RHYTHM HEALTHCARE P2 Portable Oxygen Concentrator User Guide
- June 5, 2024
- RHYTHM HEALTHCARE
Table of Contents
- RHYTHM HEALTHCARE P2 Portable Oxygen Concentrator
- Operation Procedure
- Standby
- Remove the defective sieve bed
- Disconnect the valve base from the defective sieve bed
- Connect the valve base to the new sieve bed
- Install the new sieve bed to concetrator
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
RHYTHM HEALTHCARE P2 Portable Oxygen Concentrator
Notice:
Pls turn off the device, and disconnect the power adapter before maintenance
Operation Procedure
- Remove the defective sieve bed
- Disconnect the valve base from the sieve bed column
- Connect the valve base to the new sieve bed
- Install the new sieve bed to the concentrator
Standby
Remove the battery by forcing down the latch, then slide out the battery horizontally.
- Remove the anti-slip mats installed on the screw holes on the bottom of the concentrator.
- Remove the two fixing screws on the bottom of the base by using the T10 screwdriver.
Slide upward to remove the Sieve bed compartment cover.
Remove the defective sieve bed
Pull out the silicone tube by hand or with tweezers.
Pull out the 3 data cables.
- Screw off the fixing screw on the bottom of the valve base with the T10 screwdriver.
- Pull out the sieve bed assy carefully
Disconnect the valve base from the defective sieve bed
screw off the two screws on the valve base.
Remove the MAC valve base from the sieve bed, take out the former sieve bed.
Connect the valve base to the new sieve bed
Install a new sieve bed into the unit, then install the MAC valve on the top of the sieve bed, and screw the two screws back.
Install the new sieve bed to concetrator
- Push the sieve bed assy to the end (please check carefully if the sieve bed is pushed to the end)
- Screw back the fixing screw of the sieve bed assy.
- Connect the silicone tube to the joint well.
- Connect the 3 data cables of the MAC valve to the valve ports.
- Screw on the fixing screws on the base.
- Slide down the sieve bed compartment cover to a suitable position.
- Set the anti-slip mats.
You have successfully changed the sieve bed, then Install the battery or connect the adapter to use the oxygen concentrator.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>