RHYTHM HEALTHCARE LM5BA Stationary Oxygen Concentrator User Manual
- June 11, 2024
- RHYTHM HEALTHCARE
Table of Contents
- LM5BA Stationary Oxygen Concentrator
- INTRODUCTION
- IMPORTANT SAFEGUARDS
- CONTRAINDICATIONS FOR USE
- IMPORTANT PARTS OF YOUR CONCENTRATOR
- SETTING UP YOUR OXYGEN CONCENTRATOR
- BEFORE OPERATING YOUR OXYGEN CONCENTRATOR
- OPERATING YOUR CONCENTRATOR
- CARING AND PERIODIC MAINTENANCE FOR YOUR UNIT
- ALERT INDICATION SYSTEM
- MINOR TROUBLESHOOTING
- SPECIFICATIONS
- SYMBOLS
- LIMITED WARRANTY
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
ITEM# LM5BA
STATIONARY OXYGEN CONCENTRATOR
User Manual
LM5BA Stationary Oxygen Concentrator
Smoking while using oxygen is the number one cause of fire injuries and
related deaths. You must follow these safety warnings:
Do not allow smoking, candles, or open flames in the same room with the device
or within five feet of the oxygen-carrying accessories.
Smoking while wearing an oxygen cannula may cause facial burns and possibly
death.
Removing the cannula and putting it on bedding, sofas, or other cushion
material will cause a fire when exposed to a cigarette, heat source, or flame.
If you intend to smoke, you must always turn the oxygen concentrator off,
remove the cannula and leave the room where either the cannula or mask or the
oxygen concentrator is located. If unable to leave the room, you must wait 10
minutes after you have turned the oxygen concentrator off before smoking.
INTRODUCTION
INDICATIONS FOR USE — The Oxygen Concentrator is intended for use as an oxygen
concentrator to provide supplemental low flow oxygen therapy for adult,
geriatric and pediatric in the home, nursing homes, patient care facilities,
etc. The pediatric must be treated under adult supervision. It is not intended
to sustain or support life.
Patient group: Adult, geriatric and pediatric.
Under certain circumstances, oxygen therapy can be hazardous.
Seeking medical advice before using an oxygen concentrator is advisable.
Provider Information
Provider Name:
Telephone:
Address:
Prescription Information
Name:
Oxygen liters per minute
atrest: during activity:
others:
Oxygen use per day
Hours:
Minutes:
Comments:
LM5BA Oxygen Concentrator Serial Number:
(check one)
o5-Liter
Medical Provider Information
Set-Up Person:
This instruction guide was reviewed with me, and | have been instructed on the
safe use and care of the
Oxygen Concentrator.
Signature:
Date:
IMPORTANT SAFEGUARDS
INTRODUCTION
Please read and understand this manual before operating your unit. Important
safeguards are indicated throughout this guide; pay special attention to all
safety information. If you have any questions, contact your provider.
No modification of this equipment is permitted.
For your safety, the oxygen concentrator must be used according to the
prescription determined by your provider. If you feel any discomfort or are
experiencing a medical emergency while undergoing oxygen therapy, seek medical
assistance immediately to avoid harm.
It is the responsibility of the provider supplying the LMSBA to the patient to
ensure that the device is used in a safe and effective manner in keeping with
the provider’s prescription for supplemental oxygen use.
While this manual is intended to support this goal, ultimate determination of
the patient’s safety resides with the patient’s provider.
Availability of an alternate source of oxygen is required in case of power
outage or mechanical failure.
Back-up oxygen should be available (oxygen cylinders, etc.).
Do not use your oxygen concentrator in the presence of flammable gases. This
can result in rapid burning causing property damage, bodily injuries, or
death. Do not use oil, grease, petroleum-based products or other flammable
products with the oxygen-carrying accessories or the Oxygen Concentrator. Only
water based, oxygen compatible lotions or salves should be used. Oxygen
accelerates the combustion of flammable substances.
Only use a mild detergent applied with a damp (not wet) cloth or sponge to
clean the cabinet, control panel, and power cord and then wipe all surfaces
dry. Do not allow any liquid to get inside the device.
Pay special attention to the oxygen outlet for the cannula connection to make
sure it remains free of dust, dirt, and particles.
Do not operate unit in a restricted or confined space ( i.e., a small case or
handbag) where ventilation can be limited. This can cause the oxygen
concentrator to overheat and impair performance.
The oxygen concentrator should be located in an area away from smoke,
pollutants or fumes.
The use of some oxygen administrating accessories is not specified for this
oxygen concentrator and may impair its performance. Recommended accessories
are referenced in this manual.
If the oxygen concentrator has been dropped, damaged, or exposed to water,
please contact your provider for inspection or possible repair of the device.
Do not use the oxygen concentrator if it has a damaged power cord or plug.
Do not allow the air intake or air outlet vents to become blocked. Do not drop
or insert any objects into any openings on the device. This can cause the
oxygen concentrator to overheat and impair performance.
Do not overfill the optional humidifier. Fill the optional humidifier with
water only to the level shown by the manufacturer of the humidif
If you feel any discomfort or are experiencing a medical emergency, seek
medical assistance immediately.
Operating the unit outside of the operational temperature specifications can
limit the concentrator’s ability to meet oxygen concentration specification.
Refer to the specification section of this manual for temperature limits.
The oxygen concentrator should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is unavoidable, the device should be
observed to verify normal operation.
In the event of an alarm or you observe the oxygen concentrator is not working
properly; consult the troubleshooting section in this manual. If you cannot
resolve the problem, consult your provider. To prevent voiding the warranty,
follow the manufacturer’s instructions.
Do not attempt any maintenance other than the possible solutions listed within
this manual. Do not remove covers. Only your equipment provider or a qualified
service technician should remove the covers or service this device.
Do not leave a nasal cannula on or under clothing, bed coverings or chair
cushions. If the unit is turned on but not in use, the oxygen will make the
material flammable. Turn the oxygen concentrator off when ot in use.
Always disconnect AC power supply from the wall before disconnecting the AC
power supply from the oxygen concentrator. Aiways place oxygen supply tubing
and power cords in a manner that prevents a frip hazard.
CONTRAINDICATIONS FOR USE
This device is not intended to be life-sustaining or life-supporting.
This device is not intended for newborn or infant use.
Not for use in the presence of flammable anesthetics.
The unit is not appropriate for any patient who would experience adverse
healthcare consequences as the result of such temporary interruption.
IMPORTANT PARTS OF YOUR CONCENTRATOR
Please take time to familiarize yourself with your oxygen concentrator before
operating.
Packing list. Ensure you have the following items:
- 1 x User manual
- 1 x Oxygen concentrator set
- 1 x Adapter for humidifier (optional)
- 1 x Oxygen Outlet Connector
- 1 x Optional Humidifier
If used, clean your humidifier daily to reduce the possibility of
contamination. Follow the cleaning recommendations from your provider, or the
manufacturer of the humidifier. Replace the humidifier monthly, or as
recommended by the manufacturer, or your home care provider. Do not overfill
the humidifier.
Accessories designed by other manufacturers have not been tested by Rhythm
Healthcare and are not recommended for use with Rhythm Healthcare products.
-
Power Switch
Push to switch on, then the unit turns on. Push switch
I=0 ON off, the unit stops.
O =OFF -
Monitor- Indicates the condition of system pressure statusFigure 2
• Displays wh:n the device runs. There are four indication lights on the panel.
• Power (Green indicator light)
• Normal oxygen (Green indicator light—O2 over 82%)
• Low Oxygen (Yellow indicator light — Oz less than 82%)
• Service required (Red indicator light — please turn of the unit and contact the provider) -
Operating instructions
Label of operating warnings:
1, DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THIS MANUAL, IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS AND INSTRUCTIONS, CONTACT YOUR EQUIPMENT PROVIDER BEFORE ATTEMPTING TO USE THIS EQUIPMENT; TOHERWISE, INJURY OR DAMAGE MAY RESULT.
2, Under certain circumstances, oxygen therapy can be hazardous. Seeking medical advice before using an oxygen concentrator is advisable.
3, NO SMOKING – Keep ALL sources of ignition out of the room in which this product is located and away from areas where oxygenis being delivered. Textiles, oils, and other combustibles are easily ignited and bum with great intensity in oxygen enriched air.
4, RISK OF ELECTRIC SHOCK Do not remove cover. Refer servicing to qualified service personnel. -
Circuit breaker — (selected units) resets the unit after electrical power outage or overload.
-
Flow meter knob
-
Flow meter Tum the flowrate knob to the setting prescribed by your provider.
We recommend the flowrate between 1.5L/min ~ 5L/min.
To properly read the flow meter, locate the prescribed flowrate line on the flowmeter. Next, tum the flow knob until the ball rises to the line. Now, center the ball on the L/min. line prescribed.
Never set the flow greater than 5L/min. -
Humidifier (optional)
-
Power cord
Back View
-
Cover
-
Air filter—prevents dirt, dust, and lint from entering your unit
-
Specification Label
-
Exhaust vent
The operator should read and understand this entire manual before using the
device.
The device is not intended for life support. If the provider has determined
that an interruption in the supply of oxygen, for any reason, may have serious
consequences to the user, an alternate source of oxygen ghould be available
for immediate use.
SETTING UP YOUR OXYGEN CONCENTRATOR
-
Position your unit near an electrical outlet convenient for the user.
Keep the oxygen concentrator at least 5 feet away from heat source, sparks or any sources of flame.
Do not connect to an electrical outlet controlled by a wall switch. NO other appliances should be plugged into the wall outlet. -
Position your unit at least 6 inches away from walls, draperies, or any other objects that might prevent the proper flow of air in and out of your oxygen concentrator. The oxygen concentrator should be located in an area way from pollutants or fumes.
Device operation above or outside of the voltage, temperature, humidity and/or altitude values specified may decrease oxygen concentration levels.
BEFORE OPERATING YOUR OXYGEN CONCENTRATOR
-
Before operating your unit always check to be sure the air filter (located on the back of your unit) is clean.
Proper cleaning of this filter is discussed on page 14. -
Attach the appropriate oxygen accessories to the oxygen outlet.
Oxygen Tubing Connection:
a. Connect the oxygen outlet connector onto the oxygen outlet.
b. Attach the oxygen tubing directly to the connector. (Figure 5)
Oxygen Tubing Connection with Humidifier:
Only connect the humidifier if prescribed.
-
Fill the humidifier bottle with distilled water. Do not overfill.
-
Thread the wing nut located on the top of the humidifier bottle to the oxygen outlet so that it is suspended (Figure 6). Make sure it is securely tightened.
-
Attach the oxygen tubing directly to the humidifier bottle outlet fitting. (Figure 7)
Your provider has prescribed either a nasal cannula, catheter, or face mask, in most cases, they are already attached to the oxygen tubing. If not, follow the manufacturer’s instructions for attachment. -
Remove the power cord completely from the line cord strap, make sure the power switch is in Off position, and insert the plug into the wall outlet. The unit is double insulated to guard against electric shock.
If the humidifier and tubing are assembled correctly, and the flow setting is more than 1L/min, you will feel the oxygen flow through the cannula and bubbles will rise to the water surface.
OPERATING YOUR CONCENTRATOR
Special precautions need to be taken with Oxygen Concentrators and should be
used according to the EMC information provided in this manual.
Oxygen can cause rapid burning when exposed to flames. Do not smoke while your
oxygen concentrator is running, or when you are near a person utilizing oxygen
therapy.
Keep the oxygen concentrator at least 5 feet away from hot objects, sparks, or
any sources of fire.
Improper use of the power cord and plugs can cause burns, fire, or other
electric shock hazards, do not use the unit if the power cord is damaged.
-
Press the power switch to “On” position. When unit is on, the Green Power light will illuminate, and the patient alert system —service Required light and audible signal —will briefly beep.
If the Service Required light illuminates and the unit beeps but the unit is not operating, there is no power to the unit. Refer to the Minor Troubleshooting chart on page 17 and contact your provider if hecessary.
If an audible low-frequency vibration sound is detected, the unit is not operating properly. Refer to the
Minor Troubleshooting chart on page 17 and contact your provider if necessary. -
Check the flow meter to make sure that the ball is centered on the line next to the prescribed number of your flow rate.
It is very important to follow your oxygen prescription. Do not increase or
decrease the flow of oxygen without consulting your provider.
If the flow meter knob is turned clockwise, the flow decreases (and eventually
will shut off the oxygen flow). If the knob is turned counterclockwise, the
flow increases.
CARING AND PERIODIC MAINTENANCE FOR YOUR UNIT
Before attempting any cleaning procedures, turn the unit Off.
Oxygen Humidifier (reusable bottles)
If your provider has prescribed humidification, clean your humidifier bottle
daily. Follow the instructions supplied by the manufacturer. If no cleaning
instructions were supplied, follow these steps:
- Wash the humidifier in a solution of hot water and mild dishwashing detergent.
- Soak the humidifier in a solution of one-part white vinegar to three parts hot water for 30 minutes. This solution acts as a germicidal agent.
- Rinse thoroughly with warm tap water and refill with distilled water for use. Do not overfill.
Cannula
Connect the cannula to the humidifier or oxygen outlet and position the nasal
cannula on your face, as shown below, and breathe normally through your nose
Replace the cannula and oxygen supply tubing regularly, as recommended by your provider. Clean and replace the cannula as instructed by your provider.
Ensure the cannula is fully inserted and secure. This ensures that the Oxygen Concentrator can properly detect inspiration for oxygen delivery. During inhalation, you should hear or feel oxygen flow to the prongs of the nasal cannula.
Air Filter and Oxygen Outlet Connector
The air filter and connector should be cleaned at least once a week. To clean,
follow these steps:
- Remove the air filter (located on the back of the unit). Remove the oxygen outlet connector (if used).
- Wash in a solution of warm water and mild dishwashing detergent.
- Rinse thoroughly with warm tap water and towel dry. The filter should be completely dry before reinstalling.
Do not lubricate fittings, connections, tubing, or other accessories of the
oxygen concentrator to avoid the risk of fire and burns.
Use only spare parts recommended by the manufacturer to ensure proper function
and to avoid the risk of fire and bumns.
To prevent product damage, do not attempt to operate the unit without the air
filter or while the filter is still damp.
To avoid electric shock, do not remove the concentrator cabinet. The cabinet
should only be removed by a qualified medical technician. Do not apply liquid
directly to the cabinet or utilize any petroleum-based solvents or cleaning
agents.
Clean the concentrator exterior cabinet by using a damp cloth or sponge with a mild household cleaner and wipe it dry. No need to sterilize.
Periodic Maintenance for your unit
- Clean the air filter at least once a week. The steps of cleaning see the Air Filter and Oxygen Outlet Connector.
- Pull out the air filter and clean the cabinet filters with a vacuum cleaner or wash in warm soapy water and rinse. thoroughly
- Dry the filters thoroughly before reinstallation. Do not operate the Oxygen Concentrator without the air intake filter in place. If a second filter is provided, insert the replacement filter before you clean the dirty filter. Clean the dirty filter in a warm soap and water solution then dry thoroughly prior to use.
-
Change the filter of the Inlet pre filter at least once a year. Follow these steps:
Open the safeguard then pull out the inlet pre-filter, Pull out the old filters from the Inlet Pre-Filter (two sides.
Then insert the new filter to the Inlet pre-filter on each side, insert the Inlet Pre-Filter and close the Important safeguard. -
Change the bacterial filter at least once every two years. You cannot change this by yourself, you must contact your provider to change this for you.
Accessories
The accessories below are tested by Rhythm Healthcare with Oxygen
Concentrators:
Product | Manufacturer | Model | Technical Data |
---|---|---|---|
Bubble Humidifier (200 ml Capacity) | SALTER LABS | REF7100 | ABS,PP |
Cannula (disposable) | Hsiner Co,. Ltd | 3583 | PVC |
Cannula w/ 7 ft Oxygen Tubing (disposable) | Hsiner Co,. Ltd | 3031 | PVC |
Oxygen Outlet Connector | KEY-HUB | OPO1A609A | ABS |
There are many different types of humidifiers, oxygen tubing, cannulas and
masks that can be used with this device. You should contact your provider for
recommendations on which of these devices will be best for you.
They should also give you advice on the proper usage, maintenance, and
cleaning.
The use of certain humidifiers and accessories (not specified above) may
impair the oxygen concentrator’s performance.
If the unit is not operating and the Service required light is on and the
alarm beeps there is no power to the unit. Refer to the Minor Troubleshooting
chart on page 17 and contact your provider.
All waste from Rhythm LM5SBA Oxygen Concentrator must be disposed of using the
appropriate methods specified by local authorities.
To preserve the environment, the concentrator must be disposed of using the
appropriate methods specified by local authorities.
ALERT INDICATION SYSTEM
-
The alerts of this machine include visual indication signal (yellow, red) and auditory alert. There are three alert conditions by priority: high priority, medium priority, and low priority.
-
Visual alerts
If there is any beep sound, the indication light will be illuminated. Alert Priority| Indication light color| Twinkle frequency| Duty cycle
---|---|---|---
High priority| Red| 1.4Hz-2.8Hz| 20%-60%
Medium priority| Yellow| 0.4Hz-0.8Hz| 20%-60%
Low priority| Yellow| Stays on| 100% on -
Auditory Alerts
If there are any alerts, the alarm beeps, and the sound level is between 40dBA and 85dBA.
a) High priority condition: “ Bi-Bi-Bi—-Bi-Bi” paused then after five seconds the alarm beeps again.
b) Medium priority condition: “Bi-Bi-Bi” paused then after ten seconds the alarm beeps again.
c) Low priority condition: Bi-“ paused then after 20 seconds the alarm beeps again.
Beep condition priority: high priority prior to medium priority prior to low priority. If there are more than two beep conditions at the same time, the visual and auditory alert will show the prior condition signal. -
The Alert System
No.| Alert meanings| Alert priority| Visual signal| Auditory signal| Alert type
---|---|---|---|---|---
1| After running three minutes, the oxygen concentration is less than 70%| High priority| Red| Bi-Bi-Bi–Bi-Bi
5 second interval| Technical
2| Machines shuts down| High priority| Red| Bi-Bi-Bi—Bi-Bi 5 second
interval| Technical
3| After running three minutes the oxygen concentration is between 70% and 82%| Medium priority| Yellow| Bi-Bi-Bi
10 second interval| Technical
4| The pressure is more than 20 psi or less than 3 psi| Low priority| Yellow| Bi 20 second interval| Technical -
Actions after alert
If the oxygen purity falls below the acceptable level, the green Normal Oxygen light will shut off and the yellow Low Oxygen light will illuminate. Switch to your reserve oxygen system. Refer to the Troubleshooting section in this guide on page 17 and contact your provider.
As an added safety feature should the oxygen purity continue to drop, an intermittent audible alert will sound, and the power will be cut off. Contact your provider immediately. Do not attempt any other maintenance. -
Power-off alarm will reset automatically in less than 30 seconds.
Do not change the alert indication preset and do not change the memory of the
MCU preset.
Geriatric, pediatric or any other patient unable to communicate discomfort can
require additional monitoring and/or a distributed alarm system to convey the
information about the discomfort and or the medical urgency to the responsible
care giver to avoid harm.
MINOR TROUBLESHOOTING
Symptom | Possible Cause | Remedy |
---|
A. Unit does not
operate. Power light is off when the power switch is “On .” Audible alert is
pulsing and Service Required light is flashing.| 1.Power cord not properly
inserted into wall outlet.| Check power connection at the wall outlet. On 115
volt/220volt units, check the back of the unit.
2.No power at wall outlet.| Check your home circuit breaker and reset if
necessary. Use a different wall outlet if the situation occurs again.
3.Oxygen concentrator circuit breaker activated (selected units).| Press the
concentrator circuit breaker
button (if equipped) located below the power switch. Use a different wall
outlet if the situation occurs again.
B. Unit operates; the Power light is on when the Power switch is “On”. Red
Service Required light is illuminated. Audible beep may be sounding.| 1.Air
filter is blocked.| Check the air filter. If the filter is dirty, wash it
following the cleaning instructions on page 14.
2.Exhaust is blocked.| Check the exhaust area; make sure there is nothing
restricting the unit exhaust.
3.Blocked or defective cannula, catheter, face mask or oxygen tubing.| Detach
cannula, catheter, or face mask. If proper flow is restored, clean or replace
if necessary. Disconnect the oxygen tubing at the oxygen outlet. If proper
flow is restored, check oxygen tubing for obstructions or kinks. Replace if
necessary.
4.Blocked or defective humidifier bottle.| Detach the humidifier from the
oxygen outlet. If proper flow is obtained, clean or replace humidifier.
5.Flow meter set too low.| Set flow meter to prescribed flow rate. If the
above remedies do not work, contact your provider
C. Unit operates the power light is on when power switch is “On “, audible
low-frequency vibration sound is detected.| immediately.| Turn your unit
“off”, Switch to your reserve oxygen system, and contact your provider
D. If any other problems occur with your oxygen concentrator.| | Turn your
unit Off. Switch to your reserve oxygen system and contact your provider
immediately.
F. Red Service Required light is on and an intermittent audible signal is
sounding.| 1.Flow meter is not properly set yet.| Ensure the flow meter is
properly set to the prescribed.
2.Air fitter is blocked. on page 14.| Check the air filter. If the fitter is
dirty, wash it following the cleaning instructions
3.Exhaust is blocked.| Check the exhaust area: make sure there is nothing
restricting the unit exhaust. If the above remedies do not work, contact your
provider.
SPECIFICATIONS
No. | Item | Rating |
---|---|---|
1. | Delivery Rate | 1 to 5 Umin |
2. | Outlet Pressure | 8.5 PSI ( 58.6I<Pa) |
3- | Electrical Rating | 115VAC, 60Hz<365VA 230VAC,50Hz<365VA |
4. | Oxygen Percentage | 1-5 Umin 93%±2% |
5- | Operating Environment Range | Temperature 50°F to 104°F |
Humicity: 30% to 70%
Atmospheric pressure 50-106 kpa (7.3-15.4 psi)
6.| Operating Attitude
(Tested at 69.8°F only)
0-2000 M (0-6561 ft)
2000-4000 M (6561-13123 ft)| Across the voltage range:
No degradation of performance
Not recommended/efficiency less than 85%
7.| The visible” low oxygen “indicator will activate at the following level|
Less than 82% will activate at the following level (the red “service required”
light will activate)
8.| Pressure Relief Valve| 40 psig ± 5psig (276 kPat34.5 kPa)
9-| Weight| 33.5 lbs
10.| Dimensions| 23″H x 14 .2″W x 11.4″D
11.| Operating System| Time Cyde / Pressure Swing
12.| Storage Conditions| Temperature -40°F to 158°F
Humicity 10% to 100%
Atmospheric pressure 50-106 kpa (7.3-15.4 psi)
13.| Equipment Class and Type| The type of protect against electric shock:
Class II
The degree of protect against electric shock:
TYPE B Applied part
SYMBOLS
| Warning – Describes a hazard or unsafe practice that can result in severe
bodily injury, death, or property damage
---|---
| Caution — Describes a hazard or unsafe practice that can result in minor
bodily injury or property damage
| Note – Provides information important enough to emphasize or repeat
| Consult the accompanying documents
| Type BF Equipment
| Class Il Equipment, double insulated
| Do not use oil or grease
| Do not expose to open flames, sparks or heat sources
| Do not smoke near unit
| Do not disassemble
| Consult instructions
| See manual
| Manufacturer’s serial number
| Date of manufacture
| Manufacturer
| Proper disposal of waste of electrical and electronic equipment
required
| FRAGILE
Contents of the transport package are fragile therefore it shall be handled
with care.
| THIS WAY UP
Indicates correct upright position of the transport package.
| Drip proof equipment
| Keep away from rain and water.
Transport package shall be kept away from rain.
| Non-ionizing electromagnetic radiation
Guidance and manufacturer’s declaration — electromagnetic emissions
The LM5BA is intended for use in the electromagnetic environment specified
below. The customer or the user of the LM5BA should assure that it is used in
such an environment.
Emissions test| Compliance| Electromagnetic environment-
guidance
---|---|---
RF Emissions
CISPR 11| Group 1| The device uses RF energy only for its internal function.
Therefore, its RF emissions are very
low and are not likely to cause any interference in nearby electronic
equipment.
RF Emissions
CISPR 11| Class B| The device is suitable for use in all establishments,
including domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic Emissions
IEC 61000-3-2| Class A
Voltage Fluctuations/Flicker Emissions
IEC 61000-3-3| Complies
Guidance and manufacturer’s declaration — electromagnetic immunity
The LM5BA is intended for use in the electromagnetic environment specified
below. The customer or the user of the LM5BA should assure that it is used in
such an environment.
Immunity test| IEC 60601 test level| Compliance Level|
Electromagnetic environment-guidance
---|---|---|---
Conducted RF
IEC 61000-4-6
Conducted RF
EC 61000-4-3| 3 Vrms
150 kHz to 80 MHz
3V/m
80 MHz to 2.5 GHz| 3V
3 Vim| Portable and mobile RF communications equipment should be used no
closer to any part of the LM5BA OXYGEN CONCENTRATOR including cables, than the
recommended separation distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Where pis the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d’is the recommended separation
distance in meters (m).b
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, ashould be less than the compliance level in each
frequency range. b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
is affected by absorption and reflection from structures, objects, and people.
aField strength from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the LM5BA is used exceeds the
applicable RF compliance level above, the LM5BA should be observed to verify
normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the LM5BA.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.
LIMITED WARRANTY
Rhythm Healthcare, LLC warrants the LM5BA Oxygen Concentrator under the
conditions and limitations stated below. Rhythm Healthcare, LLC warrants this
equipment to be free from defects in workmanship and materials for three (3)
years from date of factory shipment to the original purchaser, (typically the
healthcare provider) unless contractually specified otherwise. This warranty
is limited to the Buyer of new equipment purchased directly from Rhythm
Healthcare, LLC, or one of its Providers, Distributors, or Agents.
Rhythm Healthcare, LLC’s obligation under this warranty is limited to product
repair (parts and labor) at its factory or at an Authorized Service Center.
Routine maintenance items, such as filters, are not covered under this
warranty, nor does it cover normal wear and tear.
Warranty Claims Submissions
The original purchaser must submit any warranty claim to Rhythm Healthcare,
LLC or to an Authorized Service Center. Upon verification of the warranty
status, instructions will be issued. For all returns, the original purchaser
must (1) properly package the unit in a Rhythm Healthcare, LLC’s approved
shipping container, (2) properly identify the claim with the Return
Authorization Number, and (3) send the shipment freight prepaid. Service under
this warranty must be performed by Rhythm Healthcare, LLC and/or an Authorized
Service Center.
NOTE – This warranty does not obligate Rhythm Healthcare, LLC’s to provide
aloaner unit during the time that an oxygen concentrator is undergoing repair.
NOTE – Replacement components are warranted for the unexpired portion of the
original Limited Warranty. This warranty shall be voided, and Rhythm
Healthcare, LLC shall be relieved of any obligation or liability if:
- The device has been misused, abused, tampered with, or used improperly during this period.
- Malfunction results from inadequate cleaning or failure to follow the instructions.
- The equipment is operated or maintained outside the parameters indicated in the operating and service instructions.
« Unqualified service personnel conduct routine maintenance or servicing. - Unauthorized parts or components (i.e., regenerated sieve material) are used to repair or alter the equipment.
« Unapproved filters are used with the unit.
Rhythm Healthcare, LLC
3200 Tyrone Blvd N
St. Petersburg, FL 33710
Email: contactus@rhythmhc.com
Website: www.rhythmhc.com
V1:12/01/22
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>