Amico Aneroid Sphygmomanometer Instruction Manual

June 10, 2024
Amico

Aneroid Sphygmomanometer
Instruction ManualAmico Aneroid Sphygmomanometer

Intended Use

Amico Diagnostic Incorporated aneroid sphygmomanometers are used by professional healthcare providers and individuals trained in auscultatory blood pressure technique  to determine systolic and diastolic blood pressure in humans.

Contraindications

Amico Diagnostic Incorporated aneroid sphygmomanometers are contraindicated for neonate use. Do not use with neonatal cuffs or neonate patients.

General Warnings

A warning statement in this manual identifies a condition or practice which, if not corrected or discontinued immediately could lead to patient injury, illness, or death.
WARNING: Do not allow a blood pressure cuff to remain on patient for more than 10 minutes when inflated above 10 mmHg. This may cause patient distress, disturb blood  circulation, and contribute to the injury of peripheral nerves.
WARNING: If luer lock connectors are used in the construction of tubing, there is a possibility that they might be inadvertently connected to intra- vascular fluid systems,  allowing air to be pumped into a blood vessel. Immediately consult a physician if this occurs.
WARNING: Do not apply cuff to limbs used for IV infusion.
WARNING: Do not apply cuff to delicate or damaged skin.
WARNING: Patient should remain still during measurement to avoid erroneous readings.
WARNING: When using with an infant or child cuff, extra care must be taken to prevent overinflation.
With smaller cuffs (infant or child) the cuff can inflate over 300 mmHg with just two full compressions of the bulb.
To prevent discomfort or injury  to the patient and damage  to the instrument, bulb should only be  partially squeezed, so that each stroke inflates the cuff in40 mmHg to 60  mmHg increments until inflated to the desired level.
WARNING: Safety and effectiveness with neonate cuff sizes is not established.

Wall/Mobile Aneroid

  1. Connect one end of the coil tubing to the aneroid valve at the bottom of the unit.
  2. Join the cuff tube to the coil tubing.
  3. Join the inflation bulb to the other cuff tube.
  4. Verify air tight seal of system.

Pocket Aneroid

  1. Connect on cuff tube to the valve on the bottom of the pocket aneroid.
  2. Join the inflation bulb to the other cuff tube.
  3. Verify air tight seal of system.

Palm Aneroid

  1. Connect on cuff tube to the valve on side of the palm aneroid.
  2. Verify air tight seal of system.

Operation

Blood pressure measurements can be affected by the position of the patient and their physiologic condition. Before beginning a procedure, ensure that the patient rests for at  least five minutes, has support of their back and feet, and does not cross their legs. Passively support the patient’s lower arm and keep the upper arm at heart level. The  procedure needs to take place in a quiet environment with no talking. Failure to follow these recommendations can result in inaccurate blood pressure measurements.

  1. Select cuff size appropriate for the patient’s arm circumference. The applicable range, in centimeters, is printed on each cuff. NOTE: The “Artery Index Marker” on the  cuff should fall within the “Range” indicated on the cuff. If the artery index marker falls short of range, use a larger cuff to ensure accurate results. If the artery index  marker is past the range, use a smaller cuff to ensure accurate results.
  2. Wrap the cuff around the arm with the artery index marker located over the brachial artery and with the lower edge of the cuff 2.5 cm (25 mm) above the bend in the  elbow.
  3. Inflate cuff rapidly to a level 30 mmHg above estimated (or palpatory) systolic pressure.
  4. Partially open the valve to allow deflation at a rate of 2 to 3 mmHg per second. As the pressure falls, note systolic pressure and diastolic pressure detected with your stethoscope.
  5. Rapidly release the remaining pressure and record measurements.
  6. Fully exhaust cuff when not in use.

Storage

Pocket and Palm Aneroid
After use fully exhaust cuff, and open valve. Wrap cuff around aneroid and store in zippered case.
Wall and Mobile Aneroid
After use fully exhaust cuff, and open valve. Wrap cuff around bulb and store in cuff basket.

Cleaning

Aneroid Gauge, Inflation Bulb, and Valve
Wipe the aneroid gauge, inflation bulb, and valve with slightly dampened cloth or alcohol pad.
Reusable One-Piece Bladderless Cuff
Use one or more of the following methods and allow to air dry:

  • Wipe with mild detergent and water solution (1:9 solution). Rinse.
  • Wipe with Enzol per manufacturer’s instructions. Rinse.
  • Wipe with 70% isopropyl alcohol.
  • Launder with mild detergent in warm water (60 °C max), normal wash cycle. Cuff is compatible with 5 wash cycles (Reusable only). Block ports with laundering plug prior to wash.

Reusable Two Piece Cuff and Bladder
Safely clean the cuffs with a damp cloth or wash in warm water (60° C maximum) with mild detergent.
DO NOT PRESS WITH HOT IRON.
Before laundering the cuff cover:

  1. Remove the bladder from the two-piece cuffs.
  2. Place the hook and loop fasteners in the closed position.
  3. Machine launder using gentle cycle, warm water and mild detergent.
  4. Air Dry completely and reassemble components.

Calibration Check of Aneroid Sphygmomanometer

Quick Check of Calibration
At zero pressure, make certain the pointer is within the zero range on the dial. Although an unpressurized reading of zero does not guarantee accuracy at all scale points,  failure of the pointer to indicate zero (± 3 mmHg) is an obvious sign of error.
Full Check of Calibration
A full check of calibration is recommended at least every two years or according to local law:

  1. Connect the instrument under test to a high quality, known pressure standard and a 150 to 500 cc test volume using a T-connector.
  2. Pressurize gauge to slightly above 300 mmHg.
    Bleed pressure down no faster than 10 mmHg per second, stopping to check the pressure at 300, 250, 200, 150, 100, 50 and 0 mmHg. NOTE: Your ability to measure  the accuracy of a gauge depends upon the sensitivity of the pressure standard you use for the calibration procedure.

Standards

ANSI/AAMI/ISO 81060-1:2007, ANSI/AAMI SP-10:2002, EN 1060-1:1995, EN 1060-2:1995

Warranty

During the term of the warranty (3 years), Amico Diagnostic Incorporated will either repair or replace any components determined by Amico Diagnostic Incorporated to be  defective at no cost to the Customer. It shall be the Customer’s responsibility to return the Aneroid to Amico Diagnostic Incorporated. Shipping costs after the first twelve  (12) months shall be borne by the Customer. This warranty is valid only when the Aneroid has been properly installed according to Amico Diagnostic Incorporated’s specifications, used in a normal manner and serviced according to factory recommendations. This warranty does not cover failures resulting from accidents, misuse, abuse,  neglect, alteration, misapplication, improper maintenance, shipping or damages that may be attributable to acts of God. The term of this warranty begins from date of  purchase from Amico Diagnostic Incorporated.
Calibration Warranty
During the term of calibration warranty (lifetime), Amico Diagnostic Incorporated will recalibrate the aneroid at no charge should the aneroid deviate from the +/-3mmHg  accuracy specification. It shall be the Customer’s responsibility to return the Aneroid to Amico Diagnostic Incorporated. Shipping costs after the first twelve (12) months  shall be borne by the Customer.
Amico Diagnostic Incorporated does not honor verbal statements concerning the warranty. The distributor and/or dealer are not sanctioned to create verbal warranties about  the product. This warranty is inclusive and replaces all other warranties.
Amico Diagnostic Incorporated shall not be liable for any loss or damages, whether direct, incidental or consequential. All claims for warranty must first be approved by  Amico Diagnostic Incorporated’s Service Department (accessories@amico.com or  1.877.264.2697). A valid Return Goods Authorization (RGA) number must be obtained  from Amico Diagnostic Incorporated prior to commencement of any service work. Warranty work, which has not been pre-authorized by Amico Diagnostic Incorporated, will not be reimbursed.

www.amico.com

Amico Diagnostic Incorporated | 122 East Beaver Creek Road, Richmond Hill, ON L4B 1G6, Canada
Toll Free Tel: 1.877.264.2697
Tel: 905.763.7778 | Fax/Fax/Fax: 905.763.8587
Email/Courriel/Email: info@amico-accessories.com | www.amico.com
ADX-IM-ANEROID-TRILINGUAL 12.30.2021

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals