Co-Dx MPX2-R-001 Logix Smart Mpox (2-Gene) Instructions
- June 9, 2024
- Co-Dx
Table of Contents
- MANUFACTURER
- INTENDED USE
- PRODUCT DESCRIPTION
- RUO COMPONENTS
- COLLECTION, HANDLING, STORAGE, AND SHIPPING
- MATERIALS REQUIRED (NOT INCLUDED)
- BACKGROUND INFORMATION
- ACCESSSORIES (NOT INCLUDED)
- WARNINGS AND PRECAUTIONS
- SAMPLE INFORMATION
- PROCEDURE
- DATA ANALYSIS
- TROUBLESHOOTING
- REFERENCES
- LEGEND OF PACKAGE SYMBOLS
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
**Co-Dx MPX2-R-001 Logix Smart Mpox (2-Gene)
**
MANUFACTURER
Co-Diagnostics, Inc. 2401 S
Foothill Dr. Ste D Salt Lake City, UT 84109 Phone: +1
801-438-1036
Email: info@co-dx.com
Website: www.co-dx.com
INTENDED USE
The Co-Dx Logix Smart™ Mpox (2-Gene) RUO is for the simultaneous qualitative
detection of Mpox (moneybox) and other related Ortho poxviruses.
For research use only. Not for use in diagnostics procedures.
PRODUCT DESCRIPTION
The Co-Dx Logix Smart™ Mpox (2-Gene) RUO is a research-use-only multiplex
test, based on real-time polymerase chain reaction technology. It tests for
the presence or absence of deoxyribonucleic acid (DNA) of the Mpox virus and
other related Orth poxviruses.
The Co-Dx Logix Smart™ Mpox (2-Gene) RUO includes an internal control to
identify possible quantitative polymerase chain reaction (qPCR) inhibition,
confirm the integrity of the reagents, and verify the quality of the sample
extraction. The Co-Dx Logix Smart™ Mpox (2-Gene) RUO also includes a positive
control (PC) which includes synthetic DNA molecules carrying sequences that
are homologous to Mpox viruses targeted by this multiplex assay. PCs represent
a source of cross contamination. Take precautions to prevent and minimize the
risk of cross contamination.
Co Primers™ in the Co-Dx Logix Smart™ Mpox (2-Gene) RUO include the following:
- Co Primers™ that target Mpox (MPX) F8L are labelled with the Quasar® 670 fluorophore.
- Co Primers™ that target Mpox (MPX) L6R are labelled with the CAL Fluor® Orange 560 fluorophore.
- Co Primers™ that target the Human R Nase PDNA, which acts as the Internal Positive Control (IPC), are labelled with CAL Fluor® Red 610 fluorophore.
Cross Reactivity
The Co-Dx Logix Smart™ Mpox (2-Gene) RUO is expected to cross-react with some Orth poxviruses, based on In Silico analysis or wet testing. That are expected to be detected by In Silico analysis include the following:
-
Mpox
-
Variola
-
Vaccinia*
-
Cowpox
-
Camelpox
-
Akhmeta virus
-
Buffalopox
-
Alaskapox
-
The vaccinia virus is present in some forms of the smallpox vaccine. There is a chance that individuals who were recently vaccinated for smallpox could show a positive result.
RUO COMPONENTS
See Table 1 for components included in the Co-Dx Logix Smart™ Mpox (2-Gene) RUO.
Table 1
RUO Component Information
Lid Color| Component| Symbol| Catalog Number|
Description| Amount
---|---|---|---|---|---
Brown| Co-Dx Logix Smart™ Mpox Master Mix| MM| MPX2-MM-001| Proprietary
blend of CoPrimers™ and
PCR reagents
| 1×1000 µL (100 reactions)
| Co-Dx Logix Smart™ Mpox Positive Control| PC| MPX2-PC-001| Proprietary blend
of target templates| 1×1000 µL (100 reactions)
__
__
| Co-Dx Logix Smart™ Mpox No Template Control| NC| MPX2-NC-001| DNase/RNase- free water| 1×1000 µL (100 reactions)
The product code is MPX2-R-001. Contact Sales at 801-438-1036 ext. 01 or go to www.co- dx.com/contact/ to order.
COLLECTION, HANDLING, STORAGE, AND SHIPPING
See the following information for reagent storage and handling:
- The Co-Dx Logix Smart™ Mpox (2-Gene) RUO is shipped on dry ice and should arrive frozen. Contact your distributor if one or more of the components are not frozen upon arrival or are compromised during shipment.
- Store all components immediately upon arrival at a temperature between -40°C and -16°C to prevent degradation of reagents.
- Do not use expired products. Integrity of expired components cannot be guaranteed.
- Follow internal laboratory procedures for quality control when using this product.
- Protect the master mix (MM) from light.
- If you will be using reagents intermittently, freeze the reagents in multiple aliquots to ensure that less than 10 freeze/thaw cycles occur. Excessive thawing and freezing of components (specifically the MM) could affect the performance of the assay.
- Avoid storing components at temperatures between +2°C and +8°C for more than 4 hours.
- If you work in an area prone to power outages, keep a back-up generator for your freezer as well as a temperature data log system to ensure that the product remains frozen at a temperature between -40°C and -16°C.
- Dispose of the product in accordance with applicable regional, national, and local laws and regulations. This product is not biological waste. The Safety Data Sheet (SDS) for this product can be viewed from Co-Diagnostics website at Safety Data Sheets |Co-Diagnostics, Inc. (co-dx.com).
- Always follow the most recently recommended product stability data as it becomes available. This data can be found in our latest version of the Instructions for Use at Instructions For Use – Co-Diagnostics, Inc. (co-dx.com).
MATERIALS REQUIRED (NOT INCLUDED)
The following materials and devices are required but are not provided with this RUO:
- Appropriate 4-channel real-time PCR instrument, compatible with the fluorophores used in this test.
- Appropriate nucleic acid extraction system or kit
- Vortex mixer
- Centrifuge with a rotor for 2 mL reaction tubes
- Adjustable pipettes
- Disposable pipette tips with filters
- Disposable powder-free gloves
- Ice
- Biosafety cabinet, ideally BSL-2 facility
| WARNING!
Install, calibrate, and maintain all instruments properly according to the
manufacturer’s instructions and recommendations. Do not use instruments with
outdated
---|---
BACKGROUND INFORMATION
Mpox is a viral zoonosis similar to smallpox, which is also part of the
Orthopoxviral genus. Mpox primarily occurs in central and west Africa, often
in proximity to tropical rain forests, but also near urban areas.
The Mpox virus is an enveloped double-stranded DNA virus with a variety of
mammalian hosts, including rope squirrels, Gambian pouched rats, dormice, non-
human primates, among others. (Reference https://www.who.int/news-room/fact-
sheets/detail/Mpox)
Mpox virus is transmitted by close contact with lesions, body fluids,
respiratory droplets, and contaminated materials such as bedding.
The incubation period of Mpox is usually from 6-13 days but can range from
5-21 days. The infection can be divided into two periods: (Reference
https://www.who.int/newsroom/fact
sheets/detail/Monkeypox.
The invasion period (lasts between 0-5 days) characterized by fever, intense
headache, lymphadenopathy (swelling of the lymph nodes), back pain, myalgia
(muscle aches) and intense asthenia (lack of energy).
Lymphadenopathy is a distinctive feature of Mpox compared to other diseases
that may initially appear similar (chickenpox, measles, smallpox) (Reference
https://www.who.int/news-room/fact-
sheets/detail/monkeypox). The skin eruption usually begins within 1-3 days of
appearance of fever. The rash tends to be more concentrated on the face and
extremities rather than on the trunk. It affects the face (in 95% of cases),
and palms of the hands and soles of the feet (in 75% of cases). Also affected
are oral mucous membranes (in 70% of cases), genitalia (30%), and conjunctivae
(20%), as well as the cornea. The rash evolves sequentially from macules
(lesions with a flat base) to papules (slightly raised firm lesions), vesicles
(lesions filled with clear fluid), pustules (lesions filled with yellowish
fluid), and crusts which dry up and fall off. The number of lesions varies
from a few to several thousand. In severe cases, lesions can coalesce until
large sections of skin slough off (Reference https://www.who.int/news-room
/fact-sheets/detail/Monkeypox).
In May of 2022, cases of Mpox were reported to WHO from 12 member states
across three WHO regions, where Mpox virus is not endemic. These areas include
the United States, Canada, Australia, United Kingdom, Spain, and Portugal.
(Reference https://www.who.int/emergencies/disease-outbreak-
news/item/2022-DON385) -This reference is different than the above links.
ACCESSSORIES (NOT INCLUDED)
Thermocycler
Thermocyclers validated but not included with the test are displayed in Table 2.
Table 2
Thermocyclers Validated but Not Included with the Test
Thermocycler Machine | Catalog Number | Manufacturer |
---|---|---|
CoDx Box | MIC-4 | Co-Diagnostics, Inc. |
Mic qPCR Cycler | MIC-4 | BMS, Bio Molecular Systems |
Extraction System
Extraction System required but not included with the test are displayed in Table 3.
Table 3
Extraction and Automation Systems Validated with the Test
Extraction Reagent| Automation Platform (If applicable)|
Manufacturer| Sample Input Volume/Sample Elution Volume by Sample
Type
---|---|---|---
Name| Cat. Number
QIAamp DNA Mini Kit (Qiagen)| 51304 (50 extractions) __ 51306 (250
extractions)| N/A| Qiagen| Protocol Used| Sample Type| Input|
Output
Buccal Swab| dry lesion swab| 400 µL| 60 µL x 2
WARNINGS AND PRECAUTIONS
| WARNING!
Before performing any testing or running any sample, verify that all
instruments have been properly installed, calibrated, and are well maintained.
Do not use instruments with an outdated calibration.
---|---
Read this Instructions for Use document carefully before using the product. Before first use, check the components for the following:
- Integrity
- Frozenness upon arrival
- Users should ensure the following:
- Limit use of this product to personnel instructed and trained in the techniques of real-time PCR.
- Always treat samples as infectious and/or biohazardous. Use standard precautions.
- Wear protective gloves, lab coat, and eye protection when handling samples and always wear gloves when handling RUO components.
- Always use DNase/RNase-free disposable pipette tips with filters.
- Use segregated working areas for sample preparation, reaction setup, and amplification/detection activities. The workflow in the laboratory should proceed in a unidirectional workflow. To prevent contamination, change personal protective equipment (PPE) between areas.
- Store and extract positive materials (specimen, controls, and amplicons) separately from other reagents. Dedicate supplies and equipment to separate working areas and do not move them from one area to another.
- Consult the appropriate SDS for safety. The SDS for the Co-Dx Logix Smart™ Mpox (2-Gene) RUO is provided with the shipment. If not provided with the shipment, the SDS can be retrieved from Co-Diagnostics website at the following link: Safety Data Sheets | Co-Diagnostics, Inc. (co-dx.com)
- Do not open the reaction tubes/plates post amplification.
- Do not autoclave reaction tubes/plates after the PCR, since this will not degrade the amplified nucleic acid and will pose a risk to the laboratory area to contamination.
- Do not use components that have passed expiration date.
- Discard sample and assay waste according to your local safety regulations
SAMPLE INFORMATION
Sample Storage
Ensure the following when storing samples:
- Process all specimen types within 48 hours after collection, if storage is needed after 48 hours, store the samples frozen, preferably at -70°C (CDC, 2020).
- Avoid repeated freezing and thawing of any specimen. If you need to keep a specimen for retesting, aliquot the specimen in different tubes to avoid freezing and thawing cycles.
- Monitor the temperature in the storage areas and recorded temperatures regularly to identify potential fluctuations.
- Do not use domestic refrigerators/freezers with wide temperature fluctuations. Domestic refrigerators/freezers are not suitable for the storage of frozen specimens (CDC, 2020).
Sample Handling
Laboratory workers should wear appropriate PPE, which includes disposable
gloves, laboratory coat/gown, and eye protection when handling potentially
infectious specimens.
Conduct samples suspected to be or confirmed to be infected with MP ox under a
certified Class II biosafety cabinet in a BSL-2 containment facility. More
details are Provided in the Biosafety in Microbiological and Biomedical
Laboratories (BMBL) (CDC, 2009) or the WHO Laboratory Biosafety Manual (WHO,
2004).
For specific instructions on handling clinical specimens for the MP ox virus,
see also the CDC’s webpage for the Preparation and Collection of Specimens –
MP ox (Guidelines for Collecting and Handling Specimens for Mpox Testing | MP
ox | Poxvirus | CDC) (CDC, 2022) and Interim Guidance for Collection
Diagnostic Specimens from Persons with Suspect Moneybox (CDC, 2003).
The quality of the extraction of the DNA from the samples is essential for the
performance of Co-Dx Logix Smart™ MP ox (2-Gene) RUO. Perform extraction by
following manufacturer’s instructions or an internally validated protocol. The
suitability of the nucleic acid extraction procedure for use with Co-Dx Logix
Smart™ Mpox (2-Gene) RUO must be validated by the user.
| WARNING!
If your sample preparation system uses washing buffers
containing ethanol, make sure to eliminate any traces of ethanol
prior to elution of the nucleic acid. Ethanol is a strong inhibitor of
real-time PCR.
The use of carrier RNA is crucial for extraction efficiency and stability of
the extracted nucleic acid.
---|---
PROCEDURE
Real Time RT-PCR Setup
Set Up the Reagents
Perform the steps below to set up the reagents.
Clean all working surfaces with a fresh 10% bleach solution followed by a
molecular-grade alcohol or another equivalent method of cleaning that
disinfects and degrades nucleic acids.
Thaw all reagents and samples on ice, or a cold block, before starting the
setup.
Vortex all Co-Dx Logix Smart™ Mpox (2-Gene) RUO MM, PC, nuclease-free water
(used as an NC), and all sample tubes for 3 seconds before using.
Briefly spin the MM, PC, NC down before using to ensure reagents are properly
mixed and to ensure removal of any condensation or residue from the lids.
Set Up the Reaction
Perform the steps below to set up the reaction.
Collect enough reaction wells for each of the following:
- One for each NC,
- One for each sample you want to test, and
- One (or more) for each PC
Note: The example below displays the minimum number of wells needed for 5
samples.
PC 1
NC 1
Samples 5
Total wells required 7
Pipet 10 µL of MM into each well collected.
Pipet 10 µL of the NC into the appropriate wells (in addition to the 10 µL of
MM already in the well).
Note: Ensure that at least one NC is included in each run and that enough space remains for at least one PC.
Important:
- Pipette on ice, if possible.
- Perform PC pipetting and sample setup in a separate area, or at a separate time from the MM and NC.
- Change pipette tips between samples and change pipette tips after pipetting each component.
- Pipet the PC last, if possible, to avoid contamination events.
Pipet 10 µL of sample or PC into the appropriate well.
Seal the reaction plate with an optical adhesive film or seal each reaction
tube with its appropriate lid
Place the plate or tubes into the RT-PCR instrument in the correct orientation
and start the run.
PCR Instrument and Thermocycler Setup
For programming instructions questions regarding the use of other realtime PCR instruments, contact the Laboratory 801-438-1036 ext. 03 or at www.co- dx.com/contact/.
If using Co-Diagnostics Inc. Co Dx Box, contact the Laboratory (see contact
information in Section 7.4.2) for the template file for download. The template
file comes pre-programmed with the PCR instrument setup described in this
section. When not using a template, or using another device, use the settings
outlined below to program the PCR To achieve optimal performance from the
test, it is important to make sure that the instrument is compatible with the
conditions outlined below. instrument.
Define the settings as displayed in Table 4.
Table 4
Recommended Instrument Settings
Item | Setting |
---|---|
Reaction Volume | 20 µL |
Ramp Rate | Default |
Passive Reference | None |
Program PCR instrument with the cycling conditions displayed in Table 5.
Table 5
Recommended Cycling Condition Settings
Item | Stage | Cycles | Temperature | Time |
---|---|---|---|---|
Initial Denaturation | Hold | 1 | 95˚C | 2 minutes |
Amplification | Cycling | 45 | 95˚C | 3 seconds |
53˚C | 32 seconds |
Ensure that the PCR instrument being used is compatible with the fluorophores below. Some devices may not have options for the quencher. If you need help or have questions, contact Co-Diagnostics Inc. Technical Support at 801-438-1036 ext. 02 or at: support@codx.com.
Define the fluorescence detectors (dyes) as displayed in Table 6.
Table 6
Fluorescence Detector Definitions
Target | Detector Name | Reporter | Quencher |
---|---|---|---|
MPX-F8L (T4) | MPX-F8L (T4) | Quasar® 670 | BHQ®-2 |
MPX-L6R (T3) | MPX-L6R (T3) | CAL Fluor® Orange 560 | BHQ®-1 |
RNaseP (IPC) | RNaseP (IPC) | CAL Fluor® Red 610 | BHQ®-2 |
When the run is finished, ensure that the run file is saved.
DATA ANALYSIS
For basic information regarding data analysis on specific qPCR instruments, please refer to the user manual of the respective instrument.
Positive Controls
Validate the test run by checking to see that the PC has passed and that the control conditions displayed in Table 7 are met.
No-Template Controls
Validate the test run by checking to see that no-template control has passed and that the control conditions displayed in Table 7 are met.
Table 7
Required Control Conditions
Control Type| Control Name| Purpose of Control| MPX L6R
Result| MPX F8L Result| RNaseP (IPC) Result
---|---|---|---|---|---
MPX Positive Control (PC)| MPX-L6R (CF®560)| Verifies the performance of the
master mix| +| +| +
MPX-F8L (Quasar® 670)
RNaseP (IPC) (CF®610)
No Template Control (NC)| Master Mix + Water| Verifies the reagents are free
of contamination| –| –| –
If controls pass, interpret the sample results.
Invalid Test Run
If any of the controls fail, the run is invalid.
Document the run and initiate troubleshooting.
Interpretation of Results
Once the controls have passed, the unknown samples can be interpreted based on one of the following three possible outcomes:
- Positive
- Negative
- Invalid
A Positive result will show an amplification curve or cycle threshold value
for Mpox. The cut-off value should be determined by in house validation
testing. However, internal studies have shown rare primer-dimer formation or
other non-specific amplification at 45 cycles. This fact can be attributed to
the nature of the CoPrimers™ (Satterfield, 2014), (Poritz & Ririe, 2014). The
amplification of the RNaseP (IPC) shows that the extraction was successful.
A Negative result shows no amplification for Mpox; The absence of a curve for
Mpox indicates a negative result ONLY when the RNaseP (IPC) marker is
positive.
An Inconclusive result occurs if any of the controls fail. See the
Troubleshooting section.
See Table 8 for results translation
Table 8
Results Translation
| Sample Result| Co-Dx Logix Smart™ Mpox Positive Control| No
Template Control (NC) (Master Mix + Water)| Interpretation Results
---|---|---|---|---
MPX-L6R CF560 Channel| MPX-F8L Q670 Channel| Internal
Positive Control (RNaseP) CF610 Channel
Instrument Reading| +| +| +| +| –| *Mpox
Virus DNA +
+| –| +| +| –*| Mpox Virus DNA +
–| +| +| +| –**| *Mpox Virus DNA +
–| –| +| +| –| Mpox Virus DNA –
Any Result (+/-)| –| +| –| INVALID: See
Troubleshooting
+| –| –
+| +| +
Anything before 42 cycles is considered a positive analyte reading (+).
Anything at or after 42 cycles is considered a negative analyte reading (-).
*A positive result indicates a sample contains either MP ox or other orthopoxviral DNA.
Note: Samples collected from individuals who were recently vaccinated for smallpox may show a positive result because of cross-reactivity with Orth poxvirus vaccinia virus, which is present in some forms of the smallpox vaccine.
TROUBLESHOOTING
Co-Diagnostics Inc. values customer feedback and we would like to be informed of any issues with the Co-Dx Logix Smart™ Mpox (2-Gene) RUO even if the recommended steps for troubleshooting solve the issue. To give feedback please fill out the Customer Feedback form by visiting www.co- dx.com/contact/feedback/.
Stability
Real-time and accelerated shelf-life and in-use stability studies are
currently under testing. Currently, the expiration date of this product has
been established as 12 months. Do not use expired reagents, because doing so
may lead to inaccurate results.
Always use the most recent version of this document for updates as more
stability information will be added when studies are completed.
User Errors
Good Laboratory Practices for Molecular Biology Diagnostics (Viana & Wallis,
2011) are necessary for the use of this product. This product is not intended
to be used by untrained personnel.
To help prevent errors such as splashes, crossover contamination, and volume
selection, it is essential that users have some molecular biology experience
and be familiar with proper pipetting technique. Pipette tips must be replaced
after every pipetting. Gloves must be replaced often. Equipment, such as
pipettes and realtime PCR instruments, should be calibrated when applicable.
A 90-minute online training for Good Laboratory Practices for Molecular
Genetics Testing (Centers for Disease Control and Prevention, 2017) is
available at the CDC website at the following link
https://www.cdc.gov/labtraining/training-courses/goodlab-practices-molecular
genetics-testing.html
Invalid Results
Co-Dx Logix Smart™ Mpox PC not Amplifying
No amplification from the PC could be the result of one or multiple factors, such as one or more of the following:
- Pipetting errors (pipetting control into the wrong well, missing a well, pipetting inadequate amount of reagent)
- Incorrect placement of plates or tubes into the real-time PCR instrument
- Co-Dx Logix Smart™ Mpox MM or PC degradation (result of reagents being stored at a temperature above -16°C for an extended period)
- Use of expired reagents
- Wrong reagents being used
When this occurs, it is best to disregard the results from the samples and re- test by re-amplification. If the PC fails again, then an investigation should be conducted to identify possible causes for error, and the test must be reprocessed from extraction (or not, depending on the investigation results and risks identified in the process). If failure of the PC occurs a third time after re-extraction and re-amplification, open a new Co-Dx Logix Smart™ MPX PC or MM, and retest. If still failing, please contact Co-Diagnostics Inc. Technical Support by calling 801-438-1036 ext. 02 visiting support @co-dx.com.
PC RNaseP is not Amplifying in Samples
No amplification from the RNaseP channel could be the result of one or multiple factors, such as:
- Not enough nuclear material in the sample
- PCR inhibitors such as ethanol and heparin
- Incorrect extraction
- Extraction system used is not compatible or has a step that eliminates RNaseP DNA
Note: Positive amplification in the Mpox channel indicates a positive analyte result despite the lack of concurrent amplification in the IPC channel. The IPC amplification is dependent on the presence of human genomic DNA (gDNA) in the extraction sample, the amount of which is governed by the type of the sample and the extraction procedure used. Samples obtained from culture or sterile/pure sites (e.g., CSF, urine, or cell lysates) may not contain the human RNaseP gene.
When this occurs, the results should be interpreted as invalid and retesting by re-amplification should be performed. If the IPC fails again, then samples should be re-extracted and re-amplified. If it fails a third time an investigation should be conducted to identify possible causes for the error. If the cause for the error is clear, the test can either be singled out as invalid due to either PCR inhibitors being present or not enough nuclear material being present. If the cause for an error is unclear, contact Co- Diagnostics Inc. Technical Support by calling 801-438-1036 ext. 02 or contact us at support@co-dx.com.
No Template Control (NC) Showing Amplification
Amplification of MPX in a No Template Control indicates contamination in one
or more of the reagents, incorrect placement of plate or tube into the real-
time PCR instrument, or pipetting errors.
When this occurs, none of the results can be trusted and re-testing by re-
amplification should be performed. If the NC fails again, then an
investigation should be conducted to identify possible causes for error, and
the test must be reprocessed from extraction or not, depending on the
investigation results and risks identified in the process. If failure of the
NC, after re-extraction and re-amplification, happens a third time, open a new
nuclease-free water and retest. If still failing, please contact Co-
Diagnostics Inc. Technical Support by calling
801-438-1036 ext. 02 or by visiting
support@co-dx.com.
REFERENCES
CDC. (2009). Biosafety in Microbiological and Biomedical Laboratories
(BMBL) 5th Edition. Retrieved from CDC Laboratories:
https://www.cdc.gov/labs/BMBL.html
CDC. (2017, Oct 27). CDC Laboratory Training: Good Laboratory Practices
for Molecular Genetics Testing. Retrieved Mar 5, 2019, from Centers for
Disease Control and Prevention: https://www.cdc.gov/labtraining/training-
courses/good-lab-practicesmolecular-genetics-testing.html
Poritz, M., & Ririe, K. (2014, Mar). Getting things backwards to prevent
primer dimers. Journal of Molecular Diagnosis, 159-62.
doi:10.1016/j.jmoldx.2014.01.001
Satterfield, B. (2014, Mar). Cooperative primers: 2.5 million-fold
improvement in the reduction of nonspecific amplification. Journal of
Molecular Diagnosis, 163-73. doi:10.1016/j.jmoldx.2013.10.004
Viana, R. V., & Wallis, C. L. (2011). Good Clinical Laboratory Practices
(GLCP) for Molecular Based Tests Used in Diagnostic Laboratories. In D. I.
Akyar, Wide Spectra of Quality Control (pp. 29-52). InTech. Retrieved from
Good Clinical Laboratory Practice (GCLP) for Molecular Based Tests Used in
Diagnostic Laboratories |
IntechOpen
CDC. (2003, Jun 23). Interim Guidance for Collection of Diagnostic
Specimens from Persons with Suspect Monkeypox. Retrieved June 10, 2022, from
Centers for Disease Control and Prevention: Interim Guidance for Collection
of Diagnostic Specimens from Persons with Suspect Monkeypox
(aphl.org)
WHO. (2004). Laboratory Biosafety Manual. Retrieved from Emergencies
preparedness, response:
https://www.who.int/csr/resources/publications/biosafety/WHO_CDS_CSR_LYO_200
4_11/en/
CDC. (2022, Jun 9). Preparation and Collection of Specimens-Monkeypox:
https://www.cdc.gov/poxvirus/monkeypox/clinicians/prep-collection-
specimens.html
LEGEND OF PACKAGE SYMBOLS
See Table 9 for the legend of package symbols.
Table 9
Legend of Package Symbols
Icon | Definition |
---|---|
REF |
Catalog number |
Batch Code |
__
| Cap color
__
| Component
__
| Content/Volume
__
| Number
__
| Use-by-date
| Contains sufficient for X tests/reactions
__
| Protect from light
__
| Temperature limit
__
| Consult Instructions for Use
__
| Non-Sterile product – Do not sterilize
| Manufacturer
__
| Research Use Only
2401 S Foothill Dr Ste. D., Salt Lake CtJ, UT s.: 109, USA • 8014381036 • www.co.dx.com
References
- APHL
- Molecular Tests | Leading-Edge PCR Technology | Co-Dx - Co-Diagnostics, Inc.
- The Industry Leading Domain Broker - MediaOptions
- Molecular Tests | Leading-Edge PCR Technology | Co-Dx - Co-Diagnostics, Inc.
- Contact Us | Co-Dx - Co-Diagnostics, Inc.
- Feedback - Co-Diagnostics, Inc.
- Biosafety in Microbiological and Biomedical Laboratories (BMBL) 6th Edition | CDC Laboratory Portal | CDC
- Guidelines for Collecting and Handling Specimens for Mpox Testing | Mpox | Poxvirus | CDC
- Multi-country monkeypox outbreak in non-endemic countries
- Mpox (monkeypox)
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>