PYRAMID CP15AL Pulse Pre Amplifier User Manual
- June 10, 2024
- PYRAMID
Table of Contents
- Safety Information
- Models
- Scope of Supply
- Optional Items
- Intended Use and Key Features
- Specification
- Installation
- Comparison of CP15AF and CP15AL
- Circuit overview
- Setup and Calibration
- Connectors, controls and indicators
- Fault-finding
- Maintenance
- Returns procedure
- Support
- Disposal
- Revision History
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
CP15AL Pulse Pre Amplifier
User Manual
Safety Information
CP15 preamplifiers are designed for compliance with harmonized electrical
safety standard EN61010-1:2000. It must be used in accordance with its
specifications and operating instructions. Operators of the unit are expected
to be qualified personnel who are aware of electrical safety issues. The
customer’s Responsible Body, as defined in the standard, must ensure that
operators are provided with the appropriate equipment and training.
The unit is designed to make measurements in Measurement Category I as defined
in the standard.
Although the CP15 does not generate dangerous voltages, nor are they designed
to measure directly such voltages, they may be in applications where such
voltages are present. Appropriate precautions must be taken.
The unit should not be operated unless correctly assembled in its case. Only
Service Personnel, as defined in EN61010-1, should attempt to work on the
disassembled unit, and then only under specific instruction from Pyramid
Technical Consultants, Inc. or their authorized distributors.
The unit is designed to operate from +/- 12 VDC power rails, with a maximum
current requirement of 70 mV on each rail for the CP15. Up to four CP pre-
amplifiers can be powered by one C400 pulse counting detector controller. CP
pre-amplifiers can also be used in a mixture with the CP15 charge sensitive
pre-amplifier / shaping amplifier.
The CP pre-amplifier should be grounded by secure connection to a grounded
conducting surface. If the unit is mounted on an insulating surface, then one
of the four mounting screw locations must be re-assigned as a grounding
connection. Some of the following symbols may be displayed on the unit and
have the indicated meanings.
| Direct current
---|---
| Earth (ground) terminal
| Protective conductor terminal
| Frame or chassis terminal
| Equipotentiality
I| Supply ON
| Supply OFF
| CAUTION – RISK OF ELECTRIC SHOCK
| CAUTION – RISK OF DANGER – REFER TO MANUAL
Models
CP15AF| Variable gain pulse pre-amplifier optimized for use with sodium iodide
scintillation detectors.
(FMB-Oxford part CP-15N)
---|---
CP15AL| Variable gain pulse pre-amplifier optimized for use with lanthanum
bromide or lanthanum chloride scintillation detectors.(FMB-Oxford part CP-15L)
The following options for the CP15AL and CP15AF can be requested at time of
order. Jumper settings can be set in the field by trained service personnel.
-I/NI: Overall inverting or non-inverting (default is inverting so that
photomultiplier pulses give positive-going pulses)
Scope of Supply
CP15 model as specified in your order.
USB memory stick containing:
Data sheet
User manual
Test results
Optional items as specified in your order, such as power supplies, cables and
adaptors.
Optional Items
6.1 Power supplies and adaptors
PSU1212-L. +24 VDC to +/-12 VDC power supply, input for 2.1mm threaded jack,
output 4 pin Lemo 0B.304.
PSU1212-C. +24 VDC to +/-12 VDC power supply, input for 2.1mm threaded jack,
output 9 pin DSub female with pinout matching C400 connectors.
PSU1212-N. +24 VDC to +/-12 VDC power supply, input for 2.1mm threaded jack,
output 9 pin DSub female with pinout matching NIM standard pre-amp power
connectors, PSU24-25-1. Universal 24 VDC PSU, 25 W, 100-240 VAC 50-60 Hz input
via IEC C8 connector, S671K threaded jack output.
Note: If the CP15 is used with a C400, +/-12V power is provided by the C400.
6.2 Cables and adaptors
CAB-L00-10-L00 Lemo 00 coax cable assembly, 10’ (3 m). For signal output.
CAB-L00-30-L00 Lemo 00 coax cable assembly, 30’ (9 m). For signal output.
CAB-L304M-10-D9M Power cable assembly, 10’ (3 m). To power CP1x from PSU1212
or wC400.
CAB-L304M-30-D9M Power cable assembly, 30’ (9 m). To power CP1x from PSU-B12
or C400.
ADAP-LEMO-BNC Adaptor, Lemo 00 coax plug to BNC jack. For signal output if
using BNC-terminated cables.
Other lengths available on request.
Intended Use and Key Features
7.1 Intended Use
The CP15 models are intended to amplify and condition charge pulses generated
by scintillation detectors based on photomultipliers. Other detectors that
produce similar charge pulses can also use the CP15. The output is suitable
for delivery to discriminator circuits and counters, or to multichannel
analysers for pulse height spectral analysis. The C400 fast
discriminator/counter is able to connect up to four detectors and CP15 pre-
amplifiers and provides all necessary low voltage power and high voltage bias.
The CP15AF (also known as the CP-15N in FMB-Oxford Ltd installations) has gain
and filtering suited to sodium iodide (NaI(Tl)) scintillation detectors. The
burst of fast charge pulses from a single detection event is converted to a
single smooth output pulse, well-suited to discrimination or pulse height
analysis.
The CP15AL (also known as the CP-15L in FMB-Oxford Ltd installations) as gain
and filtering suited to lanthanum bromide (LaBr3) and lanthanum chloride
(LaCl3) scintillation detectors. These detectors are considerably faster than
sodium iodide detectors (and more expensive) and are preferred for higher
count rates. The CP15AL has a corresponding faster response, using filtering
that provides a single output pulse per detection event but without
compromising pulse pair resolution.
7.2 Key Features
Very compact and cost-effective unit with three DC-coupled amplification
stages.
Gain and offset control trim pots
Fully bipolar; output can be inverted if required so that output pulses are
always positive polarity.
Filtering suited to sodium iodide (CP15AF) or lanthanum halide (CP15AL)
scintillation detectors.
Tolerant of long signal input cables.
Output able to drive a 50 ohm load.
Compatible with Pyramid C400 pulse counting detector controller.
Specification
CP15AF
Conversion gain| 1.1 +/-0.05 V per pC at maximum gain setting into 50 ohm load
(output pulse height for charge in input pulse)
2.4 +/-0.1 V per pC at maximum gain setting into high impedance load (output
pulse height to charge in input pulse)
Voltage gain| 7000 nominal (+77 dB) into 50 ohm load at maximum gain
Gain adjustment range| Zero to maximum, linear 21 turn potentiometer
Pulse output rise time| 660 nsec (10% to 90%)
Typical output pulse width (59 keV X-ray, NaI scintillation detector)| FWTM
3.6 µsec
FWHM 1.9 µsec
Unloaded noise| 6 mV into 50 ohm load (@ maximum gain)
Output compliance| Up to 3.0 V into 50 ohms, 9.0 V into high impedance
Maximum line length| Able to drive at least 10 m of RG-58, RG-174 or RG-318
coaxial cable.
CP15AL
Conversion gain| 0.29 +/-0.02 V per pC at maximum gain setting into 50 ohm
load (output pulse height for charge in input pulse)
0.57 +/-0.02 V per pC at maximum gain setting into high impedance load (output
pulse height to charge in input pulse)
Voltage gain| 75 nominal (+37.5 dB) into 50 ohm load at maximum gain
Gain adjustment range| Zero to maximum, linear 21 turn potentiometer
Pulse output rise time| 80 nsec (10% to 90%)
Typical output pulse width (59 keV X-ray, LaCl3 scintillation detector)| FWTM
110 nsec
FWHM 50 nsec
Unloaded noise| 1 mV into 50 ohm load (@ maximum gain)
Output compliance| Up to 1.7 V into 50 ohms, 3.6 V into high impedance
Maximum line length| Able to drive at least 10 m of RG-58, RG-174 or RG-318
coaxial cable.
Controls and displays (CP15AF and CP15AL)
Controls| 21-turn trimpot (DC offset)
21-turn trimpot (gain control)
| Internal jumper for output polarity
---|---
Displays| Green LED (power on).
Power (CP15AF and CP15AL)
Power input| +12 (+2/-4) VDC, -12 (+4/-2) VDC 70 mA typical, each rail.
Physical (CP15AF and CP15AL)
Case| Al alloy nickel coated with mounting flange.
Case protection rating| The case is designed to rating IP43 (protected against
solid objects greater than 1 mm in size, protected against spraying water).
Weight| 0.085 kg (3.0 oz).
Operating environment| 5 to 35 C (15 to 25 C recommended to reduce drift and
offset)
< 80% humidity, non-condensing vibration < 0.2 g all axes (1 to 100 Hz)
Shipping and storage environment| -10 to 50C
< 80% humidity, non-condensing vibration < 2 g all axes, 1 to 100 Hz
Dimensions| (see figures 1 – 3).
Figure 1. CP15 case end panels. Dimensions mm
Installation
9.1 Mounting
The CP15 should be located close to the radiation detector to minimize the
effect of cable capacitance on noise levels. It may be mounted in any
orientation or may be simply placed on a level surface. A fixed mounting to a
secure frame is recommended in a permanent installation for best low signal
performance, as this can be degraded by movement and vibration. Four M2.5
tapped holes are provided in the base on a 27 mm by 30 mm rectangular pattern.
The mounting position should allow sufficient access to connectors and cable
bend radii. Leave 60mm clearance at either end for mating connectors and cable
radii.
Best performance will be achieved if the CP15 is in a temperature-controlled
environment. No forced-air cooling is required, but free convection should be
allowed around the case.
9.2 Grounding and power supply
A secure connection should be made via the mounting screws to local ground
potential. If the unit is mounted on an insulating surface, then one of the
four mounting screws can be reassigned as a grounding connection. +12 VDC and
-12 VDC power should be provided from a suitably rated power supply with the
following minimum performance:
Output voltage| +12 (+1/-2) VDC
-12 (-1/+2) VDC
---|---
Output current| <= 100 mA each rail
Ripple and noise| < 100 mV pk-pk, 1 Hz to 1 MHz
Line regulation| Better than 240 mV
9.3 Connection to equipment
9.3.1 Typical setup
The figure below shows a typical installation to read out a scintillation
counter. The connection between the detector and the CP15 should be kept
relatively short if possible. The connection from the CP15 to the readout
device can be long provided that the cable is 50 ohm impedance and terminated
in 50 ohms at the readout device.
Figure 3. Schematic CP15 installation for readout of a scintillation
detector
The charge pulse from a scintillation detector is electrons, thus negative
polarity. The CP15 can handle positive or negative input pulses, but some
readout devices might require a positive going pulse. To set the CP15 to be
overall inverting so that negative input charge gives a positive voltage, the
internal polarity jumper should be set to the default INV position.
9.3.2 Setup with the C400
The C400 is a partner product to the CP15. It provides pulse discrimination
and counting electronics for four independent channels, each of which provides
the bias, pre-amp power and signal processing for a detector system. Detector
types can be mixed. The CP15 is well-suited as a preamplifier for the C400 but
can of course be used with other pulse handling electronics also.
9.3.3 Effect of cable length
It is generally good practice to mini mise signal cable lengths as far as
practicable, in order to mitigate noise and signal distortion.
The cable from the CP15 output to the receiving electronics can be long,
provided it has the correct 50 ohm impedance and is terminated in 50 ohms at
the receiver.
The cable from the detector to the CP15 should be short if possible, although
the CP15 is relatively tolerant of longer input cables when used with
photomultiplier scintillation detectors.
Longer cable can alter the pulse amplitude and the signal level. The following
plots shows the measured amplitude and the rms noise level changed relative to
a short cable as a function of coaxial length between the scintillation
detector and the CP15.
Comparison of CP15AF and CP15AL
An ideal scintillation detector would generate a single short pulse of charge for each ionizing photon or particle that interacts in the scintillating material. In fact the interaction is more complicated. If the detector signal is examined with a very high bandwidth measurement system, it appears as a burst of charge pulses. This is particularly evident for sodium iodide detectors but is also present for lanthanum halide detectors.
Figure 6 Output for a single 59.6 keV X-ray event from a NaI detector
(left) and a LaCl3 detector (right) measured with a high bandwidth CP10B
preamp; 1 µsec per division. The role of the CP15 preamplifier is to collect
this burst of pulses into a single clean pulse that is suitable for counting
and pulse height analysis.
The CP15AF has a time constant large enough to collect all the output charge
from a sodium iodide detector into a single output pulse. It will perform the
same task for a lanthanum halide detector, but the speed advantage of the
lanthanum halide will be lost.
Figure 7 Single output for a 59.6 keV X-ray event from a NaI detector (left)
and a LaCl3 (right) measured with CP15AF preamp; 1 µsec per division.
The CP15AL is a faster circuit. It maintains the speed of the lanthanum halide
detector while providing gain and a small amount of filtering to smooth the
pulse for pulse height discrimination. If used with sodium iodide, too much of
the underlying pulse structure is still evident, so this is not a recommended
combination.
Figure 8 Single output for a 59.6 keV X-ray event from a NaI detector (left) and a LaCl3 (right) measured with CP15AL preamp; 1 µsec per division.
Circuit overview
The simplified schematics below shows the functional blocks of the two CP15 models.
Figure 10. CP15AL simplified block schematic
The voltage developed by the signal charge pulse is amplified in three DC-
coupled stages. The gains of the stages are larger in the CP15AF, but the
capacitance values in the feedback provide more low pass filtering to provide
a smooth output pulse. The CP15AL has lower total gain and less feedback
capacitance on each stage and uses fast amplifiers throughout so that the much
shorter pulses of lanthanum halide detectors are not broadened significantly,
permitting higher count rates.
A resistive divider gain control between the second and third gain stages
allows the overall gain to be adjusted. The offset potentiometer provides an
adjustable voltage to compensate any DC offset. A fourth inverting amplifier
stage has unity gain and the signal going to the output can be taken before or
after it thus providing a choice of overall inverting or non-inverting
response.
The CP15AF has a unity gain line driver matched to 50 ohms that allows large
output pulses to be driven down long 50 ohm coaxial lines.
Incoming +/-12 V power is filtered and linear regulators produce the
additional +/- 5 V rails required. A LED confirms that the -5 V supply is
present.
Setup and Calibration
12.1 Gain setting
The photomultiplier tube in a scintillation detector provides the large gain
that converts the faint light from a single detection event into a readily
measurable pulse of electrons. The amount of gain is determined by the high
voltage applied to the photomultiplier in a non-linear fashion.
Although it is possible to see low energy detection events by applying
sufficient voltage, there are limitations:
– the tube will have a maximum allowed voltage
– as the voltage is increased, random “dark current” pulses released
spontaneously become more frequent and larger, leading to unwanted background.
The adjustable gain of the CP15 allows the photomultiplier voltage to be set
at a safe level with acceptable background signal. The pulse height of the
signals of interest can then be set into a convenient range by adjusting the
gain trimpot. A clockwise rotation of the trimpot increases the gain between
zero and the maximum.
12.2 Offset correction
The offset trim pot should be adjusted as necessary to remove any residual DC
offset on the output. This can be observed by looking at the output signal on
an oscilloscope. The offset is slightly coupled to the gain setting, more
noticeably for the CP15AF. After setting the desired gain, adjust the offset
to return the baseline to zero volts. CP15 User Manual CP15_UM_221017 Page 22
of 33
12.3 Conversion gain
The conversion gain from coulombs of input charge to output pulse height can
be determined if required using a test capacitor. Mount a 1.0 pF +/- 0.05 pF
capacitor in a screened metal box, taking care to mini mise parasitic
capacitance.
A 1.0 volt peak-peak square wave with fast risetime delivered to this test box will deliver a charge of 1.0 +/- 0.05 pC to the CP15 at each transition of the square wave, producing a series of positive and negative charges which in turn give a series of positive and negative going pulses.
Figure 14 Square wave into test capacitor and resulting charge pulses measured
by the CP15.
The conversion gain at the particular trim pot gain setting of the CP15 is the
measured amplitude of the output pulses divided by the known input charge from
the test capacitor. If the conversion gain is measured at the maximum trim pot
setting, then it will reduce linearly as the trimpot is rotated anticlockwise.
12.4 Energy calibration
The amplitude of the output shaped signal will be linear with the charge in
the input pulse up until the CP15 output saturates. If the amount of charge
from the detector is also linear with the energy of the radiation, then the
system can be calibrated as an energy analyser. You will need to know at least
one point to establish the calibration.
As an example, you can calibrate a system comprising scintillation detector
and the CP15AF using a suitable radiation source. A convenient source is the
Am-241 isotope which generates a 59.6 keV gamma ray which will create a peak
at the high energy end of a pulse height spectrum. If the application uses
lower energy radiation, then the 5.9 keV X-rays from an Fe-55 source can be
used as the reference.
Unknown X-ray peaks can then be assigned their energy by simple linear scaling relative to the known peak.
Connectors, controls and indicators
13.1 Input end connectors, controls and indicators
Lemo 00 coaxial connector (NIM-CAMAC CD/N 549) for detector signal input.
Green LED illuminated when CP15 is powered.
21-turn linear trimpot adjusting DC offset of output signal.
21-turn linear trimpot adjusting gain from zero to maximum.
13.2 Output end connectors
13.2.1 Out
Lemo 00 coaxial connector for signal output. Typical mating free connector
Lemo FFA.00.250.
13.2.2 Power +/- 12 V
Lemo 0B four pin female, typical mating free connector FGG.0B.304.
Connector body and pin 3 connects directly to CP15 analog ground. Pin 2 connects to analog ground via ferrite bead.
13.3 Internal settings
We do not recommend that you open the CP15 case unless specifically instructed
to do so by your supplier or Pyramid Technical Consultants, Inc. It is
possible to damage the circuits or degrade noise performance if correct
handling precautions are not employed. The following information is provided
for reference. To open the case, remove two M2.5 hex cross head screws from
each end of the case. The top of the case can then be removed to reveal the
circuit. A 2 mm jumper near the output sets the relationship between the input
signal polarity and the output signal polarity.
INV| Output voltage is the opposite polarity to the input charge (default
configuration)
---|---
NONINV| Output voltage is the same polarity as the input charge
Fault-finding
Symptom | Possible Cause | Confirmation | Solution |
---|---|---|---|
No output pulses seen | No signal from detector. | Connect known good signal | |
source. | Correct any detector or radiation source issue | ||
Very low pulse rate | Set up oscilloscope to capture individual events. |
Increase radiation flux if possible.
| Insufficient gain| Look for pulses with oscilloscope.|
Increase gain setting. Increase detector gain if possible (photomultiplier
voltage).
Reduce threshold of pulse counting discriminator circuits if possible.
| Wrong output pulse polarity – scalers and MCAs generally require positive
polarity shaped pulses.| Check output pulse polarity with oscilloscope| Change
CP15 internal polarity jumper setting.
| No power to CP15.| Check power LED, check voltages| Correct power supply.
| Polarity jumper not in place| Check that jumper is fitted to INV or NONINV
position| Fit jumper as required.
| Missing HV bias voltage to detector| Check voltage.| Supply correct bias
voltage for the radiation detector.
| CP15 input damaged.| Use test input to inject a signal.| Arrange for repair.
Output pulses distorted| Output saturating.| Check output pulse shape with|
Use appropriate gain setting.
| | oscilloscope. Reduce gain or reduce input signal size.|
---|---|---|---
| Reflections in long output cable.| Check cable termination.| Use 50 ohm
termination at receiving device if cable is long.
Pulse heights differ by a factor of two from expected values.| Incorrect
terminating resistance on readout device for CP15 pulse output.| Check
impedance at readout device.| Use correct impedance. 50 ohms is recommended
for long transmission lines.
Output has DC offset| Offset trim is incorrect.| Adjust trimpot to remove
offset.| Adjust trimpot to remove offset.
Unexpected pulses| Cosmic rays or other background seen by detector.| Turn off
detector bias to see that pulses stop.
Move away from radiation background.| Move detector or add shielding if
possible.
If cosmic ray rate is significant for the measurement, add a coincidence detector system to reject cosmic ray events.
| Multiple pulses from detector for a single event appearing at CP15 output.| Look at individual events on CP15 output.| Use the appropriate CP15 model: CP15AF for slow detectors like NaI, CP15AL for fast detectors like LaBr3, LaCl3.
Maintenance
The CP15 does not require routine maintenance or calibration. There is risk of contamination which may degrade performance if the case is opened in an uncontrolled environment. If you need to access the unit to change jumper settings, check first with your supplier or direct with Pyramid for guidance.
Returns procedure
Returns procedure Damaged or faulty units cannot be returned unless a Returns
Material Authorization (RMA) number has been issued by your supplier or
Pyramid Technical Consultants, Inc. If you need to return a unit, contact your
supplier or Pyramid Technical Consultants at
support@ptcusa.com, stating
– model
– serial number
– nature of fault
An RMA will be issued, including details of which service center to return the
unit to.
Support
Documentation updates are available for download from the Pyramid Technical
Consultants website at www.ptcusa.com. Technical
support is available by email from
support@ptcusa.com. Please provide the model
number and serial number of your unit, plus relevant details of your
application.
Disposal
We hope that the CP15 gives you long and reliable service. The CP15 is manufactured to be compliance with the European Union RoHS Directive 2002/95/EC, and as such should not present any health hazard. Nevertheless, when your device has reached the end of its working life, you must dispose of it in accordance with local regulations in force. If you are disposing of the product in the European Union, this includes compliance with the Waste Electrical and Electronic Equipment Directive (WEEE) 2002/96/EC. Please contact Pyramid Technical Consultants, Inc. for instructions when you wish to dispose of the device.
Revision History
The release date of a Pyramid Technical Consultants, Inc. user manual can be determined from the document file name, where it is encoded yy mmdd. For example, M10_UM_080105 would be a M10 manual released on 5 January 2008.
Version | Changes |
---|---|
CP15_UM_220929 | First general release |
Pyramid Technical Consultants, Inc.
135 Beaver Street Suite 102, Waltham MA 02452 USA
US: TEL: (781) 402 1700
FAX: 781-402-1750
EMAIL: SUPPORT@PTCUSA.COM
Europe: TEL: +44 1273 492002
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>