Product Information
Specifications:
- Product Name: heliX+
- Model: ADAPTER STRAND 2 with green dye Gc
- Manufacturer: Dynamic Biosensors GmbH & Inc.
- Order Number: AS-2-Gc v5.1
- Concentration: 400 nM
- Storage: White Cap, please see expiry date on label
- For research use only
Product Description:
The heliX+ ADAPTER STRAND 2 with green dye Gc features 2 spots
with 2 different anchor sequences for DNA-encoded addressing. It
comes with a concentration of 400 nM and is designed for research
purposes.
Usage Instructions:
Preparation | MIX&RUN:
-
Mix Adapter strand 1 – Gc (400 nM) and conjugated Ligand strand
with ligand 1 (500 nM) at 1:1 ratio (v/v). -
Mix Adapter strand 2 – Gc (400 nM) and conjugated Ligand strand
with ligand 2 (500 nM) at 1:1 ratio (v/v). -
Incubate separately the two solutions of step 1 and 2 at room
temperature at 600 rpm for 30 min to ensure complete
hybridization. -
Mix solution of step 1 and 2 at 1:1 ratio (v/v). Solution is
ready to use for biochip functionalization.
Stability of the solution is related to the stability of the
ligand molecules.
FAQ:
Q: Can the heliX+ ADAPTER STRAND 2 be used for clinical
applications?
A: No, this product is for research use only and should not be
used for clinical purposes.
Q: How should I store the product to ensure maximum shelf
life?
A: Store the product with the white cap and check the expiry
date on the label. Avoid multiple freeze-thaw cycles by aliquoting
the nanolever.
“`
heliX+
User Manual
ADAPTER STRAND 2
with green dye Gc
Dynamic Biosensors GmbH & Inc. AS-2-Gc v5.1
Key Features
· Adapter strand 2 for functionalization of heliX® Adapter Chip Spot 2. ·
Compatible with heliX® Adapter Chip. · Includes Adapter strands for 50
regenerations. · Ideal for MIX&RUN sample preparation. · Adapter strand 2
carries a hydrophobic green dye (Gc) with a neutral net charge.
heliX® Adapter Chip Overview
2 spots with 2 different anchor sequences for DNA-encoded addressing.
Product Description
Order Number: AS-2-Gc
Table 1. Contents and Storage Information Material Adapter strand 2 – Gc
Cap White
Concentration Amount
400 nM
5 x 100 µL
For research use only.
This product has a limited shelf life, please see expiry date on label. To avoid many freeze thaw cycles please aliquot the nanolever.
Buffer TE40 [1]
Storage -20°C
AS-2-Gc v5.1
www.dynamic-biosensors.com
Page 1 of 3
Preparation | MIX&RUN
In-solution hybridization of adapter and ligand strands:
1. Mix Adapter strand 1 – Gc (400 nM) and conjugated Ligand strand with
ligand 1 (500 nM) at 1:1 ratio (v/v).
2. Mix Adapter strand 2 – Gc (400 nM) and conjugated Ligand strand with
ligand 2 (500 nM) at 1:1 ratio (v/v).
3. Incubate separately the two solutions of step 1 and 2 at RT at 600 rpm for
30 min to ensure complete hybridization.
4. Mix solution of step 1 and 2 at 1:1 ratio (v/v).
Solution is ready to use for biochip functionalization.
Stability of the solution is related to the stability of the ligand molecules.
Table 2. Additional material for functionalization of spot 1 and reference spot 2.
Material
Concentration Buffer Related Product Name
Adapter strand 1 – Gc
Ligand strand carrying the conjugated ligand 1
Ligand strand carrying the conjugated ligand 2
400 nM 500 nM
500 nM
TE40 [1] PE40 [
2]
Adapter strand 2 with green dye Gc heliX® Amine Coupling Kit 1
PE40 [2] heliX® Amine Coupling Kit 1
Order No AS-1-Rb HK-NHS-1
HK-NHS-1
Example
Required volume for 3 functionalizations: 100 µL with a final concentration of 100 nM.
Vial 1
Adapter strand 1 – Gc (400 nM)
25 µL
Conjugated Ligand strand with ligand 1 (500 nM)
25 µL
Vial 2
Adapter strand 2 – Gc (400 nM)
25 µL
Conjugated Ligand strand + with ligand 2 (500 nM)
25 µL
After incubation time, mix vial 1 and vial 2 to obtain 100 µL of ready-to-use DNA solution.
AS-2-Gc v5.1
www.dynamic-biosensors.com
Page 2 of 3
Contact
Dynamic Biosensors GmbH Perchtinger Str. 8/10 81379 Munich Germany
Dynamic Biosensors, Inc. 300 Trade Center, Suite 1400 Woburn, MA 01801 USA
Order Information order@dynamic-biosensors.com Technical Support support @dynamic-biosensors.com
www.dynamic-biosensors.com
Instruments and chips are engineered and manufactured in Germany. ©2024
Dynamic Biosensors GmbH | Dynamic Biosensors, Inc. All rights reserved.
[1] TE40: 10 mM Tris, 40 mM NaCl, 0.05 % Tween20, 50 M EDTA, 50 M EGTA [2] If the protein is not stable in PE40 (TE40, HE40), please check buffer compatibility with the switchSENSE® compatibility sheet.
AS-2-Gc v5.1
www.dynamic-biosensors.com
Page 3 of 3
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>