dynamic BIOSENSORS AS-2-Rc Fully Automated Laboratory Analysis System User Manual

September 26, 2024
dynamic BIOSENSORS

dynamic BIOSENSORS AS-2-Rc Fully Automated Laboratory Analysis System

Specifications

  • Product Name : heliX+ ADAPTER STRAND 2 with red dye Rc
  • Manufacturer : Dynamic Biosensors GmbH & Inc.
  • Model : AS-2-Rc v5.1
  • Key Features:
    • 2 spots with 2 different anchor sequences for DNA-encoded addressing.
  • Order Number : AS-2-Rc
  • Concentration: 400 nM
  • Storage: White cap, please see expiry date on label. Avoid many freeze-thaw cycles by aliquoting the nanolever.

Product Usage Instructions

Preparation | MIX &RUN

  1. Mix Adapter strand 1 – Rc (400 nM) and conjugated Ligand strand with ligand 1 (500 nM) at 1:1 ratio (v/v).
  2. Mix Adapter strand 2 – Rc (400 nM) and conjugated Ligand strand with ligand 2 (500 nM) at 1:1 ratio (v/v).
  3. Incubate separately the two solutions of step 1 and 2 at room temperature at 600 rpm for 30 min to ensure complete hybridization.
  4. Mix solutions of step 1 and 2 at 1:1 ratio (v/v). Solution is ready for biochip functionalization.

Key Features

  • Adapter strand 2 for functionalization of heliX® Adapter Chip Spot 2.
  • Compatible with heliX® Adapter Chip.
  • Includes Adapter strands for 50 regenerations.
  • Ideal for MIX&RUN sample preparation.
  • Adapter strand 2 carries a hydrophobic red dye (Rc) with a neutral net charge.

heliX® Adapter Chip Overview

2 spots with 2 different anchor sequences for DNA-encoded addressing.

Product Description

Order Number : AS-2-Rc

Table 1. Contents and Storage Information

  • For research use only.
  • This product has a limited shelf life, please see expiry date on label.
  • To avoid many freeze thaw cycles please aliquot the nanolever

Preparation | MIX&RUN

In-solution hybridization of adapter and ligand strands:

  1. Mix Adapter strand 1 – Rc (400 nM) and conjugated Ligand strand with ligand 1 (500 nM) at 1:1 ratio (v/v).
  2. Mix Adapter strand 2 – Rc (400 nM) and conjugated Ligand strand with ligand 2 (500 nM) at 1:1 ratio (v/v).
  3. Incubate separately the two solutions of step 1 and 2 at RT at 600 rpm for 30 min to ensure complete hybridization.
  4. Mix solution of step 1 and 2 at 1:1 ratio (v/v). Solution is ready to use for biochip functionalization. Stability of the solution is related to the stability of the ligand molecules.

Table 2. Additional material for functionalization of spot 1 and reference spot 2.

Example
Required volume for 3 functionalizations: 100 μL with a final concentration of 100 nM.

After incubation time, mix vial 1 and vial 2 to obtain 100 μL of ready-to-use DNA solution.

Contact

Dynamic Biosensors GmbH Perchtinger Str. 8/10 81379 Munich Germany

Dynamic Biosensors, Inc.
300 Trade Center, Suite 1400 Woburn, MA 01801 USA

Instruments and chips are engineered and manufactured in Germany.

©2024 Dynamic Biosensors GmbH | Dynamic Biosensors, Inc. All rights reserved.

  1. TE40: 10 mM Tris, 40 mM NaCl, 0.05 % Tween20, 50 μM EDTA, 50 μM EGTA
  2. If the protein is not stable in PE40 (TE40, HE40), please check buffer compatibility with the switchSENSE® compatibility sheet.

www.dynamic-biosensors.com

FAQ

Q: What is the shelf life of the product?
A: The product has a limited shelf life, please refer to the expiry date on the label.

Q: How should I store the product to maintain stability?
A: Store the product with a white cap to avoid exposure to light. Avoid freeze-thaw cycles by aliquoting the nanolever.

Q: How should I prepare the solution for biochip functionalization?
A: Follow the MIX&RUN preparation steps provided in the user manual to prepare the solution for biochip functionalization.