Product Information:

The 5′ Protector Premier Laboratory Hood is a benchtop-style laboratory hood designed to provide a safe and efficient workspace for laboratory experiments. It comes with a built-in exhaust blower and a sleek interior with a molded one-piece fiberglass liner. The liner is specially formulated with fiberglass- reinforced polyester that offers corrosion and fire resistance for easy cleanup. The interior has no seams, which means fewer points of deterioration for longer life. The laboratory hood is designed to meet ASHRAE 110, ASTM E84, CAN/CSA C22.2, NFPA 45, SEFA 1, and UL standards.

Product Specifications:

  • Weight: 450.0 lbs (204.0 kg)
  • Dimensions: 60.0 w x 31.7 d x 66.0 h (152.4 x 80.5 x 167.6 cm)
  • Electrical: 100-115 volts, 60 Hz, 10 amps
  • Product Subcategory: General Chemistry
  • Nominal Width: 5′
  • Blower Requirements: Integral blower included
  • Enclosure Height: 66.0
  • Lighting: LED
  • Style: Benchtop

Required Accessories (Not Included):

  • Remote blower (except for built-in blower models)
  • Ductwork
  • SpillStopperTM Work Surface
  • Base cabinet or stand

Optional Accessories:

  • Service Fixture Kits (some models include service fixtures)
  • Electrical Duplex Kits (some models include electrical duplex)
  • Guardian Airflow Monitor Kits
  • Distillation Grid Kits
  • Sash Stop Kits
  • Ceiling Enclosures and Rear Finish Panel Kits

Product Usage Instructions

  • Before using the laboratory hood, ensure that all required accessories are available and properly installed. Place the laboratory hood on a flat surface and connect it to a power source.
  • Turn on the LED lighting and adjust it as needed. If using the built-in blower, make sure it is properly ventilating the laboratory hood. Place the experiment inside the hood and conduct it as needed.
  • After completing the experiment, turn off the LED lighting and the blower, and remove the experiment from the hood.
  • Clean the interior of the laboratory hood with appropriate cleaning agents and dispose of any waste materials in accordance with laboratory regulations.

Overview

Protector Premier Laboratory Hoods incorporate a sleek interior with a molded one-piece fiberglass liner, the signature feature of Labconco’s leading line of fume hoods since 1961. The one-piece liner of specially formulated, fiberglass-reinforced polyester offers corrosion and fire resistance and easy clean up. Without seams, the interior has fewer points of deterioration for longer life.
These hoods are suitable for providing personal protection needed for general chemistry applications involving fumes and vapors. They meet the SEFA-1 standard of a low velocity, high-performance hood and may be safely operated as low as 60 fpm. Choose from hoods designed to be used with a remotely- located blower or hoods that include a built-in blower — the only high- performance hood with a built-in blower available anywhere. Explosion-proof hoods with built-in blowers are available for laboratories with potentially explosive/flammable atmospheres (Class 1 Div I or II).
SEE OUR SUSTAINABILITY INITIATIVES
Catalog Number: 100500040

Specifications

  • Weight: 450.0 lbs
  • Weight metric: 204.0 kg
  • Dimensions: 60.0″ w x 31.7″ d x 66.0″ h
  • Dimensions metric: 152.4 x 80.5 x 167.6 cm
  • Electrical: 100-115 volts, 60 Hz, 10 amps
  • Product Subcategory: General Chemistry
  • Nominal Width: 5′
  • Blower Requirements: Integral blower included
  • Conformance: ASHRAE 110, ASTM E84, CAN/CSA C22.2, NFPA 45, SEFA 1, UL Enclosure Height: 66.0″
  • Lighting: LED
  • Style: Benchtop

Description

Features (All Models)

  • By-pass airflow design
  • Eco-Foil™ Air Foil reduces energy consumption by 7-10% compared to flat air foils while its aerodynamic curve allows air to sweep the work surface for maximum containment. Clean-Sweep™ openings pull inflow air from under the air foil forcing air into non-turbulent air streams.
  • Cord-Keeper™ Slots on left and right side of air foil keep electrical cords out of the way
  • Glacier white powder-coated steel exterior is durable and corrosion-resistant
  • One-piece molded fiberglass liner and pre-set baffle(s) with flame spread less than 25 per ASTM E84 Tempered safety glass vertical-rising sash with powder-coated sash handle and Clean-Sweep™ openings 37.5″ (95.3 cm) high sightline from the work surface to header panel allows taller users comfortable viewing while standing
  • Removable front and side panels, and front service access panels for access to plumbing and electrical wiring
  • Pre-wired LED lighting with vapor-proof design and ADA-compliant light and blower switches. Lighting rated to 50,000 hours. (Not on models with built-in explosion proof blower)
  • On models for use with remote blower, 12.8″ ID exhaust connection(s)

Built-In Blower Models Feature

  • Built-in belt-driven, corrosion-resistant exhaust blower with adjustable sheave, molded thermoplastic housing and non-sparking, coated aluminum impeller
  • 10.8″ ID exhaust connection on 4′ models; 12.8″ ID exhaust connection on 5′ and 6′ models

Fixtured Models May Feature

  • Two pre-plumbed service fixtures with forged brass valves, lower right side with brass tubing for gas and lower left side wit copper tubing for cold water. Components for converting either or both fixtures to air and vacuum are provided. Inlet tubing is not provided.
  • One pre-wired GFCI electrical duplex receptacle on lower right side and, on 8′ models only, one additional pre-wired GFCI electrical duplex receptacle on lower left side

Required Accessories (Not Included)

  • Remote blower (except for built-in blower models)
  • Ductwork
  • SpillStopper™ Work Surface
  • Base cabinet or stand

Optional Accessories

  • Service Fixture Kits (some models include service fixtures)
  • Electrical Duplex Kits (some models include electrical duplex)
  • Guardian Airflow Monitor Kits
  • Distillation Grid Kits
  • Sash Stop Kits
  • Ceiling Enclosures and Rear Finish Panel Kits

For Product Assistance

  • Ryan Caton
    Sales Representative

  • Eleanor Watkins
    Customer Service Representative

Labconco Corporation
8811 Prospect Avenue Kansas City, MO 64132-2696
Phone:816-333-8811LABCONCO-100500040
5-Protector-Premier-Laboratory-Hood-FIG- \(3\)

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

LABCONCO User Manuals

Related Manuals