INOR 70ADA00051 Adapter mA
GENERAL INORMATION
Adapter mA part no. 70ADA00051 is an adapter to convert mA signals to mV
signals (shunt resistor), intended to be mounted on the input side of
transmitter IPAQ R330.
With the adapter mounted, it gives the possibility to connect mA input signals
to the transmitter IPAQ R330. The transmitter will then work as an active
isolation transmitter for mA input signals with a 4-20 mA output in a 2 wire
connection. It is suitable for direct connection to the input modules of a
control system. Thanks to the programmable input range and the low input
impedance, it meets almost any need for current signal isolation and load
amplification.
Specifications for IPAQ R330 with adapter mA
ADAPTER CONNECTION
- Remove input terminal marked 1 to 4.
- Connect 70ADA00051 instead.
CONFIGURATION calculation How to set up the transmitter
Step 1
Calculation of transmitter Measuring Range to configure
- Use the calculation formula to calculate your measuring range for the transmitter in mV.
- IStart = Start value of input current range in mA
- IEnd = End value of input current range in mA
- LRV = Lower range value to set in mV
- URV = Upper range value to set in mV
Calculation formula
- LRV = IStart x 1
- URV = IEnd x 10
Example 1
- Input: 4…20 mA
- LRV = 4 x10 = 40
- URV = 20 x 10 = 200
Example 2
- Input: 0…55 mA
- LRV = 0 x10 = 0
- URV = 55 x 10 = 550
Step 2
Configure the transmitter via a configuration tool, PC Software Con Soft or
Smart phone App INOR Connect.
CONFIGURATION via PC
Configuration via PC Software ConSoft example 4…20 mA
- Change sensor to Volt.
- Enter LRV and URV for measuring range
- Write to transmitter.
CONFIGURATION via App
Configuration via Smart phone App INOR Connect (example 4…20 mA)
- Change Sensor type to Volt (Pic 2)
- Enter LRV and URV for range
- Write to transmitter.