AND FG-D-CWP Waterproof Digital Platform Scale Instruction Manual

June 9, 2024
AND

FG-D-CWP Waterproof Digital Platform Scale

FG-D-CWP
Weighing Indicator
For Waterproof Digital Platform Scale
1WMPD4004854

This Manual and Marks

All safety messages are identified by the following, “WARNING” or “CAUTION”, of ANSI Z535.4 (American National Standard Institute: Product Safety Signs and Labels). The meanings are as follows:

WARNING CAUTION

A potentially hazardous situation which, if not avoided, could result in death or serious injury. A potentially hazardous situation which, if not avoided, may result in minor or moderate injury.

This is a hazard alert mark.

© 2022 A&D Company Ltd. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any language in any form by any means without the written permission of A&D Company Ltd.
The contents of this manual and the specifications of the instrument covered by this manual are subject to change for improvement without notice.
Windows, Word and Excel are registered trademarks of the Microsoft Corporation.

Contents
1. COMPLIANCE ……………………………………………………………………………………………………………………….. 4 1.1. COMPLIANCE WITH FCC RULES ……………………………………………………………………………………………. 4 1.2. CLASSIFICATION OF PROTECTION PROVIDED BY ENCLOSURES ……………………………………………………. 4
2. INTRODUCTION ……………………………………………………………………………………………………………………. 5 3. OUTLINE AND FEATURES……………………………………………………………………………………………………… 5 4. ITEMS INCLUDED………………………………………………………………………………………………………………….. 6 5. INSTALLATION AND PRECAUTIONS………………………………………………………………………………………. 7
5.1. PRECAUTIONS FOR INSTALLING THE INDICATOR ……………………………………………………………………….. 7 5.2. LOAD CELL CONNECTION …………………………………………………………………………………………………… 7 5.3. HOW TO CONNECT THE LOAD CELL ………………………………………………………………………………………. 9 5.4. VERIFYING LOAD CELL OUTPUT AND INPUT SENSITIVITY …………………………………………………………. 10 5.5. SETTING UP THE INDICATOR …………………………………………………………………………………………….. 11 5.6. SENSITIVITY ADJUSTMENT AND PARAMETER SETTINGS ………………………………………………………….. 11 6. PART NAMES …………………………………………………………………………………………………………………….. 12 6.1. DISPLAY AND SYMBOLS…………………………………………………………………………………………………….. 12 6.2. OPERATION KEYS ……………………………………………………………………………………………………………. 13 7. SENSITIVITY ADJUSTMENT…………………………………………………………………………………………………. 14 7.1. SENSITIVITY ADJUSTMENT ITEMS ………………………………………………………………………………………… 14 7.2. SETTING OF THE WEIGHING CAPACITY / MINIMUM DIVISION / DECIMAL POINT POSITION …………………. 16 7.3. SENSITIVITY ADJUSTMENT USING A WEIGHT …………………………………………………………………………. 17 7.4. CORRECTING FOR GRAVITATIONAL ACCELERATION ………………………………………………………………… 18 8. SETTING FUNCTIONS OR UNITS FOR USE ………………………………………………………………………….. 19 8.1. SETTING OF THE CF FUNCTION ………………………………………………………………………………………….. 19 8.2. SETTING UNITS FOR USE ………………………………………………………………………………………………….. 20 8.3. CF FUNCTION LIST ………………………………………………………………………………………………………….. 21 9. OPTIONS …………………………………………………………………………………………………………………………….. 22 9.1. OPTION LIST…………………………………………………………………………………………………………………… 22 10.MAINTENANCE ……………………………………………………………………………………………………………………. 23 10.1. REPAIR …………………………………………………………………………………………………………………………. 23 10.2. ERROR DISPLAY ……………………………………………………………………………………………………………… 23 10.3. ERROR CODE TABLE………………………………………………………………………………………………………… 24 11.SPECIFICATIONS ………………………………………………………………………………………………………………… 25 11.1. SPECIFICATIONS LIST……………………………………………………………………………………………………….. 25 11.2. DIMENSIONS …………………………………………………………………………………………………………………… 26 11.3. GRAVITY ACCELERATION MAP ……………………………………………………………………………………………. 27
3

1. Compliance
1.1. Compliance with FCC rules
Please note that this equipment generates, uses and can radiate radio frequency energy. This equipment has been tested and has been found to comply with the limits of a Class A computing device pursuant to Subpart J of Part 15 of FCC rules. These rules are designed to provide reasonable protection against interference when this equipment is operated in a commercial environment. If this unit is operated in a residential area, it might cause some interference and under these circumstances the user would be required to take, at his own expense, whatever measures are necessary to eliminate the interference. (FCC = Federal Communications Commission in the U.S.A.)
1.2. Classification of Protection Provided by Enclosures
This equipment complies with the IP Code IEC 60529. The meaning of the IP67 code is as follows:
IP : International Protection. 6 : Protection against ingress of dust and airborne particles. 7 : Protection against water and liquids, up to and included submersion of up to one meter.
4

2. Introduction
Thank you for purchasing an A&D FG-D-CWP weighing indicator (for the IP67 waterproof digital platform scale). This instruction manual mainly describes installation. Please refer to it when connecting or installing the FG-D-CWP to a weighing platform (load cell). Also, the FG-D-CWP has the same functions as that of the indicator for the FG-CWP series. Although this manual covers some of the functions and how to use the series, the following instruction manual explains all of them. [The instruction manual of the FG-CWP series] The instruction manual for understanding and fully utilizing the waterproof digital platform scale, the FGCWP series, can be downloaded from the A&D website (https://www.aandd.jp).
3. Outline and Features
The FG-D-CWP is an electronic indicator that is appropriate for composing an IP67 waterproof digital platform scale by connecting to a weighing platform (load cell), and it has the following features.
The indicator made of all stainless steel and has waterproof/dustproof performance compliant with IP67. (Intrusion of water into the device is prevented even when submerged to a depth of 1 m for 30 minutes.) Operating up to four 350 load cells is available. Available range selection, for the display resolution, is 1/20 to 1/100,000. Free setting of the weighing capacity, within the display resolution of 1/100,000, is available. Available weighing units are kg (kilogram), g (gram), Ib (pound), oz (ounce), t (ton), and pcs (pieces for the counting mode). It has similar specifications and functions as an indicator of the FG-CWP series.
5

4. Items Included
The following items are packed with the product. FG-D-CWP Model label (1083039230-1) Indicator

Attach here

Write the scale’s ratings, etc.

Quick start guide

FG-D-CWP

(This document is included in the packing contents for the balance.)

Quick Start Guide For Water proof platform scale

Engli sh (EN)

Guide de démarrage rapide Pour balance plate-forme étanche

Français (FR)

Schne llstartanleitung Für wasserdichte Plattformwaage

Deut sc h (DE)

Guía de inicio rápido Para báscula de plataforma a prueba de agua
Guida introduttiva Per bilancia a piattaforma impermeabile

Español (ES) It ali ano (IT) (CT)

1WMPD4004769

AC adapter

Rubber packing (To be used when the diameter of load cell cable is from 3.5 to 7.0)

(Image diagram)

6

5. Installation and Precautions
5.1. Precautions for Installing the Indicator
DANGER
Ground the scale, so that the user will not be subjected to an electric shock. To prevent electric shock, do not handle the USB cable with wet hands. The USB cable is not waterproof. Install the scale in an area where it will not get wet. Do not install the scale where flammable or corrosive gas is present. Do not put excessive force on cables. The scale is heavy. Use caution, as lifting may cause it to fall over. Use shielded load cell cables.
Please read “Precautions for Installing the Scale” in [Instruction Manual: FG- CWP Series] on the A&D website (https://www.aandd.jp) for other precautions, cautions as an assembled scale, storage and cleaning.

5.2. Load Cell Connection

FG-D-CWP has two connection configurations: the 6-wire configuration and the 4-wire configuration. We recommend connecting using the 6-wire configuration for optimum accuracy and stability. The applicable diameter of load cell cable is from 3.5 to 9.0 mm.
Terminal block for the load cell Weidmüller/LSF-SMT 3.50/07/180 3.5SN BK TU

Inside the load cell panel

Terminal No. 1 2 3 4 5 6 7

Terminal No.

1

SIG-

2

SIG+

3

EXC-

4

SEN-

5

SEN+

6

EXC+

7

SHIELD

Function Load cell input (-) Load cell input (+) Load cell excitation voltage (-) Sensing input (-) Sensing input (+) Load cell excitation voltage (+) Shield

7

Load cell

Cable
Load cell input (-) Load cell input (+) Load cell excitation voltage (-) Sensing input (-) Sensing input (+) Load cell excitation voltage (+) Shield

Cable Shield

(A) 6-wire configuration connection to load cell (recommended)

(B) 4-wire configuration connection to load cell

Method

Advantage

Disadvantage

Remark

6-wire configuration (recommended)

Less error even if the length of load cell cable is long or a thin load cell cable is used. Less error even if multiple load cells are used.

Slightly complicated to wire.

When using a summing box, it is strongly recommended wiring be done with the 6-wire configuration.

4-wire Easy to wire.
configuration

The accuracy of temperature counting is affected by resistance of the conductor wire for the load cell cable. It is affected by contact resistance of the connector.

Using a long load cell cable or multiple load cells tends to cause error.

Precautions for connecting with the 4-wire configuration Make sure to do the following when connecting with the 4-wire configuration.
Short-circuit between EXC+ and SEN+, and between EXC- and SEN-. When using a long a load cell cable, use one with a cross-sectional area that is as large as possible. Also, use a cable that is as short as possible.

Ref.: Connection examples of connecting to the SCB series, Dust & Waterproof Precision Weighing Platform (with the 4-wire configuration)

Load cell terminal number
1 2 3, 4 5, 6 7

Color of cable core for the SCB series
White Red Blue Green Yellow

Function of terminals for SCB series
Excitation – (Input terminal) Excitation + (Input terminal)
Signal – (Output terminal) Signal + (Output terminal)
Shield

8

5.3. How to Connect the Load Cell
1. Disconnect the power cable of the scale (indicator) from the plug.
2. Remove the 4 M3 screws from the load cell panel on the backside of the indicator, and open the load cell panel.

Load cell panel

3. Loosen the lock of the cable clamp connecting to the load cell panel, and pass the load cell cable through. Fit the rubber packing in the load cell cable.

Load cell cable Lock

Cable clamp

The applicable diameter of load cell cable is from 3.5 to 9.0mm. If the diameter of cable is from 3.5 to 7.0, replace the cable clamp with the accessory rubber packing that is packed with the product.
If the diameter of cable is from 7.0 to 9.0, the cable clamp does not need to be replaced.

Rubber packing

Load cell panel

4. Refer to “5.2. Load Cell Connection”, and connect the cable core to the terminal block of the load cell. Hold down the white button on each terminal in the terminal block of the load cell when connecting or disconnecting the wire. (Additional tool is not required.)
5. Once the load cell is connected, make sure that wiring has been done correctly.
6. Fasten the lock of the cable clamp, and secure the load cell cable.
7. Secure and close the load cell panel with the 4 screws. To maintain water- proof performance, tighten the cable clamp firmly when connecting the load cell cable. And tighten the bolts on the rear panel firmly. The recommended torque is 0.4Nm (4kgfcm).
9

Connect or disconnect the wire while pressing the white button

5.4. Verifying Load Cell Output and Input Sensitivity
The input sensitivity is 0.1 µV/division or more. When designing a weighing scale, the following calculation formula should be satisfied. The input sensitivity requires a change in output voltage, from the weighing device, to change the
display value as a one digit step. In order to achieve a scale with stable performance, select the input sensitivity as large as possible.

Calculating formula:

E x B x d

0.1

A x n

A: Rated capacity of load cell B: Rated output (mV/V) d: Minimum division E: Excitation voltage (mV) = 5000 mV for FG-D-CWP n: Number of load cells

Example of calculation:
In the case of designing a scale with a capacity of 60 kg and minimum division of 0.005 kg, using one load cell with a rated capacity of 100 kg and rated output of 1 mV/V:

Rated capacity of Load cell: Rated output: Minimum division: Excitation voltage: Number of load cells:

A = 100 kg B = 1 mV/V d = 0.005 kg E = 5000 mV n = 1

5000 x 1 x 0.005

100 x 1

= 0.25 > 0.1

The above indicates there is no problem with this design.

10

5.5. Setting Up the Indicator
1. Insert the USB plug into the AC adapter included with this product and connect it to the power supply. Confirm that local voltage and receptacle type are correct. You can also use except for the AC adapter included with this product. This product can also be powered from a USB port or mobile battery. Note: Not all operations are guaranteed. The USB cable serves as both a power and a communication.

If using the scale in a location where static electricity is
easily generated or where it may get wet, secure the grounding wire with the protective grounding terminal (screw).
Protective Grounding Terminal

Rear of Display Unit

5.6. Sensitivity Adjustment and Parameter Settings
When connecting to a load cell (weighing platform), sensitivity adjustment needs to be performed. For details, refer to “7. Sensitivity Adjustment”. As needed, configure the function settings. For details, refer to “8. Setting Functions or Units for Use”.

11

6. Part Names
Display Unit (SUS304)

Display unit knob Serial Number

Load Cell Panel Load Cell Connector
USB Plug USB Cable (Approx. 3 m)
6.1. Display and Symbols
Display Unit
Weighing Condition Conditions for Each Function

Comparator LED
Weighing Units

Display/symbol

Description

STABLE NET

This is lit when the weighing value is stable, indicating that the scale is in the proper condition for reading weighing values. This is lit when weight of the container (tare) is subtracted.

ZERO

This is lit when the scale is at the zero point (reference point for weighing).

Weighing Units ” “, ” “, ”

“, ” ” , ” ” and ” ” is lit.

This is lit when the display is held.

12

This is a function to detect impact to the mass sensor section and to display the impact level. Turns on when the connection with the wireless communication receiver is established.
While the comparator function is being used, the weighing value is compared using the preset threshold values and the indicator displays the result.
For details on the method of operation and various functions, refer to the [Instruction Manual: FG-CWP Series] on our website https://www.aandd.jp.

6.2. Operation Keys
Operation key ON/OFF Key Shows or hides the display.

Description

SAMPLE Key

Selecting ”

” as the unit activates individual weight setting mode.

Pressing and holding (for about 2 seconds) activates the comparator threshold setting

mode.

PRINT Key

Outputs the weighing value as data.

Increases the value of the digit flashing when setting.

MODE Key
Switches the weighing unit. Shifts to the right the digit flashing when making settings.

ZERO Key
Based on the zero point weighed at the time of turning on the display, by pressing the ZERO key when the weight value is stable within ±2 % of the weighing capacity, the zero point indicator turns on as soon as the display becomes zero. TARE Key The weight of the item on the weighing pan is subtracted as the weight of a container (tare weight).

13

7. Sensitivity Adjustment
The description of sensitivity adjustment in this manual is different from that in the instruction manual for the FG-CWP series. Please read the description below.

7.1. Sensitivity Adjustment Items
Sensitivity adjustment mode has the following three functions. Setting of the weighing capacity, minimum division, and decimal point position. Sensitivity adjustment using a weight Correcting for gravitational acceleration
There are two ways to enter sensitivity adjustment mode as shown below.

Method 1 (using operation keys)

1. Confirm that the scale(indicator) is in weighing mode (display shows measurement unit).

2. Press and hold the TARE key for 4 sec. or longer. Release the key when

is

displayed. (Then go to “7.2. Setting of the Weighing Capacity / Minimum Division / Decimal

Point Position”, “7.3. Sensitivity Adjustment Using a Weight”, or “7.4. Correcting for

Gravitational Acceleration”.)

Method 2 (using the CAL switch)
1. Confirm that the scale(indicator) is in weighing mode (display shows measurement unit).

2. Remove the protective cover from the four screws of the load cell panel on the rear of the display unit, and then loosen the four screws and open the load cell panel.

3. Press the CAL switch to display

.

(Then go to 7.2. Setting of the Weighing Capacity /

Minimum Division / Decimal Point Position”, “7.3.

Sensitivity Adjustment Using a Weight”, or “7.4.

Correcting for Gravitational Acceleration”)

CAL switch Load cell panel
Rear of display unit

14

Relations between items in sensitivity adjustment mode and key operations are shown in the diagram below.

Weighing mode

Method 1 Press and hold

Press

Method 2

In sensitivity adjustment mode (transition method)

Press

Setting of the weighing capacity, minimum division, and decimal point

Press

Press

Sensitivity adjustment using a weight

Press

Press

Press Press

Press

Setting of the CF function

Press

or

Display OFF
Note: Use the MODE key to select the item to execute, and then execute with the TARE key. Note: To end sensitivity adjustment mode, press the ZERO key or CAL switch. The display
turns off. (You can also turn it off using the ON/OFF switches.)

15

7.2. Setting of the Weighing Capacity / Minimum Division / Decimal Point Position
Set the weighing capacity, minimum division, and decimal point position. Be sure to perform these setting first when connecting to a load cell (base unit). The indicator does not require setting each time, if nothing is to be changed.

1. Refer to “7.1. Sensitivity Adjustment Items”, to enter

the calibration mode. The indicator displays

.

2. Press the TARE key to enter the setting mode of the weighing capacity, minimum division, and decimal point position.
3. The indicator displays the currently set minimum division and calibration unit. (Display example: – kg )
The blinking digit indicates the minimum division. Press the PRINT key to select , , or . The unit indicates the sensitivity adjustment unit. Press the
SAMPLE key to select either “kg” or “lb”.

Press
Press (Select)

Press the TARE key to store the currently set minimum division and calibration unit, and proceed to the next step.

Press

4. After displaying

, the indicator displays

the currently set weighing capacity and decimal

point position. (Display example:

kg )

Set the weighing capacity and decimal point

position using the following keys:

MODE Shift the blinking digit to the right.

PRINT Increases the number (+1) at the blinking digit.

SAMPLE Shift the decimal point position.

Press (Select)

Press the TARE key to store the currently displayed weighing capacity and decimal point position, and proceed to the next step.
Available setting range of the resolution is 100,000 or less. The resolution is the value that divides the weighing capacity by the minimum division. (Decimal point position is ignored.)

5. After displaying

, the indicator displays

,

and the setting is complete.

After setting the weighing capacity, minimum division,

and decimal point position, be sure to perform

“7.3. Sensitivity Adjustment Using a Weight”.

Press

16

7.3. Sensitivity Adjustment Using a Weight
This function adjusts the scale (indicator) for accurate weighing. Sensitivity adjustment must be done when the FG-D-CWP is initially connected to a load cell (base unit) or the combination is changed. Sensitivity adjustment may also be required according the changes in the environment.
When the scale (indicator) is initially installed. When the scale (indicator) is moved to a location far away. When the ambient environment has greatly changed. In periodic sensitivity adjustment

Note: Prepare a weight. (A weight equivalent to the weighing capacity of the scale to be adjusted is recommended. However, the value for the weight can be set (changed).)

1. Turn on the scale and supply it with power for 30 minutes or longer.

Change the parameter for the function table item

or

place something on the weighing pan to keep the automatic

power off function from activating.

2. Refer to “7.1. Sensitivity Adjustment Items”.

is

displayed when you enter sensitivity adjustment mode.

Press

3. Press the MODE key to display

.

4. Press the TARE key to display

.

Make sure there is nothing on the weighing pan and wait for

the stability indicator to light up.

Press

5. Press the TARE key to save the zero point on the scale and display the value for the weight.
The value for the weight is the same as the weighing capacity. (Initial state)
To adjust just the zero point, turn off the scale to end without performing step 6.

Press
Press (Select)

6. To adjust with a weight value different than the weighing capacity, change the value with the keys below.

Key

Content

Moves the blinking digit

Weight
Weighing Pan

Increases the blinking digit by 1
A weight equivalent to the weighing capacity of the scale is ordinarily used. If using a different weight, it should be at least 2/3 the weighing capacity.

Press

17

7. Place on the weighing pan a weight of the value displayed. Wait for the stability indicator to light up.

8. Press the TARE key to end sensitivity adjustment.

After displaying

, the setting is complete.

will be displayed.

9. To end setting, press the ZERO key or CAL key. The display turns off. Note: If the scale (indicator) is moved to a distant location after sensitivity adjustment, adjust the gravitational acceleration value to that of the area where it will be adjusted next. Then adjust the scale. See the next chapter for setting gravitational acceleration.

7.4. Correcting for Gravitational Acceleration
When relocating a scale (indicator), you need to perform sensitivity adjustment with a weight. If a weight is not available, you can adjust the scale (indicator) by correcting the gravitational acceleration. Refer to the acceleration map on the last page, and change the gravitational acceleration value recorded in the scale (indicator) to the one for appropriate location.
Note: If sensitivity adjustment is done using a weight at the place where the scale (indicator) is used, it does not need to be corrected for gravitational acceleration.

1. Refer to “7.1. Sensitivity Adjustment Items”.

is

displayed when you enter sensitivity adjustment mode.

2. Press the MODE key twice to display the gravitational acceleration parameter (Example: “G 9.7985”).

3. Press the TARE key to enter the gravitational acceleration setting mode.

4. To change the gravitational acceleration parameter displayed, use the keys below.

Key

Content

Moves the blinking digit

Increases the blinking digit by 1

Press
Press
Press (Select)

twice

5. Press the TARE key to display

and save the

parameter. After displaying the gravitational acceleration

parameter (Example: “G 9.7985”), the setting is complete.

6. To adjust using a weight, return to step 2 of “7.3. Sensitivity Adjustment Using a Weight”. To end setting, press the ZERO key or CAL key. The display turns off.

Press

18

8. Setting Functions or Units for Use
The indicator has CF function settings to specify the indicator performance. After installing the indicator, the indicator does not require resetting or changing of this CF function as far as it is used normally. If you want to maintain only the needed units, set using this function. The parameters set in the CF function settings and unit setting are maintained even if the power is turned OFF.
8.1. Setting of the CF Function

Item

Parameter

1. Refer to “7.1. Sensitivity Adjustment Items”, to enter the sensitivity

adjustment mode. The indicator displays

.

2. Press the MODE key three times to display

.

Press

3. Press the TARE key to display the setting item.

4. Select the setting item and parameter using the following keys: MODE Proceed to next setting item. PRINT Increases the number (+1) at the blinking digit.
(Change the parameter) When the currently stored set is changed, the STABLE indicator ” ” turns off. Changing the parameter using the PRINT key at CF1, 2, 8, 12 or 13 is not available.

Press
Press (Select)

5. Press the TARE key to store the setting. After displaying

, the indicator displays

and proceeds to

the unit setting mode.

To proceed, without storing the CF function setting, press

the ZERO key. After displaying

, proceed to

the use unit setting mode. (Display example:

pcs )

Press

6. When the unit setting is to be performed, go to step 5 of “8.2.

Setting the unit for use”.

When the unit setting is not to be performed, press the ZERO1 key.

After displaying

, the setting is complete.

To adjust to the designed scale specification, set other items by following the procedure described above.

3 times

19

8.2. Setting Units for Use

1. Refer to “7.1. Sensitivity Adjustment Items”, to enter the sensitivity

adjustment mode. The indicator displays

.

2. Press the MODE key three times to display

.

Press

3 times

3. Press the TARE key to display the setting item.

Press

4. To proceed without storing this setting item, press the ZERO key.

After displaying

, proceed to the unit setting mode.

(Display example:

pcs )

Press

5. Select the unit to be used, using the following keys. MODE Select the unit (Shift the blinking unit to the right)

PRINT Set whether the selected unit is available (0) or not available (1). (Change the number of the blinking digit to “0” or “1”.)
About the unit setting mode Each of the display digits are assigned to a unit setting. (The setting at the first and fourth digit is not available.)

Display assignment

7

6

5

4

3

2

1

Unit

pcs g

kg

Set disabled

oz

lb

t

The unit during selection is indicated.

With each digit, if setting to “0”, its unit is not available,

and if setting to “1”, its unit is available.

(Only when set to “1”, is a unit available.)

Even if a unit is stored, availability for use depends on

the weighing capacity and resolution.

Example) When setting to

pcs “, only “pcs”

and “kg” are available as units.

6. Press the TARE key to store the setting. After displaying

,

the indicator displays

, and the setting is complete.

If you want to proceed without storing the setting of

the unit setting mode, press the ZERO key.

After displaying

, the setting is complete.

Press (Select)
Press

20

8.3. CF Function List

Item Used internally

Parameter

Description

Not available for setting change

Used internally

Not available for setting change

Zero tracking
Power on zero range
Calculation of tare Zero-point mark during tare subtraction
Tare clear by zero operation Used internally Minus and printing limit Auto tare function
Stable detection range
Used internally

Not available for setting of function

(It is fixed to

)

Available for setting of function None

±10% of weighing capacity ±50% of weighing capacity ±100% of weighing capacity
Tare by displayed division Tare by internal count Do not illuminate the zero- point mark during tare subtraction Illuminate the zero-point mark during tare subtraction only when the net weight is zero Illuminate the zero-point mark during tare subtraction Tare is not cleared by zero operation Tare is cleared by zero operation

When

,

has no

function.

Calibration zero reference

Not available for setting change
Non-limit Limit has function

No auto tare function

Auto tare function

Not available for setting of function

(It is fixed to

)

Available for setting of function Not available for setting change

Refer to below note.

When

,

any of ,

, or

has no function.

When

,

has

no function.

21

Item Used internally

Parameter

Description

Not available for setting change

Saving zero value and tare weight Display limit of ” ”
Zero tracking width/ time

Do not save
Save
Weighing capacity + 8d Weighing capacity + 8d + (capacity/30) Weighing capacity + 8d + (capacity/15) Weighing capacity + 8d + (capacity/6) Weighing capacity + 8d + (capacity/3) 0.5d / 0.5 seconds 0.5d / 1.0 seconds 0.5d / 1.5 seconds 0.75d / 0.5 seconds 0.75d / 1.0 seconds 0.75d / 1.5 seconds 1.0d / 0.5 seconds 1.0d / 1.0 seconds 1.0d / 1.5 seconds

Saves zero value and tare weight even after the power turns off.
Zero tracking: The function to track the zero point and keep the display zero

: Factory settings (which differ depending on the destination market) d = minimum division The settings are not changeable for CF1, 2, 8, 12, or 13.

Note: About minus and printing limit setting (

)

When setting

to ” “, the indicator has the following limit.

When the total weight is -19d or less, the indicator displays – .

When the total weight is minus, the indicator does not perform the printing.

9. Options

9.1. Option List

Option name FG-27CWP AD-8541-PC AD-8931 AX-TB301

Contents Bluetooth communication interface Bluetooth dongle for PC connection Bluetooth External Display AC adapter

Refer to the optional instruction manual for details.

22

10. Maintenance
Refer to “5. Installation and Precautions” regarding use. Refer to “10.3. Error Code Table” and corresponding mode for displayed error code. Refer to “7. Sensitivity Adjustment” regarding precision weighing. Periodically check the accuracy of weighing. Adjust the scale (indicator), if it is moved to another location or the environment has changed.

10.1. Repair
Do not disassemble / assemble the indicator without an authorized service engineer. Doing so may cause an electric shock or damage to the indicator, etc. In this case, repair is not covered under warranty. Contact your local A&D dealer if your indicator needs service or repair.

10.2. Error Display

In this situation Nothing displayed. Scale does not turn on.
The scale does not display zero at first.
is displayed.
is displayed.
is displayed.

Confirm these items Is the main power the correct voltage? Is the main power cord properly connected? Check around the weighing pan. Is there anything on the weighing pan? Refer to “7. Sensitivity Adjustment” and perform zero point sensitivity adjustment.
Weighing error that meaning “Overloaded”.
Indicates that the sample weight is too light to set the unit weight in the counting mode.
Sensitivity adjustment error that meaning “Too light”.

is displayed.
is displayed and does not proceed.
Fixed display.

Displayed when the main power drops.

The weighed value is unstable due to drift, vibration or other factors. A breeze or vibration may be affecting measurement. Check around the weighing pan. Check the connection of the load cell cable. No zero display when the display is turned on. Remove anything that is on the pan. Perform zero point sensitivity adjustment.

Did you use the “hold function” ? Refer to the [FG-CWP Series

Instruction Manual] on the A&D website (https://www.aandd.jp) and

change the function

.

Turn off scale and turn it on again.

23

10.3. Error Code Table

When any of the following errors are displayed, try turning the display off and on again.

Display

Content

is displayed The temperature sensor has failed.

is displayed The memory (circuit) has failed.

is displayed The internal circuitry has failed.

is displayed The mass sensor has failed.

24

11. Specifications

11.1. Specifications List

Specifications are subject to change without notice.

Input sensitivity

0.1 V/d min. (d: minimum division)

Input signal range

-16 mV to 16 mV

Load cell excitation voltage 5 V ± 10 %

Load cell drive capacity

Up to 4 x 350 ohm loadcells

Temperature coefficient Zero ±0.02 V/°C typ. ±0.1 V/°C max.

Temperature coefficient Span ±3 ppm/°C typ. ±15 ppm/°C max.

Linearity

±0.01 % of full scale

Maximum display resolution 100,000

Display

7 segment LCD with backlight (Character height 30 mm)

Display update

Approx. 10 times/second

Operating Consitions

-10 °C to 40 °C, 85 % R.H. or less (non-condensing)

Power Supply

AC adapter or mobile battery or USB port USB Cable Length: Approx. 3 m (USB cable is also used for communication.)

Dimensions

Width 240 mm × Depth 76 mm × Height 155 mm

Weight (approximately)

1.6 kg (Includes stand)

25

11.2. Dimensions
FG-D-CWP
Unit: mm
26

11.3. Gravity Acceleration Map

Values of gravity at various locations

Amsterdam Athens Auckland NZ Bangkok Birmingham Brussels Buenos Aires Cape Town Chicago Copenhagen Cyprus Frankfurt Glasgow Havana Helsinki Jakarta Kolkata (Calcutta) Kuwait Lisbon London (Greenwich) Los Angeles Madrid

9.813 m/s2 9.800 m/s2 9.799 m/s2 9.783 m/s2 9.813 m/s2 9.811 m/s2 9.797 m/s2 9.796 m/s2 9.803 m/s2 9.816 m/s2 9.797 m/s2 9.811 m/s2 9.816 m/s2 9.788 m/s2 9.819 m/s2 9.781 m/s2 9.788 m/s2 9.793 m/s2 9.801 m/s2 9.812 m/s2 9.797 m/s2 9.802 m/s2

Manila Melbourne Mexico City Milan Moscow New York Oslo Ottawa Paris Rio de Janeiro Rome San Francisco Singapore Stockholm Sydney Taichung Taipei Tokyo Vancouver, BC Washington DC Wellington NZ Zurich

9.784 m/s2 9.800 m/s2 9.786 m/s2 9.807 m/s2 9.816 m/s2 9.802 m/s2 9.819 m/s2 9.807 m/s2 9.810 m/s2 9.788 m/s2 9.803 m/s2 9.800 m/s2 9.780 m/s2 9.819 m/s2 9.796 m/s2 9.789 m/s2 9.790 m/s2 9.798 m/s2 9.810 m/s2 9.801 m/s2 9.803 m/s2 9.808 m/s2

27

1. Back cover: A&D Global

A&D Company, Limited

3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013, JAPAN

Telephone: [81] (3) 5391-6132

Fax: [81] (3) 5391-1566

A&D ENGINEERING, INC.

47747 Warm Springs Blvd, Fremont, California 94539, U.S.A.

Tel: [1] 800-726-3364

Weighing Support:[1] 888-726-5931

Inspection Support:[1] 855-332-8815

A&D INSTRUMENTS LIMITED

Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire OX14 1DY United Kingdom

Telephone: [44] (1235) 550420

Fax: [44] (1235) 550485

A&D AUSTRALASIA PTY LTD

32 Dew Street, Thebarton, South Australia 5031, AUSTRALIA

Telephone: [61] (8) 8301-8100

Fax: [61] (8) 8352-7409

A&D KOREA Limited

..()

633 () 817 07331

( 817, Manhattan Bldg., 33. Gukjegeumyung-ro 6-gil, Yeongdeungpo-gu, Seoul, 07331 Korea )

: [82] (2) 780-4101

: [82] (2) 782-4264

OOO A&D RUS

OOO ” ”

:121357, , ., . , 17

: 117545, , . , . , .3, .6, . 8

( 121357, Russian Federation, Moscow, Vereyskaya Street 17 )

.: [7] (495) 937-33-44

: [7] (495) 937-55-66

A&D Instruments India Private Limited

D-48

( D-48, Udyog Vihar, Phase- , Gurgaon – 122016, Haryana, India )

: [91] (124) 4715555

: [91] (124) 4715599

A&D SCIENTECH TAIWAN LIMITED. A&D

5 4

( 4F No.5 Ching Tao East Road, Taipei Taiwan R.O.C. )

Tel : 886 2322-4722

Fax : 886 2392-1794

A&D INSTRUMENTS (THAILAND) LIMITED () 168/16 1 12110
( 168/16 Moo 1, Rangsit, Thanyaburi, Pathumthani 12110 Thailand )
Tel : [66] 20038911

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

AND User Manuals

Related Manuals