dynamic BIOSENSORS AS-2-Rc Fully Automated Laboratory Analysis System User Manual
- September 26, 2024
- dynamic BIOSENSORS
Table of Contents
dynamic BIOSENSORS AS-2-Rc Fully Automated Laboratory Analysis System
Specifications
- Product Name : heliX+ ADAPTER STRAND 2 with red dye Rc
- Manufacturer : Dynamic Biosensors GmbH & Inc.
- Model : AS-2-Rc v5.1
- Key Features:
- 2 spots with 2 different anchor sequences for DNA-encoded addressing.
- Order Number : AS-2-Rc
- Concentration: 400 nM
- Storage: White cap, please see expiry date on label. Avoid many freeze-thaw cycles by aliquoting the nanolever.
Product Usage Instructions
Preparation | MIX &RUN
- Mix Adapter strand 1 – Rc (400 nM) and conjugated Ligand strand with ligand 1 (500 nM) at 1:1 ratio (v/v).
- Mix Adapter strand 2 – Rc (400 nM) and conjugated Ligand strand with ligand 2 (500 nM) at 1:1 ratio (v/v).
- Incubate separately the two solutions of step 1 and 2 at room temperature at 600 rpm for 30 min to ensure complete hybridization.
- Mix solutions of step 1 and 2 at 1:1 ratio (v/v). Solution is ready for biochip functionalization.
Key Features
- Adapter strand 2 for functionalization of heliX® Adapter Chip Spot 2.
- Compatible with heliX® Adapter Chip.
- Includes Adapter strands for 50 regenerations.
- Ideal for MIX&RUN sample preparation.
- Adapter strand 2 carries a hydrophobic red dye (Rc) with a neutral net charge.
heliX® Adapter Chip Overview
2 spots with 2 different anchor sequences for DNA-encoded addressing.
Product Description
Order Number : AS-2-Rc
Table 1. Contents and Storage Information
- For research use only.
- This product has a limited shelf life, please see expiry date on label.
- To avoid many freeze thaw cycles please aliquot the nanolever
Preparation | MIX&RUN
In-solution hybridization of adapter and ligand strands:
- Mix Adapter strand 1 – Rc (400 nM) and conjugated Ligand strand with ligand 1 (500 nM) at 1:1 ratio (v/v).
- Mix Adapter strand 2 – Rc (400 nM) and conjugated Ligand strand with ligand 2 (500 nM) at 1:1 ratio (v/v).
- Incubate separately the two solutions of step 1 and 2 at RT at 600 rpm for 30 min to ensure complete hybridization.
- Mix solution of step 1 and 2 at 1:1 ratio (v/v). Solution is ready to use for biochip functionalization. Stability of the solution is related to the stability of the ligand molecules.
Table 2. Additional material for functionalization of spot 1 and reference spot 2.
Example
Required volume for 3 functionalizations: 100 μL with a final concentration of
100 nM.
After incubation time, mix vial 1 and vial 2 to obtain 100 μL of ready-to-use DNA solution.
Contact
Dynamic Biosensors GmbH Perchtinger Str. 8/10 81379 Munich Germany
Dynamic Biosensors, Inc.
300 Trade Center, Suite 1400 Woburn, MA 01801 USA
- Order Information order@dynamic-biosensors.com
- Technical Support support@dynamic-biosensors.com
- www.dynamic-biosensors.com
Instruments and chips are engineered and manufactured in Germany.
©2024 Dynamic Biosensors GmbH | Dynamic Biosensors, Inc. All rights reserved.
- TE40: 10 mM Tris, 40 mM NaCl, 0.05 % Tween20, 50 μM EDTA, 50 μM EGTA
- If the protein is not stable in PE40 (TE40, HE40), please check buffer compatibility with the switchSENSE® compatibility sheet.
FAQ
Q: What is the shelf life of the product?
A: The product has a limited shelf life, please refer to the expiry date on
the label.
Q: How should I store the product to maintain stability?
A: Store the product with a white cap to avoid exposure to light. Avoid
freeze-thaw cycles by aliquoting the nanolever.
Q: How should I prepare the solution for biochip functionalization?
A: Follow the MIX&RUN preparation steps provided in the user manual to
prepare the solution for biochip functionalization.
References