GOSSEN METRAWATT METRAHIT T-COM Plus Cable Multimeter for Measurement of
Symmetrical Copper Cable Units
The product being described is the METRAHITT-COM PLUS Cable Multimeter. It is a device used for measuring symmetrical copper cable units. The package includes the following items:
- 1 Cable multimeter
- 1 Protective rubber cover
- 1 Cable set KS21-T (1000 V CAT III) consisting of:
- 1 two-core measurement cable (yellow/blue), 2 m long with test probes
- 1 earth connection line (black), 2 m long with test probe
- 1 Short-form operating instructions in German/English
- 1 CD-ROM with operating instructions in German/English
- 2 Batteries (1.5 V, type AA) were inserted into the unit
Please note that detailed operating instructions are available for download in PDF format at www.gossenmetrawatt.com.
The short-form instructions provided are not a substitute for the detailed instructions. In terms of safety, it is important to follow the safety instructions provided in the user manual. The maximum rated voltage for the device is 1000 V, and it falls under Measuring Category CAT III. The maximum rated current is 1 A with the safety cap applied and 16 A without the safety cap applied. It is crucial to observe the maximum values of the electrical safety of the device.
When using the METRAHITT-COM PLUS Cable Multimeter, please keep in mind the following usage instructions:
- Read and thoroughly understand the detailed operating instructions before using the device. The short-form instructions provided are not comprehensive enough.
- Always use original fuses for the current measuring ranges. The input of the current measuring ranges is fitted with a fuse. Refer to the label on the housing or technical data section for the correct fuse type.
- Make sure to insert batteries into the instrument before operating it. Without batteries, dangerous currents or voltages will not be indicated, and the instrument may be damaged.
- Do not operate the device with the fuse or battery compartment cover removed.
Following these instructions will help ensure the proper and safe usage of the METRAHITT-COM PLUS Cable Multimeter.
The symbol indicates parameter settings that are only described in the detailed operating instructions.
Standard Equipment
- 1 Cable multimeter
- 1 Protective rubber cover
- 1 Cable set KS21-T (1000 V CAT III) consisting of:
- 1 ea. two-core measurement cable (yellow/blue), 2 m long with test probes,
- 1 ea. earth connection line (black) 2 m long with test probe
- 1 Short-form operating instructions in German/English
- 1 CD-ROM with operating instructions in German/English *
- 2 Batteries 1.5 V, type AA inserted in the unit
Instrument Service
When you need service, please contact:
Gossen Metrawatt GmbH Product Support Hotline
- Telephone +49 911 8602-0
- Phone +49 91 81 602-0
- Fax +49 149 81 602-709
- E-mail support@gossenmetrawatt.com
All trademarks, registered trademarks, logos, product names, and company names are the property of their respective owners.
Gossen Metrawatt
Gossen Metrawatt GmbH Südwestpark 15 90449 Nürnberg
- Germany Phone +49 911 8602-0
- Fax +49 911 8602-669
- E-Mail info@gossenmetrawatt.com www.gossenmetrawatt.com
Detailed operating instructions are available for download on the Internet at www.gossenmetrawatt.com.
Standard Equipment
- 1 Cable multimeter
- 1 Protective rubber cover
- 1 Cable set KS21-T (1000 V CAT III) consisting of:
- 1 two-core measurement cable (yellow/blue), 2 m long with test probes
- 1 earth connection line (black), 2 m long with test probe
- 1 Short-form operating instructions in German/English
- 1 CD-ROM with operating instructions in German/English
- 2 Batteries (1.5 V, type AA) were inserted in the unit
Application of measuring cable set KS21-T
Maximum Rated Voltage | 1000 V | 1000 V |
---|---|---|
Measuring Category | CAT III | CAT II |
Maximum Rated Current | 1 A | 16 A |
with a safety cap applied | · | — |
without a safety cap applied | — | · |
Please observe the maximum values of the electrical safety of the device.
Safety Instructions
In order to maintain the flawless condition of the instrument, and to ensure its safe operation, it is imperative that you read the operating instructions thoroughly and carefully before placing your instrument into service, and that you follow all instructions contained therein
Observe the following safety precautions:
- The multimeter may not be used in potentially explosive atmospheres.
- The multimeter may only be operated by persons who are able to recognize contact hazards and take the appropriate safety precautions. Contact hazards exist wherever voltages of more than 33 V (RMS value) and/or 70 V DC occur.
- The maximum voltage allowable according to standard between the voltage inputs or all inputs towards earth respectively is equal to 600 V, category II / 300 V, category III.
- Attention: Unexpected voltages may occur at defective devices, capacitors,…!
- The insulation of the measurement cables may not be damaged, cables and plugs may not be interrupted!
- No measurements may be made in electrical circuits with corona discharge (high-voltage)! Special care is required when measurements are made in HF electrical circuits where dangerous pulsating voltages may be present.
- Measurements under moist ambient conditions are not permissible.
- Do not overload the measuring ranges beyond their allowable capacities!
- The input of the current measuring ranges is fitted with a fuse. Use original fuses only, see label on the housing or technical data section!
- Only operate the instrument with batteries inserted. Otherwise dangerous currents or voltages will not be indicated and your instrument may be damaged.
- The device may not be operated with the fuse or battery compartment cover removed.
Switching on / off / Light on
Battery Test
Battery Replacement
- 2 Batteries: IEC LR6
Disconnect the instrument from the measuring circuit before opening the battery compartment lid! Turn the slotted head screw counter-clockwise for this purpose. Observe the correct polarity of the batteries! When refitting the battery compartment lid the side with the guide hooks must be inserted first. Then turn the slotted head srew clockwise.
Select Measuring Function
Measuring Range Selection
Measuring Inputs
Repair and Replacement Parts Service DAkkS Calibration Lab and Rental Instrument Service Neem desgewenst contact op met:
When you need service, please contact: GMC-I Service GmbH Service center Beuthener Strasse 41 90471 Nürnberg Germany
- Phone +49 911 817718-0
- Fax +49 911 817718-253
- E-Mail service@gossenmetrawatt.com www.gmci-service.com
Product Support
Neem desgewenst contact op met: When you need support, please contact: Gossen
Metrawatt GmbH Product Support Hotline
- Telefoon +49 911 8602-0
- Phone +49 911 8602-0
- Fax +49 911 8602-709
- E-mail support@gossenmetrawatt.com
All trademarks, registered trademarks, logos, product names, and company names are the property of their respective owners.
Gossen Metrawatt GmbH Südwestpark 15 90449 Nürnberg Germany
- Phone +49 911 8602-0
- Fax +49 911 8602-669
- E-Mail info@gossenmetrawatt.com www.gossenmetrawatt.com
DATA-Hold/-Compare
MIN/MAX memory
Zero Balancing
Resistance Measurement
Temperature Measurement
Diode Testing
Direct Voltage / Pulsating Voltage Measurement
AC Voltage – Frequency, without/with Low-pass Filter
Capacitance Cable Length Measurement
DC/Pulsating/Alternating Current/Frequency Measurement
Measurement with Clip-on Current Transformer
Device and Measuring Parameters
Loop Resistance Measurement Continuity Testing
Insulation Resistance Measurement Measuring Inputs
Capacitance (low disturbance) galv. Signature detection
Technical Data
Meetbereik
Measuring Range
| Meetfouten / Measuring Error| Overbelastbaarheid
Overload capacity
---|---|---
| /
300 mV| ±(0,5% + 3 D)
| ±1,5% + 3 D
(> 300 D)
| 600 V DC AC
eff Sinus/ sine
| __
__
Continue
continuous
3 V| __
±(0,5% + 1 D)
| __
±1,5% + 3 D
(> 100 D)
30 V
300 V
600 V
1 Hz … 300,00 kHz| | ±(0,5% + 1 D)| 600 V 3)| max. 10 s
| | /| |
300 mA| ±(0,5% + 5 D)| __
±1,5% + 5 D
(> 100 D)
| __
0,3 A
| __
Continue
continuous
3 mA| __
±(0,5% + 3 D)
30 mA
300 mA
1 A| ±(0,5% + 5 D)| 1,6 A| 5 min
1 Hz … 30,00 kHz| | ±(0,5% + 1 D)| 4)|
| ±(… % v. MW/rdg. + … D)|
300 W| 0,5 + 3 2)| __
__
__
600 V DC AC
eff/rms Sinus/ sine
| __
__
__
__
max. 10 s
3 kW| __
0,5 + 1
30 kW
300 kW
3 MW
30 MW| 2 + 5
RSL 300 W| 3 + 5
RSL 3 kW| 3 + 5
300 W| 3 + 5
5,100 V| 2 + 5
__| | |
30 nF| ±(3% + 6 D) 2)| 600 V DC AC
eff Sinus/ sine
| __
__
max. 10 s
300 nF| __
±(1% + 6 D)
3 mF
30 mF
300 mF| ±(5% + 6 D)
- at + 40 C
- with function ZERO active
- Power limiting: 6 · 106 V · Hz (max. 600 V 1 kHz)
- maximale stroomwaarden
- max. current values see current measuring range
Insulation Resistance Measurement
Measuring Range| UISO| M easuring Error
±( % v. MW/rdg + D)
---|---|---
6 … 110 V 1)| Ri ca. 120 kW| 3 + 30
5 … 310,0 kW| 10//100 V| 3 + 5
0,280 … 3,100 MW| 10//100 V| 3 + 5
02,80 … 31,00 MW| 10//100 V| 5 + 5
028,0 … 310,0 MW| 100 V| 5 + 5
Interference voltage measurement
Electrical Safety
- Protection class II
- EN 61010-1:2010
- Measuring Category CAT II / CAT III
- Nominal Voltage 600 V / 300 V
- Pollution degree 2
- Test Voltage 3,5 kV~ / per IEC 61010-1/EN 61010-1
- Protection/ Housing:
- IP54 pressure equalization by means of housing
- Extract from table on the meaning of IP codes
Electromagnetic Compatibility EMC
- Interference Emission
- EN 61326-1:2013class B
- Interference Immunity
- EN 61326-2-1:2013
Ambient Conditions
- Accuracy range 0C … + 40 C
- Operating temperature –10C … + 50C
- Storage temperature
- without battery – 25C … + 70C
- relative humidity 40 … 75 % no condensation allowed
- Elevation up to 2000 m maximum
Fuse
- FF: 1,6 A / 700 V AC DC 6,3 mm x 32 mm
- breaking capacity: 50 kA (min. 10 kA)
If you use other fuses than the one indicated above you forfeit your product guarantee.
Internal Fuse Test
Fuse Replacement
- Disconnect the instrument from the measuring circuit before opening the fuse compartment lid! Turn the (captive) slotted head screw counter-clockwise for this purpose. Remove the fuse with the flat end of the fuse compartment lid.
- When refitting the fuse compartment lid the side with the guide hooks must be inserted first. Then turn the slotted head srew clockwise.
References