Co-Dx Logix Smart SARS-CoV-2 Direct Saliva User Guide

June 3, 2024
Co-Dx

Quick Start Guide
for Using the
Logix Smart™ SARS-CoV-2 Direct Saliva RUO Product
(COVDS-R-003 and -004)


Plate the Master Mix

Perform in Clean Area

a. Load the Co-Dx Box software.
b. In the Clean area, clean work surfaces with water > 10% bleach > water > 70% alcohol (in that order), or another equivalent method of cleaning that disinfects and degrades nucleic acids.Co-Dx Logix Logix Smart Influenza A
and B and SARS-CoV-2 RUO fig 2c. Thaw the master mix (MM) and the nuclease-free water (NFW) (for the no-template control [NC]), on ice.Co-Dx Logix Logix Smart
Influenza A and B and SARS-CoV-2 RUO fig 3 d. Briefly vortex and centrifuge the MM and NFW reagents .Co-Dx Logix Logix Smart Influenza A and B and SARS-CoV-2 RUO fig
4 e. Add 10 µL of MM into every well that will be used (1 well for NC, 1 well for positive control [PC], and 1 well for each sample).

Co-Dx Logix Logix Smart Influenza A and B and SARS-CoV-2 RUO fig
5

f. Pipet 10 µL of NFW into only the NC well and cap the NC well.Co-Dx Logix
Logix Smart Influenza A and B and SARS-CoV-2 RUO fig 6

g. Place reagents back into the freezer.

Plate Samples and Positive Control (PC)

Perform in Positive Area

a. Load the Co-Dx Box software.
b. In the Positive area, clean work surfaces with water > 10% bleach > water > and 70% alcohol (in that order).

c. Thaw the PC on ice.
d. Briefly vortex and centrifuge all reagents.
e. Perform sample extraction, either using RNA extraction or lysis.
f. Add 10 µL of the sample extract to the appropriate wells and cap the sample extract wells.
Co-Dx Logix Logix Smart Influenza A and B and SARS-CoV-2 RUO fig
12g. Add 10 µL PC to the PC well and cap the PC well.Co-Dx Logix Logix Smart Influenza A and B and SARS-CoV-2
RUO fig 13h. Place capped tubes in the thermocycler, in appropriate locations and fill the rest of the carousel with blank tubes.Co-Dx Logix Logix Smart Influenza A and B and SARS-CoV-2
RUO fig 14i. (Optional): Cap all tubes and place samples and reagents back into the freezer.Co-Dx Logix Logix
Smart Influenza A and B and SARS-CoV-2 RUO fig 15

Start the Run

Perform in Co-Dx Box Software
a. In the Co-Dx Box software, click New. b. Select an assay template and click OK.

Co-Dx Logix Logix Smart Influenza A and B and SARS-CoV-2 RUO fig
17

c. Under Run Setup, select Samples and type sample information in the appropriate order.
Co-Dx Logix Logix Smart Influenza A and B and SARS-CoV-2 RUO fig
18d. Click Information and in the Notes field, type any needed information.Co-Dx Logix Logix Smart
Influenza A and B and SARS-CoV-2 RUO fig 19 e. Select the appropriate thermocycler and from the drop-down select Start Run. f. Verify the tube clamp is in place and the appropriate reaction volume is listed, and select Start.Co-Dx Logix Logix
Smart Influenza A and B and SARS-CoV-2 RUO fig 21The run initializes and a progress bar gives an estimated time to completion.