INOR APAQ C130TC Programmable 2 Wire Transmitter for Thermocouple Instruction Manual

June 12, 2024
INOR

INOR APAQ C130TC Programmable 2 Wire Transmitter for Thermocouple Instruction Manual
INOR APAQ C130TC Programmable 2 Wire Transmitter for
Thermocouple

GENERAL INFORMATION

APAQ C130TC is a low cost, non-isolated, easy-to-use 2-wire transmitter, for thermocouple input, intended to be used in industrial environment.

The transmitter is configured from a mobile device via NFC communication, by using the app INOR Connect. Inor Connect is available for free download on App Store, Google Play Store and Huawei AppGallery.

CONFIGURATION

CONFIGURATION

Before making a configuration of APAQ C130TC you need to do following:
Make sure that you have a mobile device with NFC communication activated. Download the app INOR Connect to your mobile device.

Required versions:

iOS:
iOS 13 or later and  Iphone 7 or later for NFC

Android:
Android 4.4 or later

Configuration procedure:

  1. Launch the app by clicking on the App icon or holding your mobile device against the transmitter on the part of the device where NFC is located (only possible with Android).

  2. Click on “Read Configuration” and hold your mobile device against the transmitter as explained in the first section.

  3. In the app you can edit the following:
    a. Sensor type
    b. Measuring range
    c. Upscale or downscale sensor error indication
    d. Filter level
    e. TAG-number
    f. Password

  4. Once you have set the desired values in the app, you transfer it to the transmitter by clicking the transfer button and holding the mobile device against the transmitter until a green check box appears confirming that the transfer has been completed

OUTPUT LOAD DIAGRAM

OUTPUT LOAD DIAGRAM

Maximum load depending on power supply Formula for the max. permissible output load for APAQ C130TC:
Load R =(U-6)/0.022

DIMENSIONS

DIMENSIONS

INSTALLATION

APAQ C130TC is designed to fit inside connection heads type DIN B or larger. The larger center hole, diameter 7 mm/0.28 inch, facilitates the pulling through of the sensor leads or an insert tube, greatly simplifying the mounting procedure.

Connect input, output and power supply acc. to “CONNECTIONS”. A convinient way to install the transmitter is to use the INOR mounting kits for in-head and DIN rail mounting.
In order to minimize measuring errors make sure the connecting screws are tightened enough, 0.5-0.6 Nm.

MOUNTING

How to mount on DIN-rail Adapter

Step 1

Head mounting kit
MOUNTING

Step 2
MOUNTING

CONNECTIONS

CONNECTIONS

ORDERING INFORMATION

APAQ C130 70C1300211
Head mounting kit 70ADA00017
DIN-rail Adapter + Screw (10 pcs) 70ADA00027

The transmitters are delivered with either a default factory setting or configured according to customer’s specification.

DATA (short form)

Input 2-wire connection
TC type B – Ft3ORh-Ft6fth (IK 60584) 0…+1820 °C / +32…+3308 °F
TC type E – NiCr-CuNi (IEC 60584) -270…+1090 °C / -454…+1832 °F
TC type l – Fe-CuNi (IEC 60584) -210…+1200 °C / -346…+2192 °F
TC type K – NiCr-NiAl (IEC 60584) -270…+1300 °C / -454…+2372 °F
TC type N – NiCrSi-NiSi (IK 60584) -270…+1300 °C I -454…+2372 °F
TC type R – Pt13Rh-Pt (IK 60584) -50…+1750 °C / -58…+3182 °F
TC type S – Pt1ORh-Pt (IEC 60584) -50…+1750 °C/ -58…+3182 OF
TC type T – Cu-CuNi (IEC 60584) -270…+400 °C / -454…+752 °F
Input impedance 10 MO
Maximum wire loop resistance 1n-head bansmilter (TC sensor): 5 IQ
Cold ]unction Compensation Internal or fixed
Output 4-20 mA, temperature linear
Sensor failure indication Upscale (1.21.0 mA) or downscale (S3.6 rr
NAMUR Compliance Current limitations and failure currents acc to NAMUR

NE 43
Adjustable filtering level| 0.4 to 26 sec.
Warm-up time| After a max. 20 minutes the accuracy specifications are reached(due to the internal cold junction)
Ambient temperature|
Storage and operation| -40 to +85 °C I -40 to +185 °F
Galvanic isolation| None
Power supply| 6 to 32 VDC
Accuracy|
TC type B| See Data Sheet
TC type R, 5, T| Max. of ±2.0 °C or 0.2% of span
| Max. of ±3.6 °F or ±0.2% of span
TC type E, 3, K| Max. of ±I °C or
0.2% of span
| Max. of 1.8 °F or ±0.2% of span
TC type N (-100…+1300 °C)| Max. of
1 °C or 0.2% of span
| Max. of
1.8 °F or ±0.2% of span
TC type N (-270…400 °C)| 2.0 °C / 16 °F
CJC ao:uracy| Typical 1 °C / 1.8 °f (max *3 °C / ±5.4 °F) within ambient temperature rang
Mounting| DIN B-head or larger
Configuration| App INOR Connect via NFCT”
Terminal| Snje/stratedvins rrec1Srrm2,max05Nrnloc
RoHS, China RoHS| Directive: 2011/65/EU and 2015/863/EU Harmonized standard: EN IEC 63000 China RoHS 2
Factory settings| Type K
Input| 0 to 1000 °C
Output| 4 to n mA
Sensor error indication| Upscale

LIMITED WARRANTY

INOR Process AB, or any other affiliated company within the Inor Group (hereinafter jointly referred to as ”Inor”), hereby warrants that the Product will be free from defects in materials or workmanship for a period of five (5) years from the date of delivery (”Limited Warranty”). This Limited Warranty is limited to repair or replacement at Inor’s option and is effective only for the first end-user of the Product. Upon receipt of a warranty claim, Inor shall respond within a reasonable time period as to its decision concerning:

  1. Whether Inor acknowledges its responsibility for any asserted defect in materials or workmanship; and, if so,
  2. the appropriate cause of action to be taken (i.e. whether a defective product should be replaced or repaired by Inor)

This Limited Warranty applies only if the Product:

  1. is installed according to the instructions furnished by Inor;
  2. is connected to a proper power supply;
  3. is not misused or abused; and
  4. there is no evidence of tampering, mishandling, neglect, accidental damage, modification or repair without the approval of Inor or damage done to the Product by anyone other than Inor.

This Limited Warranty is provided by Inor and contains the only express warranty provided

INOR SPECIFICALLY DISCLAIMS ANY EXPRESS WARRANTY NOT PROVIDED HEREIN AND ANY IMPLIED WARRANTY, GUARANTEE OR REPRESENTATION AS TO SUITABILITY FOR ANY PARTICULAR PURPOSE, PERFORMANCE, QUALITY AND ABSENCE OF ANY HIDDEN DEFECTS, AND ANY REMEDY FOR BREACH OF CONTRACT, WHICH BUT FOR THIS PROVISION, MIGHT ARISE BY IMPLICATION, OPERATION OF LAW, CUSTOM OF TRADE OR COURSE OF DEALING, INCLUDING IMPLIED WARRANTIES OF MER-CHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. EXCEPT AS PROVIDED HEREIN, INOR FURTHER DISCLAIMS ANY RESPONSIBILITY FOR LOSSES, EXPENSES, INCONVENIENCES, SPECIAL, DIRECT, SECONDARY OR CONSEQUENTIAL DAMAGES ARISING FROM OWNERSHIP OR USE OF THE PRODUCT.

Products that are covered by the Limited Warranty will either be repaired or replaced at the option of Inor. Customer pays freight to Inor, and Inor will pay the return freight by post or other “normal” way of transport. If any other type of return freight is requested, customer pays the whole return cost.

This product should not be mixed with other kind of scrap, after usage. It should be handled as an electronic/electric device.

INOR Process AB , PO Box 9125, SE-200 39 Malmö, Sweden,
Phone: +46 40 312 560, Fax: +46 40 312 570, E-mail: [email protected]

INOR Transmitter OY , Unikkotie 13, FI-01300 Vantaa, Finland,
Phone:+358 10 421 7900, Fax: +358 10 421 7901, E-mail: [email protected]

INOR Transmitter GmbH , Am See 24, D-47279 Duisburg, Germany,
Phone: +49-203 7382 762 0, Fax: +49-203 7382 762 2, E-mail: info @inor-gmbh.de

KROHNE Temperature Division INOR , 55 Cherry Hill Drive,
Beverly, MA 01915, United States