Sonic Driver POCKET-UFM Ultrasonic Flowmeter Instruction Manual
- September 16, 2024
- Sonic Driver
Table of Contents
- POCKET-UFM Ultrasonic Flowmeter
- Specifications
- Product Information
- Product Usage Instructions
- 1. Using the Quick Start Sequence
- 2. Sensor Positioning
- 3. Battery Life
- 4. Cleaning and Storage
- 5. Connecting Flow Transducers
- Q: What should I do if any item on the packing list is missing
- Q: What type of battery should I use for the UFM?
POCKET-UFM Ultrasonic Flowmeter
“`html
Specifications
- Made in: Britain
- Product: POCKET-UFM Ultrasonic Flowmeter
- Version: 5.0
- Date: 1st July 2024
- Copyright: Sonic Driver Ltd 2024
Product Information
The Sonic Driver POCKET-UFM is a clamp-on ultrasonic flowmeter
that utilizes advanced Digital Signal Processing (DSP) and transit
time measurement techniques to provide accurate and reliable flow
velocity measurements on liquids in closed pipes. The device
features an intuitive User Interface (UI) for easy operation and
display of transit time measurements.
Product Usage Instructions
1. Using the Quick Start Sequence
Follow the Quick Start Sequence outlined in the manual to set up
and start using the POCKET-UFM.
2. Sensor Positioning
Ensure proper positioning of the flow transducers on the pipe
according to the guidelines provided in the manual for accurate
measurements.
3. Battery Life
Use a high-capacity primary battery of at least 1200 mAh for
optimal performance. To change the battery, slide the cover of the
battery compartment located at the rear of the UFM.
4. Cleaning and Storage
After use, clean the UFM and sensors with a tissue or soft cloth
to remove excess coupling gel. Store the UFM and transducers in the
supplied carry case when not in use.
5. Connecting Flow Transducers
Connect the flow transducers to the Lemo connectors at the top
of the UFM as shown in figure (5) in the manual.
FAQ
Q: What should I do if any item on the packing list is missing
or damaged?
A: Contact Service as mentioned in Appendix A of the manual for
assistance.
Q: What type of battery should I use for the UFM?
A: It is recommended to use a high capacity primary battery of
at least 1200 mAh for optimal performance.
“`
Made in Britain
POCKET-UFM Ultrasonic Flowmeter Installation Manual
Version 5.0
1st July 2024 Copyright Sonic Driver Ltd 2024
Contents
1.0 Introduction
· 1.1 Transit Time Measurement · 1.2 Packing List · 1.3 General Precautions ·
1.4 Cleaning · 1.5 Storage · 1.6 Fitting/Changing Internal Battery · 1.7
Connecting The Flow Transducers · 1.8 Mounting The Flow Transducers · 1.9
Turning the UFM On
2.0 Using the Quick Start Sequence
· 2.1 Transducer Menu
· 2.1.1 Type · 2.1.2 Mounting
· 2.2 Pipe Menu
· 2.2.1 Outer Diameter · 2.2.2 Wall Thickness · 2.2.3 Material · 2.2.4 Sound
Velocity
· 2.3 Liner Menu
· 2.3.1 Material · 2.3.2 Sound Velocity · 2.3.3 Thickness
· 2.4 Fluid Menu
· 2.4.1 Type · 2.4.2 Temperature
3.0 Sensor Positioning
· 3.1 Optimising Transducer Mounting Location · 3.2 Upstream and Downstream
Pipe Runs · 3.3 Transducer Mounting · 3.4 Transducer Spacing
4.0 Error Codes
5.0 Icons
6.0 Battery Life
7.0 Specification
8.0 Product Identification
9.0 Service
10.0 Limited Warranty and Disclaimer
Appendix A Contact Details Appendix B Table of fluid properties Appendix C
Table of pipe and lining material properties Appendix D Table of speed of
sound in water Appendix E Table of typical pipe roughness values
1.0 Introduction
Congratulations on choosing the Sonic Driver TM POCKET-UFM TM clamp-on
ultrasonic flowmeter, figure (1).
Figure (1) The Sonic Driver POCKET-UFM. The ultrasonic flowmeter (UFM) uses
advanced Digital Signal Processing (DSP) and transit time measurement
techniques (Sonic Driver TM) to make accurate and reliable clamp-on ultrasonic
flow velocity measurements on liquids flowing in closed pipes. Using
information about the installation, entered by the user, using the meters
intuitive and easy to us menu driven User Interface (UI) the UFM can display;
· Flow velocity · Volumetric flow rate · Mass flow rate · Heat quantity flow
rate With the addition of a wall thickness measuring probe the UFM can also
function as a pipe wall thickness gauge (WTG).
1.1 Transit Time Measurement
The principle of flow measurement using ultrasonic clamp-on transit time
measurement is simple, see figure (2).
Figure (2) The principle of transit time flow measurement. Two ultrasonic
transducers are coupled or clamped to the outside of the pipe at a
predetermined distance apart. Ultrasonic pulses travel between the transducers
through the pipe wall and the fluid within the pipe. If the fluid is flowing
then it takes slightly longer for the ultrasound to travel against the flow
(upstream time T_up) than with the flow (downstream time T_down), see figure
(2). In a typical installation the individual times measured upstream and
downstream are just a few hundred microseconds, the difference between them is
typically measured in tens of nanoseconds. This very small time difference
(T_up – T_down) is measured by the UFM and is directly proportional to the
flow velocity (V) of the fluid. Knowing the pipe internal cross sectional area
the UFM can calculate volumetric flow rate in many common engineering units. A
further knowledge of the density of the fluid allows the UFM to calculate mass
flow rate. Finally, a knowledge of inlet and outlet fluid temperature and
Specific Heat Capacity of the fluid allows the UFM to calculate heat flow
rate. All of these rates can be totalled and positive, negative and net values
displayed.
1.2 Packing List
Figure (3) UFM Carry Case and Contents. Within the UFM carry case, see figure (3) you should find;
Item POCKET-UFM PEEK/Stainless Steel Flow Transducer Wall Thickness (WTG)
Probe Ball Chain Mounting Clamp Tape Measure Coupling Gel
Battery
Quantity 1 2
Optional 2 1 1 1
Table (1) Packing List.
If any item on the packing list is missing or has been damaged in transit
contact Service, see Appendix A.
1.3 General Precautions
The content of this manual has been carefully checked and is believed to be
accurate.
Sonic Driver Ltd assumes no responsibility for any inaccuracies that may be
contained in this manual.
In no event will Sonic Driver Ltd be liable for direct, indirect, special,
incidental or consequential damages resulting from any defect or omission in
this manual, even if we are advised of the possibility of such damages.
Sonic Driver Ltd reserves the right to make improvements to its manuals,
instructions and products at any time, without notice or obligation. The
latest revisions may be found on the company web site, see appendix A.
The UFM is a precision measuring instrument and should be handled and operated
with care;
· Before operating the UFM for the first time read the installation manual and
operating instructions fully.
· Only use the UFM in the way and for the purpose that it is intended. · Do
not subject the UFM to bumps and shocks such as caused by dropping the UFM. ·
Keep the UFM and its transducers and probes clean. · Only use the UFM within
its ambient temperature and stated level of Ingress Protection. · Avoid
excessive stress and bending of transducer cables and connectors. · Avoid
striking the clear display window and keypad with sharp objects.
1.4 Cleaning
Wipe the UFM and sensors with tissue or soft cloth after use, remove excess
coupling gel.
1.5 Storage
When not is use the UFM and transducers should ideally be cleaned and stored
in the carry case they were supplied in.
1.6 Fitting/Changing Internal Battery
The UFM is powered by a single 9 V internal battery. Sonic Driver recommend a
high capacity primary battery of at least 1200 mAh be used.
To fit or change the battery slide the cover of the battery compartment in the
rear of the UFM, see figure (4).
Figure (4) Battery compartment. Lower capacity or use of rechargeable battery
will decrease the run time of the UFM.
1.7 Connecting The Flow Transducers
Connect the flow transducers to the Lemo connectors at the top of the UFM, see
figure (5).
Figure (5) Flow transducer connection. Red flow transducer is mounted on the
pipe downstream, black flow transducer is mounted upstream. When using the WTG
probe then connector orientation is not important.
1.8 Mounting The Flow Transducers
Mount the flow transducers on the pipe using the chain clamps supplied, see
figure (6). Ensure that the arrow on the labels (arrowhead and flights) on the
flow transducers is pointing in the direction of flow. Use coupling gel
between the transducers and the pipe to give good ultrasonic contact. Measure
the spacing of the transducers using the tape measure provided, note that
spacing is measured between the front faces of the transducers. Ensure the
transducers are facing each other and aligned axially along the pipe.
Figure (6) Flow transducer mounting, spacing is 25 mm between front faces. In
figure (7) and figure (8) the transducers are misaligned and twisted, as a
result the UFM will make poor flow measurements.
Figure (7) Misaligned transducers.
Figure (8) Twisted transducers.
When using the WTG probe it is possible to simply hold the probe on the pipe
by hand, making good contact using coupling gel.
1.9 Turning the UFM On
To power on a POCKET-UFM press and hold the On key.
As soon as the UFM is switched on a self-diagnostic program will start. This
program fully tests both the UFM hardware and software.
If an error is detected an error message will be displayed prompting user
action. If the error persists contact customer support, see appendix A.
Error codes and their meanings can be found in the Main Menu under the
Diagnostics Menu.
In order to turn the UFM off from the Main Menu press and hold the “</Off”
key.
A progress bar decrements to turn the UFM off.
If the “</Off” key is released before the progress bar fully decrements the
UFM returns to the Measurement screen.
The “</Off” key has no function when in Measurement Mode.
If there is no keypad activity in User Interface (UI) mode for 60 seconds then
the UFM will automatically power off. Automatic power off does not apply in
Measurement mode.
The LCD backlight will flash when the battery is low and needs to be changed.
2.0 Using the Quick Start Sequence
Once powered on the UFM will be in UI mode displaying the Main Menu.
The Main Menu allows the user to select a group of parameters to edit or a
meter function;
· Quick Start · Installation · System Settings · Diagnostics · Load/Save
Setups
The Quick Start function takes the user through the minimum sequence of
parameters needed to get the UFM measuring reliably;
· Transducer Type · Transducer Mounting · Pipe Diameter · Pipe Wall Thickness
· Pipe Material · Liner Material · Fluid · Fluid Temperature
Each parameter is described below in the relevant section.
After selecting Transducer Type and Transducer Mounting it is also possible to
jump to the end of the sequence by pressing the Run/Stop key.
2.1 Transducer Menu
This menu allows the user to change transducer settings.
2.1.1 Type
The user is prompted to select the type of sensors mounted on the pipe from a
list;
· DS05 · DS10 · DS20 · DS40 · DM10 (Default) · DM20 · DN40 · Flow Other · WTG
DM sensors are Sonic Driver standard PEEK/stainless steel design. DN sensors
are Sonic Driver small pipe design. DS sensors are Sonic Driver large pipe
design.
If Flow Other is selected then the user will be prompted to enter detailed
transducer specific information.
The ability to select Flow Other is intended for use when using the UFM with
special sensors supplied by Sonic Driver.
Wall Thickness Gauging (WTG) is for use with dual element 5 MHz wall thickness
probes.
When operating as a wall thickness gauge the user can scroll between 3 display
modes by repeatedly pressing the “Enter” key, the modes are;
· Full display of wall thickness measurement and all measurement diagnostics.
· A-Scan showing the raw received ultrasonic signal. · Processed signal
showing the measured signal corresponding to the wall thickness.
2.1.2 Mounting
The user is prompted to select the sound path in the pipe from a list;
· Auto · Z · V · N · W · 5 · 6 (Default) · 7 · 8
· 9 · 10 · 11 · 12
Selecting Auto means that the UFM determines for itself which sound path to
use. Ideally choose a number of passes that results in a path length in the
fluid of 100 mm or greater.
· Z or 1 pass, most common on large diameter pipes, typically 100 mm or more
in diameter. If the UFM suggests a negative spacing then this is measured as
in figure (9).
Figure (9) Z or 1 pass, demonstrating a negative transducer spacing. · V or 2
passes, the most commonly used method, this is the simplest to install as both
sensors
are on the same side of the pipe, see figure (10).
Figure (10) V or 2 passes.
· N or 3 passes, used on small diameter pipes. · W or 4 passes, used on the
smallest diameter pipes, see figure (11).
Figure (11) W or 4 passes. · 5 to 11 and 12, etc. It may be that on the
smallest diameter pipes then the recommended transducer spacing at 12 passes
is not sufficient to allow the transducers to be coupled on the same side of
the pipe, using an even number of passes as they still touch. In this case it
is unavoidable to couple the transducers on opposite sides of the pipe using
an odd number of passes, for example 9 or 11 passes. Red flow transducer is
mounted on the pipe downstream, black flow transducer is mounted upstream, see
figure (5). When using the WTG probe then connector orientation is not
important.
2.2 Pipe Menu
The following parameters allow the user to specify the pipe.
2.2.1 Outer Diameter
The user is prompted to enter a value for the pipe outer diameter. Allowed
values are ranged 10.0 to 6500.0 mm, default 56.0 mm.
2.2.2 Wall Thickness
The user is prompted to enter a value for the pipe wall thickness. Allowed
values are ranged 0.5 to 100.0 mm, default 1.8 mm.
2.2.3 Material
The user can select the pipe material from a list;
· Carbon Steel · Stainless 304 · Stainless 316 · Cast Iron · Ductile Iron ·
Copper · PVC (Default) · Lead · Nylon · PE · Aluminium · Asbestos · Fibre
Glass · Other
If Other is selected then the user is prompted to enter the transverse (shear)
speed of sound in the pipe material, see below. Otherwise the transverse speed
of sound in the pipe material is read from a database held in the UFM.
2.2.4 Sound Velocity
Appearance of this parameter is context driven. If the user selected Other
from the list of pipe materials then the user is prompted to enter the
transverse speed of sound in the pipe material. Otherwise the speed of sound
in the pipe material is read from a database held in the UFM.
Allowed values are ranged 500 to 7000 m/s, default 1060 m/s (PVC).
2.3 Liner Menu
This menu allows the user to change pipe lining settings.
2.3.1 Material
The user can select a pipe liner material from a list;
· None (Default) · Cement · Epoxy · Glass · PP · Teflon · Rubber · Other
The list allows no liner (None) to be selected.
2.3.2 Sound Velocity
Appearance of this parameter is context driven. If the user selected Other
from the list of liner materials then the user is prompted to enter the
transverse speed of sound in the liner material. Otherwise the speed of sound
in the liner material is read from a database held in the UFM.
Allowed values are ranged 500 to 7000 m/s, default 0 m/s (None).
2.3.3 Thickness
Appearance of this parameter is context driven. If the user selected a liner
then the user is prompted for the thickness of the liner. Allowed values are
ranged 0.5 to 100.0 mm, default 0.0 mm (None).
2.4 Fluid Menu
This menu allows the user to change fluid settings.
2.4.1 Type
The user can select the fluid in the pipe from a list;
· Water (Default) · Sea Water · Kerosene · Petrol · Fuel Oil · Crude Oil ·
Freon R134a · Freon R22 · Diesel Oil · Castor Oil · F-76 Fuel Oil · Novec 1230
· Glycol/Water · Alcohol · Other
If the user selected Other from the list of fluid types then the user is
prompted to enter various other context driven parameters including; Fluid
Sound Velocity, Fluid Kinematic Viscosity and Fluid Density.
2.4.2 Temperature
The user is prompted to enter the temperature of the fluid in the pipe.
Allowed values are ranged -20 to +150 degC, default 18 degC.
Changing Fluid Temperature causes Fluid Sound Velocity, Fluid Kinematic
Viscosity, Fluid Density and Fluid Specific Heat Capacity to be recalculated.
3.0 Sensor Positioning
After completing entry of all parameters in the Quick Start function sequence
the user is prompted to confirm Transducer Type and is then taken to the
Sensor Positioning screen. Using the entered parameters the UFM calculates and
gives the required transducer spacing on the pipe. The screen also displays
the number of passes selected by the user or calculated by the UFM if in Auto
Passes mode. Clamp the transducers on the pipe using the supplied chain clamps
and coupling gel, measure the spacing using the tape measure provided. The
user is then presented with a sensor positioning screen showing,
· Graph of received ultrasonic signal · Signal to Noise Ratio · ATA/ETA If the
parameters entered are all correct then the graph should appear as in figure
(12).
Figure (12) Ideal Sensor Positioning. However, if the user has an incomplete
knowledge of the pipe then the screen may look like figure (13) or figure
(14).
Figure (13) Non-ideal Sensor Positioning, transducers too far apart.
Figure (14) Non-ideal Sensor Positioning, transducers too close together.
In figure (13) the received signal is too far to the right, the user should
slide the transducers closer together.
In figure (14) the received signal is too far to the left, the user should
slide the transducers further apart.
The most common cause of an incomplete or incorrect spacing is a lack of
knowledge about the pipe wall thickness.
As long as the vertical line acting as arrival marker is between the 2 guide
marks, resulting in an ATE/ETA value between 97 and 103 % then the UFM will
measure accurately. It is acceptable to reposition the transducers to adjust
their spacing by +/- 5 mm to optimise the positioning screen, the arrival
marker will move on the screen accordingly.
SNR should peak and be above 24 dB.
3.1 Optimising Transducer Mounting Location
For the best results ensure that,
· Ideally the transducers are mounted on bare pipe material, for metal pipes
this should be metal free from dust, rust and paint.
· Consider a location away from internal corrosion, sediment and streams of
entrained air, do not mount the transducers top to bottom on the pipe, mount
at 2 or 10 o’clock.
· Avoid mounting the transducers either on or opposite axial welds along the
pipe. · Ensure the transducers are aligned axially along the pipe. · Mount the
transducers away from bends, valves and other inserted instrumentation. ·
Observe were practical the advised upstream and downstream straight sections,
see below,
figure (15). · Ensure the pipe will always be full at the point of
installation, ideally mount the transducers
at a low point in the system. · If mounting the transducers on a vertical pipe
section ensure the flow direction is upwards in
the section. · Composite pipes can have de-laminations in their wall
thickness, this type of pipe is
notoriously bad when installing a UFM. · Ensure the temperature at the
transducer location is within the transducers rated range.
· Ideally the fluid should be free of particulates and bubbles, in the limit
then an alternative method such as Doppler flow measurement may be required.
· Pipe linings that are not bonded properly or are not conductive of
ultrasound (rubber) will cause measurement problems.
· Porous pipes, such as concrete can cause measurement problems. · Using
information from Standard Pipe Tables can be inaccurate, it is always best to
measure
the pipe outer diameter and wall thickness. · No matter how accurate the meter
is at making a velocity measurement, an inaccurate
knowledge of the internal cross-sectional area of the pipe will lead to
inaccuracy in the conversion to volumetric flow rate. 3.2 Upstream and
Downstream Pipe Runs Ideally the UFM transducers should be installed on as
long a section of straight pipe as is possible, see figure (15).
Figure (15) Upstream and downstream pipe lengths. Considering a pipe with an
outer diameter of D then if possible ensure at least 10D upstream distance
between the transducers and a bend in the pipework. In the case of an upstream
Valve then if possible ensure at least 20D upstream. In the case of an
upstream Pump then if possible ensure at least 30D upstream. In all cases
ideally ensure 5D downstream exists before a bend or obstruction in the
pipework. 3.3 Transducer Mounting Locate an optimum position on the pipe
following the advice above. Use Coupling gel. Apply adequate couplant and
ensure no gap exists between the transducer and the pipe surface. Banding or
clamping is required to keep the transducers in place. It is recommended to
use the
chain clamp supplied or 10 mm wide jubilee clips.
3.4 Transducer Spacing
Given that all information regarding the installation has been entered
accurately and the advice above has been followed then the UFM will measure
reliably and accurately.
This is confirmed by,
· A strong received signal strength · A high SNR value · Value of ATA/ETA
close to 100 %
If is acceptable to make small adjustments to the transducer spacing to
optimise the received signal strength and ATA//ETA.
However, if large adjustments are necessary then the importance of wall
thickness should be considered. Wall thickness is typically the parameter
about which the user last the least knowledge.
If all other avenues have been explored, including recoupling the transducers
at several different locations on the pipework then adjusting the wall
thickness parameter may help.
4.0 Error Codes
As soon as the UFM is switched on a self-diagnostic program will start. This
program fully tests both the UFM hardware and software.
If an error is detected an error message will be displayed prompting user
action. If the error persists contact customer support, see appendix A.
Error codes and their meanings can be found in the Main Menu under the
Diagnostics Menu.
· MATH, there is a problem with the UFM central maths processor. · STACK,
there is a stack overflow problem. · ADDR, there is a memory addressing
problem. · OSC, the UFM core oscillator is suffering a problem. · SYSTEM, a
general system wide fault has occurred. · SPI2, there is a problem with the
internal SPI data bus. · I2C1, there is a problem with the internal I2C data
bus. · FRAM, there is a problem with the UFM external FRAM memory.
5.0 Icons
The UFM has a series of icons along the top of the LCD.
The meanings of these icons are from left to right as they appear;
· Maple leaf, unreliable measurement, check installation. · Display, LCD
backlight on/off. · Loudspeaker, Audio buzzer on/off. · Bell, I2C1 error.
· Clock, SYSTEM or timing error. · Warning Triangle, MATH, STACK, ADDR or OSC
error. · Spinner, SPI2 or FRAM error. In addition to showing icons, depending
on the flow regime, see figure (16) the UFM displays; · Zero, no flow · L,
laminar · Tr, transition · T, turbulent in the top-left when in measurement
mode. Laminar flow is generally regarded to exist for Reynolds Number less
than 2300, transition is in the range 2300 to 4000 and turbulent flow
typically has a Reynolds Number greater than 4000.
Figure (16) Laminar (a) and turbulent (b) flow.
6.0 Battery Life
The POCKET-UFM is powered by a single high capacity 9 Vdc PP3 primary cell
1200 mAh. Under normal operating conditions the battery life of the UFM is in
excess of 12 hours continuous measurement. If using a rechargeable cell or
primary cell with lower capacity then the run time of the UFM will be reduced.
If using a rechargeable cell then use the external charger recommended by that
cell’s manufacturer. The UFM has an auto off feature to preserve battery life,
if no keypad entry is made for 60 seconds when operating in the UI mode then
the meter will auto power off. This feature is not available when in
measurement mode
The LCD backlight will flash when the battery is low and needs to be changed.
7.0 Specification
The UFM has the following specification, features and performance;
· Pipe outer diameters ranging from 10 to 6500 mm. · Temperature range for
pocket meter -10 to +65 degC. · Weight 230 g (including internal PP3 battery).
· Dimensions 185 x 85 x 25 mm. · IP54 enclosure. · Full 128 x 68 pixel
graphics display with backlight. · Full 17 key tactile keypad with audio
feedback. · Integral wall thickness gauging, 2 to 24 mm. · Battery life up to
12 hours continuous use on a standard 9 V PP3 battery for easy
replacement.
Features
· Light weight and small for easy, one handed operation. · Intuitive
installation using menu driven UI. · Full set of instrument and measurement
diagnostics. · Signal oscilloscope for sensor positioning and diagnostics. ·
Internal database of pipe, fluid and lining materials. · Wall thickness
gauging as standard using wall thickness probe. · Heat quantity measurement
(if inlet and outlet temperatures are known). · All measured values can be
totalled. · Ability to save up to 10 configurations for different measurement
points.
Performance
· Measurement principle ultrasonic transit time difference. · Flow velocity
range 0.01 to 25 m/s. · Resolution 0.25 mm/s. · Repeatability 0.15% of
measured value. · Accuracy +/- 0.5 to +/- 3.0 % of measured value for velocity
greater than 0.2 m/s, depending
on application. · Turn down ratio 1/100. · Measurement rate 1 Hz as standard.
· Gas/solids < 10 % of volume.
Quantity and units of measurement
· Flow velocity · Volumetric flow rate · Mass flow rate · Energy flow rate ·
Volume · Mass · Energy
Transducers
Figure (17) PEEK/stainless steel transducers. · 3 operating frequencies to
cover the range of pipe outer diameter 10 to 6500 mm. · DM10, 1 MHz for pipes
of 25 to 750 mm. · DS10, 1 MHz for pipes of 50 mm and over. · DM20, 2 MHz for
pipes less than 25 mm. · DN40, 4 MHz for pipes less than 15 mm. · Dimensions
60 x 30 x 35 mm and 40 x 20 x 25 mm for locations with limited access. ·
Material stainless steel and PEEK. · Temperature range -10 to +80 degC. ·
Ingress Protection rated IP54, with IP68 option. · Cable length 1.5 m as
standard. · Matched pairs for accurate zero flow measurement.
8.0 Product Identification
Each UFM, pair of flow transducers and wall thickness probe come with a unique
Identification code.
In the case of the UFM this is written into the software and can be read using
the UI.
In the event of a need to contact Sonic Driver please have these codes
available to quote.
9.0 Service
The UFM is a sophisticated measuring instrument and contains no user
serviceable parts.
For all operational problems please contact our service department by
telephone or email, see Appendix A.
Sonic Driver do offer a software upgrade service. Please contact the factory
for information about the latest software.
10.0 Limited Warranty and Disclaimer
Sonic Driver Ltd warrants to the end purchaser, for a period of one year from
the date of shipment from our factory, that all new products manufactured by
it are free from defects in materials and workmanship.
This warranty does not cover products that have been damaged due to normal
use, misapplication, abuse, lack of maintenance, or improper installation.
Sonic Driver obligation under this warranty is limited to the repair or
replacement of a defective product, if the product is inspected by Sonic
Driver Ltd and found to be defective. Repair or replacement is at the
discretion of Sonic Driver Ltd.
If the product is outside of the warranty period a purchase order must be
received from the end purchaser before repair work will start.
The product must be thoroughly cleaned and any contamination removed before it
will be accepted for return.
The purchaser must determine the applicability of the product for its desired
use and assumes all risks in connection therewith. Sonic Driver Ltd assumes no
responsibility or liability for any omissions or errors in connection with the
use of its products.
Sonic Driver Ltd will under no circumstances be liable for any incidental,
consequential, contingent or special damages or loss to any person or property
arising out of the failure of any product, component or accessory.
All expressed or implied warranties, including the implied warranty of
merchantability and the implied warranty of fitness for a particular purpose
or application are expressly disclaimed and shall not apply to any products
sold or services rendered by Sonic Driver Ltd.
The above warranty supersedes and is in lieu of all other warranties, either
expressed or implied and all other obligations or liabilities.
No agent or representative of Sonic Driver Ltd has any authority to alter the terms of this warranty in any way.
Appendix A Contact Details
Telephone: +44(0)7971 273000
Postal Address: Sonic Driver Ltd, Lochiel, Llaneilian Road, Amlwch, Gwynedd, LL68 9HU, UK.
Email: service@sonic-driver.com
Website: www.sonic-driver.com
Appendix B Table of fluid properties
Fluid Water 20 (degC) Acetone Methanol Ethanol Alcohol Glycerine Petrol Diesel Toluene Coal Oil Petroleum Pine Oil Castor Oil Glycol Peanut Oil
Longitudinal Speed of Sound (m/s) 1482 1190 1121 1166 1440 1923 1250 1385 1170 1420 1290 1280 1502 1620 1472
Appendix C Table of pipe and lining material properties
Material Steel ABS Aluminium Copper Cast Iron GRP Glass PVC Cement Teflon Rubber
Shear Speed of Sound (m/s) 3206 2286 3046 2270 2460 3430 3276 1060 4190 1450 1600
Appendix D Table of speed of sound in water
Temperature (degC) 0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 99
Speed of Sound (m/s) 1402.3 1426.1 1447.2 1465.8 1482.3 1496.6 1509.0 1519.7 1528.8 1536.3 1542.5 1547.3 1550.9 1553.4 1554.7 1555.1 1554.4 1552.8 1550.4 1547.1 1543.9
Appendix E Table of typical pipe roughness values
When a fluid flows through a pipe then the pipes own internal roughness is important when considering friction losses.
Pipe manufacturers often quote a pipe roughness value for their products.
Some typical figures are given below.
Pipe Material Concrete Cast Iron Galvanized Iron Asphalted Cast Iron Commercial or Welded Steel PVC, Glass and other drawn tubing
Peak to Trough Roughness (mm) 0.3 – 3.0 0.26 0.15 0.12 0.045 0.0015
By default the Sonic Driver flowmeter uses a figure of 0.01 mm as a good compromise for most common pipes.