MIKSTER INDU 20 Temperature Recording System User Manual

September 2, 2024
MIKSTER

INDU 20 Temperature Recording System

“`html

Technical Specifications

  • Power supply:

  • Power consumption:

  • Housing:

  • Dimensions: One-piece Front Panel 134x134x65mm

  • Mounting hole: 90x90mm

  • Weight: 500g

  • Display:

  • Keyboard:

  • Temperature measurement range:

  • Temperature measurement resolution:

  • Inputs: 1x analogue PT100/PT500/PT1000, 2x control (alarm
    signaling or keyboard lockout)

  • Outputs: 1x analogue 0/4..20mA, 4x relay

  • Other features: Internal vacuum sensor

  • IP rating: IP65 (front) IP20 (rear)

  • RTC: Yes

  • Communication: 1xRS485

  • Registration: 100000 records*

  • Operating conditions: AUTOSTART according to RTC, with advanced
    programming (up to 10 days)

  • Types of temperature control: 2 types of bistable governors and
    the PID governor

  • Ending of the process: Set time or manually

Product Usage Instructions

3. Start Up

NOTE: In case of power failure, the controller
remembers the mode of operation and returns to this mode upon power
supply restart (unless the time set in the 48 Setup cell
expires).

3.1 Operating Panel

The operating panel includes:

  • RTC CLOCK/PROCESS TIME
  • NEGATIVE PRESSURE VALUE
  • KEY OF THE AUTOSTART MODE
  • KEY OF THE INFO MODE
  • KEY OF THE START MODE

3.4 Autostart Mode

The autostart mode is used to enter the START mode with a time
delay.

To set the autostart moment:

  1. Setup – HMD
  2. Activation after a predetermined number of hours and minutes
    (F47 SETUP – HM)

To exit the AUTOSTART mode, press the AUTO/START key again. You
can transition directly from AUTOSTART to START mode by pressing
the START key when the controller is in START mode.

Frequently Asked Questions (FAQ)

Q: How do I edit the set parameters of the process?

A: To enter the mode of editing set parameters, press the
corresponding key. The mode is indicated by the pulsation of the
diode on the key EDIT. Use the keys for correction of parameters.
The parameter being edited will pulsate, and confirmation and
proceeding to the next editing field is done by the OK key.
Pressing EDIT repeatedly means exiting from the editing mode.

“`

EN
INDU 20
USER MANUAL v1.6(19)EN
ul. Wojkowicka 21, 41-250 Czelad, Poland tel. +48 32 763 77 77 fax: +48 32 763 75 94 mikster.eu

CONTENTS

User manual INDU 20 v1.6(19)EN

CONTENTS …………………………………………………………………………………………………………………………………………………………………..2 1. TECHNICAL SPECIFICATIONS………………………………………………………………………………………………………………………………..3 2. DIMENSIONS ………………………………………………………………………………………………………………………………………………………..4 3. START UP …………………………………………………………………………………………………………………………………………………………….4
3.1 OPERATING PANEL……………………………………………………………………………………………………………………………………….4 3.2 EDIT MODE ­ CHANGING OF THE SET VALUES ………………………………………………………………………………………………5 3.3 INFO MODE…………………………………………………………………………………………………………………………………………………..5 3.4 AUTOSTART MODE ……………………………………………………………………………………………………………………………………….6 3.5 START MODE………………………………………………………………………………………………………………………………………………..6 4. FUNCTION OF RELAYS ………………………………………………………………………………………………………………………………………….7 5. GOVERNOR OF THE “TEMPERATURE APPROACH” …………………………………………………………………………………………………7 6. SELECTION OF SETTINGS OF THE PID GOVERNOR ………………………………………………………………………………………………..7 7. FUNKCJE SERWISOWE DOSTPNE DLA UYTKOWNIKA …………………………………………………………………………………………8 8. ALARMS ……………………………………………………………………………………………………………………………………………………………….8 9. CONTROLLER SETUP ……………………………………………………………………………………………………………………………………………9 10. EXAMPLE OF APPLICATION* ………………………………………………………………………………………………………………………………….14

| 2

www.mikster.eu

User manual INDU 20 v1.6(19)EN

1. TECHNICAL SPECIFICATIONS

Power supply:
Power consumption:
Housing:
Dimensions:

230VAC option: 24VAC, 110VAC ± 10% VAC 3 W
One-piece Front Panel 134x134x65mm

Mounting hole: Weight: Display:
Keyboard: Temperature measurement range: Temperature measurement resolution: Inputs:
Outputs:
Other features: IP rating:
RTC: Communication:

90x90mm
500g
LED ½ ” x 4 digits LED ½ ” x 2 digits LED ½ ” x 2 digits Micro switch buttons
­ 30°C ÷ 400°C
0,1°C od ­ 9,9°C do 99,9°C 1°C in other ranges
1x analogue PT100/PT500/PT1000 2x control (alarm signaling or keyboard lockout)
1x analogue 0/4..20mA 4x relay
Internal vacuum sensor
IP65 (front) IP20 (rear)
Yes
1xRS485

Registration:
Operating conditions:

100000 records*
Temperature: 0°C ÷ 55°C Humidity: 5%RH ÷ 85%RH

AUTOSTART: according to RTC, with the possibility of an advanced programming (up to 10 days) of the controller switching on.
Types of the temperature control: 2 types of bistable governors and the PID governor.
Ending of the process can be done either by the set time or manually.

  • recording module, R – version

www.mikster.eu

| 3

2. DIMENSIONS 134mm

User manual INDU 20 v1.6(19)EN
65mm

134mm

80mm
3. START UP
After the mains are connected the controller is automatically activated. After the welcoming words the device displays in turn: current hour and minute, measurement in channel 1 ­ vacuum in %, set value of the rotational speed of the drum. Three horizontal lines indicate the lack or failure of a measuring element. Diodes on the keys signal the mode of the device (e.g. editing or autostart mode). Horizontal lines at the left side of the displayed measured value signal the governor’s mode: controlled output causes that the diode lights. Diodes on the keys indicate the controller’s mode. Possible modes of operation are: AUTOSTART, START, INFO and EDIT. In the STOP mode, when the START mode ended, the word ,,STOP” is displayed on the screen ­ instead of an hour and minute.
NOTE: in the case of power failure the controller remembers the mode of operation and returns to this mode at the power supply restart. (unless the time set in the 48 Setup cell expires).
3.1 OPERATING PANEL

RTC CLOCK/ PROCESS TIME
NEGATIVE PRESSURE VALUE
ROTATIONAL SPEED ­ SET VALUE

KEY OF THE AUTOSTART MODE
KEY OF THE INFO MODE
KEY OF THE START MODE

| 4

www.mikster.eu

User manual INDU 20 v1.6(19)EN 3.2 EDIT MODE ­ CHANGING OF THE SET VALUES

In order to enter the mode of editing the set parameters of the process the key should be pressed mode is indicated by pulsation of the diode on the key EDIT.

. Entering into the edit

Keys

are used for correction of parameters. Parameter being edited is pulsating, confirmation and proceeding

to the next editing field is done by the OK key. Consecutive pressing of the key EDIT means exit from the editing mode.

The set parameters are co:

– START mode duration (number of hours and minutes)

– Negative pressure ­ set value

– Rotational speed ­ set value

3.3 INFO MODE

Single pressing of the

causes the display of information in dependence of the controller’s mode of operation::

For the AUTOSTART mode Depending on the parameter in the 47 Setup cell: At selecting HMD ­ hour, minute and twenty-four hours delay before the START mode commences, At selecting HM ­ number of hours and minutes remaining to the START mode

Consecutive information, for all modes, are the same: · temperature measurement: the upper display shows the current temperature measurement in channel 1, the lower reads TE, · the upper display shows PSET, the middle one the negative pressure (set value), the lower display indicates the rotational speed of the drum, · current date: starting from the upper display – one can read the year, month and day, · current time: starting from the upper display – one can read hour, minute and second.

Proceeding into the next (previous) information can be done by keys should be pressed again.

. To exit the INFO mode the INFO key

www.mikster.eu

| 5

3.4 AUTOSTART MODE The autostart mode is used for entering the START mode with a time delay.

User manual INDU 20 v1.6(19)EN

Pressing

causes transition to the parameters editing of this mode.

There are two possibilities of setting the moment of the controller AUTOSTART:

Activation at the predetermined hour and minute ­ with the possibility of adding the twenty-four hour delay (F47

SETUP – HMD).

Activation after the predetermined number of hours and minutes (F47 SETUP – HM)

To exit the AUTOSTART mode the AUTO/START key should be pressed again.

There is also a possibility of a direct transition from the AUTOSTART into the START mode. Single pressing of the START key is needed.

3.5 START MODE When all set parameters are introduced (see: EDIT MODE ­ CHANGING OF THE SET VALUES) the process can begin, which means

the controller should be in the START mode. Beginning and ending of the START mode is done by pressing the key:

.

For the typical settings of the controller, entering the START mode activates governors and the countdown of the process time starts.

The display shows the number of hours and minutes remaining to the end of the process.

The end of the process is indicated by the internal sound signal of the controller and by controlling of the relay output REL5 (unless it is declared in the SF81 Setup as the temperature governor). If the cell SF81 value is set to 2 relays REL4, REL5 work in the Left/Right turnover.

SF81 2 REL4 and REL5 carries out the functions, which illustrate the chart.
SF75 TL – time relay attach turnover LEFT (default = 10sek.) SF76 TR – time relay attach turnover RIGHT (default = 10sek.) SF77 Tp ­ pause time (default = 20sek) Pressing the OK key switches off the sound signal.

| 6

www.mikster.eu

User manual INDU 20 v1.6(19)EN
4. FUNCTION OF RELAYS
REL 2: is responsible for the increase of the negative pressure and the pump controlling (governor: simple hysteresis). REL 3: is responsible for the decrease of the negative pressure (governor: reversed hysteresis). REL 4: controlled in the START mode. REL 5: temperature regulation or alarm event signaling.
5. GOVERNOR OF THE “TEMPERATURE APPROACH”

Parameters:
Tz ­ set temperature value. Tza ­ temperature of the governor activation; up to this temperature the output is controlled (heating). When this temperature is reached the governing algorithm starts.
Dout ­ state at the digital output (high state corresponds to the heaters being ON).
6. SELECTION OF SETTINGS OF THE PID GOVERNOR
To obtain an access to the settings of the PID governor coupled with the particular measuring channel the MINUS key should be pressed and held and then the INFO key. If the temperature governor is selected on the relay (REL 5) the upper display shows the information that the governor can be adjusted ­ the OK key should be pressed. The edition of the selected parameter is being done on the middle display (pulsating value). Increasing of the parameter’s value can be done by the PLUS key, while decreasing by the MINUS key. Proceeding to the next parameter and confirmation of changes is done by pressing OK. Exit from the edition mode – by pressing the EDIT key.
Regulation is being done on the bases of:
To ­ sampling period, Pr ­ amplification of the proportional element, Ti ­ integration constant (doubling time), Td ­ differential constant (advancing time), TS ­ set temperature, Writing 0 value for the integrating or differentiating element causes its switching off. Due to that feature, it is possible to obtain the arbitrary governing algorithm.

www.mikster.eu

| 7

User manual INDU 20 v1.6(19)EN

7. FUNKCJE SERWISOWE DOSTPNE DLA UYTKOWNIKA

Cell number

Description

F0

Setting of the real time clock. Next parameter is reached by the OK key.

F1

Change of the access code for the user

Range 0..9999

For 0 value ­ checking of the access code is switched off

F2

Information concerning the current program version

F3

Switching ON/OFF of the keyboard click

OFF ­ switching off

ON ­ switching on

To enter the user’s mode the MINUS key should be pressed and held, the PLUS key should be pressed and held. Those functions are

accessible when the access code is given. The access code is introduced by To switch off the checking of the access code its value should be set as zero. As a rule the access to the user’s settings is switched off.
8. ALARMS
Controller INDU 20 signals 11 alarm events: · Err1 Failure of the internal sensor of negative pressure, · Err2 Failure or lack of the measuring element in the channel 1 (temperature), · Err3 · Err4 Permissible MAX negative pressure exceeded, · Err5 Permissible MAX temperature exceeded, · Err6 · Err7 Permissible MIN negative pressure exceeded, · Err8 Permissible MIN temperature exceeded, · Err9 · Err10 Alarm from the control input 1, · Err11 Alarm from the control input 2.

keys.

The first step in an alarm activation is the setting of the time delay [seconds] in SETUP (cells 71..73) between an alarm event and the alarm activation. Then the activation of the selected alarms in SETUP (cells 60..70) should be done.
Alarms are signaled on the display by the word Err with the adequate number, by internal howler, and – when in Setup cell 81 the relay REL 5 is in the mode of alarm signaling – by controlling of its output.
Alarm should be confirmed by the OK key. If the cause of an alarm is not removed, the controller ­ after the time delay of the alarm activation – will again signal the alarm.

| 8

www.mikster.eu

User manual INDU 20 v1.6(19)EN

9. CONTROLLER SETUP

Nr

Default

value

Range

Description

SF0

1

0..99

Address in the MODBUS network

SF1

0

0..4

Transmission rate:

0 – 9600, 1 – 19200, 2 – 38400, 3 – 57600, 4 – 115200

SF2

1

0..12

Type of the measuring input for channel 1 · 0 ­ PT-500 · 1 ­ PT-100 · 2 ­ PT1000
· 3 ­ 0..20mA · 4 ­ 4..20mA · 5 ­ thermocouple s · 6 ­ thermocouple b · 7 ­ thermocouple r · 8 ­ thermocouple t · 9 ­ thermocouple j · 10 ­ thermocouple e · 11 ­ thermocouple k · 12 ­ thermocouple n * – current input version ** – thermocouple servicing version

SF3

SF4

SF5

0°C

-99,0..999°C Value corresponding to 0 mA for channel 1

for 0..20 mA

SF6

100°C

-99,0..999°C Value corresponding to 20 mA for channel 1

for 0..20 mA

SF7

SF8

SF9

SF10

SF11

0°C

-99,0..999°C Value corresponding to 4 mA for channel 1

for 4..20 mA

SF12

100°C

-99,0..999°C Value corresponding to 20 mA for channel 1

for 4..20 mA

SF13

SF14

SF15

SF16

SF17

0,0°C

-20,0..20°C Temperature correction for channel 1

www.mikster.eu

User settings | 9

User manual INDU 20 v1.6(19)EN

Nr

Default

value

SF18

SF19

SF20

On

SF21

On

Range

Description

On / Off
On / Off

Negative pressure governor:

Off ­ always

On ­ only in the START mode

Temperature governor:

User settings

SF22

SF23

0

SF24

99

SF25

SF26

SF27

SF28

SF29

1

0..99%
0..99%
0..3

Off ­ always

On ­ only in the START mode

The lowest negative pressure value, which could be set by the user

The highest negative pressure value, which could be set by the user

Temperature governor coupled to relay REL 5

· 0 ­ simple hysteresis · 1 ­ reversed hysteresis · 2 ­ approaching hysteresis · 3 ­ PID

SF30

SF31

1

SF32

1

SF33

1

SF34

SF35

0

SF36

1,0°C

SF37

1,0°C

SF38

50°C

SF39

0..5 0..5 0..5
0..5 0,0..5,0°C 0,0..5,0°C -99..999°C

Lower hysteresis for the governor coupled to REL2 Lower hysteresis for the governor coupled to REL3 Lower hysteresis for the governor coupled to REL5 Upper hysteresis for the governor coupled to do REL2 Upper hysteresis for the governor coupled to do REL3 Upper hysteresis for the governor coupled to do REL5 Set temperature –

| 10

www.mikster.eu

User manual INDU 20 v1.6(19)EN

Nr

Default

value

SF40

SF41

20°C

SF42

SF43

SF44

1

SF45

SF46

1

SF47

HMD

SF48

5

SF49

1

SF50

1

SF51

°C

SF52 1 [min]

SF53

1

SF54

99

SF55 150°C

SF56

SF57

0

SF58

-99°C

SF59

SF60

Off

SF61

Off

Range

Description

0..99°C
0..100 sek 0..1
HMD/HM
0..10 godz.
1..360 min 1..360 min
°C/°F 0..99 [min] 0 .. 1
0..99% -99..999°C
0..99% -99..999°C
On / Off On / Off


Temperature of activation (Tza) of the governor coupled to REL 5 for the ,, temperature approach” algorithm


Time delay of the governor [seconds] coupled to REL 5

Recording: · 0 ­ continuous recording · 1 ­ recording only in the START mode
AUTOSTART mode ­ parameters format: · HMD ­ hour, minute and twenty-four hours delay before the START of the process · HM ­ number of hours and minutes remaining to the START mode
Maximal time (in hours) after which (when there was a power failure) the controller does not return to the START mode
Rate of the measurements recording
Rate of alarms recording
Temperature unit
Sound signal – duration. Note! When 0 value entered, the signal can be deleted by the OK key!
Alarm output mode of operation: · 0 ­ interrupted signal · 1 ­ continuous signal
Maximal permissible negative pressure (alarm)
Maximal permissible temperature (alarm)

Minimal permissible negative pressure (alarm)
Minimal permissible temperature (alarm)

Alarm activation ­ failure of the negative pressure sensor
Alarm activation ­ failure of the temperature sensor

User settings

www.mikster.eu

| 11

User manual INDU 20 v1.6(19)EN

Nr

Default

value

SF62

SF63

Off

SF64

Off

SF65

SF66

Off

SF67

Off

SF68

SF69

0

SF70

0

SF71

60

SF72

60

SF73

60

SF74

0

SF75

10

SF76

10

SF77

20

SF78

SF79

1

SF80

99

SF81

0

Range

Description

On / Off Alarm activation – Max negative pressure exceeded

On / Off Alarm activation – Max temperature exceeded

On / Off Alarm activation – Min negative pressure exceeded

On / Off Alarm activation – Min temperature exceeded

0..4

Control input 1

0 – alarm switched off

1 ­ alarm, when inputs 6-8 shorted

2 ­ alarm, when inputs 6-8 not shorted

3 ­ keyboard lockout when inputs 6-8 shorted

4 – keyboard lockout when inputs 6-8 not shorted

0..4

Control input 2

0 – alarm switched off

1 ­ alarm, when inputs 7-8 shorted

2 ­ alarm, when inputs 7-8 not shorted

3 ­ keyboard lockout when inputs 7-8 shorted

4 – keyboard lockout when inputs 7-8 not shorted

0..999 sek Time delay in the alarm signaling, when sensors are faulty

0..999 sek

Time delay in the alarm signaling, when the permissible settings are exceeded

0..999 sek

Time delay in the alarm signaling, when there is an alarm on inputs

0..9999

Change of the access code to SETUP

0 value ­ checking the code switched off

0..9999 sek Attach a turnover time of the relay LEFT

0..9999 sek Attach a turnover time of the relay RIGHT

0..9999 sek Pause time

-99.100

Negative pressure offset for the governor coupled to REL2

0..99

Set speed value corresponding to 20mA on the current output

0..2

Mode of operation of the REL5 relay :

· 0 ­ alarms signaling,

· 1 – temperature governor

· 2 – relays REL4, REL5 operate in a mode

RIGHT/LEFT turnover

User settings

| 12

www.mikster.eu

User manual INDU 20 v1.6(19)EN

Nr

Default

value

SF82

1

SF83

1

SF84

1

SF85

0

SF86

1

SF87

0

Range

Description

0..1

Response to alarm: measuring sensors faulty:

0 – signaling,

1 – process stops

0..1

Response to alarm: settings exceeded

0 – signaling,

1 – process stops

0..1

Response to alarm from the control input

0 – signaling,

1 – process stops

0..1

Type of the current output:

0: 0..20 mA

1: 4..20 mA

0..1

State of the relay output REL3 (aerating valve) when the

controller is not in the START mode

0: REL3 not controlled

1: REL3 controlled (aeration)

0..1

Unit processing time

0 ­ hours:minutes HH:MM

1 – minute:second MM:SS

User settings

www.mikster.eu

| 13

10. EXAMPLE OF APPLICATION*

User manual INDU 20 v1.6(19)EN

14
PORT1 (RS-485)

POWER

30

SUPPLY

+ 28

  • The example of application should give the user a general idea only, and can not be considered the project of the control system neither as total nor as its part.

| 14

www.mikster.eu

User manual INDU 20 v1.6(19)EN
NOTES