KELLER K-114 Interface Converter Owner’s Manual

June 9, 2024
KELLER

Manual
Interface Converter K-114KELLER K 114 Interface
Converter-

K-114 Interface Converter

General information
The K-114 interface converter is used to convert a USB signal to a serial RS485 half-duplex signal, for connection to any desired computer with a USB port. The K-114 converter and the connected transmitters are powered via the USB connection. An external power supply unit can also be connected to power multiple instruments. The K-114 converter is preferred for use with KELLER products.
With the K-114 interface converter, you can… convert a USB signal to RS 485 (half-duplex) measure the applied signal voltage (0…12 VDC) e.g. the output signal voltage from a pressure transmitter  measure the power consumption of connected end consumers in the 0…40mA range e.g. the power consumption of connected end consumers or the power output signal from a pressure transmitter The K-114 interface converter offers you visual status and configuration displays (LED) electrical isolation between the computer and the converter  diagnostic tools

KELLER -icon

Pin assignment for converter / connections

K-114
plug-in screw terminal| K-114-A
Fischer plug-in connector S 103 A054-130| K-114-B
Binder female cable connector Serie 680 5-pole
---|---|---
| |

Technical specifications

Symbol| Parameter| Condition| Min.| typ.| Max.| Unit
---|---|---|---|---|---|---
Current consumption, K-114| | No connection| 30| 43| 55| mA
Power consumption, K-114| | No connection| 150| 215| 275| mW
Supply for end consumer(s)| U-Out| No power supply unit| 11,2| 12.| 12,5| VDC
Supply for end consumer(s)| I-Out| With power supply unit| –| | 150| mA
External supply| | Power supply unit| 12| 15| 20| VDC
Input voltage| U-In| K-114-A / K-114-B| 0| | 12| VDC
Accuracy of input voltage| U-In| R! >- 30 k()| | 0,2| 0,3| %FS
Current measurement| I-Out| | 0| | 40| mA
Accuracy of current measurement| I-Out| | | 0,2| 0,3| %FS
Data transmission rate| slow| Max. transmission distance 20 km| –| –| 250| kbps
| high| Max. transmission distance 1 km| –| | 3| Mbps
K-114 device| Protection Class| | IP40| –| –| –
Device fuse (USB)| Fl| No power supply unit| | 0,5| | A
Storage and operating temperatures| | | -10| 20| 50| °C

Description

The K-114 communicates with the connected devices via an RS485 (half-duplex mode) bus. The devices connected to the K-114 are supplied via the PC’s USB output or via an external power supply unit (K-114 socket). KELLER products operate with fail-safe drivers that generate a logical «high» at the reception output in case of short-circuited, open or terminated inputs, in order to avoid invalid signal statuses. KELLER products also have a slew rate limitation which limits the edge steepness (i.e. slew rate) of the driver output. This prevents high-frequency emissions from devices and data lines. Up to a maximum of 128 transmitters can be connected to this RS485 master bus.

Typical application

KELLER K 114 Interface Converter-fig2

Overview of products

Product Connection Product number Products supported
K-114 Plug-in screw terminal 309010. All digital KELLER products Series 3X,

Series 4X, DCX*
K-114A| Fischer plug| 309010.| DCX-16 /-22-25PVDF / -38, LEO-Record, LEX1, ARC1
K-114B| Binder female cable connector| 309010.| Series 30X / 40X, LEO3, EV-120, dV22-PP, dV2-PS, Castello
K-114M| M12 female cable connector| 309010.| DCX-18 only (communication and charging cable)

  • no voltage input
    ** requires additional cable option

System requirements for K-114_Config software

Processor min. Pentium 75 MHz
Screen resolution min. 1024 x 768
Main (working) memory min. 16 MB RAM
Free hard disk space min. 20 MB (recommended)
Internet connection recommended (required for support)
Operating system Windows XP Windows 7

Factory configuration for the K-114

Converter bus address 253 (cannot be changed)
Baud rate 9600 baud
Echo off off
Bias network off
Termination High Speed off
High Speed off

These settings are generally recommended to ensure troublefree operation of KELLER products.
Functional overview for the K-114

KELLER K 114 Interface Converter-fig3

Echo off
Data sent from the PC (TX) aren’t received by the PC.
Bias network
Prevents undefined bus levels when line drivers are inactive.
➞ Greater immunity to interference
Termination
Prevents reflections on the signal lines.
High Speed
Disables the RS485 driver’s slew-rate limitation.
This allows communication at higher transmission speeds (> 250 kbps). Unlike high speed mode, the standard mode (< 250 kbps) reduces reflections on the signal lines and provides better EMC behaviour.
KELLER devices operate at transmission speeds < 250 kbps, so this function is not activated ex factory.

LED display

KELLER K 114 Interface Converter-fig 4

Power

Ready to operate
Flashing rapidly, error ➞ U-USB < 4,5 VDC ➞ Connect power supply unit
Flashing slowly, error ➞ U-UOUT < 11,2 VDC ➞ Connect power supply unit
Error

RX/TX
Shows the status of the transmission line (TX) and the reception line (RX).
TX, sending data via the RS485 bus
RX, receiving data via the RS485 bus

Echo off
Shows the status of the Echo off function.
Echo off disabled (➞ echo switched on)
Echo off enabled (➞ echo switched off)

Bias on
Shows the status of the Bias on function.
Bias resistors (560 Ω) for RS485A and B are enabled
No bias resistors are connected
KELLER products use fail-safe RS485 drivers which output a valid signal even with an undefined level.
For this reason, it is not mandatory to enable this function.
Termination on
Termination resistor (120 Ω) connected
Termination resistor not connected

High Speed
Shows the status of the High speed function.**
High speed enabled
High speed disabled

**IMPORTANT: On battery-powered devices, this function can lead to operating errors ➞ Recommendation: do not enable the function.
RoHS
This product is compliant with EC directive 2002/95/EC on the Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment (RoHS).
Disposal
**** This symbol on the product or the product documentation indicates that the product must not be disposed of as normal household waste at the end of its useful life. In order to prevent potential damage to the environment or to health as a result of uncontrolled waste disposal, this product must be separated from other waste and must be correctly recycled in order to ensure the sustainable use of the raw materials.

Help with fault analysis

KELLER - Fig

Installing the K-114_Config software

First, install the K-104 / K-114 driver on your computer and then run the K-114_Config software.
(The software can be downloaded free of charge at www.keller- druck.com)
Running the K-114_Config software
Run K-114_Config and select the appropriate COM port. Then go to the K-114 Configuration view (top right).
Tip: When the program is open, Comport is selected automatically as soon as the K-114 is plugged into the computer.

KELLER - Fig 1

Voltage and current measurements with K-114_Config

KELLER - Fig 2

No. Symbol Function Description
1 I-OUT Current supply — external consumer No valid display during

operation with external current supply
2| U-USB| USB voltage supply|
3| U-OUT| Voltage supply — external consumer| Switched-in voltage supply is indicated via U-OUT
4| U-IN| Voltage input| Range: 0…12 VDC

Warnings and other messages in the softwareKELLER - Fig
3

U-USB is below minimum
This message appears if the U-USB voltage is below 4,8 VDC. Fault-free functioning of the converter is no longer guaranteed. Supply the converter from an external power supply unit.
U-OUT outside of range
This message appears if U-OUT is less than 11,2 VDC. Supply the converter from an external power supply unit.
U-IN outside of measurement range
The U-IN measurement range is from 0…12 VDC. The exclamation mark warns you that the upper limit of the measuring range that can be displayed (> 16 VDC) has been reached.
The voltage for measurement that is actually applied (U-IN) can therefore be greater than the value for U-IN that is displayed.
Connect power supply unit
If the difference between U-OUT and U-IN is less than 3 VDC, the «Connect power supply unit» message appears.

Getting support

KELLER - Fig 4

If faults or errors occur while you are working with the converter, use the Help with fault analysis section to rectify the problem. If problems persist after you have worked through this section, you can request assistance from KELLER by telephone. For contacts, visit: www.keller-druck.com
The «Download remote maintenance software» function automatically launches the download of a Remote Desktop Program. Once the download has completed, run the program. After you enter the connection code, our technical support team will log into your computer and offer you on-the-spot help directly on your screen.
Recording K-114 measurement data
The values measured for I-OUT, U-IN, U-OUT and U-USB can be recorded and saved via Software ControlCenterSeries30 (CCS30). (Read the CCS30 manual on this subject.)
To view the K-114 controller’s measured values via the software, you must enter bus address 253 in the CCS30 and let the device search. Use the «New measurement» function key to view and save the measurement data.

KELLER - Fig 5

Declaration of Conformity
Herewith we declare, that the following product Converter K-114 meet the basic requirements, which are established in the guidelines of the European Community Directive EMC 2014/30/EU Directive RoHS 2011/65/EG under following conditions

  • Cable lengths of less than 3 m
  • For use inside buildings
    As criteria, the following norms for this converter are applied.

EN 61326-1:2013 EN 61000-6-2:2011 EN 61000-6-4:2011
This declaration is given for the manufacturer:
St. Gallerstrasse 119, CH-8404 Winterthur in full responsibility by
KELLER GmbH, Schwarzenegger 17, DE-79798 Jeannette
March

Managing Owen with legally effective signature

KELLER AG für Druckmesstechnik
CH-8404 Winterthur
** +41 52 235 25 25
info@keller-druck.com
KELLER Ges. für Druckmesstechnik mbH
DE-79798 Jestetten
**
+49 7745 9214 0
eurocenter@keller-druck.com
Version | Edition
12/2022
www.keller-druck.com

Documents / Resources

| KELLER K-114 Interface Converter [pdf] Owner's Manual
K-114 Interface Converter, K-114, K-114 Converter, Interface Converter, Converter
---|---

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals