ASSEMBLY GUIDE gev. 02
M162 LCR METER DIY KIT

MODEL: M162
PCB VERSION: MAIN: 109-16202-008
ANALOG: 109-16201-00F
FIRMWARE: 113-16202-060

REQUIRED TOOLS

  1. 25W Soldering Iron
  2. Soldering Wire (Rosin-core Recommended)
  3. Digital Multimeter
  4. Screwdriver -Phillis #1
  5. Flush (wire) cutter
  6. Tweezers
  7. Micro-USB power cable

Getting Started

TEST THE MAIN BOARD

  1. Connect the USB Cable to the Micro-USB port on the Main Board and a USB power source.
  2. Ensure your MI62 main board boots up correctly. When powered on, the LED (D1) will blink twice. The display will then  through two splash screens and enter into a state similar to the below illustration.
  3. The values of the numbers are random. Please ignore them ana just ensure the display is normal.

If your MI62 main board Goes not pOWer Up, of powers up with a blank screen, please contact us at supportajyetech.com. Do ot solder any parts onto the board if you encounter any issues as this will void the warranty.

Main Board Assembly

STRIP SOCKETS

JYETECH M162 LCR Meter DIY Kit - STRIP SOCKETS

 Js 12xstop socket
J8,J9 2x1stop socket

Analog Board Assembly

JYETECH M162 LCR Meter DIY Kit - Main bord1

Ensure the polarity of the capacitors are installed correct. The positive pole goes into the holes of the- square pads,
PIN HEADERS

J5 12×1 pin header
J8,J9 2x! pin header
PUSH BUTTONS

sw1.sw2, sw3 6 2×6 2×9 5 mm atwitch
CONTACT ARMS

JYETECH M162 LCR Meter DIY Kit - CONTACT ARMS

Use the provided tool to hold the arms in place. with your fingers or a clp. Solder at the bottom side.

JYETECH M162 LCR Meter DIY Kit - Assembly8

CLOSE JP1

(Note: skip this step if the battery charger board is installed.)
JP1

Verify Voltages

  1. Apply 5V power supply via the micro-USB connector on the analog board.
  2. Spot check the voltages on the board at the test points labeled to ensure they measure to the voltages in the chart below.

REFERENCE

Test Points Voltage
Min Typical
V+ 3.60V
VS+ 3.60V
AV+ 3.10V
V-  
VS-  
AV- -2.20V
V1  
V2  
V3  
V4 1.40V

Quick Test

  1. Attach the main board to the analog board.
  2. Apply power supply via the micro-USB connector on the analog board and slide the power switch to ON position.
  3. Once the device starts up, hold down RCL button 10 bring up the menu. Select the “Default” option to set all the parameters to their Default values
  4. Test measurement with the provided resistors (10 & 1MQ), capacitor (01, and inductor (MH), or with your own com- ponentes.

Note: The provided resistors are 1% tolerance. The capacitor and inductor tolerance can be 10% or higher. If the measured values are away from their labeled value they might not reflect a problem of the meter. Further confirmation is required.

Enclosure Assembly

  1. Identify the ends of the standoffs.

  2. Assemble the two side pieces and back piece. Ensure the. standoffs are at the correct orientation.
    Insert the assembled main boarg-analog board with the. analog board fitted into the standoff sits near the. bottom ends.

  3. Assemble the front piece. Ensure the standoffs are at the corret orientation,

  4. Assemble the top and bottom pieces with screws. Remember to attach the 3 caps to the pushbuttons before securing the top piece.

  5. This is how your M126 LCR Meter should look once enclosure assembly is complete:

© 2022 JYETech Inc.
[email protected] 
jyetech.com

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Jyetech User Manuals

Related Manuals