nosiboo Pro Electric Nasal Aspirator User Manual

June 3, 2024
nosiboo

nosiboo-Pro-Electric-Nasal-Aspirator-LOGO

nosiboo Pro Electric Nasal Aspirator

nosiboo-Pro-Electric-Nasal-Aspirator-PRODUCT

NOTEBOOK PRO IS A HYGIENIC MEDICAL DEVICE

You may only exercise your right of withdrawal if the device is undamaged and the protective foil on the Colibri head is intact.

Guiding and applicable legislation: Directive 2011/83/EU § 16 (e) of the European Parliament and of the Commission, Austrian Republic: FAGG: § 18 (1) 5, Hungary: 45/2014. (II. 26.) Government decree § 29 (1) (e).
Register your device and we will send you a coupon as a gift that entitles you to a discounted purchase. You may redeem the coupon in the Nosiboo Pro manufacturer’s online store. www.nosiboo.eu/productregistration
Package contents

  • Colibri head
  • Power cord
  • User manual
  • Device
  • Tube
  • Spare removable nose tip
  • Quick guide
  • Foreword

Dear customer,
Congratulations on buying our KSCI, PTPA, Red Dot Design and German Design awarded Nosiboo Pro nasal aspirator! Please read this user manual carefully so that you’ll be able to use the nasal aspirator for years and can prevent illnesses for a long time.

  • Please read the user manual carefully before using the device!
  • Attract Kft. is an ISO 13485 certified medical device manufacturer. This product is an MDD Legacy Device.
    Actor EUDAMED No.: HU-MF-000004252

Warning and safety signs

  • You may see the following signs in the user manual or on the label of the device:
  • MANUFACTURED BY ATTRACT KFT. 7622 Pécs, Siklósi út 1/1 – Hungary
  • T +36 72 551 642 customerservice@nosiboo.com T +36 72 551 643 www.nosiboo.eu

Explanation

  • Type BF applied parts (Increased protection against electric shock)

  • Class I device (double insulation)

  • Class lla medical device

  • Name and address of the manufacfurer

  • Read the user manual!

  • For home use only

  • Reference (order number)

  • Serial number

  • Global Trade Item Number (UDI-DI)

  • Lot number (production identifier)

  • Protected against solid objects over 12.5 mm and vertically falling drops of a wafer.

  • Keep dry!

  • Warning!

  • Hazardous waste

Important safety instructions

Warning: If possible hazardous situations below are not avoided, user injury or even a fatal accident may occur. Safety instructions regarding general use

  • The device can be used exclusively for removing the nasal mucus from a child’s nose. It cannot be used for any other purpose!
  • It is FORBIDDEN to use the device in the bathroom or any other humid or wet environment!
  • Stop using the device if liquid or solid has been drawn into the housing. Contact a qualified service centre.
  • The device should only be used by adults.
  • While in use, the device should be in a stable position, standing on a flat surface.
  • Do not lay the child on their back and do not block the free nasal opening while using the device!
  • Do not leave the device unattended near children or people with reduced physical or mental abilities!
  • Do not leave the device unattended near large pets (dog, cat) that can knock it over!
  • The power cord can easily wind round the child’s neck!
  • The device cannot be used in oxygen-rich environments (in clinical circumstances)!
  • To prevent overuse and for the safety of your child, the device switches off automatically after 15 minutes. After the forced switch-off, the device can be turned on again. In case of overheating, the electronics will prevent re-use for approx. 50 minutes.
  • For more information please read Chapter 4 of the User Manual.
  • Make sure that all units are appropriately attached before using the device!
  • Do not use the device without the air mesh.
  • Before use, double-check if you have inserted the (dry) collared tube into the Colibri head after cleaning.
  • Do not use the Colibri head and the device without the collared tube.
  • Do not use more than 5 ml saline solution because if the solution gets into the tube, it can cause damage to the device.
  • The connection point of the Colibri head should point downwards while in use.
  • Stop using the device if mucus or saline solution has been drawn into the tube, disconnect it from the mains, then empty the Colibri head, clean and dry the Colibri head and the tube, check the air mesh and replace it if it is wet or clogged.
  • Make sure that the housing, the tube and the Colibri head are not damaged or cracked! Do not use the device if you see that it is damaged! Replace the damaged part! (List of spare parts can be found in section “Accessories, spare parts”.)
  • If rashes or redness appear on the nose of the child after using the device, do not use the device again! Consult a physician.
  • Use the device exclusively with the tube and Colibri head supplied by the manufacturer!
  • Do not attach the Colibri head to any other devices.
  • Safety instructions regarding connecting the device to the mains
  • Protect the power cord against damage! Do not break it, do not place heavy items on it because a damaged cord may cause electric shock.
  • Keep the power cord away from heat (e.g. furnace, heater and convector) because it can cause damage to the isolation.
  • Only use the device if the tube and the Colibri head are dry.
  • Do not remove the mounting screws of the housing. Do not reconstruct the device for other purposes.
  • The use of the device is forbidden if there is a breakage or crack on the housing or on the power cord due to misuse.
  • In case of unexpected failure or malfunction, immediately switch off the device and interrupt the power supply by unplugging it from the mains. Safety instructions regarding the next use of the device
  • Always disinfect the Colibri head after use!
  • Do not leave the parts of the dismounted Colibri head within reach of children because they may swallow these.
General Precautions
  • Read and follow the instructions in the “Important information about reducing electromagnetic interferences” section!
  • Read and follow the instructions in the “Environment” section if you would like to dispose of the device or any part of it.
Intended use

The Notebook Pro nasal aspirator was designed to remove the nasal mucus from the child’s nose. The vacuum and airflow needed for the aspiration is created by an aspirating device designed specifically for this purpose. The Colibri head, which is inserted into the nose, is connected to this device via a tube. The Colibri head has a patented design that prevents mucus from entering the device.

Intended purpose:

  • to ensure nasal breathing in infants
  • to prevent the complications of childhood common cold and to accelerate the healing process by aspirating the nasal mucus
  • to relieve childhood allergy-like symptoms by removing the nasal mucus.

Patients to be treated:

  • children who are (typically under the age of 6) unable to remove the nasal mucus by blowing their noses. Illness, a condition requiring treatment (indication):
  • to eliminate clogging of the nose
  • treatment of common cold and auxiliary treatment of superinfected paranasal sinuses
  • supplementary treatment of the allergic infectious condition.

Conditions in which the device cannot be used (contraindication):

  • haemophilia
  • if the blood vessels of the nasal septum’s mucous membrane widen – in case of frequent nosebleed
  • while taking blood-thinning medicines
  • in case of congenital malformations (e.g. cleft palate).

Suggested user:

  • Age: adult over 18
  • Sex: not applicable (man or woman)
  • Knowledge: should have general knowledge according to their age
  • Education: min. primary school education, no need for special knowledge to use the device.
  • Important: knowledge of the User manual included in the package.

Product introduction

Figure 1 – Nosiboo Pro nasal aspiratornosiboo-Pro-Electric-Nasal-
Aspirator-FIG-5

  1. Power cord
  2. On/Off control button
  3. Assembled housing
  4. Air mesh
  5. Connecting unit
  6. Tube
  7. Collared tube
  8. Nozzle housing
  9. Nose tip
  10. Nozzle housing with connection
  11. Colibri head

First-time operation

Assembling the Colibri head

  1. Clean and dry the parts of the Colibri head (see sections “Cleaning” and “Maintenance and storage”).

  2. Insert the collared tube into the nozzle housing that is supplied with a connection.

  3. Place the nozzle member that has no connection in the right position.

  4. Push the flexible nozzle end on the joint collar.

  5. Attach one end of the tube to the Colibri head.
    Attaching the device

  6. Make sure the air mesh behind the connecting unit is in place.

  7. Attach the free end of the tube to the device housing.

  8. Attach one end of the 230 V mains cord to the plug on the lower part of the device housing, then the other end to the socket. The device is ready for use.

Instructions for use

Before use, the device should be placed in such a way that the plug of the power cord can be easily pulled out from the socket! You can turn the device on or off by pushing and control the intensity of the suction by turning the button on top of the housing. Adjustability guarantees that after selecting the appropriate suction power, the parent cannot cause any secondary deformation in the nose. Before turning on the device, find the right airflow rate fit for your child’s age based on Figure 2 (Adjusting the suction power).
Figure 2 – Adjusting the suction powernosiboo-Pro-Electric-Nasal-
Aspirator-FIG-6

The minimum value of the scale is suitable for newborns, the middle value for babies and the maximum value for small children and for preschoolers. The effectiveness of the device is highly influenced by the thickness of the mucus, the extent of the disease and the unique anatomy of the child. If the mucus is thin, you can use the device at a lower setting, if it is thick, you can use the device at a higher setting than usually. The device can be operated exclusively by adults. Do not give the device ready for use to the child! Hold the child in a stable position while using the device. Sit your child onto your lap and hold them or sit them into a straight position supporting their back with a pillow. You may also sit them into a resting chair with a tilted back. Do not lay your child on their back and do not block the free nasal opening! Make sure that the vents at the back of the device are free!
Warning! Hot air may exit the device through the vents!
Unplug the device after it has completely shut off and the cooling phase is over!

Insert the previously disinfected and dried Colibri head into the child’s nose, and then turn on the device by pushing the control button (the device operates at the setting used last time).
Figure 3 – Correct positioning of the Colibri headrushnosiboo-Pro-
Electric-Nasal-Aspirator-FIG-7

the tip of the Colibri head into the nose. The tip should point slightly downwards and outwards towards the eyeball. In this position, you can gently move the Colibri head. With this method you can reach the opening leading to the maxillary sinus and you will not hurt the mucous membrane of the nasal concha or the capillaries of the nasal septum. In rare instances, it is possible that the nostrils of a newborn are too small for inserting the nose tip. In this case, place the nose tip in a way that the nostril is completely covered. The nasal aspirator’s efficiency is enough for cleaning the nasal passages properly. The connection of the flexible tube should point downwards while in use. The patented design of the Colibri head can hold up to 5 ml thin liquid and prevents the mucus from entering the flexible tube. Dry and very thick mucus can be dissolved by saline solutions. Do not use more than 5 ml of these at a time because if liquid gets into the tube it can damage the device! After using the solution, wait a few minutes, remove the mucus running out with a tissue. Only after this should the nasal aspirator be used. Nose drops and other medicinal products suggested by physicians should be used after the aspiration!

The suggested length of the aspiration is 10 seconds/nasal opening.

If you have to interrupt the aspiration, switch off the device. After the motor completely shuts down, you can restart it. The device switches off after 15 minutes of continuous use but it can be restarted. In case of overuse or if not providing sufficient ventilation, the thermal switch switches off the device and will prevent re-use for approx. 50 minutes for safety reasons. Do not use the nasal aspirator for nasal washing!

  • The Nosiboo Pro nasal aspirator can be used several times a day. It is recommended to clear the airways before sleeping time and feeding to ensure the peaceful breathing of the child.
  • Do not cause pain to the child; introduce the device and the method in a tolerant way.
  • Always switch off the device after use with the control button.
  • The motor will operate for a while to cool down after you have switched off the device.
  • Do not unplug the device from the mains until it completely shuts down!
  • Do not cover the vents on the back of the device while in use!
  • Unplug the device after it has completely shut off and the cooling phase is over.

Recommended power range:

  • 0.04 – 0.08 bar. Use the lower values of the range in case of infants and the upper values in case of 4-year-old or older children.
  • The suction power is highly influenced by the thickness of the mucus and the extent of illness.

Troubleshooting malfunctions

If you suspect malfunctioning of the Nosiboo Pro device, please read the next two chapters. As the first step, the Troubleshooting guide provides suggestions for resolving concrete problems. In the chart, you can find the problems and the troubleshooting sug-gestions. If these suggestions do not solve the problem, the Malfunction management guide helps you/your merchant/distributor in localizing the malfunction as soon as possible. Please study these chapters in detail in case of a malfunction. The manufacturer shall not be held responsible for the failure of products from unauthorised import sources!

Troubleshooting guide

Malfunction

  • The device cannot be switched on.
  • The device cannot be switched on.
  • The suction intensity decreased.
  • The suction intensity decreased.
  • Liquid or mucus got into the tube.
  • Liquid or mucus got into the tube.

Reason

  • The plug is not connected.
  • The device switched off due to overuse for security reasons.
  • The Colibri head got blocked or was incorrectly assembled.
  • The air mesh wore out.
  • The Colibri head was not held in the cor-rect position.
  • The connection point of the tube did not point downwards.
  • Too much saline solution was used.

Recommendation

  • Connect the power cord to the device then to the 230 V socket.
  • For the sake of your child’s health, stop using the nasal aspirator.
  • The device can be used again in appr. 50 minutes.
  • Take it apart and clean it. After drying it, assemble it based on the user manual.
  • Replace the air mesh based on the user manual’s „Maintenance and storage” section.
  • Switch off the device and clean and dry the Colibri head and the tube. Check the air mesh and replace it if necessary.
  • Reduce the volume of the saline solution to less than 5 ml. Switch off the device and clean it, then dry the Colibri head and the tube.
  • Check the air mesh and replace it if necessary.

If the device does not work properly in spite of performing the instructions of the troubleshooting guide, check the Malfunction management guide with the distributor/merchant.
Malfunction management guide
The guarantee is valid for 24 months after the purchase date. It will be only validated if the device was used according to its intended purpose and does not bear signs of damage. In the next step the existence of malfunctioning has to be cleared. And if the malfunc-tion occurred due to the user’s own fault or if it was no fault of his own. The manufacturer provides 24 months warranty in case of use according to the intended purpose. The following checks can be conducted by any user at home or with the distributing merchant or their trustee after presenting the filled in warranty card or the receipt. The distributor or the legal entity charged with failure assessment is obliged to fill in the warranty card supplied with the User manual or provide their own warranty document, record the steps of trou-bleshooting in detail and forward it to the manufacturer within 24 hours.

Troubleshooting Visual inspection

  • The entireness of the label on the bottom of the device.
  • Other signs of damage on the housing, scratches, bumps, bruises, strains, scratches around the screws.
  • Traces of contamination on or around the air mesh. Damage of the opening behind the air mesh, grit in the device or traces indicating that grit has been drawn into the device.
  • Any kind of contamination of the tube

Manual inspection

  • Before operating the device, check the connection of the power cord on the device.
  • Any kind of failure on the control button.

Powering up/turning on

  • Powering up the device happens with the power cord supplied with the device.
  • Powering up the device is possible but after turning on the device irregularity occurs.

Detailed description

  • Signs of damage on the label indicate opening of the device housing.
  • Signs of damage on the device housing indicate opening of the device housing.
  • The excessive contamination of the air mesh, dirt getting into the device indicate the misuse of the device.
  • Contamination getting into the tube indicates misuse of the device.
  • If the connection (not the power cord) on the device is caved in, can be moved, can be removed, turning on and operating the device is life-threatening! Such malfunctions are due to external impact and are not covered by the warranty.
  • If the control button on top of the device is damaged, jammed, moves freely, or fell off, the malfunction is due to external impact and is not covered by the warranty. Turning on and operating the device is life-threatening!
  • If the device does not start to operate under these conditions, the replacement of the power cord is suggested. Such malfunctions are due to external impacts and are not covered by the warranty.
  • If you experience “unusual” phenomena while operating the device, e.g. low suction power, smoke emission, strong smell, the device buzzes or clacks unusually loud, it shakes, it shorted out, cannot be adjusted via the control button and does not turn off after 15 minutes.

Warranty tasks

  • Due to the misuse of the device, the warranty is not valid.
  • Due to the misuse of the device, the warranty is not valid.
  • Due to the misuse of the device, the warranty is not valid.
  • Due to the misuse of the device, the warranty is not valid.
  • Due to the misuse of the device, the warranty is not valid.
  • Due to misuse of the device, the warranty is not valid
  • Due to the misuse of the device, the warranty is not valid.
  • If you experience “unusual” phenomena while operating the device and the device does not bear traces of physical damage included in the first two points, the customer’s device is eligible for a warranty.

Warranty tasks
The distributor of the Nosiboo Pro device is obliged to promptly handle customer complaints based on the aspects included in the chapters Troubleshooting guide and Malfunction management guide of the User manual. The distributor of the device is obliged to keep a written record about the troubleshooting. This can be done within the framework of the distributor’s internal documen-tations or in the warranty card supplied with the User manual. It has to contain the detected malfunction, the date and the unique serial number to be found on the label on the bottom of the device.

Specialist service station contact details

  • Attract Kft.
  • H-7622 Pécs, Siklósi út 1/1 – Hungary
  • Tel.: +36 72 551 642 (Mo-Fr: 8 am – 4 pm)
  • E-mail:customerservice@nosiboo.com

Maintenance and storage

Only after the cooling process has stopped should the power cord be unplugged from the mains and from the device! Disconnect the tube from the device and the Colibri head. There is an air mesh behind the connecting unit which prevents solid objects and dust from entering the device. In the event of an unexpected drop in performance, it is recommended that you check the air mesh. If the air mesh is dirty or clogged, replace it with a spare air mesh. (A list of spare parts can be found in the “Accessories, spare parts” section.)
Figure 1 – Replacing the air mesh

  • Clean the housing with a dry, soft cloth; do not use any detergents or chemicals!
  • Keep the device in a closed space where it is protected from dust and damage (e.g. in a cabinet).
  • If you have more children, we recommended to use a separate Colibri head for each child and to store the Colibri heads separated.
  • Do not store near a heat source!

Cleaning

  1. Remove the tube and the nose tip, then take apart the head housing.

  2. Wash the collared tube, the nose tip, and the head housing in lukewarm water and dishwashing detergent and let them dry.
    Figure 4 – Cleaning of the Colibri head.

  3. Clean the device housing with a dry, soft cloth; do not use any liquids, detergents or chemicals.

  4. Wipe the tube with a cloth if necessary. Don’t wash the tube.

Important information about reducing electromagnetic interferences

The home wireless network equipment, mobile phones and wireless phones may affect the operation of medical equipment, as electro-magnetic interference can occur between devices. Electromagnetic interference may cause harmful interference in the operation of medical devices and therefore emergency situations may occur. Due to the above mentioned reasons, medical devices must not interfere with any other equipment. To guarantee the electromagnetic compatibility, the necessary requirements were fixed in standards. The standards define the protection level of the medical device and the maximum value of the electromagnetic radiation emitted by these. The Nosiboo Pro nasal aspirator complies with the requirements of EN 60601-1-2:2007. In spite of that, you should follow the instructions below while using the device:

  • Do not use your cell phone or wireless phone or any other device emitting strong electromagnetic radiation near the device!Otherwise, the device may malfunction which can cause an emergency.
  • It is recommended to keep at least 5 m distance between the devices! Check the proper operation of the device if this distance is shorter!

Environment

Do not dispose of the device at the end of the product life with household waste because electronic products qualify as hazardous waste. This is indicated by the following sign on the product packaging: The Colibri head does not present a serious risk to the environment so it can be disposed of as household waste.

Specifications

  • Adjustable motor performance: 200-390 W
    Differential pressure: 0.04-0.08 bar

  • Max. airflow: 54 l/min

  • Power supply: 230 V, 50-60 Hz

  • IP class: IP21

  • Protection class: II

  • Type of applied parts: BF

  • Applied parts: the complete housing

  • Operation type: no continuous operation

  • Operation length: 15 min/appr. off 50 min

  • Duration of continuous use: max 15 min.

  • Tripping temperature of the thermal switch: 95 °C

Accessories, spare parts

  • Parts can be ordered from the Specialist service station: Attract Kft.
  • H-7622 Pécs, Siklósi út 1/1 – Hungary
  • Tel.: +36 72 551 642 (Mo-Fr: 8 am – 4 pm)
  • E-mail:customerservice@nosiboo.com
  • Average housing temperature: 30-35 °C after 1 min continuous use Noise level: max. 70 dB
  • Temperature range: -10 °C… +40 °C
  • Relative humidity: max. 75%
  • The expected life of the housing (electrical parts): 3 years
  • The expected life of the Colibri head: 3 months
  • Diameter of the nasal insertion end of the Colibri head: min. 5.6 mm Length of the nasal insertion end of the Colibri head: min. 32 mm Weight (housing): 972 g
  • Dimensions (housing, W x H x D): 200 x 155 x 175 mm
  • Dimensions (complete package): 180 x 180 x 225 mm
  • Notebook Pro Accessory Set:
  • Accessory pack consisting of Nosiboo Pro spare parts
  • ACS-01-01; ACS-01-02; ACS-01-03
  • https://shop.nosiboo.eu/

Accessories, spare parts

Notebook Pro Accessory Set: Accessory pack consisting of Nosiboo Pro spare parts https://shop.nosiboo.eu/

nosiboo-Pro-Electric-Nasal-Aspirator-FIG-4

International warranty card

The guarantee is valid for 24 months after the purchase date. It will be only validated if the device was used according to its intended purpose and does not bear signs of damage. Warranty is validated at the authorised distributor of the device. (The authentic distributor invoice has to be presented.) The distributor of the Nosiboo Pro device is obliged to promptly handle customer complaints based on the aspects included in chapters Troubleshooting guide and Malfunction management guide of the User manual. The distributor of the device is obliged to keep a written record about the troubleshooting. This can be done within the framework of the distributor’s internal documen-tation or in the warranty card supplied with the User manual. It has to contain the perceived malfunction, the date and the unique serial number to be found on the label on the bottom of the device.

To be completed by the distributor

  • Name and address of distributor:
  • Serial number of the Nosiboo Pro nasal aspirator: (to be found on the bottom of the device, on the label)
  • Purchase date:
  • Type of malfunction:
  • Date of malfunction:
  • Countersignature of distributor: (date, signature, and seal)
    • To be completed by the service
  • Date of receiving:
  • Serviced parts:
  • New serial number:
  • Date of return:
  • Countersignature and seal of service:
  • The warranty period for the replaced parts is 24 months.

To be completed by the distributor

  • Name and address of distributor:
  • Serial number of the Nosiboo Pro nasal aspirator: (to be found on the bottom of the device, on the label)
  • Purchase date:
  • Type of malfunction:
  • Date of malfunction:
  • Countersignature of distributor: (date, signature and seal)

To be completed by the service

  • Date of receiving:
  • Serviced parts:
  • New serial number:
  • Date of returning:
  • Countersignature and seal of service:
  • Warranty period of the replaced parts is 24 months.

The expressed warranty does not affect the consumer’s implied warranty rights!
Nosiboo Pro is a hygienic medical device. you may only exercise your right of withdrawal if the device is undamaged and the protective foil on the nose tip is intact. Due to hygienic reasons, we do not replace the device if the packaging has been opened. Guiding and ap-plicable legislation: Directive 2011/83/EU § 16 (e) of the European Parliament and of the Commission, Austrian Republic: FAGG: § 18 (1) 5, Hungary: 45/2014. (II. 26.) Government decree § 29 (1) (e) Manufacturer and specialist service station: Attract Kft. Hungary, H-7622 Pécs, Siklósi út 1/1 Tel: +36 72 551 642 (Mo-Fr: 8 am – 4 pm)
E-mail: customerservice@nosiboo.com: www.nosiboo.eu

Documents / Resources

| nosiboo Pro Electric Nasal Aspirator [pdf] User Manual
Pro Electric Nasal Aspirator, Pro, Electric Nasal Aspirator, Nasal Aspirator, Aspirator
---|---
| nosiboo Pro Electric Nasal Aspirator [pdf] User Guide
Pro Electric Nasal Aspirator, Pro, Electric Nasal Aspirator, Nasal Aspirator, Aspirator

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals