Sonel F-xA Flexible Coil User Manual
- June 9, 2024
- Sonel
Table of Contents
Sonel F-xA Flexible Coil
Wire Connection
- Rogowski coil
- fastener
- connection cable
- plug
Description
The arrow located on the closing unit indicates the current flow direction. It
is assumed that the current is flowing in the positive direction if it is
flowing from the source to the receiver. Such probe orientation is required
for a correct power measurement.
The output signal is supplied by a 2.5-meter lead with a pin adapted for the
socket in the meter.
F-xA coil
F-1A, F-2A and F-3 flexible coils (Rogowski coil) are used to measure the
alternating current in the range from 40 Hz up to 20 kHz and current up to
3000 A (10 kA peak at 50 Hz)
The only difference between the F-1A, F-2A and F-3A flexible coils is the coil
size. The electrical parameters are identical.
The output signal if voltage proportional to the derivative of the measured
current, with the sensitivity equal to 38.83 mV/1000 A for 50 Hz and 46.6
mV/1000 A for 60 Hz.
F-xA1 coil
Flexible clamps (Rogowski coil) F-1A1, F-2A1 and F-3A1 are probes of a double-
increased sensitivity comparing to F-1A/F-2A/F-3A, which translates into
halved measuring range for current up to 1500 A
The output signal is voltage proportional to the measured current derivative
at the sensitivity of 77.66 mV/1000 A for 50 Hz and 93.19 mV/1000 A for 60 Hz
F-xA6 coil
Flexible probes (Rogowski coil) F-1A6, F-2A6 and F-3A6 are clamps of a double-
reduced sensitivity comparing to F-1A/F-2A/F-3A, which translates into doubled
measuring range for current up to 6000 A.
The output signal is voltage proportional to the measured current derivative
at the sensitivity of 19.415 mV/1000 A for 50 Hz and 23.3 mV/1000 A for 60 Hz
Safety
NOTE!
Do not use non-insulated coils for conductors with a potential exceeding 1000
V with respect to the ground, in systems with the measurement category higher
than III and systems of measurement category IV with potential exceeding 600
V.
Operation
To measure the current, open the coil slightly by turning the clasp, clamp them on the conductor with flowing current and roughly center the conductor relative to the coil loop. Pay attention to the direction of the arrow, to avoid the power measurement error.
NOTE!
Follow the instructions for mechanical limitations of the coils. Mechanical
damage to the coil is not covered by the warranty.
**** Do not pull by force
Do not bend forcibly
Do not pack in a bent state
Cleaning and maintenance
**NOTE!**
Apply only maintenance methods specified by the manufacturer in this manual.
Before cleaning, disconnect the coil from the tested circuit and the meter.
The coil may be cleaned with a soft, damp cloth using all-purpose detergents.
Do not use any solvents.
Dismantling and utilization
Worn-out electric and electronic equipment should be gathered selectively,
i.e. it must not be placed with waste of another kind.
Worn-out electric equipment should be sent to a collection point in accordance
with the law of waste electrical and electronic equipment.
Before the equipment is sent to a collection point, do not dismantle any
elements. Observe the local regulations concerning disposal of packages.
Reference conditions
a) temperature ……………………………………………+18°C …+22°C
b) conductor position ………..centered in relation to the coil loop
c) continuous magnetic field…………………. <40 A/m (earth field)
d) alternating magnetic field …………………………………………none
e) external electric field………………………………………………..none
Technical data
Basic technical data
a) measuring range
- F-xA …………………….1 A…3000 A (10000 A peak at 50 Hz)
- F-xA1 …………………….1 A…1500 A (5000 A peak at 50 Hz)
- F-xA6 …………………..1 A…6000 A (20000 A peak at 50 Hz)
b) output level
-
F-xA ………………………………………38.83 mV/1000 A (50 Hz)
………………………………………………… 46.6 mV/1000 A (60 Hz) -
F-xA1 …………………………………….77.66 mV/1000 A (50 Hz)
………………………………………………..93.19 mV/1000 A (60 Hz) -
F-xA6 …………………………………..19.415 mV/1000 A (50 Hz)
…………………………………………………… 23.3 mV/1000 (60 Hz)
c) accuracy………………………………±1% in the measuring range
d) additional error due to the position of the conductor. ±1% max.
When using the coil with a SONEL meter, total measurement accuracy of the measuring system of the meter and coil is specified in the manual of a given meter. It is not the sum of the accuracy of the meter and accuracy of the coil.
Other technical data
a) insulation type acc. to IEC 61010-1……………………….. double
b) measurement category acc. to IEC 61010-1 ..III 1000 V, IV 600 V
c) pollution degree…………………………………………………………… 2
d) ingress protection acc. to IEC 60529………………………….IP67
e) coil diameter……………………………………………………… 8.4 mm
f) fastener diameter (maximum)……………………………….. 36 mm
g) coil circuit
- F-1A, F-1A1, F-1A6………………………………………….. 120 cm
- F-2A, F-2A1, F-2A6…………………………………………… 80 cm
- F-3A, F-3A1, F-3A6…………………………………………… 45 cm
h) inner diameter of coil after closing
- F-1A, F-1A1, F-1A6……………………………………………. 38 cm
- F-2A, F-2A1, F-2A6…………………………………………… 25 cm
- F-3A, F-3A1, F-3A6…………………………………………… 14 cm
i) weight
- F-1A, F-1A1, F-1A6………………………………………. ca. 200 g
- F-2A, F-2A1, F-2A6……………………………………….. ca. 170 g
- F-3A, F-3A1, F-3A6………………………………………. ca. 140 g
j) length of coil cable ………………………………………………… 2.5 m
k) operating temperature………………………………. -30°C…+80°C
l) storage temperature…………………………………. -40°C…+80°C
m) the product meets requirements of standards……IEC 61010-1
………………………IEC 61010-031, IEC 61010-2-031, IEC 61010-2-032
Manufacturer
he manufacturer, which also provides guarantee and post-guarantee services:
Customer Support
SONEL S.A.
Wokulskiego 11
58-100 Świdnica
Poland
tel. +48 74 858 38 60
fax +48 74 858 38 09
e-mail:export@sonel.pl
web page:www.sonel.pl
Manufactured in Italy for SONEL S.A.
Documents / Resources
|
Sonel F-xA Flexible
Coil
[pdf] User Manual
F-xA Flexible Coil, F-xA, Flexible Coil, Coil
---|---
|
Sonel F-xA Flexible
Coil
[pdf] User Manual
F-xA, F-xA1, F-xA6, F-xA Flexible Coil, F-xA, Flexible Coil, Coil
|
Sonel F-xA Flexible
Coil
[pdf] User Manual
F-xA, F-xA Flexible Coil, Flexible Coil, Coil
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>