10W High Reliable Green Medical On Board Type
Product Information
- Product Name: 10W High Reliable Green Medical On Board Type
- Model: MFM-10 series
- User’s Manual: TPTC004
Product Usage Instructions
- Power Consumption: The product is designed to have a low power consumption, even when there is no load connected.
- Installation: Follow the installation instructions provided in the user manual to ensure proper installation of the product.
- Power Supply: Connect the product to a suitable power supply according to the specifications mentioned in the user manual.
- Operation: Once the product is connected to the power supply, it is ready to use. Ensure that the load is within the specified range for optimal performance.
- Maintenance: Regularly inspect the product for any signs of damage or wear. Follow the maintenance instructions provided in the user manual for proper care and longevity of the product.
- Safety Precautions: Adhere to all safety precautions mentioned in the user manual while using the product to prevent any accidents or injuries.
- Troubleshooting: In case of any issues or malfunctions, refer to the troubleshooting section of the user manual for possible solutions or contact customer support for assistance.
For detailed information and additional instructions, please refer to the complete user’s manual (TPTC004) provided with the product.
10W High Reliable Green Medical On Board Type
Features
- 1.65″X0.88″ compact size
- Medical safety approved (2 × MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
- Suitable for BF application with appropriate system consideration
- No load power consumption<0.075W
- Extremely low leakage current
- Wide operating temp. range -30 ~ +85°C
- EMI class B for class II configuration
Protections
Short circuit / Overload / Over voltage / Over temperature
- No minimum load required
- 3 years warranty
Applications
- Portable medical device
- Mobile clinical workstation
- Medical computer monitor
- Medical examination instrument
GTIN CODE
MW Search: https://www.meanwell.com/serviceGTIN.aspx
Description
MFM-10 is a 10W high density and small size (4222.320.5mm) AC/DC on board type medical grade power supply series. It features the operation for 80~264VAC, a low no load power consumption less than 0.075W, a high efficiency up to 84%, Class II (no FG) double insulation, outstanding dissipation, 5G anti-vibration, high EMC performance, 4KVAC isolation, etc. The design observes IEC/BS EN/EN60601-1 and ANSI/AAMI ES60601-1 version three with 2×MOPP level and ultra-low leakage current (<80 u A). It is very suitable for BF (patient contact) type medical device or relevant equipment.
Model Encoding
SPECIFICATION
MODEL| MFM-10-3.3| MFM-10-5| MFM-10-12| MFM-10-15|
MFM-10-24
---|---|---|---|---|---
OUTPUT
| DC VOLTAGE| 3.3V| 5V| 12V| 15V| 24V
RATED CURRENT| 2.5A| 2A| 0.85A| 0.67A| 0.42A
CURRENT RANGE Note.2| 0 ~ 2.5A| 0 ~ 2A| 0 ~ 0.85A| 0 ~ 0.67A| 0
~ 0.42A
PEAK CURRENT| 2.75A| 2.2A| 0.94A| 0.74A| 0.46A
RATED POWER| 8.3W| 10W| 10.2W| 10W| 10W
PEAK LOAD(10sec.) Note.3| 9W| 11W| 11.3W| 11.1W| 11W
RIPPLE & NOISE (max.) **Note.4| 120mVp-p| 100mVp-p| 180mVp-p|
180mVp-p| 200mVp-p
VOLTAGE TOLERANCE Note.5| ±2.5%| ±2.5%| ±2.5%| ±2.5%| ±2.5%
LINE REGULATION| ±0.3%| ±0.3%| ±0.3%| ±0.3%| ±0.3%
LOAD REGULATION| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%
SETUP, RISE TIME| 1000ms, 30ms/230VAC 1000ms, 30ms/115VAC at
full load
HOLD UP TIME (Typ.)**| 40ms/230VAC 8ms/115VAC at full load
INPUT
| VOLTAGE RANGE Note.6| 80 ~ 264VAC
FREQUENCY RANGE| 47 ~ 440Hz
EFFICIENCY (Typ.)| 78%| 81%| 83%| 83%| 84%
AC CURRENT (Typ.)| 0.3A/115VAC 0.2A/230VAC
INRUSH CURRENT (Typ.)| COLD START 25A/115VAC 45A/230VAC
LEAKAGE CURRENT (max.) Note.7| Touch current <80μA/264VAC
PROTECTION
| OVERLOAD| 110% ~ 180% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is
removed
OVER VOLTAGE| 3.8 ~ 5V| 5.75 ~ 6.8V| 13.8 ~ 16.2V| 17.3 ~ 20.3V| 27.6 ~
32.4V
Protection type : Shut off o/p voltage, clamping by zener diode
OVER TEMPERATURE| Protection type : Shut down o/p voltage, recovers
automatically after temperature goes down
ENVIRONMENT
| WORKING TEMP.| -30 ~ +85℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +100℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 50℃)
SOLDERING TEMPERATURE| Wave soldering: 265℃,5s (max.); Manual soldering:
390℃,3s (max.)
VIBRATION| 10 ~ 500Hz, 5G 10min./1cycle, period for 60min. each along X,
Y, Z axes
OPERATING ALTITUDE Note.8| 5000 meters
SAFETY & EMC
(Note 9)
| SAFETY STANDARDS| IEC60601-1, BS EN/EN60601-1, EAC TP TC 004,UL ANSI/AAMI ES60601-1(3.1 version), CAN/CSA-C22 3rd Edition approved ;
Design refer to BS EN/EN60335-1(by request)
ISOLATION LEVEL| Primary-Secondary: 2xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC
ISOLATION RESISTANCE| I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION
| Parameter| Standard| Test Level / Note
Conducted| BS EN/EN55011 (CISPR11)| Class B
Radiated| BS EN/EN55011 (CISPR11)| Class B
Harmonic Current| BS EN/EN61000-3-2| Class A
Voltage Flicker| BS EN/EN61000-3-3| —–
EMC IMMUNITY
| BS EN/EN55035, BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| BS EN/EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 3, 1KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| BS EN/EN61000-4-11| 100% dip 1 periods, 30% dip 25
periods,
100% interruptions 250 periods
OTHERS
| MTBF| 9314.1K hrs min. Telcordia SR-332 (Bellcore) ; 1756.2K hrs
min. MIL-HDBK-217F (25℃)
DIMENSION| 4222.320.5mm (LWH) or 1.65″*0.88″0.80″ inch
PACKING| 0.018Kg; 270pcs/5.8Kg/0.94CUFT
NOTE
|
-
All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 C of ambient temperature.
-
No minimum load required.
-
33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power
-
Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1f & 47 . f parallel capacitor.
-
Tolerance: includes set up tolerance, line regulation and load regulation.
-
Derating may be needed under low input voltages. Please check the derating curve for more details.
-
Touch current was measured from primary input to DC output.
-
The ambient temperature derating of 3.5°C/1000m with fanless models and of 5°C/1000m with fan models for operating altitude higher than 2000m(6500ft).
-
The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets MC directives. For guidance on how to perform these MC tests, please refer to “EMI testing of component power supplies.” (as available on http://www.meanwell.com) > Product Liability
Disclaimer : For detailed information. please refer to https://[www.meanwell.com/serviceDisclaimer.aspx](http://www.meanwell.com/serviceDisclaimer.aspx)
Block Diagram
Derating Curve
Output Derating VS Input Voltage
Mechanical S Specification
Installation Manual
Please refer to : http://www.meanwell.com/manual.html
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>