college park Horizon HD Prosthetic Foot Device Instructions
- June 12, 2024
- college park
Table of Contents
- college park Horizon HD Prosthetic Foot Device
- Package Contents
- Tools Recommended
- Key Components
- PRODUCT DESCRIPTION
- TECHNICAL SPECIFICATIONS
- FIRMNESS CATEGORIES
- MOUNTING
- ASSEMBLY AND DISASSEMBLY (for sock replacement)
- WARRANTY INSPECTION /MAINTENANCE INFORMATION
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
college park Horizon HD Prosthetic Foot Device
Package Contents
- (1) Horizon HD Foot
- (1) Foot Shell
- (1) CPI Sock
- (1) Horizon Wedge Kit
- (1) Cosmetic Attachment Plate (CAP)
Tools Recommended
- (1) 4mm Hex Key
- (1) Foot Horn
This diagram is to help familiarize you with the unique parts of the Horizon HD. These parts are referenced in the instructions and used when speaking with a technical service representative.
Key Components
- A. Integrated Pyramid
- B. Housing
- C. Toe spring
- D. Heel Spring
- E. Permanent Wedge
- F. Rounded Heel Wedge (Optional)
- G. Nut Plate
- CPI Sock (not shown)
- Foot Shell (not shown)
PRODUCT DESCRIPTION
This prosthetic foot device is constructed with an integrated pyramid, housing, two composite springs, and heel wedge. The toe spring is secured to the housing and heel spring with fasteners.
INTENDED USE
The Horizon HD is a prosthetic foot designed to replace one or more functions
of the biologic human foot.
INDICATIONS
- Lower limb amputations
CONTRAINDICATIONS
- None known
PROTECTIVE COVER ON DOME
- Remove the protective cover on dome after alignment is completed and before patient leaves clinic.
TECHNICAL SPECIFICATIONS
FOOT SIZE | WEIGHT LIMIT | BUILD HEIGHT | **FOOT WEIGHT*** |
---|---|---|---|
21-25 cm | 330 lbs / 150 kg | **** |
2.3 in / 5.8 cm
| ****
559 g
26-30 cm| 365 lbs / 165 kg
- 26cm foot w/shell
GAIT MATCHING® GUIDELINES
The gait match determines the firmness of the foot based on the user’s specifications (foot size, patient weight, and impact level).
FIRMNESS CATEGORIES
Refer to the chart below to determine the correct firmness category.
Note: Incorrect category selection may result in poor device function.
Contact College Park Technical Service if you have questions about category
selection.
WEIGHT LBS| 0-140| 141-180| 181-220| 221-275|
276-330| 331-365
---|---|---|---|---|---|---
WEIGHT KG| 0-63| 64-81| 82-100| 101-125| 126-150| 151-166
SIZE CM| 21-30| | | | | 26-30
LOW-MODERATE IMPACT| 1| 2| 3| 4| 5| 6
HIGH IMPACT| 2| 3| 4| 5| 6| 6
MOUNTING
- Use only high quality proximal endoskeletal components.
ASSEMBLY AND DISASSEMBLY (for sock replacement)
- Use the Foot Horn to don and doff the foot shell. Remove the CPI Sock and replace as needed. Any further disassembly or modification of components will void the warranty.
STATIC ALIGNMENT
For optimal function, balance the patient’s weight evenly between the heel and toe.
- The Horizon HD was designed with a 3/8” (10 mm) heel rise.
- The load line divides the foot at 1/3 heel lever and 2/3 toe lever.
DYNAMIC ADJUSTMENTS
DESIRED RESULT | ALIGNMENT CHANGE | COMPONENT CHANGE |
---|---|---|
Firmer Toe Response | Plantarflex the Horizon or move the load line posterior |
None
Softer Toe Response| Dorsiflex the Horizon or move the load line anterior|
None
Firmer Heel Response| Dorsiflex the Horizon or move the load line anterior|
Install Rounded Heel Wedge
Softer Heel Response| Plantarflex the Horizon or move the load line posterior|
Remove Rounded Heel Wedge
- To install or remove toe and heel wedges, refer to the Horizon Wedge Kit Instructions.
WARNING
- Do not expose this product to corrosive materials, salt water or pH extremes.
- Failure to follow these technical instructions or use of this product outside the scope of its Limited Warranty may result in injury to the patient or damage to the product.
RESIDUAL RISK STATEMENT
- NOTICE OF RESIDUAL RISK
- During fitting process, ensure that CPI sock does not become pinched between foot and endoskeletal componentry.
WARRANTY INSPECTION /MAINTENANCE INFORMATION
High patient weight and/or impact level may require more frequent inspections. We recommend you visually inspect the following applicable parts for excessive wear and fatigue at each warranty inspection.
- Composites and Adapters
- CPI Sock
- Foot Shell
- Heel Wedge (if applied)
WARRANTY INSPECTION SCHEDULE FOR HORIZON HD: ANNUALLY.
TECHNICAL ASSISTANCE / EMERGENCY SERVICE 24-7-365
College Park’s regular office hours are Monday through Friday, 8:30 am – 5:30
pm (EST). After hours, an emergency Technical Service number is available to
contact a College Park representative.
LIABILITY
The manufacturer is not liable for damage caused by component combinations
that were not authorized by the manufacturer.
CAUTION
College Park products and components are designed and tested according to the
applicable official standards or an in-house defined standard when no official
standard applies. Compatibility and compliance with these standards are
achieved only when College Park products are used with other recommended
College Park components.
This product has been designed and tested based on single patient usage. This
device should NOT be used by multiple patients.
CAUTION
If any problems occur with the use of this product, immediately contact your
medical professional. The prosthetist and/or patient should report any serious
incident* that has occurred in relation to the device to College Park
Industries, Inc. and the competent authority of the Member State in which the
prosthetist and/or patient is established.
- ‘Serious incident’ is defined as any incident that directly or indirectly led, may have led, or might lead to any of the following; (a) the death of a patient, user, or other person, (b) the temporary or permanent serious deterioration of a patient’s, user’s, or other person’s state of health, (c) a serious public health threat.
800.728.7950 | 586.294.7950 | college-park.com.
COLLEGE PARK INDUSTRIES, INC
- 27955 College Park Dr., Warren, MI 48088 USA
EMERGO EUROPE
- Prinsessegracht 20, 2514 AP The Hague, Netherlands
Australian Sponsor
- EMERGO AUSTRALIA
- Level 20, Tower II, Darling Park, 201 Sussex Street,
- Sydney, NSW 2000 Australia
HorizonHD
- COMPONENTES DE ÓRTESES E PRÓTESES EXTERNAS
ANVISA Registro : 80117580371
- IMPORTADOR: EMERGO BRAZIL IMPORT IMPORTAÇÃO DE PRODUTOS MÉDICOS HOSPITALARES LTDA. AVENIDA
- Francisco Matarazzo, 1.752, Salas 502/503,
- Agua Branca, Sao Paulo-SP, CEP – 05001-200
MADE IN THE USA
- ©2022 College Park Industries, Inc. All rights reserved.