KELLER ECO2 Digital Manometer
ECO2
Digital Manometer with Min.-/Max.-Display.
Description
Digital manometer with Min.-/Max.- pressure indication.
The technical data of the digital manometer can be taken from the
corresponding data sheet or from the agreed specifications.
Turn-On and Functions
ECO2 has two operating keys. The left key (SELECT) serves to select the
functions and the pressure units.
The right key (ENTER) activates the selected function or pressure unit. The
right key is also used to switch between the Min.- and Max.- pressure value.
Turn-on:
Pressing the SELECT key turns the instrument on.
The instrument first displays the full-scale pressure range (top display) and
the software version (year/week).
The instrument is then ready for use and indicates the actual pressure (top
display) and the last measured Max. pressure value (bottom display).
The instrument has the following functions:
RESET:
Min.-/Max.-value are set to the actual pressure.
OFF:
Turns off the instrument.
MANO:
Releases the following functions:
ZERO SET:
Sets a new pressure zero reference.
ZERO RES:
Sets the pressure zero to factory setting.
CONT on:
Deactivates the automatic turn-off function.
or
CONT off:
Activates the automatic turn-off function (the instrument turns off 15 minutes
after the last key operation),
…followed by the unit selection: bar, mbar, hPa, kPa, MPa, PSI, kp/cm2
Example: Setting a new pressure unit (mbar):
- Turn on the instrument by shortly pressing SELECT.
- Wait for the instrument’s measuring mode (≈ 3 s).
- Press the SELECT-key 3 times: MANO appears.
- Press ENTER: ZERO SEt appears.
- Press SELECT: ZERO rES appears.
- Press SELECT: CONT on or CONT off appears.
- Press SELECT: bar appears.
- Press SELECT: mbar appears.
- Press ENTER: The new pressure unit (mbar) is set. The instrument returns to the measuring mode.
Display of the Minimum Value
When in the measuring mode (Display: Actual Pressure and Max. pressure value), you may display the Min. pressure value for 5 seconds by shortly pressing the ENTER-key.
Notes
- The functions and units can also be called up by keeping the SELECT-key depressed. Releasing the key enables the displayed function or unit to be activated with the ENTER-key.
- If the selected function or unit is not activated within 5 seconds with the ENTER-key, ECO2 returns to the measuring mode without changing any settings.
- Turning ECO2 on and off does not influence any of the previous settings.
- If the CONT on function is activated, it is indicated with a flashing sign on the display (OFF flashes when CONT on is set).
- If a pressure can not be represented on the display, OFL (overflow) or UFL (underflow) appears on the display.
- If the actual pressure goes beyond the measuring range, the last valid pressure value starts flashing on the display (overload warning).
- Temperatures outside of 0…50 °C could impair the readability of the display.
Installation
The installation must be carried out by qualified personnel only.
ECO2 has 7/16-20 UNF male thread and is delivered complete with an O-ring seal
swivel fitting which allows ECO2 to be rotated 360°.
Only minimal torque is need-ed to seal up to 700 bar. Process pressure
connection is G1/4 male.
The swivel adapter can be tightened with a max. torque of 50 Nm.
Battery Change / Battery Life
When the battery starts weakening, a low battery warning (BAT LOW) will appear in the display.
Battery change:
Open the battery compartment and change the battery (type CR2430).
When reassembling, make sure that the O-ring remains imbedded in the cover.
The battery life is approximately 1000 hours at continuous operation.
Please note: This manometer is equipped with a battery (Type CR2430)
installed.
Please use a coin for opening the battery box to prevent da-mage to the
battery cover.
Dispose of discharged batteries properly, where they are to be picked up by a
qualified waste management company.
Place replacement battery between the contact springs, paying attention to the
polarity (positive pole facing up).
Close the cover plate by hand, if possible.
Ranges / Calibration
The ZERO-function allows to set any pressure value as a zero reference.
The factory setting of the pres-sure zero for the ranges ≤ 61 bar absolute is
at vacuum (0 bar absolute).
For relative pressure measurements, activate «ZERO SEt» at ambient pressure.
Instruments with ranges over 200 bar are calibrated at 1 bar abs as a zero
reference.
General Safety
Instructions
When installing and operating the digital manometer, attention should be paid to the corresponding safety regulations.
Only mount the digital mano-meter onto unpressurized sys-tems.
On pressure ranges ≥ 61 bar, the pressure connections could show residual
hydraulic oil.
Please also note the corresponding data sheet.
Accessories, Spare Parts
- Battery Renata CR2430, Lithium 3,0 V / Order Number /557005.0001
- Protective rubber covering/ Order Number /309030.0001
- Carrying bag / Order Number / 309030.0003
EU / UK DECLARA TION OF CONFORMITY
Herewith we declare, that the following products
Digital Manometer ECO2 comply with the requirements of the following EU/UK
Directives:
Directive EMC 2014/30/EU
Directive RoHS 2011/65/EU and Commis-sion Delegated Directive (EU) 2015/863
UKSI 2016:1091
UKSI 2012:3032
The Digital Manometer ECO2 comply with the following standards:
- EN IEC 61000-6-1:2019
- EN IEC 61000-6-2:2019
- EN IEC 61000-6-3:2021
- EN IEC 61000-6-4:2019
- EN 61326-1:2013
- EN 61326-2-3:2013
This declaration is given for the manufacturer:
KELLER Druckmesstechnik AG, St. Gallerstrasse 119, CH-8404 Winterthur
issued by:
KELLER Gesellschaft für Druckmesstechnik mbH, Schwarzwaldstrasse 17, DE-79798
Jestetten
Jestetten, 14.09.2022
Bernhard Vetterli
Leiter Entwicklung | Technical Director Responsable développement
Matthias Schlimper
Qualitäts Manager | Quality Manager Responsable qualité
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>