AQUALABO NC-POR-C-00284 Log-Aqua Autonomous Communicating Datalogger User Guide

June 4, 2024
AQUALABO

AQUALABO NC-POR-C-00284 Log-Aqua Autonomous Communicating Datalogger

Access to the web configuration interface

You have received an email invitation to log into your account user interface. Click on “Accept invitation”. Enter your name, first name, phone number and a password to connect.

User interface presentation

Data visualisation

  • Select the Visualisation Menu
  • By clicking on the black arrow to the left of each device, a drop-down menu will appear on the screen. The list of physico-chemical parameters, the last sample, the last transmission, the battery level as well as the last alerts are displayed.
  • Click on the desired parameter.
    • The value of the last sample as well as the last transmission are displayed. On the right is the data graph.
    • You can select the desired date range via the cursor or by defining specific dates via the Date range /Custom range menu
    • Extract the graphic in PNG format using the download link at the bottom right

Edit your profile / Modify the language of the interface

  • Click on the bottom left on the name of your profile / Select Profile / Modify profile
  • You can change your name, enter your email, phone number and change the language with the drop-down menu.

Reports Extraction

  • Select the Reports menu

  • In Sequence History, you can select a device and desired settings from the drop-down menus, as well as date ranges.

  • Click on “Export in CSV format”. You can also save the report configuration so that you don’t have to reselect every parameters every
    time.

  • Open the downloaded .csv file. It should look like the screenshot below.

  • To convert this file and put the data in column form, select column A.

  • Click on Data → Convert

  • Choose the “Delimited” function then check “Comma” and “Interpret identical separators…”. Click on Next

  • On the next window, do not touch anything, simply click on Finish

  • You get a file that you can rework as you wish in Excel, saving it as an .xls or .xlsx file

Digital sensors calibration

  • The calibration of PONSEL digital sensors is carried out via a 4200 Module and the CALSENS software. The software was provided to you with the devices upon receipt of your rental equipment. It is not possible to do the calibration with an ODEON portable device.

  • Physically connect your probes to the 4200 module using the fischer connector bases for this purpose. There is no need for a connection cable because the sensors are already with the correct connectors. Connect the module to your PC using the USB cable.

  • Start the calibration procedure on CALSENS and using the digital sensor user manuals via the “calibration” chapter (user manuals available on our website www.aqualabo.fr ).
    Remarks

    • The middle age of the digital sensors must be 1 (to be notified at the end of the calibration).
    • In the event of several OPTOD sensors connected to the same LOG-AQUA, set the middle age 1 for the sensors and check that the SDI-12 addresses are indeed 1 (for sensor 1), 2 (for sensor 2) and 3 (for sensor 3 if there is a third of the same type).
    • If a STACSENSE sensor is connected, you can check that the Modbus address is indeed 70.