KELLER 36KyX Level Probes Instruction Manual

June 9, 2024
KELLER

36KyX Level Probes
Instruction Manual

36KyX Level Probes

OPERATING INSTRUCTIONS LEVEL PROBES

SERIES
26C, 26Y, 27Y 26X, 26Xi, 27X, 36KyX, 36XS 36XW, 36XiW, 36XiW-CTD 46X
and various other customised series

CONTACT

KELLER
Druckmesstechnik AG
St. Gallerstrasse 119
CH-8404 Winterthur
Tel. +41 52 235 25 25
info@keller-druck.com| Keller Gesellschaft
für Druckmesstechnik mbH
Schwarzenegger 17
DE-79798 Jestetten
Tel. +49 7745 9214 0
eurocenter@keller-druck.com
---|---

You can find details of other international KELLER subsidiaries and agents at https://keller-druck.com/en/company/subsidiaries

GENERAL INFORMATION AND SAFETY INSTRUCTIONS

These operating instructions contain important information on using level probes correctly.The quick installation guide enclosed will be sufficient for experienced specialist personnel. Please read these operating instructions carefully for a detailed explanation of how to install the device and put it into operation. Follow the safety information in these operating instructions. You must also comply with national legislation, standards and regulations. These operating instructions form an integral part of the device and must always be accessible to the relevant staff.
Disclaimer
KELLER accepts no liability in case of improper use, damage or modification to the device or failure to observe this manual.
SUBJECT TO TECHNICAL ALTERATIONS!
1.1. SYMBOLS USED

SYMBOL WARNING NATURE AND SOURCE OF THE DANGER
DANGER Danger of death or injury to staff.
WARNING Potentially hazardous situation that could result in serious

injuries or even death.
| NOTE| Tips and information for users.

1.2. INTENDED USE

| DANGER| In the event of incorrect use!
---|---|---

  • Level probes are used to convert a hydro static pressure into a standardised electrical signal.
  • The level probes in the various series referred to have been developed for applications measuring the relative or absolute pressure of water or fill levels depending on the type of probe.
  • Make sure that the device is suitable for your correspond ding application. Please get in touch with your direct sales contact if you are unsure of any thing.
  • Please refer to the relevant data sheet and agreed specifications for information about the level probe’s features.
  • The technical specifications listed in the data sheet are only binding insofar as no other agreements have been made.
  • The liquids used as measuring media must be compatible with the materials specified in the data sheet that come into contact with the media.

1.3. STAFF QUALIFICATIONS
The product must only be assembled, installed, put into operation, operated, maintained, taken out of operation and disposed of by technically trained staff.

PRODUCT IDENTIFICATION

| SWISS MADE| **
---|---|---
Type: PR-26Y
Range: 0…10 mH2O
Output: 4…20 mA
Supply: 8…32 VDC| WH: n.c.
RD: OUT/GND
BK: +Vs
P/N: 222625.0001| 15/22**

The label and the specifications on the product serve to identify it. The laser engraving or adhesive labels typically contain the following details:

  • Product number
  • Type designation
  • Pressure range
  • Output signal
  • Supply voltage
  • Electrical connections
  • Production date

Products with a serial number display this on the metal housing.

MOUNTING

| DANGER| Danger of death by electrocution, pressure discharge or leaking media!
The device must only be installed on systems when they are not pressurized or connected to a power supply.
---|---|---
| DANGER| Danger of death in the event of incorrect installation!
The device must only be installed by specialist personnel who have read and understood these operating instructions.
Follow the national standards and safety regulations during the assembly and installation of the device.

3.1. INSTALLATION AND SAFETY INFORMATION
Only operate the level probe within its technical performance limits. You can find these on the label, in the data sheet or in the specifications. The device must not be operated for long periods in its overload range as this can damage it.
The measuring medium must not be permitted to ice up.
Ensure that the material used for the selected level probe, including the associated seals, is suitable for the medium to be measured.
As standard, level probes are delivered without accessories such as mounting materials.
Avoid electrochemical voltage potentials.
DO NOT use the device to perform safety functions.
Do not remove the packaging until immediately before installation to avoid damaging the diaphragm and/or the thread.
The output signal may be dependent on the installation position. Always tighten the transmitter to the manufacturers’ instructions
Feed the cable output on pressure transmitters with cable into a dry terminal box.
Avoid excessive tensile loads on the cable connections.
Make sure the holes in the end cap are free from impurities/dirt.

3.3. MECHANICAL INSTALLATION
Installation of screw able level probe with pressure connection

Before installation, make sure that the seal surfaces on the device and the measuring point are clean and undamaged.
Make sure that the seal is intact and is in its intended position. Check that the seal specifications correspond with the environment where it will be used.
Make sure that you do not damage the diaphragm. Front flush versions are to be treated with particular caution.
Use suitable tools for installing the device.
Tighten the device with the appropriate torque.

Follow the instructions below to ensure optimum installation:

  1. Make sure that the seal surface seals correctly.
  2. Screw the device into the corresponding thread by hand.
  3. Using a suitable torque wrench on the spanner flats, tighten the device into place.

Install the device in such a way that it cannot be exposed to any prohibited mechanical tension. This could have a negative effect on the zero point and/or the characteristic lines.

3.4. ELECTRICAL INSTALLATION
Connect the device according to the electrical connections on the label or the specification.
Prevent the cable from kinking.
The following minimum bending radius’s must be adhered to when installing the cable:
Minimum bending radius for cables without capillary:
permanent installation: 6-fold cable diameter
flexible usage: 12-fold cable diameter
Minimum bending radius for cables with capillary:
15-fold cable diameter
Ensure adequate earthing via the cable shield.
For reference devices (relative pressure versions, PR), make sure there is adequate, dry ventilation.
Make sure the cable is mounted securely, e.g. by using an additional retaining cable, for cable lengths > 50 m.

CONNECTION DIAGRAMS

KELLER 36KyX Level Probes-fig2| KELLER 36KyX
Level Probes-fig3
---|---
KELLER 36KyX Level Probes-fig4| KELLER 36KyX
Level Probes-fig5

May vary for specific series. Please see the relevant communication protocol for additional information.
COMMUNICATION PROTOCOLS
You can find links to these via the relevant product on the KELLER website at www.keller-druck.com.

COMMISSIONING

| WARNING| Before operating the device for the first time, check whether the device has been installed properly.
---|---|---
| WARNING| The device may only be operated by qualified personnel who have read and under stood the operating instructions.
| WARNING| The device may only be operated within the specifications. See the technical data sheet or the agreed specifications.

TROUBLESHOOTING

Common installation errors:

DIFFERENT ZERO POINT SIGNAL
Possible cause:

  • Diaphragm damaged
  • Ambient temperature too high/low

Action:

  • Contact the manufacturer and, if necessary, replace the device
  • Keep to the permitted temperatures given in the data sheet

SIGNAL SPAN DECLINES/IS TOO LOW
Possible cause:

  • Mechanical overload
  • Abrasive/aggressive medium; corrosion on the diaphragm/pressure connection

Action:

  • Contact the manufacturer

CONSISTENT OUTPUT SIGNAL WHEN PRESSURE CHANGES
Possible cause:

  • Mechanical overload
  • Electrical fault

Action:

  • Replace the device; if the fault occurs again, contact the manufacturer

NO OUTPUT SIGNAL
Possible cause:

  • No supply voltage
  • Device connected incorrectly
  • Broken cable

Action:

  • Check the supply voltage
  • Correct connection in accordance with the electrical connections

SIGNAL SPAN FLUCTUATING
Possible cause:

  • Source of EMC interference nearby (e.g. pump, frequency converter, etc.)

Action:

  • Remove the source of interference
  • Shield the source of interference properly

SERVICING AND REPAIRS

6.1. MAINTENANCE
KELLER products require no maintenance and, if used in accordance with the specifications, are fault free.
6.2. RECALIBRATION
Depending on the conditions in the area in which the device is used, its zero point or signal span may change over time, causing an incorrect output signal.
We recommend that highly accurate devices with an accuracy of < 0,1 %FS are periodically re calibrated or have their zero point adjusted.
6.3. RETURNS
Before returning a device for re calibration or repair, it must be thoroughly cleaned and securely packaged. For faulty devices, use the KELLER returns form and describe the fault in as much detail as possible.
If your device has come into contact with harmful substances, you must state this on the returns form.
If you return a device without mentioning contact with harmful substances and our repair department suspects that this is the case, the device will not be repaired until the facts have been established.

DISPOSAL

To dispose of the device, either return it to the supplier or dispose of it in a professional manner in accordance with the corresponding EU directive.

WARRANTY TERMS

The warranty is limited to 12 months from the date of delivery. KELLER guarantees that the products are free of manufacturing and material defects and that they comply with the specifications confirmed in writing.

EU DECLARATION OF CONFORMITY

The device supplied complies with the legal requirements. The relevant directives and harmonised standards are listed in the EU Declaration of Conformity applicable to the product. You can find the EU Declarations of Conformity by the relevant product at www.keller-druck.com.

Subject to technical alterations.
450565.0123, Version 09/2022