BOSCAROL Pediatric Ob Spinal Board
Class I medical device in accordance with the European Medical Device Regulation 2017/745
IFU PEDIATRIC SPINAL BOARD_ED01_REV01-2023_EN
2 – 12
PRODUCED BY:
OSCAR BOSCAROL SRL
Via Enzo Ferrari 29
39100 Bolzano
ITALY
Tel. +39 0471 932893
Fax: +39 02 57760140
info@boscarol.it
www.boscarol.it
Information on the manufacturer and the medical device:
-
Oscar Boscarol applies a quality management system according to international standards ISO 13485 and ISO 9001
-
Medical devices in the “Immobilisation” category (in all their configurations) comply with European Regulation MDR 2017/745 and bear the CE marking in accordance with risk class I (see Annex VIII of the Regulation)
-
The medical device meets the safety and performance requirements (GSPR) described in Annex I of the European Regulation 2017/745
About these operating instructions: -
This document contains important information for a safe, effective and compliant use of the medical device
-
Use the information reported to train users and confirm their training
-
This manual may not be modified (even partially). Only the manufacturer of the device may make changes
-
These instructions must always accompany the device. We recommend using the electronic version and making it available on operators’ PDAs, tablets and mobile phones
These operating instructions apply to the following devices:
MEANING OF SYMBOLS AND PICTOGRAMS
Symbols used in these operating instructions to draw the reader’s attention
| Danger: important safety information on the correct use of the
pediatric spinal board to prevent injury to the operator or patient and/or
damage to the device itself
---|---
| Warnings: information requiring special attention
| Notes or information for proper use and to prevent damage to the device
or others. Activate the correct preventive measures
1.| List of actions to be performed: follow them step by step
| These operating instructions
|
Required maintenance service (please contact the manufacturer and/or its authorised service centres)
Symbols used on the device and accessories of the device
| Use the pediatric spinal board only within the specified temperature
range. Using the pediatric spinal board outside these limits may decrease its
functional performance and damage it
---|---
| Limits of use with regard to humidity
| Read these operating instructions carefully and completely
| Indicates the need for the user to consult these operating instructions for
information such as warnings and precautions that cannot be displayed on the
medical device in question
| CE marking in accordance with European MDR 2017/745 for class I medical
devices
| Manufacturer
| Device reference code
| Please read the operating instructions in other languages available on
the indicated website
| Indicates that the pediatric spinal board is a medical device
****| Example of UDI-DI and UDI-PI code of medical device:
(01) Identification of manufacturer and associated device
(11) Production date
(20) Product variant
(10) Lot number
INTENDED USE
Device name | Pediatric OB spinal board |
---|
__
Primary use
| Professional-type medical device intended for the immobilisation and
transport, in a horizontal position, of pediatric patients with spinal or limb
injuries
Other Uses| No uses other than the primary use are envisaged
Medical purpose| Ensures correct, total and safe immobilisation of the
spine and patient during transfer
Part of application in the human body| __
Entire human body
---|---
__
Type of patients
| Pediatric patients of both sexes, with a total weight of less than 40 kg
and height between 60 and 120 cm.
Application time on the same patient| __
Temporary’ use (maximum 60 minutes of consecutive use)
Information on use| · Follow the procedures approved by the relevant Emergency Medical Service for the immobilisation, positioning and transport of the patient.
· The device can be used simultaneously with other medical devices applied to the patient (e.g. a cervical collar)
· The transport of a patient must always be carried out by professional rescuers who are
trained and familiar with lifting and transport techniques
__
Misuse
| The manufacturer accepts no liability in the event of use of the product under the conditions considered below as improper use:
· Handling by unqualified personnel
· Transport of patients with pathologies defined as not compatible by medical personnel
· Failure of operators to comply with existing safety regulations/procedures
· Modifications or tampering not authorised by the manufacturer
· Failure to comply with what is specified in this user manual.
WARNINGS, PRECAUTIONS AND IMPORTANT INFORMATION
Read carefully
These operating instructions have been prepared using simple, easy-to- understand language. In case of difficulty in interpreting what is written, please contact the manufacturer for further clarification.
__
Phone +39 0471 93 28 93info@boscarol.it
__
LATTICE
| The pediatric OB spinal board is constructed and manufactured without the
use of latex. The materials used are latex-free, however, it cannot be ruled
out that latex or traces of it may have come into contact throughout the
production chain
DEVICE CONTAMINATED| Warning in case of device contamination:
Transporting a patient with the pediatric spinal board may be a source of
contamination. For this reason the device must be cleaned and disinfected
after each use to eliminate any residual risk. Follow the instructions in this
user manual.
In case of doubt, please contact Boscarol’s technical service department by sending an e-mail to info@boscarol.it or calling +39 0471 932893
IMPORTANT INFORMATION TO KNOW BEFORE USE
The pediatric spinal board has been designed and tested to comply with the requirements of the Medical Device Regulation 2017/745. The pediatric spinal board is a medical device of risk class I according to Annex VIII of the Regulation.
| Should the user or patient become aware of a danger in use, a side
effect, an accident caused by the device, or a critical issue (operational and
constructional) not addressed in these instructions for use, he or she must
immediately report it to the manufacturer
at:raq@boscarol.it
PERIODIC SAFETY INSPECTION| **Preventive maintenance and periodic
safety inspection:**
The pediatric spinal board must be checked periodically (a full functional check at least once a week is recommended)
There is no periodic safety inspection on the device
In case of failure, the device must be removed and replaced
---|---
__ LIFETIME| __
The pediatric OB spinal board has a lifespan of 5 years from the date of manufacture if stored and used in accordance with these instructions for use
__
__
__
Responsibility of operators/users
| | · Take precautions in case of direct contact of the device with the patient’s body and always interpose sterile drapes (biocompatibility)
· The pediatric OB spinal board is designed for emergency medical service and must therefore be ready for use at any time and in any situation
· Follow the procedures approved by the Emergency Medical Service for positioning, immobilising and transporting the patient
· Replace the device immediately if it is found to be partially damaged or deformed upon inspection. The pediatric spinal board must be stored in a place inaccessible to children
· The device must be stored without placing material and/or other heavy objects on it that could deform or damage it
· Dispose of packaging in accordance with current regulations and ensure that it is out of the reach of children
· Tampering with, alterations and modifications of the device are not permitted without the manufacturer’s consent
· Operators must be trained and knowledgeable in occupational safety rules and
regulations (use of PPE)
CONTRAINDICATIONS AND SIDE EFFECTS (POSSIBLE DURING USE)
The use of this device, if carried out as described in this user manual and by appropriately trained personnel familiar with emergency medical doctrine, has no particular other contraindications or side effects
__CONTRAINDICATIONS| · Direct contact of the patient’s skin
(injured or intact) with the device. A sterile drape must always be interposed
to prevent allergies, irritation or other phenomena related to the type of
material used in manufacture
---|---
__SIDE EFFECTS|
· Phenomena of skin irritation and reactivity due to direct patient contact with the surface of the device
· Uncontrolled patient reactions related to nervous stress
PEDIATRIC OB SPINAL BOARD
The pediatric spinal board arrives ready for use and does not need to be assembled. After receiving it, ensure its structural integrity. In particular, carefully check the efficiency of the safety belts.
TYPE OF CONSTRUCTION
The picture below shows the table and its structure.
Made entirely of plastic laminate (ABS) with radiotranslucent and impermeable
characteristics , it has 14 comfortable handles (7 handles on each long side)
around its perimeter for easy grip of the board and 2 safety belts for safe
immobilisation of the patient. In addition, the pediatric OB spinal board is
designed to accommodate the pediatric head restraint, an accessory that is
often necessary for the correct immobilisation of the patient, quickly and
safely.
A label bearing all manufacturer and product identification data, the CE marking and the UDI number, which complies with the requirements of European Regulation MDR 2017/745, is attached to the device. This must never be removed or covered.
USE OF THE DEVICE
Attention
Follow the procedures approved by the relevant Emergency Medical Service
for the immobilisation, positioning and transport of the patient.
The procedures below are drawn up on the basis of generic usage information
The user manual must be read and made available to the rescuer at all times.
For this reason, it should be kept (also in digital format on one’s
smartphone) near or with the device.
The product is intended for professional use; it is recommended that operators
have specific experience in the care sector, particularly in the use of spinal
boards, and are in a suitable physical condition, i.e. have the muscular
strength and coordination to guarantee the highest level of safety for the
patient.
In order to use it correctly, it is good to practise several times using
weights and sizes that simulate different patients.
To operate safely and efficiently, the table requires at least two qualified
and trained operators.
In all cases, the control and co-ordination of table movements is the
responsibility of a qualified operator, while other operators/additional
helpers must contribute to the movements following the instructions of the
responsible operator.
- Before picking up the patient with the spinal board, stabilise the damaged parts and apply appropriate cervical or rescue collars as additional support for the head and neck.
- While one operator holds the patient’s neck and head in a neutral position, the other operators can proceed to apply the pediatric spinal board.
- Before laying the patient down (if possible), place a sterile drape in between.
- When operators have access to both sides of the patient, use the pediatric spinal board by applying the TRONK ROTATION METHOD:
- Turn the patient on his side.
- Slide the pediatric spinal board closer to the patient.
- Rotate the patient over the pediatric spinal board.
- Secure the patient on the pediatric spinal board using the safety belts before starting to move it.
- If it is inadvisable to use the trunk rotation method, the four operators should proceed according to the VERTICAL LIFTING METHOD:
- Bring the pediatric spinal board closer to the patient.
- While one operator supports the patient’s head and neck, the others lift the patient just enough to slide the pediatric spinal board under the patient’s back.
- Secure the patient on the pediatric spinal board using the safety belts before starting to move it.
- Take all measures to avoid worsening the patient’s condition
- Once the pediatric spinal board has been lifted with the patient properly secured to it, it is possible to proceed with its transfer following orders given by the operations manager. Rescuers should be coordinated and have good physical endurance as well as general control of their strength.
Lifting and carrying the patient requires care and attention. Never overestimate your own strength and, if necessary, always request the help of other rescuers
Attention
- Always use seat belts, checking their tightness each time they are used;
- All operators must face the patient during transport;
- At least one operator must never leave the patient for as long as he or she remains on the table.
DEVICE CLEANING AND MAINTENANCE
Cleaning
After each use, the medical device must be cleaned and disinfected. These
operations assume greater importance if the patient’s pathological condition
is unknown and there may be a risk of direct contamination. The user must
always take protective measures and means to avoid contact with contaminated
devices.
Attention
- Failure to carry out cleaning operations may result in the risk of cross-infection due to the presence of secretions and/or residues.
- During all control and sanitising operations, the operator must wear appropriate personal protective equipment, such as gloves, goggles, etc.
After removing all substances from the device with lukewarm water and neutral
detergent, use a non-abrasive sponge to remove any encrustations and/or any
traces of blood and/or secretions left by the patient.
Disinfect the surface of the pediatric spinal board with products that are
suitable for this purpose and that do not have a solvent or corrosive action
on the materials making up the device. Do not use metal brushes, steel wool or
blades of any kind to remove fouling as this may damage the surface of the
device.
Rinse thoroughly with lukewarm water, making sure to remove all traces of
detergent, which could deteriorate it or compromise its integrity and
durability.
The use of high-pressure water must be avoided.
Before storing the device make sure it is completely dry.
Maintenance
Establish a programme of maintenance and periodic checks, identifying a contact person. The person entrusted with the maintenance of the device must guarantee the basic requirements laid down by the manufacturer.
Attention
During all control, maintenance and sanitisation operations, the operator
must wear appropriate personal protective equipment, such as gloves, goggles,
etc.
Routine maintenance
The checks to be carried out before and after each commissioning are as
follows:
- General functionality of the device
- State of cleanliness of the device (please note that failure to clean can lead to the risk of cross-infection)
- Absence of cuts, holes, tears or abrasions on the entire structure
- State of wear and tear
- Integrity of handles (are they torn or show signs of tearing)
The frequency of checks is determined by factors such as legal requirements, type of use, frequency of use, environmental conditions during use and storage. Please note that it is necessary to carry out the cleaning described under “Cleaning” and the functionality check before and after each use.
Periodic revision
There are no scheduled periodic overhauls at the manufacturer’s premises or at
a centre authorised by the manufacturer, but the cleaning and checks indicated
in the respective sections “Cleaning” and “Routine maintenance” are
prescribed.
STORAGE OF THE DEVICE
Attention
Periodically check the integrity of the components of the device and the
hygienic conditions of the storage place, in order to enjoy maximum efficiency
and safety when in use.
The device, before being stored, must be thoroughly checked and replaced if
tears or loss of mechanical integrity are detected.
The pediatric OB spinal board can be used and stored in the temperature range
of -10 to +50 °C. At very low or too high values the material in contact with
the patient may become extremely cold or hot causing complications for the
patient (hypothermia and hyperthermia). Take all necessary measures to contain
these effects and limit them as far as possible.
The pediatric spinal board must be stored clean and dry in a closed
environment, taking care to protect it from exposure to sunlight, dust and
dirt in order to guarantee its hygienic conditions. The device must also not
be exposed to or come into contact with sources of combustion and flammable
agents and/or substances, chemical agents, which could alter its safety
characteristics.
It is also recommended that the board be stored in a place where it can be
easily and safely reached by first-aid personnel.
DEMOLITION OF THE DEVICE
The device can be disposed of in accordance with national and local regulations for the disposal of plastic-based substances. All materials are REACH-compliant and contain no hazardous substances.
TECHNICAL SERVICE AND SPARE PARTS
The device cannot be repaired. In case of breakage or even partial damage it must be replaced with a new one. Should you require additional information to that provided in these operating instructions, please contact the manufacturer by telephone on +39 0471 932893 or by sending an email to info@boscarol.it.
TECHNICAL DATA AND COMPLIANCE FOR THE PEDIATRIC OB SPINAL BOARD
Classification of the medical device (in accordance with MDR 2017/745) | I |
---|---|
Basic UDI number (in conformity with MDR 2017/745) | 805240088IMMGK |
Technical Specifications | Device made of plastic laminate (ABS) and equipped |
with 14 carrying and lifting handles and 2 safety
belts for correct patient restraint
Risk assessment (technical documentation)| ISO 14971:2019
Lifespan| 5 years from date of manufacture
Spinal board dimensions
Pediatric OB spinal board| Dimensions: 135x35x5cm| Weight: 3.65kg
Conditions of storage and use
| Temperature range for transport, use and storage of the device| -10 to 50°
C (14 to 122 °F)
__| Humidity range accepted by the device for transport, use and
storage| 5 ÷ 95 % R.H. n.c.
__| For further technical information or information on the use and
storage of the device, please contact the
manufacturer(info@boscarol.it).
__| On all measurements the tolerance is ±5 cm
Declaration of conformity
| The declaration of conformity is kept by the manufacturer together with all
traceability data applicable to materials and production processes. You can
request a copy from the manufacturer by sending an email to:
info@boscarol.it
__| The device contains no metal parts or ferrous alloys
ACCESSORIES
Boscarol supplies the complete range of medical equipment and accessories for emergency rescue. In particular, for the use of the pediatric OB spinal board the customer can request the following accessory:
- Universal OB pediatric headband
For the specific codes of the above article, please consult www.boscarol.it or send an e-mail to: info@boscarol.it
GUARANTEE
Oscar Boscarol warrants the pediatric spinal board for a period of 1 year from
the date of purchase from the original distributor or manufacturer. The
company guarantees that the device is free of defective materials and/or
defects due to manufacturing processes.
The guarantee does not cover: normal wear and tear of the device,
discolouration and any other cosmetic irregularities that do not affect the
operation of the unit.
If, during the entire 1-year warranty period, the product is found to be
defective, it must be sent to Oscar Boscarol S.r.l. (Ltd) with a note
describing the defect. Oscar Boscarol S.r.l. (Ltd) will repair or replace the
defective parts and/or the entire unit at its discretion. All shipping costs
shall be borne by the customer.
Warranty Conditions:
To benefit from the guarantee, the registration form in the product
documentation must be completed and returned by post, fax or e-mail to the
following address:
OSCAR BOSCAROL SRL V. E. Ferrari, 29 – 39100 BOLZANO, ITALY
Fax: +39 0257760142 – E-mail: info@boscarol.it
To validate the guarantee process, the customer must provide evidence of the
following documentation:
- copy of the invoice and/or purchase receipt containing the batch number of the device and the date of purchase
- confirmation by the manufacturer or a person representing the manufacturer that it is indeed a fault due to the production process or defective components since their procurement
- absence of tampering, modifications and/or anything not conforming to the original product
In terms of the safety, reliability and functionality of the spinal board, Oscar Boscarol S.r.l. can only be held liable if:
- all planned technical operations, repairs, modifications and checks have been carried out by Oscar Boscarol S.r.l. or an authorised service centre
- the device has been and is being used correctly, strictly following the instructions given in these operating instructions
With reference to what is described in these warranty conditions, Oscar Boscarol S.r.l. cannot be held responsible for any direct or indirect accidental damage, if modifications, repairs, unauthorised technical interventions have been carried out on the device or any of its parts have been damaged by accident or improper use. There are no other expressed or limited warranties of merchantability, fitness or otherwise on the pediatric spinal board other than those described in this user guide.
Printed in Italy by Oscar Boscarol Srl (Ltd)
IFU PEDIATRIC SPINAL BOARD_ED01_REV01-2023_EN
Original language of editing: Italian
https://www.boscarol.it/ita/eifu.php
Emergency Medical Systems
Via Enzo Ferrari 29, 39100 Bolzano IT
Tel. +39 0471932893 – Fax: +39 0257760140
info@boscarol.it
www.boscarol.it