PASCO PS-3521 Nitrate ISE Probe User Manual
- June 13, 2024
- PASCO
Table of Contents
PS-3521 Nitrate ISE Probe
OVERVIEW
Included Equipment | Item | Included Equipment | Item |
---|---|---|---|
Nitrate ISE Replacement Module | 1 | Nitrate Standard Solution 1000 ppm | 5 |
Nitrate ISE (Cl-) Probe | 2 | Nitrate Reference Fill Solution | 6 |
Storage Bottle | 3 | Nitrate Ionic Strength Adjuster (ISA) | 7 |
BNC Male-to-Female Cable | 4 |
Introduction
This Nitrate Ion Selective Electrode (ISE) Probe is a combination electrode
that includes both reference and sensing half cells in one body housing. It is
used for
measuring nitrate ion (NO3+) activity in aqueous samples.
Technology: Replaceable Polymer/PVC Membrane (Nitrate ISE Replacement Module PS-3552)
Overview
- Size and Material Body (Epoxy): 12 mm OD x 155 mm Length Cap (ABS): 16 mm OD x 57 mm Lengt
- Reference Double-junction Ag/AgCl, ceramic pin junction, refillable, sodium chloride electrolyte
- Features Replaceable Sensor Module
- Range 0.4 to 62,000 mg/L
- Reproducibility ±2%
- Slope -54 to -59 mV/decade @ 25°C typical
- pH Range 2.5 to 11
- Operating Temperature 0 to 50°C
- Interferences ClO4 -, ClO3 -, I-,
- Application/Note Very popular for environmental, industrial, agriculture, food/beverage
Required Equipment
- PASCO Wireless pH/ISE/ORP Sensor (PS-3204) or
- PASSPORT High Precision pH/Temperature with ISE/ORP Amplifier (PS-2107) with a PASCO interface (see www.pasco.com for information) or
- Ion meter or pH/mV meter
- PASCO data collection software (see www.pasco.com)
- Wash bottle filled with distilled or de-ionized water
- Lab wipes
- Pipettes: 1 mL, 10 mL and 100 mL
- Clean beakers
- SE-7287 100 ml Beakers (12 pack) or SE-7288 1000 ml Beakers (6 pack)
Included Required Solutions
- Reference Fill Solution
- 1000 ppm Nitrate Standard Solution
- Ionic Strength Adjuster (ISE) (SC-3536 Nitrate ISE Solutions
Electrode Preparation
Replace the Module Blank
This electrode performs better when the reference junction is kept wetted
during transit and storage. However, the sensor module is best stored and
transported dry. Therefore, the electrode is shipped with a plastic Module
Blank inside that must be removed prior to use.
- Unscrew the cap of the storage solution bottle (also known as the “Soaker Bottle”) and remove the electrode. Unscrew the plastic Module Blank from the end of the electrode. Replace the blank with the included Nitrate ISE Replacement Module.
- The Module Blank can be retained for long-term storage.
- Caution: Do not touch the PVC sensor membrane on the end of the Nitrate ISE Replacement Module with your fingers.
- Thoroughly rinse the electrode with distilled water. Wipe carefully with a clean lab wipe.
- Return the electrode to the storage bottle when it is not in use.
Fill the Electrode with Electro
- The reference (outer) chamber must be filled with
- Reference Fill Solution and remain open during testing
-
Slide the sleeve of the electrode FastFil cap down to uncover the fill hole. (see fig. 1)
-
Shake the electrode downward like a thermometer to remove any air bubbles trapped inside.
- The surface of the
Reference Fill Solution in the reference chamber must be above the inner junction. This is approximately 3″ from the electrode tip.
- The surface of the
-
Rinse the electrode with Dl water, and blot dry. Do not rub dry.
-
Soak the electrode in Dl water for 10 minutes, then in a diluted Nitrate Standard Solution for two hours prior to calibration or use.
Software Help
See the SPARKvue Help or PASCO Capstone Help for information about
collecting, displaying, and analyzing data
- In SPARKvue, select the HELP button in any screen including the Home Screen.
- In PASCO Capstone, select PASCO Capstone Help from the Help menu, or press F1.
Checking Electrode Operation (Slope
- Connect your electrode to the input connector on the Wireless pH /ISE/ORP Sensor (or other measuring device). Place 100 mL of DI water into a 150 mL beaker. Add 2 ml Nitrate ISA to the DI water and stir throughly.
- Rinse the electrode with DI water, blot it dry, and place it in the beaker with the prepared solution.
- Pipet 1 mL of 1000 ppm Nitrate Standard Solution into the beaker. Stir thoroughly, then record the potential (E1) in mV when a stable reading is displayed.
- Pipet 10 mL of the same standard solution into the beaker. Stir thoroughly. When a stable reading is displayed, record the potential (E2) in mV.
- The difference between the first and second potential readings, (E2-E1), is defined as the electrode slope. The normal range for the slope is -56±4 mV at 25°C.
Troubleshooting
If the electrode slope is not within the normal range, the following procedure may restore the electrode.
- Soak the electrode in a diluted standard solution for 2 hours before use.
- Repeat the “Checking Electrode Operation” procedure again. Note: All standard solutions should be prepared fresh. For best performance use ISA in all solutions.
Periodically check the Reference Fill Solution level in the reference chamber
of the electrode. The solution level must be higher than the inner junction
which is visible as a white ceramic pin on the inner body.
If the electrode slope is still outside the normal range after this procedure,
you must replace the sensing module.
Reading a Sample with the Electrode
Various procedures may be used to determine the concentration of a sample. The
most common is the Direct Calibration method, which is described below.
Contact PASCO’s Technical Support department for details of other methods.
In Direct Calibration a series of standard solutions of differing
concentrations is used to calibrate the electrode. Then each sample requires
only a single measurement device reading, which is compared with the
calibration readings to obtain the sample concentration. ISA is added to all
solutions to ensure the samples and the standards have the same ionic
strength.
Set up:
- Prepare the electrode as described in “Electrode Preparation” and “Checking Electrode Operation”. Connect the electrode to the sensor or measuring device.
- Prepare two standard solutions that differ in concentration by a factor of ten and bracket the expected sample concentration range. For example, if your expected sample concentration is 5 mg/L Calcium, you should use a 1 mg/L low Standard Solution and a 10 mg/L high Standard Solution.
NOTE: Temperature compensation is not typical of ISE measurements – for best performance, try to ensure that the standards are within 5 to 10°C of the sample.
Measurement
If using a measuring device with direct concentration reading capability see the software online help or the measuring device’s instruction manual for specific information:
- Place 100 mL of the low standard into a 150 mL beaker. Add 2 mL of ISA. Stir thoroughly.
- Rinse the electrode with Dl water, blot it dry, and place it in the beaker. Wait for a stable reading, and then adjust the measuring device to display the value of the standard. Refer to the software online help or the measuring device’s instruction manual for the adjustment procedure.
- Measure 100 mL of the high standard into a second 150 mL beaker. Add 2 mL of ISA and stir.
- Rinse the electrode with Dl water, blot it dry, and place it in the second beaker. Wait for a stable reading, and then adjust the measuring device to display the value of the second standard.
- Pipette 100 mL of the sample into a 150 mL beaker. Add 2 mL of ISA. Stir thoroughly.
- Rinse the electrode with Dl water, blot dry, and place it in the sample beaker. Wait for a stable reading and the sample concentration will be displayed on the measuring device.
- Determine the sample concentration using the data analysis tools in the software.
Electrode Storage
Short Term (overnight or the weekend)
- Rinse the electrode thoroughly with Dl water and place the tip in a diluted standard solution between measurements. Slide the FastFil sleeve upward to close the fill hole.
- Refill the reference before putting the electrode back into service
Long Term
Refill the electrode with Reference Filling Solution and close the fill hole.
Remove the sensing element and store in the glass vial or replace the cap to
protect the sensing element.
Follow procedures in “Electrode Preparation” and
“Checking Electrode Operation” before using the electrode again.
Electrode Cleaning
Cleaning should only be attempted if troubleshooting methods fail. The PVC
membrane is a delicate sensor and should not be brushed or otherwise
contacted. It can be rinsed vigorously under warm water to remove debris.
Soaking for 10-15 minutes in DI water might be useful in extracting other
contaminants.
Recondition the electrode by soaking in a low-standard solution immediately after any cleaning method.
Replacement Items
- SC-3527 Nitrate ISE Solutions: 60 mL Standard Solution, 30 mL Reference Fill Solution, 30 mL ISA Solution
- PS-3552 Nitrate ISE Replacement Module
Limited Warranty
For a description of the product warranty, see the PASCO catalog.
Copyright
This PASCO scientific instruction sheet is copyrighted with all rights
reserved. Permission is granted to non-profit educational institutions for
reproduction of any part of this manual, providing the reproductions are used
only in their laboratories and classrooms, and are not sold for profit.
Reproduction under any other circumstances, without the written consent of
PASCO Scientific, is prohibited. Rev: 10/17
Trademarks
- PASCO, PASCO scientific, PASCO Capstone, PASPORT and
- SPARKvue are trademarks or registered trademarks of PASCO Scientific, in the United States and/or in other countries. All other brands, products, or service names are or may be trademarks or service marks of, and are used to identify, products or services of, their respective owners. For more information visit www.pasco.com/legal.
Product End-of-Life Disposal Instructions
This electronic product is subject to disposal and recycling regulations that
vary by country and region. It is your responsibility to recycle your
electronic equipment per your local environmental laws and regulations to
ensure that it will be recycled in a manner that protects human health and the
environment. To find out where you can drop off your waste equipment for
recycling, please contact your local waste recycle/disposal service, or the
place where you purchased the product.
The European Union WEEE (Waste Electronic and Electrical Equipment) symbol (to the right) and on the product or its packaging indicates that this product must not be disposed of in a standard waste container.
Technical Support
For assistance with any PASCO product, contact
PASCO at:
- Address: PASCO Scientific 10101 Foothills Blvd. Roseville, CA 95747-7100
- Phone: 916-462-8384 (worldwide) 800-772-8700 (U.S)
- Email: techsupp@pasco.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>