Co-Dx MPX2-R-001 Logix Smart Mpox (2-Gene) Instructions

June 9, 2024
Co-Dx

**Co-Dx MPX2-R-001 Logix Smart Mpox (2-Gene)

**

MANUFACTURER

Symbol Co-Diagnostics, Inc. 2401 S Foothill Dr. Ste D Salt Lake City, UT 84109 Phone: +1 801-438-1036
Email: info@co-dx.com
Website: www.co-dx.com

INTENDED USE

The Co-Dx Logix Smart™ Mpox (2-Gene) RUO is for the simultaneous qualitative detection of Mpox (moneybox) and other related Ortho poxviruses.
For research use only. Not for use in diagnostics procedures.

PRODUCT DESCRIPTION

The Co-Dx Logix Smart™ Mpox (2-Gene) RUO is a research-use-only multiplex test, based on real-time polymerase chain reaction technology. It tests for the presence or absence of deoxyribonucleic acid (DNA) of the Mpox virus and other related Orth poxviruses.
The Co-Dx Logix Smart™ Mpox (2-Gene) RUO includes an internal control to identify possible quantitative polymerase chain reaction (qPCR) inhibition, confirm the integrity of the reagents, and verify the quality of the sample extraction. The Co-Dx Logix Smart™ Mpox (2-Gene) RUO also includes a positive control (PC) which includes synthetic DNA molecules carrying sequences that are homologous to Mpox viruses targeted by this multiplex assay. PCs represent a source of cross contamination. Take precautions to prevent and minimize the risk of cross contamination.

Co Primers™ in the Co-Dx Logix Smart™ Mpox (2-Gene) RUO include the following:

  • Co Primers™ that target Mpox (MPX) F8L are labelled with the Quasar® 670 fluorophore.
  • Co Primers™ that target Mpox (MPX) L6R are labelled with the CAL Fluor® Orange 560 fluorophore.
  • Co Primers™ that target the Human R Nase PDNA, which acts as the Internal Positive Control (IPC), are labelled with CAL Fluor® Red 610 fluorophore.
Cross Reactivity

The Co-Dx Logix Smart™ Mpox (2-Gene) RUO is expected to cross-react with some Orth poxviruses, based on In Silico analysis or wet testing. That are expected to be detected by In Silico analysis include the following:

  • Mpox

  • Variola

  • Vaccinia*

  • Cowpox

  • Camelpox

  • Akhmeta virus

  • Buffalopox

  • Alaskapox

  • The vaccinia virus is present in some forms of the smallpox vaccine. There is a chance that individuals who were recently vaccinated for smallpox could show a positive result.

RUO COMPONENTS

See Table 1 for components included in the Co-Dx Logix Smart™ Mpox (2-Gene) RUO.

Table 1

RUO Component Information

Lid Color| Component| Symbol| Catalog Number| Description| Amount
---|---|---|---|---|---
Brown| Co-Dx Logix Smart™ Mpox Master Mix| MM| MPX2-MM-001| Proprietary blend of CoPrimers™ and

PCR reagents

| 1×1000 µL (100 reactions)
| Co-Dx Logix Smart™ Mpox Positive Control| PC| MPX2-PC-001| Proprietary blend of target templates| 1×1000 µL (100 reactions)
__

__

| Co-Dx Logix Smart™ Mpox No Template Control| NC| MPX2-NC-001| DNase/RNase- free water| 1×1000 µL (100 reactions)

The product code is MPX2-R-001. Contact Sales at 801-438-1036 ext. 01 or go to www.co- dx.com/contact/ to order.

COLLECTION, HANDLING, STORAGE, AND SHIPPING

See the following information for reagent storage and handling:

  • The Co-Dx Logix Smart™ Mpox (2-Gene) RUO is shipped on dry ice and should arrive frozen. Contact your distributor if one or more of the components are not frozen upon arrival or are compromised during shipment.
  • Store all components immediately upon arrival at a temperature between -40°C and -16°C to prevent degradation of reagents.
  • Do not use expired products. Integrity of expired components cannot be guaranteed.
  • Follow internal laboratory procedures for quality control when using this product.
  • Protect the master mix (MM) from light.
  • If you will be using reagents intermittently, freeze the reagents in multiple aliquots to ensure that less than 10 freeze/thaw cycles occur. Excessive thawing and freezing of components (specifically the MM) could affect the performance of the assay.
  • Avoid storing components at temperatures between +2°C and +8°C for more than 4 hours.
  • If you work in an area prone to power outages, keep a back-up generator for your freezer as well as a temperature data log system to ensure that the product remains frozen at a temperature between -40°C and -16°C.
  • Dispose of the product in accordance with applicable regional, national, and local laws and regulations. This product is not biological waste. The Safety Data Sheet (SDS) for this product can be viewed from Co-Diagnostics website at Safety Data Sheets |Co-Diagnostics, Inc. (co-dx.com).
  • Always follow the most recently recommended product stability data as it becomes available. This data can be found in our latest version of the Instructions for Use at Instructions For Use – Co-Diagnostics, Inc. (co-dx.com).

MATERIALS REQUIRED (NOT INCLUDED)

The following materials and devices are required but are not provided with this RUO:

  • Appropriate 4-channel real-time PCR instrument, compatible with the fluorophores used in this test.
  • Appropriate nucleic acid extraction system or kit
  • Vortex mixer
  • Centrifuge with a rotor for 2 mL reaction tubes
  • Adjustable pipettes
  • Disposable pipette tips with filters
  • Disposable powder-free gloves
  • Ice
  • Biosafety cabinet, ideally BSL-2 facility

| WARNING!
Install, calibrate, and maintain all instruments properly according to the manufacturer’s instructions and recommendations. Do not use instruments with outdated
---|---

BACKGROUND INFORMATION

Mpox is a viral zoonosis similar to smallpox, which is also part of the Orthopoxviral genus. Mpox primarily occurs in central and west Africa, often in proximity to tropical rain forests, but also near urban areas.
The Mpox virus is an enveloped double-stranded DNA virus with a variety of mammalian hosts, including rope squirrels, Gambian pouched rats, dormice, non- human primates, among others. (Reference https://www.who.int/news-room/fact- sheets/detail/Mpox) Mpox virus is transmitted by close contact with lesions, body fluids, respiratory droplets, and contaminated materials such as bedding.

The incubation period of Mpox is usually from 6-13 days but can range from 5-21 days. The infection can be divided into two periods: (Reference https://www.who.int/newsroom/fact sheets/detail/Monkeypox.
The invasion period (lasts between 0-5 days) characterized by fever, intense headache, lymphadenopathy (swelling of the lymph nodes), back pain, myalgia (muscle aches) and intense asthenia (lack of energy).
Lymphadenopathy is a distinctive feature of Mpox compared to other diseases that may initially appear similar (chickenpox, measles, smallpox) (Reference https://www.who.int/news-room/fact- sheets/detail/monkeypox). The skin eruption usually begins within 1-3 days of appearance of fever. The rash tends to be more concentrated on the face and extremities rather than on the trunk. It affects the face (in 95% of cases), and palms of the hands and soles of the feet (in 75% of cases). Also affected are oral mucous membranes (in 70% of cases), genitalia (30%), and conjunctivae (20%), as well as the cornea. The rash evolves sequentially from macules (lesions with a flat base) to papules (slightly raised firm lesions), vesicles (lesions filled with clear fluid), pustules (lesions filled with yellowish fluid), and crusts which dry up and fall off. The number of lesions varies from a few to several thousand. In severe cases, lesions can coalesce until large sections of skin slough off (Reference https://www.who.int/news-room /fact-sheets/detail/Monkeypox).
In May of 2022, cases of Mpox were reported to WHO from 12 member states across three WHO regions, where Mpox virus is not endemic. These areas include the United States, Canada, Australia, United Kingdom, Spain, and Portugal. (Reference https://www.who.int/emergencies/disease-outbreak- news/item/2022-DON385) -This reference is different than the above links.

ACCESSSORIES (NOT INCLUDED)

Thermocycler

Thermocyclers validated but not included with the test are displayed in Table 2.

Table 2

Thermocyclers Validated but Not Included with the Test

Thermocycler Machine Catalog Number Manufacturer
CoDx Box MIC-4 Co-Diagnostics, Inc.
Mic qPCR Cycler MIC-4 BMS, Bio Molecular Systems
Extraction System

Extraction System required but not included with the test are displayed in Table 3.

Table 3

Extraction and Automation Systems Validated with the Test

Extraction Reagent| Automation Platform (If applicable)| Manufacturer| Sample Input Volume/Sample Elution Volume by Sample Type
---|---|---|---
Name| Cat. Number
QIAamp DNA Mini Kit (Qiagen)| 51304 (50 extractions) __ 51306 (250 extractions)| N/A| Qiagen| Protocol Used| Sample Type| Input| Output
Buccal Swab| dry lesion swab| 400 µL| 60 µL x 2

WARNINGS AND PRECAUTIONS

| WARNING!
Before performing any testing or running any sample, verify that all instruments have been properly installed, calibrated, and are well maintained. Do not use instruments with an outdated calibration.
---|---

Read this Instructions for Use document carefully before using the product. Before first use, check the components for the following:

  • Integrity
  • Frozenness upon arrival
  • Users should ensure the following:
  • Limit use of this product to personnel instructed and trained in the techniques of real-time PCR.
  • Always treat samples as infectious and/or biohazardous. Use standard precautions.
  • Wear protective gloves, lab coat, and eye protection when handling samples and always wear gloves when handling RUO components.
  • Always use DNase/RNase-free disposable pipette tips with filters.
  • Use segregated working areas for sample preparation, reaction setup, and amplification/detection activities. The workflow in the laboratory should proceed in a unidirectional workflow. To prevent contamination, change personal protective equipment (PPE) between areas.
  • Store and extract positive materials (specimen, controls, and amplicons) separately from other reagents. Dedicate supplies and equipment to separate working areas and do not move them from one area to another.
  • Consult the appropriate SDS for safety. The SDS for the Co-Dx Logix Smart™ Mpox (2-Gene) RUO is provided with the shipment. If not provided with the shipment, the SDS can be retrieved from Co-Diagnostics website at the following link: Safety Data Sheets | Co-Diagnostics, Inc. (co-dx.com)
  • Do not open the reaction tubes/plates post amplification.
  • Do not autoclave reaction tubes/plates after the PCR, since this will not degrade the amplified nucleic acid and will pose a risk to the laboratory area to contamination.
  • Do not use components that have passed expiration date.
  • Discard sample and assay waste according to your local safety regulations

SAMPLE INFORMATION

Sample Storage

Ensure the following when storing samples:

  • Process all specimen types within 48 hours after collection, if storage is needed after 48 hours, store the samples frozen, preferably at -70°C (CDC, 2020).
  • Avoid repeated freezing and thawing of any specimen. If you need to keep a specimen for retesting, aliquot the specimen in different tubes to avoid freezing and thawing cycles.
  • Monitor the temperature in the storage areas and recorded temperatures regularly to identify potential fluctuations.
  • Do not use domestic refrigerators/freezers with wide temperature fluctuations. Domestic refrigerators/freezers are not suitable for the storage of frozen specimens (CDC, 2020).
Sample Handling

Laboratory workers should wear appropriate PPE, which includes disposable gloves, laboratory coat/gown, and eye protection when handling potentially infectious specimens.
Conduct samples suspected to be or confirmed to be infected with MP ox under a certified Class II biosafety cabinet in a BSL-2 containment facility. More details are Provided in the Biosafety in Microbiological and Biomedical Laboratories (BMBL) (CDC, 2009) or the WHO Laboratory Biosafety Manual (WHO, 2004).
For specific instructions on handling clinical specimens for the MP ox virus, see also the CDC’s webpage for the Preparation and Collection of Specimens – MP ox (Guidelines for Collecting and Handling Specimens for Mpox Testing | MP ox | Poxvirus | CDC) (CDC, 2022) and Interim Guidance for Collection Diagnostic Specimens from Persons with Suspect Moneybox (CDC, 2003).
The quality of the extraction of the DNA from the samples is essential for the performance of Co-Dx Logix Smart™ MP ox (2-Gene) RUO. Perform extraction by following manufacturer’s instructions or an internally validated protocol. The suitability of the nucleic acid extraction procedure for use with Co-Dx Logix Smart™ Mpox (2-Gene) RUO must be validated by the user.

| WARNING!
If your sample preparation system uses washing buffers
containing ethanol, make sure to eliminate any traces of ethanol
prior to elution of the nucleic acid. Ethanol is a strong inhibitor of
real-time PCR.
The use of carrier RNA is crucial for extraction efficiency and stability of the extracted nucleic acid.
---|---

PROCEDURE

Real Time RT-PCR Setup

Set Up the Reagents

Perform the steps below to set up the reagents.

Clean all working surfaces with a fresh 10% bleach solution followed by a molecular-grade alcohol or another equivalent method of cleaning that disinfects and degrades nucleic acids.
Thaw all reagents and samples on ice, or a cold block, before starting the setup.
Vortex all Co-Dx Logix Smart™ Mpox (2-Gene) RUO MM, PC, nuclease-free water (used as an NC), and all sample tubes for 3 seconds before using.
Briefly spin the MM, PC, NC down before using to ensure reagents are properly mixed and to ensure removal of any condensation or residue from the lids.

Set Up the Reaction

Perform the steps below to set up the reaction.

Collect enough reaction wells for each of the following:

  • One for each NC,
  • One for each sample you want to test, and
  • One (or more) for each PC

Note: The example below displays the minimum number of wells needed for 5 samples.
PC 1
NC 1
Samples 5
Total wells required 7

Pipet 10 µL of MM into each well collected.
Pipet 10 µL of the NC into the appropriate wells (in addition to the 10 µL of MM already in the well).

Note: Ensure that at least one NC is included in each run and that enough space remains for at least one PC.

Important:

  • Pipette on ice, if possible.
  • Perform PC pipetting and sample setup in a separate area, or at a separate time from the MM and NC.
  • Change pipette tips between samples and change pipette tips after pipetting each component.
  • Pipet the PC last, if possible, to avoid contamination events.

Pipet 10 µL of sample or PC into the appropriate well.
Seal the reaction plate with an optical adhesive film or seal each reaction tube with its appropriate lid
Place the plate or tubes into the RT-PCR instrument in the correct orientation and start the run.

PCR Instrument and Thermocycler Setup

For programming instructions questions regarding the use of other realtime PCR instruments, contact the Laboratory 801-438-1036 ext. 03 or at www.co- dx.com/contact/.

If using Co-Diagnostics Inc. Co Dx Box, contact the Laboratory (see contact information in Section 7.4.2) for the template file for download. The template file comes pre-programmed with the PCR instrument setup described in this section. When not using a template, or using another device, use the settings outlined below to program the PCR To achieve optimal performance from the test, it is important to make sure that the instrument is compatible with the conditions outlined below. instrument.
Define the settings as displayed in Table 4.

Table 4

Recommended Instrument Settings

Item Setting
Reaction Volume 20 µL
Ramp Rate Default
Passive Reference None

Program PCR instrument with the cycling conditions displayed in Table 5.

Table 5

Recommended Cycling Condition Settings

Item Stage Cycles Temperature Time
Initial Denaturation Hold 1 95˚C 2 minutes
Amplification Cycling 45 95˚C 3 seconds
53˚C 32 seconds

Ensure that the PCR instrument being used is compatible with the fluorophores below. Some devices may not have options for the quencher. If you need help or have questions, contact Co-Diagnostics Inc. Technical Support at 801-438-1036 ext. 02 or at: support@codx.com.

Define the fluorescence detectors (dyes) as displayed in Table 6.

Table 6

Fluorescence Detector Definitions

Target Detector Name Reporter Quencher
MPX-F8L (T4) MPX-F8L (T4) Quasar® 670 BHQ®-2
MPX-L6R (T3) MPX-L6R (T3) CAL Fluor® Orange 560 BHQ®-1
RNaseP (IPC) RNaseP (IPC) CAL Fluor® Red 610 BHQ®-2

When the run is finished, ensure that the run file is saved.

DATA ANALYSIS

For basic information regarding data analysis on specific qPCR instruments, please refer to the user manual of the respective instrument.

Positive Controls

Validate the test run by checking to see that the PC has passed and that the control conditions displayed in Table 7 are met.

No-Template Controls

Validate the test run by checking to see that no-template control has passed and that the control conditions displayed in Table 7 are met.

Table 7

Required Control Conditions

Control Type| Control Name| Purpose of Control| MPX L6R Result| MPX F8L Result| RNaseP (IPC) Result
---|---|---|---|---|---
MPX Positive Control (PC)| MPX-L6R (CF®560)| Verifies the performance of the master mix| +| +| +
MPX-F8L (Quasar® 670)
RNaseP (IPC) (CF®610)
No Template Control (NC)| Master Mix + Water| Verifies the reagents are free of contamination| | |

If controls pass, interpret the sample results.

Invalid Test Run
If any of the controls fail, the run is invalid.
Document the run and initiate troubleshooting.

Interpretation of Results

Once the controls have passed, the unknown samples can be interpreted based on one of the following three possible outcomes:

  • Positive
  • Negative
  • Invalid

A Positive result will show an amplification curve or cycle threshold value for Mpox. The cut-off value should be determined by in house validation testing. However, internal studies have shown rare primer-dimer formation or other non-specific amplification at 45 cycles. This fact can be attributed to the nature of the CoPrimers™ (Satterfield, 2014), (Poritz & Ririe, 2014). The amplification of the RNaseP (IPC) shows that the extraction was successful.
A Negative result shows no amplification for Mpox; The absence of a curve for Mpox indicates a negative result ONLY when the RNaseP (IPC) marker is positive.
An Inconclusive result occurs if any of the controls fail. See the Troubleshooting section.

See Table 8 for results translation

Table 8

Results Translation

| Sample Result| Co-Dx Logix Smart™ Mpox Positive Control| No Template Control (NC) (Master Mix + Water)| Interpretation Results
---|---|---|---|---
MPX-L6R CF560 Channel| MPX-F8L Q670 Channel| Internal Positive Control (RNaseP) CF610 Channel
Instrument Reading| +| +| +| +| | *Mpox Virus DNA +
+| | +| +| –*|
Mpox Virus DNA +
| +| +| +| –**| *Mpox Virus DNA +
| | +| +| | Mpox Virus DNA
Any Result (+/-)| | +| | INVALID: See Troubleshooting
+| |
+| +| +

Anything before 42 cycles is considered a positive analyte reading (+). Anything at or after 42 cycles is considered a negative analyte reading (-).
*A positive result indicates a sample contains either MP ox or other orthopoxviral DNA.

Note: Samples collected from individuals who were recently vaccinated for smallpox may show a positive result because of cross-reactivity with Orth poxvirus vaccinia virus, which is present in some forms of the smallpox vaccine.

TROUBLESHOOTING

Co-Diagnostics Inc. values customer feedback and we would like to be informed of any issues with the Co-Dx Logix Smart™ Mpox (2-Gene) RUO even if the recommended steps for troubleshooting solve the issue. To give feedback please fill out the Customer Feedback form by visiting www.co- dx.com/contact/feedback/.

Stability

Real-time and accelerated shelf-life and in-use stability studies are currently under testing. Currently, the expiration date of this product has been established as 12 months. Do not use expired reagents, because doing so may lead to inaccurate results.
Always use the most recent version of this document for updates as more stability information will be added when studies are completed.

User Errors

Good Laboratory Practices for Molecular Biology Diagnostics (Viana & Wallis, 2011) are necessary for the use of this product. This product is not intended to be used by untrained personnel.
To help prevent errors such as splashes, crossover contamination, and volume selection, it is essential that users have some molecular biology experience and be familiar with proper pipetting technique. Pipette tips must be replaced after every pipetting. Gloves must be replaced often. Equipment, such as pipettes and realtime PCR instruments, should be calibrated when applicable.
A 90-minute online training for Good Laboratory Practices for Molecular Genetics Testing (Centers for Disease Control and Prevention, 2017) is available at the CDC website at the following link https://www.cdc.gov/labtraining/training-courses/goodlab-practices-molecular genetics-testing.html

Invalid Results

Co-Dx Logix Smart™ Mpox PC not Amplifying

No amplification from the PC could be the result of one or multiple factors, such as one or more of the following:

  • Pipetting errors (pipetting control into the wrong well, missing a well, pipetting inadequate amount of reagent)
  • Incorrect placement of plates or tubes into the real-time PCR instrument
  • Co-Dx Logix Smart™ Mpox MM or PC degradation (result of reagents being stored at a temperature above -16°C for an extended period)
  • Use of expired reagents
  • Wrong reagents being used

When this occurs, it is best to disregard the results from the samples and re- test by re-amplification. If the PC fails again, then an investigation should be conducted to identify possible causes for error, and the test must be reprocessed from extraction (or not, depending on the investigation results and risks identified in the process). If failure of the PC occurs a third time after re-extraction and re-amplification, open a new Co-Dx Logix Smart™ MPX PC or MM, and retest. If still failing, please contact Co-Diagnostics Inc. Technical Support by calling 801-438-1036 ext. 02 visiting support @co-dx.com.

PC RNaseP is not Amplifying in Samples

No amplification from the RNaseP channel could be the result of one or multiple factors, such as:

  • Not enough nuclear material in the sample
  • PCR inhibitors such as ethanol and heparin
  • Incorrect extraction
  • Extraction system used is not compatible or has a step that eliminates RNaseP DNA

Note: Positive amplification in the Mpox channel indicates a positive analyte result despite the lack of concurrent amplification in the IPC channel. The IPC amplification is dependent on the presence of human genomic DNA (gDNA) in the extraction sample, the amount of which is governed by the type of the sample and the extraction procedure used. Samples obtained from culture or sterile/pure sites (e.g., CSF, urine, or cell lysates) may not contain the human RNaseP gene.

When this occurs, the results should be interpreted as invalid and retesting by re-amplification should be performed. If the IPC fails again, then samples should be re-extracted and re-amplified. If it fails a third time an investigation should be conducted to identify possible causes for the error. If the cause for the error is clear, the test can either be singled out as invalid due to either PCR inhibitors being present or not enough nuclear material being present. If the cause for an error is unclear, contact Co- Diagnostics Inc. Technical Support by calling 801-438-1036 ext. 02 or contact us at support@co-dx.com.

No Template Control (NC) Showing Amplification

Amplification of MPX in a No Template Control indicates contamination in one or more of the reagents, incorrect placement of plate or tube into the real- time PCR instrument, or pipetting errors.
When this occurs, none of the results can be trusted and re-testing by re- amplification should be performed. If the NC fails again, then an investigation should be conducted to identify possible causes for error, and the test must be reprocessed from extraction or not, depending on the investigation results and risks identified in the process. If failure of the NC, after re-extraction and re-amplification, happens a third time, open a new nuclease-free water and retest. If still failing, please contact Co- Diagnostics Inc. Technical Support by calling 801-438-1036 ext. 02 or by visiting support@co-dx.com.

REFERENCES

CDC. (2009). Biosafety in Microbiological and Biomedical Laboratories (BMBL) 5th Edition. Retrieved from CDC Laboratories: https://www.cdc.gov/labs/BMBL.html
CDC. (2017, Oct 27). CDC Laboratory Training: Good Laboratory Practices for Molecular Genetics Testing. Retrieved Mar 5, 2019, from Centers for Disease Control and Prevention: https://www.cdc.gov/labtraining/training- courses/good-lab-practicesmolecular-genetics-testing.html
Poritz, M., & Ririe, K. (2014, Mar). Getting things backwards to prevent primer dimers. Journal of Molecular Diagnosis, 159-62. doi:10.1016/j.jmoldx.2014.01.001
Satterfield, B. (2014, Mar). Cooperative primers: 2.5 million-fold improvement in the reduction of nonspecific amplification. Journal of Molecular Diagnosis, 163-73. doi:10.1016/j.jmoldx.2013.10.004
Viana, R. V., & Wallis, C. L. (2011). Good Clinical Laboratory Practices (GLCP) for Molecular Based Tests Used in Diagnostic Laboratories. In D. I. Akyar, Wide Spectra of Quality Control (pp. 29-52). InTech. Retrieved from Good Clinical Laboratory Practice (GCLP) for Molecular Based Tests Used in Diagnostic Laboratories | IntechOpen
CDC. (2003, Jun 23). Interim Guidance for Collection of Diagnostic Specimens from Persons with Suspect Monkeypox. Retrieved June 10, 2022, from Centers for Disease Control and Prevention: Interim Guidance for Collection of Diagnostic Specimens from Persons with Suspect Monkeypox (aphl.org)
WHO. (2004). Laboratory Biosafety Manual. Retrieved from Emergencies preparedness, response: https://www.who.int/csr/resources/publications/biosafety/WHO_CDS_CSR_LYO_200 4_11/en/
CDC. (2022, Jun 9). Preparation and Collection of Specimens-Monkeypox: https://www.cdc.gov/poxvirus/monkeypox/clinicians/prep-collection- specimens.html

LEGEND OF PACKAGE SYMBOLS

See Table 9 for the legend of package symbols.

Table 9

Legend of Package Symbols

Icon Definition
REF Catalog number
Batch Code

__

| Cap color

__

| Component

__

| Content/Volume

__

| Number

__

| Use-by-date
| Contains sufficient for X tests/reactions

__

| Protect from light

__

| Temperature limit

__

| Consult Instructions for Use

__

| Non-Sterile product – Do not sterilize

Symbol

| Manufacturer

__

| Research Use Only

2401 S Foothill Dr Ste. D., Salt Lake CtJ, UT s.: 109, USA • 8014381036 • www.co.dx.com

Co-Dx Logo

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Co-Dx User Manuals

Related Manuals