APOGEE SQ-202X Original Quantum Sensor
CERTIFICATE OF COMPLIANCE
EU Declaration of Conformity
This declaration of conformity is issued under the sole responsibility of the
manufacturer: Apogee Instruments, Inc. 721 W 1800 N Logan, Utah 84321 USA for
the following product(s):
Models: SQ-204X
Type: Quantum Sensor
The object of the declaration described above is in conformity with the
relevant Union harmonization legislation:
- 2014/30/EU Electromagnetic Compatibility (EMC) Directive
- 2011/65/EU Restriction of Hazardous Substances (RoHS 2) Directive
- 2015/863/EU Amending Annex II to Directive 2011/65/EU (RoHS 3)
Standards referenced during compliance assessment:
- EN 61326-1:2013 Electrical equipment for measurement, control and laboratory use – EMC requirements
- EN 50581:2012 Technical documentation for the assessment of electrical and electronic products with respect to the restriction of hazardous substances
Please be advised that based on the information available to us from our raw
material suppliers, the products manufactured by us do not contain, as
intentional additives, any of the restricted materials including lead (see
note below), mercury, cadmium, hexavalent chromium, polybrominated biphenyls
(PBB), polybrominated biphenyls (PBDE), bis(2-Ethylhexyl) phthalate (DEHP),
butyl benzyl phthalate (BBP), dibutyl phthalate (DBP), and di isobutyl
phthalate (DIBP). However, please note that articles containing greater than
0.1% lead concentration are RoHS 3 compliant using exemption 6c.
Further note that Apogee Instruments does not specifically run any analysis on
our raw materials or end products for the presence of these substances, but
rely on the information provided to us by our material suppliers.
Signed for and on behalf of:
Apogee Instruments, June 2021
Bruce Bugbee President
Apogee Instruments, Inc.
INTRODUCTION
Radiation that drives photosynthesis is called photosynthetically active
radiation (PAR) and is typically defined as total radiation across a range of
400 to 700 nm. PAR is often expressed as photosynthetic photon flux density
(PPFD): photon flux in units of micromoles per square meter per second
(µmol m-2 s-1, equal to microEinsteins per square meter per second) summed
from 400 to 700 nm (total number of photons from 400 to 700 nm). While
Einsteins and micromoles are equal (one Einstein = one mole of photons), the
Einstein is not an SI unit, so expressing PPFD as µmol m-2 s-1 is preferred.
The acronym PPF is also widely used and refers to the photosynthetic photon
flux. The acronyms PPF and PPFD refer to the same variable. The two terms have
co-evolved because there is not a universal definition of the term “flux”.
Some physicists define flux as per unit area per unit time. Others define flux
only as per unit time. We have used PPFD in this manual because we feel that
it is better to be more complete and possibly redundant. Sensors that measure
PPFD are often called quantum sensors due to the quantized nature of
radiation. A quantum refers to the minimum quantity of radiation, one photon,
involved in physical interactions (e.g., absorption by photosynthetic
pigments). In other words, one photon is a single quantum of radiation.
Typical applications of quantum sensors include incoming PPFD measurement over
plant canopies in outdoor environments or in greenhouses and growth chambers,
and reflected or under-canopy (transmitted) PPFD measurement in the same
environments. Apogee Instruments SQ-100X series quantum sensors consist of a
cast acrylic diffuser (filter), interference filter, photodiode, and signal
processing circuitry mounted in an anodized aluminum housing, and a cable to
connect the sensor to a measurement device. Sensors are potted solid with no
internal air space, and are designed for continuous PPFD measurement in indoor
or outdoor environments. SQ-100X series sensors output an analog voltage that
is directly proportional to PPFD. The voltage signal from the sensor is
directly proportional to radiation incident on a planar surface (does not have
to be horizontal), where the radiation emanates from all angles of a
hemisphere. . SQ-100X series sensors output an analog
voltage that is directly proportional to PPFD. The voltage signal from the
sensor is directly proportional to the radiation incident on a planar surface
(does not have to be horizontal), where the radiation emanates from all angles
of a hemisphere.
SENSOR MODELS
This manual covers the amplified output quantum sensor model SQ-204X (listed in bold below). Additional models are covered in their respective manuals.
Sensor model number and serial number are located on the bottom of the sensor. If you need the manufacturing date of your sensor, please contact Apogee Instruments with the serial number of your sensor.
SPECIFICATIONS
Calibration Traceability
Apogee SQX series quantum sensors are calibrated through side-by-side
comparison to the mean of transfer standard quantum sensors under a reference
lamp. The reference quantum sensors are recalibrated with a 200 W quartz
halogen lamp traceable to the National Institute of Standards and Technology
(NIST).
Spectral Response Cosine Response
Directional (cosine) response is defined as the measurement error at a
specific angle of radiation incidence. Error for Apogee SQ-100x series quantum
sensors is approximately ± 2 % and ± 5 % at solar zenith angles of 45° and
75°, respectively.
Mean cosine response of five SQ-100X series quantum sensors. Cosine response measurements were made by direct side-by-side comparison to the mean of seven reference SQ-500 quantum sensors.
DEPLOYMENT AND INSTALLATION
Mount the sensor to a solid surface with the nylon mounting screw provided. To accurately measure PPFD incident on a horizontal surface, the sensor must be level. An Apogee Instruments model AL-100 Leveling Plate is recommended to level the sensor when used on a flat surface or being mounted to surfaces such as wood. To facilitate mounting on a mast or pipe, the Apogee Instruments model AL-120 Solar Mounting Bracket with Leveling Plate is recommended.
To minimize azimuth error, the sensor should be mounted with the cable pointing toward true north in the northern hemisphere or true south in the southern hemisphere. Azimuth error is typically less than 1 %, but it is easy to minimize by proper cable orientation.
In addition to orienting the cable to point toward the nearest pole, the sensor should also be mounted such that obstructions (e.g., weather station tripod/tower or other instrumentation) do not shade the sensor. Once mounted, the blue cap should be removed from the sensor. The blue cap can be used as a protective covering for the sensor when it is not in use.
CABLE CONNECTORS
Apogee sensors offer cable connectors to simplify the process of removing sensors from weather stations for calibration (the entire cable does not have to be removed from the station and shipped with the sensor). The ruggedized M8 connectors are rated IP68, made of corrosion-resistant marine-grade stainless steel, and designed for extended use in harsh environmental conditions.
Instructions
Pins and Wiring Colors: All Apogee connectors have six pins, but not all pins are used for every sensor. There may also be unused wire colors inside the cable. To simplify datalogger connection, we remove the unused pigtail lead colors at the datalogger end of the cable. If a replacement cable is required, please contact Apogee directly to ensure ordering the proper pigtail configuration.
Alignment: When reconnecting a sensor, arrows on the connector jacket and
an aligning notch ensure proper orientation.
Disconnection for extended periods: When disconnecting the sensor for an
extended period of time from a station, protect the remaining half of the
connector still on the station from water and dirt with electrical tape or
other methods.
Tightening: Connectors are designed to be firmly finger-tightened only. There is an o-ring inside the connector that can be overly compressed if a wrench is used. Pay attention to thread alignment to avoid cross-threading. When fully tightened, 1-2 threads may still be visible.
WARNING: Do not tighten the connector by twisting the black cable or sensor head, only twist the metal connector (blue arrows).
OPERATION AND MEASUREMENT
Connect the sensor to a measurement device (meter, datalogger, controller)
capable of measuring and displaying or recording a voltage signal with a range
of 0-2.5 V (SQ-202X) or 0-5 V (SQ-205X) to cover the entire range of PPFD from
the sun. In order to maximize the measurement resolution and signal-to-noise
ratio, the signal input range of the measurement device should closely match
the output range of the quantum sensors. The amplification circuit requires a
power supply of 2.5-24 VDC (SQ-202X) or 5-24 VDC (SQ-205X). NOTE: To prevent
sensor damage, DO NOT connect the sensor to a power source greater than 24
VDC.
Wiring for SQ-202X
Sensor Calibration
Amplified quantum sensor models have a standard PPFD calibration factor of
exactly: SQ-202X: 1.6 µmol m-2 s -1 per mV SQ-205X: 0.8 µmol m-2 s -1 per mV
Multiply this calibration factor by the measured mV signal to convert sensor
output to PPFD in units of µmol m-2 s -1 : Calibration Factor (0.8 µmol m-2 s
-1 per mV) * Sensor Output Signal (mV) = PPFD (µmol m-2 s -1 )
Spectral Errors
Apogee SQ-100X series sensors can measure PPFD for sunlight and electric light with a single calibration factor. However, errors occur in various light sources due to changes in spectral output. If the light source spectrum is known then errors can be estimated and used to adjust the measurements. The weighting function for PPFD is shown in the graph below, along with the spectral response of Apogee SQ-100X series quantum sensors. The closer the spectral response matches the defined PPFD spectral weighting functions, the smaller spectral errors will be. The table below provides spectral error estimates for PPFD measurements from light sources different than the calibration source. The method of Federer and Tanner (1966) was used to determine spectral errors based on the PPFD spectral weighting functions, measured sensor spectral response, and radiation source spectral outputs (measured with a spectroradiometer). This method calculates spectral error and does not consider calibration, cosine, and temperature errors. Federer, C. A., and C. B. Tanner, 1966. Sensors for measuring light available for photosynthesis. Ecology 47:654-657. McCree, K. J., 1972. The action spectrum, absorptance and quantum yield of photosynthesis in crop plants. Agricultural Meteorology 9:191-216.
Spectral Errors for PPFD and YPFD Measurements with Apogee SQ-100X Series Quantum Sensors
Quantum sensors can be a very practical means of measuring PPFD and YPFD from multiple radiation sources, but spectral errors must be considered. The spectral errors in the table above can be used as correction factors for individual radiation sources.
Underwater Measurements and Immersion Effect
When a quantum sensor that was calibrated in air is used to make underwater
measurements, the sensor reads low. This phenomenon is called the immersion
effect and happens because the refractive index of water (1.33) is greater
than air (1.00). The higher refractive index of water causes more light to be
backscattered (or reflected) out of the sensor in water than in air
(Smith,1969; Tyler and Smith,1970). As more light is reflected, less light is
transmitted through the diffuser to the detector, which causes the sensor to
read low. Without correcting for this effect, underwater measurements are only
relative, which makes it difficult to compare light in different environments.
The SQ-100X series sensors have an immersion effect correction factor of 1.15.
This correction factor should be multiplied to measurements made underwater.
Further information on underwater measurements and the immersion effect can be
found at http://www.apogeeinstruments.com/underwater-par-measurements/.
MAINTENANCE AND RECALIBRATION
Blocking of the optical path between the target and detector can cause low readings. Occasionally, accumulated materials on the diffuser of the upward- looking sensor can block the optical path in three common ways:
- Moisture or debris on the diffuser.
- Dust during periods of low rainfall.
- Salt deposit accumulation from the evaporation of sea spray or sprinkler irrigation water.
Apogee Instruments quantum sensors have a domed diffuser and housing for
improved self-cleaning from rainfall, but active cleaning may be necessary.
Dust or organic deposits are best removed using water, or window cleaner, and
a soft cloth or cotton swab. Salt deposits should be dissolved with vinegar
and removed with a cloth or cotton swab. Salt deposits cannot be removed with
solvents such as alcohol or acetone. Use only gentle pressure when cleaning
the diffuser with a cotton swab or soft cloth to avoid scratching the outer
surface. The solvent should be allowed to do the cleaning, not mechanical
force. Never use abrasive material or cleaner on the diffuser. Although Apogee
sensors are very stable, nominal calibration drift is normal for all research-
grade sensors. To ensure maximum accuracy, recalibration every two years is
recommended. Longer time periods between recalibration may be warranted
depending on tolerances. See the Apogee webpage for details regarding return
of sensors for recalibration (http://www.apogeeinstruments.com/tech-support-
recalibration-repairs/). To determine if your sensor needs recalibration, the
Clear Sky Calculator
(www.clearskycalculator.com) website
and/or smartphone app can be used to indicate the PPFD incident on a
horizontal surface at any time of day at any location in the world. It is most
accurate when used near solar noon in spring and summer months, where accuracy
over multiple clear and unpolluted days is estimated to be ± 4 % in all
climates and locations around the world. For best accuracy, the sky must be
completely clear, as reflected light from clouds causes PPFD to increase above
the value predicted by the clear sky calculator. Measured values of PPFD can
exceed values predicted by the Clear Sky Calculator due to reflection from
thin, high clouds and edges of clouds, which enhances incoming PPFD. The
influence of high clouds typically shows up as spikes above clear sky values,
not a constant offset greater than clear sky values. To determine
recalibration need, input site conditions into the calculator and compare PPFD
measurements to calculated values for a clear sky. If sensor PPFD measurements
over multiple days near solar noon are consistently different than calculated
values (by more than 6 %), the sensor should be cleaned and re-leveled. If
measurements are still different after a second test, email
calibration@apogeeinstruments.com to
discuss test results and possible return of sensor(s).
Clear Sky Calculator for quantum sensors. Site data are input in blue cells in middle of page and an estimate of PPFD is returned on right-hand side of page.
TROUBLESHOOTING AND CUSTOMER SUPPORT
Independent Verification of Functionality
Apogee SQ-200X series quantum sensors provide an amplified voltage output that
is proportional to incident PPFD. A quick and easy check of sensor
functionality can be determined using a DC power supply and a voltmeter. Power
the sensor with a DC voltage by connecting the positive voltage signal to the
red wire from the sensor and the negative (or common) to the black wire from
the sensor. Use the voltmeter to measure across the white wire (output signal)
and black wire. Direct the sensor head toward a light source and verify the
sensor provides a signal. Increase and decrease the distance from the sensor
head to the light source to verify that the signal changes proportionally
(decreasing signal with increasing distance and increasing signal with
decreasing distance). Blocking all radiation from the sensor should force the
sensor signal to zero.
Compatible Measurement Devices (Dataloggers/Controllers/Meters)
SQ-200X series quantum sensors are calibrated with a standard calibration
factor of 1.6 µmol m-2 s -1 per mV (SQ212X) or 0.8 µmol m-2 s -1 per mV (SQ-
215X), yielding a sensitivity of 0.6 mV per µmol m-2 s -1 (SQ-212X) or 1.25 mV
per µmol m-2 s -1 (SQ-215X). Thus, a compatible measurement device (e.g.,
datalogger or controller) should have a resolution of at least 0.6 mV to
provide PPFD resolution of 1 µmol m-2 s -1 .
Effect of Cable Length
When the sensor is connected to a measurement device with high input
impedance, sensor output signals are not changed by shortening the cable or
splicing on an additional cable in the field. Tests have shown that if the
input impedance of the measurements device is greater than 1 mega-ohm there is
negligible effect on the calibration, even after adding up to 100 m of cable.
All Apogee sensors use shielded, twisted-pair cable to minimize
electromagnetic interference. For best measurements, the shield wire must be
connected to an earth ground. This is particularly important when using the
sensor with long lead lengths in electromagnetically noisy environments.
Modifying Cable Length
See Apogee webpage for details on how to extend sensor cable length:
(http://www.apogeeinstruments.com/how-to-make-a-weatherproof-cable-
splice/).
Unit Conversion Charts
Apogee SQX series quantum sensors are calibrated to measure PPFD in units of
µmol m-2 s -1. Units other than photon flux density (e.g., energy-flux
density, illuminance) may be required for certain applications. It is possible
to convert the PPFD value from a quantum sensor to other units, but it
requires spectral output of the radiation source of interest. Conversion
factors for common radiation sources can be found on the Unit Conversions page
in the Support Center on the Apogee website (http://www.apogeeinstruments.com
/unit-conversions/; scroll down to Quantum Sensors section). A spreadsheet to
convert PPFD to energy flux density or illuminance is also provided in the
Unit Conversions page in the Support Center on the Apogee website
(http://www.apogeeinstruments.com/content/PPFD-to-Illuminance-
Calculator.xls)
RETURN AND WARRANTY POLICY
RETURN POLICY
Apogee Instruments will accept returns within 30 days of purchase as long as
the product is in new condition (to be determined by Apogee). Returns are
subject to a 10 % restocking fee.
WARRANTY POLICY
-
What is Covered
All products manufactured by Apogee Instruments are warranted to be free from defects in materials and craftsmanship for a period of four (4) years from the date of shipment from our factory. To be considered for warranty coverage an item must be evaluated by Apogee. Products not manufactured by Apogee (spectroradiometers, chlorophyll content meters, EE08-SS probes) are covered for a period of one (1) year. -
What is Not Covered
The customer is responsible for all costs associated with the removal, reinstallation, and shipping of suspected warranty items to our factory. The warranty does not cover equipment that has been damaged due to the following conditions:
- Improper installation or abuse.
- Operation of the instrument outside of its specified operating range.
- Natural occurrences such as lightning, fire, etc.
- Unauthorized modification.
- Improper or unauthorized repair.
Please note that nominal accuracy drift is normal over time. Routine recalibration of sensors/meters is considered part of proper maintenance and is not covered under warranty.
Who is Covered
This warranty covers the original purchaser of the product or other party who
may own it during the warranty period.
What Apogee Will Do
At no charge Apogee will:
- Either repair or replace (at our discretion) the item under warranty.
- Ship the item back to the customer by the carrier of our choice.
Different or expedited shipping methods will be at the customer’s expense.
How To Return An Item
-
Please do not send any products back to Apogee Instruments until you have received a Return Merchandise Authorization (RMA) number from our technical support department by submitting an online RMA form at www.apogeeinstruments.com/tech-support-recalibration-repairs/. We will use your RMA number for tracking of the service item. Call 435-245-8012 or email techsupport@apogeeinstruments.com with questions.
-
For warranty evaluations, send all RMA sensors and meters back in the following condition: Clean the sensor’s exterior and cord. Do not modify the sensors or wires, including splicing, cutting wire leads, etc. If a connector has been attached to the cable end, please include the mating connector – otherwise, the sensor connector will be removed in order to complete the repair/recalibration. Note: When sending back sensors for routine calibration that have Apogee’s standard stainless-steel connectors, you only need to send the sensor with the 30 cm section of cable and one-half of the connector. We have mating connectors at our factory that can be used for calibrating the sensor.
-
Please write the RMA number on the outside of the shipping container.
-
Return the item with freight pre-paid and fully insured to our factory address shown below. We are not responsible for any costs associated with the transportation of products across international borders.
Apogee Instruments, Inc.
721West 1800 North Logan, UT
84321, USA -
Upon receipt, Apogee Instruments will determine the cause of failure. If the product is found to be defective in terms of operation to the published specifications due to a failure of product materials or craftsmanship, Apogee Instruments will repair or replace the items free of charge. If it is determined that your product is not covered under warranty, you will be informed and given an estimated repair/replacement cost.
PRODUCTS BEYOND THE WARRANTY PERIOD
For issues with sensors beyond the warranty period, please contact Apogee at
techsupport@apogeeinstruments.com to discuss repair or replacement options.
OTHER TERMS
The available remedy of defects under this warranty is for the repair or
replacement of the original product, and Apogee Instruments is not responsible
for any direct, indirect, incidental, or consequential damages, including but
not limited to loss of income, loss of revenue, loss of profit, loss of data,
loss of wages, loss of time, loss of sales, accruement of debts or expenses,
injury to personal property, or injury to any person or any other type of
damage or loss. This limited warranty and any disputes arising out of or in
connection with this limited warranty (“Disputes”) shall be governed by the
laws of the State of Utah, USA, excluding conflicts of law principles and
excluding the Convention for the International Sale of Goods. The courts
located in the State of Utah, USA, shall have exclusive jurisdiction over any
Disputes. This limited warranty gives you specific legal rights, and you may
also have other rights, which vary from state to state and jurisdiction to
jurisdiction, and which shall not be affected by this limited warranty. This
warranty extends only to you and cannot by transferred or assigned. If any
provision of this limited warranty is unlawful, void or unenforceable, that
provision shall be deemed severable and shall not affect any remaining
provisions. In case of any inconsistency between the English and other
versions of this limited warranty, the English version shall prevail. This
warranty cannot be changed, assumed, or amended by any other person or
agreement.
APOGEE INSTRUMENTS, INC. | 721 WEST 1800 NORTH, LOGAN, UTAH 84321, USA TEL:
435-792-4700 | FAX:
435-787-8268 | WEB:
APOGEEINSTRUMENTS.COM
Copyright © 2021 Apogee Instruments, Inc.
References
- How to Make a Weatherproof Cable Splice
- Recalibration and Repair | Apogee Instruments
- Underwater PAR Measurements | Apogee Instruments
- Unit Conversions | Apogee Instruments
- Clear Sky Calculator | Apogee Instruments Inc.
- Clear Sky Calculator | Apogee Instruments Inc.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>