Sonic Driver POCKET MOBILE-UFM Ultrasonic Clamp on Flow Meter Instruction Manual
- September 17, 2024
- Sonic Driver
Table of Contents
POCKET MOBILE-UFM Ultrasonic Clamp on Flow Meter
“`html
Product Information
Specifications
-
Product Name: POCKET MOBILE-UFM Ultrasonic Clamp-on
Flowmeter -
Version: 3.0
-
Manufacturer: Sonic Driver Ltd
-
Copyright: 2024
Product Usage Instructions
1. General Precautions
The UFM is a precision measuring instrument and should be
handled and operated with care.
2. App Installation
-
Open Google Play Store on your Android device.
-
Search for “Sonic Driver Flowmeter” and select the Sonic Driver
MOBILE-UFM app. -
Tap on “Install” and follow the on-screen instructions to
complete the installation.
3. Hardware Connection
-
Ensure the UFM device is turned on and its screen lock is
off. -
Connect the UFM to your phone or tablet.
-
Allow permissions when prompted, and the mobile app will
attempt to connect to the UFM automatically.
4. The User Interface
The App display includes a top bar with company logo, name,
communications globe, and settings icon. Below that, there are
three tabs: CONNECT, INSTALL, and MEASURE.
CONNECT Tab
This tab provides options for managing device connection
settings such as Device Information, USB Device, Test Pattern, Baud
Rate, Data Bits, Parity, Stop Bits, Flow Control, Slave ID, and
Connection Indicator.
INSTALL Tab
This tab allows input of Pipe Information (Type, Outer Diameter,
Wall Thickness, Inner Wall Roughness, Number of Passes), Fluid
Information (Type, Temperature), Transducer Information (Type, Flow
Minimum Cut-Off, Flow Damping), Transducer Spacing, Zero Tracking,
and Zero Calibration.
MEASURE Tab
This tab displays Signal Display, Measurements, Diagnostics
(Signal Amplitude, Signal Noise, SNR, Amplifier Gain, Delta Time
Difference, Upstream Transit Time, Delta Time Offset, ATA/ETA), and
Totaliser information.
About and Exit
“`
Made in Britain
POCKET MOBILE-UFM Ultrasonic Clamp-on Flowmeter Operating Instructions
Version 3.0
1st July 2024 Copyright Sonic Driver Ltd 2024
Contents
1.0 Introduction
2.0 General Precautions
3.0 App Installation
4.0 Hardware Connection
5.0 The User Interface
5.1 Connect Tab
5.1.1 Device Information 5.1.2 USB Device 5.1.3 Test Pattern 5.1.4 Baud Rate
5.1.5 Data Bits 5.1.6 Parity 5.1.7 Stop Bits 5.1.8 Flow Control 5.1.9 Slave ID
5.1.10 Connection Indicator
5.2 Install Tab
5.2.1 Pipe Information
5.2.1.1 Type 5.2.1.2 Pipe Outer Diameter 5.2.1.3 Pipe Wall Thickness 5.2.1.4
Pipe Inner Wall Roughness 5.2.1.5 Number of Passes
5.2.2 Fluid Information
5.2.2.1 Type 5.2.2.2 Fluid Temperature
5.2.3 Transducer Information
5.2.3.1 Type 5.2.3.2 Flow Minimum Cut-Off 5.2.3.3 Flow Damping
5.2.4 Transducer Spacing 5.2.5 Zero Tracking 5.2.6 Zero Calibration
5.3 Measure Tab
6.0 About 7.0 Exit
5.3.1 Signal Display 5.3.2 Measurements 5.3.3 Diagnostics
5.3.3.1 Signal Amplitude 5.3.3.2 Signal Noise 5.3.3.3 SNR 5.3.3.4 Amplifier
Gain 5.3.3.5 Delta Time Difference 5.3.3.6 Upstream Transit Time 5.3.3.7 Delta
Time Offset 5.3.3.8 ATA/ETA
5.3.4 Totaliser
1.0 Introduction
Congratulations on choosing the Sonic Driver POCKET MOBILE-UFM TM clamp-on
ultrasonic flowmeter, figure (1).
Figure (1) Sonic Driver POCKET MOBILE-UFM The ultrasonic flowmeter (UFM)
simply connects to your Android mobile smart phone or tablet device running
the Sonic Driver App and turns it into an accurate and reliable clamp-on
ultrasonic flowmeter The UFM makes flow velocity measurements using advanced
Digital Signal Processing (DSP) and transit time measurement techniques. Using
information about the installation, entered by the user, using the App and its
intuitive and easy to us menu driven User Interface (UI) the UFM can display;
· Flow velocity (m/s, ft/s) · Volumetric flow rate (l/min, m3/hr, GPM, GPH,
ft3/min) · Mass flow rate (kg/min) · Totaliser (l) Values are averaged and
have flow profile compensation applied.
2.0 General Precautions
The content of this manual has been carefully checked and is believed to be
accurate.
Sonic Driver Ltd assumes no responsibility for any inaccuracies that may be
contained in this manual.
In no event will Sonic Driver be liable for direct, indirect, special,
incidental or consequential damages resulting from any defect or omission in
this manual, even if we are advised of the possibility of such damages.
Sonic Driver Ltd reserves the right to make improvements to its manuals,
instructions and products at any time, without notice or obligation. The
latest revisions may be found on the company web site, www.sonic-driver.com
The UFM is a precision measuring instrument and should be handled and operated
with care;
· Before operating the UFM for the first time read the installation manual and
operating instruction fully.
· Only use the UFM in the way and for the purpose that it is intended. · Do
not subject the UFM to bumps and shocks such as caused by dropping the UFM. ·
Keep the UFM and its transducers and probes clean. · Only use the UFM within
its ambient temperature and stated level of Ingress Protection. · Avoid
excessive stress and bending of transducer cables. · Always connect the UFM to
the mobile device with the device screen unlocked, section 4.0. · Always use
the Exit screen to correctly exit the App, section 12.0, figure (13). Failure
to do
so will result in multiple instances of the App running, which will each
attempt to gain control and use of the mobile device’s single communication
port. Reliable communication between the mobile device and UFM will fail. ·
Unreliable communication between the mobile device and UFM will cause the
communication globe to flash orange, section 5.1.10. In addition Device
Information, section 5.1.1, USB Device, section 5.1.2 and Test Pattern,
section 5.1.3 will become erratic. · It is advised to turn Do Not Disturb on
to avoid background tasks interrupting the totaliser.
3.0 App installation
On your Android phone or tablet open Google Play Store, or visit Google Play
Store on a web browser, figure (2).
Figure (2) Google play Store Search or browse for the content “sonic driver
flowmeter”, figure (3).
Figure (3) Search for “sonic driver flowmeter” Scroll through the list and tap
the “Sonic Driver MOBILE-UFM” item, the top item in figure (3).
Tap Install or the items price, figure (4).
Figure (4) Install and pay Follow the on-screen instructions to complete the
transaction and get the content.
4.0 Hardware Connection
Before connecting the UFM to your phone or tablet ensure that the device is
turned on and its screen lock is off.
Connect the UFM to your phone or tablet.
The device will automatically identify the UFM and request that you allow
permissions. You may tick the Always Allow box.
The mobile App will now automatically run and attempt to connect to the UFM.
5.0 The User Interface
The App display consists of a standard top bar with company logo, name,
communications globe and Settings (3 vertical dots icon) top right Below the
top bar is a bar with 3 tabs; CONNECT, INSTALL and MEASURE, figure (5).
Figure (5) App display Settings gives About and Exit options. 5.1 Connect Tab
This tab shows useful information about the connection between the mobile
device and the UFM. 5.1.1 Device information Device information consists of;
· Company Name · Model Name. · Serial Number uniquely assigned during
manufacture. · Hardware and Software version numbers. 5.1.2 USB Device The
serial communication port on the phone or tablet to which the UFM has
connected, for example /dev/bus/usb/001/002.
5.1.3 Test Pattern Upon successful connection the test pattern 1234.0 should appear. The following settings are designed for the MOBILE-UFM. They should not be changed. The ability to change the settings is intended for future product enhancements. 5.1.4 Baud Rate Default is 115200. 5.1.5 Data Bits Default is 8. 5.1.6 Parity Default is Even. 5.1.7 Stop Bits Default is 1. 5.1.8 Flow Control Default is None. 5.1.9 Slave ID Default is 1. 5.1.10 Connection Indicator This communications indicator globe will flash green when communications are taking place correctly. If communication is interrupted, then it will flash orange. 5.2 Install Tab This tab collects all of the information necessary to fully install the UFM on a pipe, see figures (6). 5.2.1 Pipe Information This menu allows the user to change pipe settings, figure (6).
Figure (6) Pipe Information (metric) By sliding the switch top right of the
screen from “mm” to “in” the entry of values can be made in imperial units or
translated into imperial units, see figure (7).
Figure (7) Pipe Information (imperial) Other relevant values, such as sensor
spacing and fluid temperature are also converted, see figure (8).
Figure (8) Pipe Information (imperial)
5.2.1.1 Pipe Type
The user can select the pipe material from a list;
· Carbon Steel · Stainless · Copper · PVC (Default) · Cast Iron · Ductile Iron
· HDPE
The transverse speed of sound in the pipe material is read from a database
held in the UFM.
5.2.1.2 Pipe Outer Diameter
The user is prompted to enter a value for the pipe outer diameter. Allowed
values are ranged 15.0 to 500.0 mm, default 56.0 mm.
5.2.1.3 Pipe Wall Thickness
The user is prompted to enter a value for the pipe wall thickness. Allowed
values are ranged 0.5 to 100.0 mm, default 1.8 mm.
5.2.1.4 Pipe Inner Wall Roughness
The user is prompted to enter a value for the peak to trough height of the
roughness on the inside surface of the pipe. Allowed values are ranged 0.001
to 2.00 mm, default 0.010 mm.
This value is used in flow profile correction calculations. It is unlikely the
user will change this value. 5.2.1.5 Number of Passes The user is prompted to
enter the number of times the sound path crosses the pipe. Allowed values are
1 to 16 passes, the default is 4 passes. Ideally choose a number of passes
that results in a total path length in the fluid of 100 mm or greater.
· 1 pass, most common on large diameter pipes, sensors are on opposite sides
of the pipe. · 2 passes, the most commonly used method, simplest to install as
both sensors are on the
same side of the pipe. · 3 passes, used on small diameter pipes. · 4 passes,
used on the lowest diameter pipes. · 5 to 15 and 16, etc. It may be that on
small diameter pipes then the recommended transducer spacing at 16 passes is
not sufficient to allow the transducers to be coupled on the same side of the
pipe, an even number of passes as they touch. In this case it is unavoidable
to couple the transducers on opposite sides of the pipe using an odd number of
passes, for example 15 passes. 5.2.2 Fluid Information This tab allows the
user to change fluid settings, figure (9).
Figure (9) Fluid Information
5.2.2.1 Type The user can select the fluid in the pipe from a list;
· Water (Default) · Petrol · Diesel · Glycol/Water The fluid longitudinal
sound velocity, kinematic viscosity and density are read from a database held
within the UFM. 5.2.2.2 Fluid Temperature The user is prompted to enter the
temperature of the fluid in the pipe. Allowed values are ranged -20 to +150
degC, default 18 degC. Changing Fluid Temperature causes fluid longitudinal
sound velocity, kinematic viscosity and density to be recalculated. 5.2.3
Transducer Information This menu allows the user to change transducer
settings, figure (10).
Figure (10) Transducer Information 5.2.3.1 Type The user is prompted to select
the type of sensors mounted on the pipe from a list;
· DN40
· DM10 (Default) · DM20
DM sensors are Sonic Driver standard PEEK/stainless steel design. DN sensors
are Sonic Driver small pipe design.
5.2.3.2 Flow Minimum Cut-Off
If the flow velocity falls below the low flow cutoff value, the measured flow
velocity and calculated flow rate indication is driven to zero. This function
can prevent the flow meter from reading flow after a pump is shut down but
there is still circulating liquid creating movement in the pipe.
Generally, 0.025 m/s is recommended as the low flow cutoff point. The low flow
cutoff value has no relation to the measurement results once the velocity
increases over the low flow cutoff value.
The user is prompted to enter a value in m/s below which the meter reports
flow as zero. Allowed values are ranged 0.000 to 1.000 m/s, default 0.025 m/s.
Note. This absolute value is applied to both positive and negative flow as a
+/- band either side of zero.
5.2.3.3 Flow Damping
The user is prompted to enter a display damping or averaging time.
Allowed values are ranged 1 to 255 seconds, the default is 10 seconds.
The damping time can be adjusted to stabilise the flow value being displayed.
Essentially, it is a type of signal filter applying an RC time constant.
Increasing the damping increases the stability. However, the measurement
displayed can be slightly delayed due to over damping. Too much damping may
also result in no response to real time fluctuations, especially when flow
rate fluctuates wildly.
Therefore, damping should be kept at a minimum and increased just enough to
reduce the fluctuation to an acceptable degree.
5.2.4 Transducer Spacing
After entering the required parameters above, the spacing between the ends of
the 2 transducers is calculated and displayed. Check the value displayed and
space the transducers accordingly.
The Installation Manual that accompanies these Operating Instructions gives
more details regarding mounting, coupling and spacing the transducers
correctly.
5.2.5 Zero Tracking
When the measured delta time falls below a lower limit then the flow being
measured is assumed by the UFM to be zero.
The user can turn a tracking algorithm On/Off so that such a small offset
value is tracked/trended to zero.
By default zero tracking is always turned On. At extremely low flows the meter
can mistakenly identify a flow from for example a real leak as an erroneous
offset. To avoid the leak being tracked off turn zero tracking Off. 5.2.6 Zero
Calibration If the Zero calibration function is enabled by ticking this box
then a zero flow calibration will be made. It is vitally important to ensure
that there is zero flow during this procedure. When the process is complete
the user is required to untick this option. By default Set Zero calibration is
turned Off. NOTE. PIPE MUST BE FULL AND FLOW MUST BE ZERO. 5.3 Measure Tab The
Measurement tab shows a signal display graph, measurement values and
diagnostics, figures (9,10,11) The UFM calculates and displays the following
measurements and diagnostics once per second. 5.3.1 Signal Display The signal
display graph is a useful diagnostic to confirm correct transducer
installation, figure (11).
Figure (11) Signal Display
5.3.2 Measurements The UFM flow measurement values are shown below the signal
display graph.
Flow Velocity – displayed in m/s, ft/s Flow Rate Volumetric – displayed l/min,
m3/hr, GPM, GPH, ft3/min Flow Rate Mass – displayed in kg/min Flow Rate Heat –
future product enhancement Values are averaged and have flow profile
compensation applied. 5.3.3 Diagnostics Scrolling down the screen will show
diagnostic values, figure (12,13).
Figure (12,13) Diagnostics 5.3.3.1 Signal Amplitude Signal strength indicates
the detected strength of the sonic signal in decibels (dB). Signal strength is
indicated by numbers from typically -25 to +55.0. Normally, the stronger the
signal strength detected the better and more reliable the flow measurement is,
as well as the more stable the measurement value obtained. Adjust the
transducer positioning to the best position, within limits and check to ensure
that enough sonic coupling compound is applied during installation to obtain
the maximum signal strength.
The UFM normally requires signal strength over 0.0 dB to measure reliably. If
the signal strength detected is too low (is zero or negative), the transducer
installation position and the transducer mounting spacing should be adjusted
and the pipe should be re-inspected. If necessary, change the mounting method.
5.3.3.2 Signal Noise
Noise indicates the level of extraneous sonic and electrical noise being
detected in dB. Noise is indicated by numbers from typically -25.0 to +55.0.
The UFM normally requires noise strength below +10.0 dB to measure reliably.
5.3.3.3 SNR
SNR indicates the quality of the sonic signal detected. SNR is indicated by
numbers from typically 1.0 to 99.0, in dB.
1.0 represents the minimum SNR whilst 99.0 represents the maximum.
Normally, the transducer position should be adjusted and coupling compound
application should be checked until the SNR detected is as large as possible.
The UFM normally requires SNR over 12.0 dB to measure reliably.
5.3.3.4 Amplifier Gain
Gain indicates the amount of electronic gain being used by the UFM receiver
amplifier. Gain is indicated by numbers from typically 0.0 to 81.0, in dB.
0.0 represents the minimum gain whilst 81.0 represents the maximum.
Old pipes, attenuating pipes, corrosion, attenuating fluids, etc. can require
the UFM to automatically turn up its gain.
The UFM normally operates with a gain typically around 30.0 to 60.0 dB.
5.3.3.5 Delta Time Difference
Delta Time is the difference in time between the two absolute transit times
through the fluid. Delta Time is of the order of tens of nanoseconds because
the absolute upstream and downstream times are so close together in value.
These values can help indicate the accuracy and condition of the installation.
The measurement calculations in the UFM are based upon these two values.
Therefore, when transit time difference fluctuates widely, the flow and
velocities fluctuate accordingly. This is usually accompanied by a signal
strength and/or signal to noise ratio (SNR) that is too low and varying. This
may be the result of poor pipe installation conditions, inadequate transducer
installation, or incorrect parameter input. Generally, fluctuations should be
less than ± 20 %.
5.3.3.6 Upstream Transit Time The absolute upstream transit time through the
fluid in the pipe and the absolute downstream transit time through the fluid
in the pipe are usually of the order of hundreds of microseconds. They are
very nearly identical. For this reason, transit time is simply displaying the
absolute upstream transit time through the fluid in the pipe. 5.3.3.7 Delta
Time Offset dT Offset is the offset value currently being used by the zero
tracking function. This is usually in the range ± 2.5 ns. 5.3.3.8 ATA/ETA This
is a measure of the ratio of the measured transit time to that which is
expected given the parameters entered by the user during installation. It
indicates if the transducer mounting and spacing is accurate. The normal
transit time ratio should be 100 ± 3 % if the installation is correct. It is
acceptable to have to move one of the transducers up to ± 5 mm to achieve a
figure of 100 %. If more movement is necessary, then one of the pipe
parameters is probably incorrect. This is most likely to be the value entered
for pipe wall thickness as this is often taken from tables or it is an
estimated value. 5.3.4 Totaliser At the bottom of the Measure Tab it is
possible to; START, STOP and RESET a totaliser in units of litres, see figure
(14).
Figure (14) Totaliser Functionality
Since mobile devices are complex devices with many possible functions running
in the background it is advisable to turn the smart device’s “Do Not Disturb”
function on to avoid background tasks interrupting the totaliser. If such an
event should occur then a red circle with exclamation mark will appear next to
the totalised value to warn the user.
6.0 About
The About screen shows useful information about the App and a link to the
Sonic Driver website, figure (15).
Figure (15) About
7.0 Exit
Use this screen to correctly exit the App, figure (16). Failure to do so will
result in multiple instances of the App running, which will each attempt to
gain control and use of the mobile devices single communication port.
Sonic Driver
Figure (16) Exiting the App