KELLER HPX High Pressure Calibrator Instruction Manual
- June 9, 2024
- KELLER
Table of Contents
- KELLER HPX High Pressure Calibrator
- View
- Notes on the operating instructions
- Description of the device
- Scope of delivery
- Intended use
- General safety warnings
- Operating the HPX calibrator
- Reset
- Description of the functions
- Menu navigation for calibrators
- Commissioning
- Maintenance / disposal
- Disposal
- Software for calibrators
- Spare parts and accessories for KELLER pressure calibrators
- EU Declaration of Conformity
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
KELLER HPX High Pressure Calibrator
View
- Port for interface cable (K-104-A or K-114-A) (PC connection / RS485)
- Device base
- Display
- SELECT and ENTER buttons
- Oil chamber
- Recirculation pipe
- Screwed sealing plug
- Manual booster pump
- Pressure connection for test object, without overpressure valve (700 bar)
- Drain valve
- Screw compressor
Notes on the operating instructions
- The operating instructions are in-tended for specialist workers and trained personnel.
- Before each stage of work, read the relevant notes and warnings care-fully, and keep to the sequence as stated.
- Pay particular attention to the sec-tion on «General safety warnings».
If you have any problems or questions, please contact your supplier or consult KELLER directly.
Description of the device
General description
The high pressure calibrator enables pressure to be generated by means of the
integrated pressure pump, up to 700 bar relative. The measurement technology
incor-porated into this device allows ac-curate measurement and documen-tation
of the characteristic of a test object that is connected to it. The measured
pressure progression can be displayed, evaluated and saved with a computer
monitoring program (CCS30). The calibrator is operated with the two function
buttons SELECT and ENTER, located directly below the display. The calibrator
itself is powered by a 3,0 V battery, but power can also be supplied
externally via the K-114-A interface converter. Test ob-jects (transmitters or
pressure switch-es) must be supplied from an external source.
Pressure range for the display
High pressure calibrators are them-selves calibrated with the ambient air
pressure as the zero point reference. The Zero function (Set Zero) allows any
desired pressure value to be set as the new zero point reference. To reset the
pressure zero point to the factory setting, use the RES Zero function (reset
zero).
Commissioning
A pressure-resistant connection for the test object is required in order to
use the high pressure calibrator. The pressure connection for the test object
is already screwed to the pressure distributor of the high pressure calibra-
tor so that it is pressure resistant when it leaves the factory, and it must
not be dismantled. Recommended torque for the test object pressure connection:
30 Nm
IMPORTANT!
Nothing must adhere to the surface of the test object (no oil, grease, water,
etc). Impurities could pass through the adapter to reach the high pressure
calibrator and damage it.
Proof pressure
If the pressure exceeds the measuring range by more than 20 %, the measur-ing
cell or the mechanism of the high pressure calibrator may be destroyed.
Recalibration
The recalibration cycle depends on the conditions of use. Recommended
recalibration cycle: 1 year.
Scope of delivery
- 1 calibrator (including oil filling)
- 1 carrying case
- 1 test record (5 points)
- 1 test object adapter G 1/4”–G 1/8”
- 1 test object adapter G 1/4”–G 1/2”
- 1 Allen key
- 1 spare battery, type CR2430 (3,0 V)
- 1 set of operating instructions
- 1 USB interface converter, K-114-A
Intended use
The high pressure calibrator (HPX) may only be used to generate pres-sure with the type HLP 22 BP hydrau-lic oil that is supplied with the product. Use of the calibrator with other media will damage it. The operational safety of the device supplied is guaranteed only if it is used as intended. The limit values as stated (see page 19: «Tech-nical data») must never be exceeded. Before installing the high pressure calibrator, check that it is suitable for your applications.
General safety warnings
The current national regulations on accident prevention and workplace
safety must be followed whenever work is carried out. Internal regula-tions
issued by the operator must be followed, even if they are not mentioned in
these instructions.
Never use the high pressure calibra-tor together with an external pressure
source. Do not remove any connected compo-nents (e.g. test objects) when the
high pressure calibrator is under pressure. Open the screwed sealing plug
before removing parts. Do not use Teflon tape to seal the pressure connection.
Residues of Teflon tape could penetrate the high- pressure calibrator and
damage it. Only use the adapters and seals that are available as accessories.
Do not store the calibrator under pres-sure: only store the high pressure
cali-brator with the drain valve open. Avoid the action of force of any kind
on the high pressure calibrator and its operating controls. Do not use high
pressure calibrators if they are damaged or faulty.
Operating the HPX calibrator
Operating the high pressure calibrator is described starting on page 16.
Connect the test object
You can connect your test object to the high pressure calibrator via the pres-
sure connection (9).
Pressure generation
When using the calibrator, the screwed sealing plug (7) must be opened (2
turns), so that overpressure cannot build up in the oil reservoir. Use the
manual booster pump (8) to set the pressure to about 10 bar. You can use the
screw compressor (11) to increase or reduce the pressure.
Release pressure
- Open the screw compressor (11) completely
- Open the drain valve (10)
IMPORTANT! Do not open if there is high pressure in the system
If you can no longer reach the desired pressure, please consult the section on “Maintenance” to find out how to vent the system.
Zeroing the device
Open the drain valve (10) to release any pressure that may have built up. If
the pressure display does not show zero, perform a zeroing procedure (Set
Zero) and then close the drain valve.
Information about the display
If no pressure can be shown on the display, it will show OFL (overflow) or UFL
(underflow). If pressure outside the device’s measuring range is applied, the
last valid pressure value that was measured will flash on the display
(overload warning).
Reset
Description of the functions
Menu navigation
If the selected function or unit is not ac-tivated by pressing the ENTER
button within 5 seconds, the display will return to measuring mode without
changing a setting.
SELECT button
The SELECT button positioned on the front is used to switch the device on, to select a function and to select the various pressure units.
ENTER button
The ENTER button positioned on the front is used to activate the se-lected
function or pressure unit on the device. You can also press the ENTER button
to switch between the minimum and maximum pressure values measured thus far.
Menu navigation for calibrators
Commissioning
Switch the device on
Press the SELECT button to switch the device on. Initially, the device shows
the pressure range calibrated in the factory (top) and the software version
(year / week).
Switch the device off
- Keep the SELECT button pressed down until the display shows OFF.
- Press the ENTER button to execute the shutdown.
- The settings made previously are retained when you switch the device on and off.
Display mode
Display mode is the calibrator’s basic mode. The upper part of the display
shows the pressure unit and the pres-sure that is currently measured. The
lower part of the display shows the last function that was used, either the
min./ max. display or the Leak function.
Using the functions
Written descriptions of the individual functions are given below (in addition
to the diagram above).
Selecting functions
The individual sub-functions are called up from the MANO menu. Keep the SELECT
button pressed until MANO is shown, and press ENTER to ac-tivate. You can now
use SELECT to choose the function you want, and ENTER to execute the function.
De-pending on the current setting, the first function to be shown is either
min/max disp or LEAK disp.
Leak measurement function
Leak mode is used to determine the pressure change over a defined period,
which can be adjusted. The unit to be tested must be connected to the high
pressure calibrator on the pressure side.
Start leak measurement
Activate the MANO menu. The dis-play shows Leak dISP. Press the ENTER button
and then the SELECT button. Press ENTER to confirm Leak Start. The leak
measurement starts, and the display alternates be-tween the current leak time
and the pressure change measured thus far.
Active leak measurement
During leak measurement, the lower part of the display alternates each sec-ond
between the measurement time that has now elapsed [mm:ss] and the pressure
change measured thus far.
End leak measurement early
To end a leak measurement early, press the ENTER button and confirm the «Leak
Stop» display by pressing ENTER.
Leak measurement completed
If the leak measurement time has elapsed or if the measurement was manually
ended ahead of time, the display alternates between the elapsed leak
measurement time and the measured pressure change.
Set leak measurement time
The leak measurement time is preset to 10 minutes in the factory, and it can
only be changed with the «Mano Con-fig» software.
MANO / «Continuous» function
Automatic switch-off function (the de-vice switches off automatically 15 mi-
nutes after a button was last pressed). Leak measurements are canceled by the
automatic switch-off function if the measurement time is more than the switch-
off time.
- Cont on: Disables the automatic switch-off function
- Cont oFF: Enables the automatic switch-off function
If the «Continuous» function is en-abled, Cont flashes on the display.
Maintenance / disposal
Venting the pressure system
Release the pressure completely and then open the drain valve (10) and the
screwed sealing plug (7). Screw the screw compressor (11) in completely. Pump
steadily with the manual boos-ter pump (8) to clear the system of air. When no
more bubbles come out of the recirculation pipe (6), close the drain valve
(10).
Changing the oil
We recommend that you have KELLER change the oil. The entire system is cleaned
at the same time. Only use type HLP 22 BP hydraulic oil.
Battery
The pressure calibrator is powered by a 3 V button-cell battery (behind the
display). If the battery is low, the battery symbol on the display lights up.
Replacing the battery
Please switch the device off. Turn the display section ring beyond the limit
stop until it is released from the hous-ing section (turn through about 180°).
Open the battery compartment and change the battery (type CR 2430).
Disposal
This product must not be disposed of as normal household waste at the end of its useful lifetime. To prevent pos-sible damage to the environment or to health due to uncontrolled waste disposal, this product must be separated from other waste and recycled correctly in order to ensure sus-tainable use of the raw materials.
Software for calibrators
The USB interface converter (K-114-A) enables communication between the calibrator and a computer. Before you connect the interface converter to the computer, install driver K-104 / K-114 (can be downloaded free of charge at www.keller-druck.com)
Settings on the high pressure calibrator with the ManoConfig software
Device settings such as the leak measurement time or the switch-off time for
the pressure calibrator can be adjusted using the «ManoConfig» software.
Record measurements with the CCS30 software
The CCS30 software records the data measured by the pressure calibra-tor, and
shows them in both graphic and tabular form. Measured data can be saved or
exported for further pro-cessing. You will find more informa-tion about the
software in the CCS30 manual.
Step-by-step software installation
Install from www.keller-druck.com:
- K-104 / K-114 driver
- (CCS30) Control Center Series 30 3.) ManoConfig (if desired)
Spare parts and accessories for KELLER pressure calibrators
EU Declaration of Conformity
Herewith we declare, that the follow-ing product or product range
High pressure calibrator HPX
meet the basic requirements, which are established in the guidelines of the European Community:
Directive EMC 2014/30/EU Directive RoHS 2011/65/EU and Commission Delegated Directive (EU) 2015/863 As criteria, the following norms for this high pressure calibrator HPX are applied:
- EN 61000-6-1:2007
- EN 61000-6-2:2005
- EN 61000-6-3:2011
- EN 61000-6-4:2011
- EN 61326-1:2013
- EN 61326-2-3:2013
This declaration is given for the manu-facturer:
KELLER Druckmesstechnik AG, St. Gallerstrasse 119, CH-8404 Winterthur in full responsibility by KELLER Gesellschaft für Druckmesstechnik mbH, Schwarzwaldstrasse 17, DE-79798 Jestetten
KELLER Druckmesstechnik AG
St. Gallerstrasse 119 • CH-8404 Winterthur Tel. +41 52 235 25 25 • Fax +41 52
235 25 00 www.keller-druck.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>