MIKSTER Loggicar-C Transport Recorder User Manual

June 10, 2024
MIKSTER

Loggicar-C Transport Recorder

Product Information

The Transport Temperature and Humidity Logger is a device used
to monitor and record the temperature and humidity levels during
the transportation of sensitive products. It is equipped with
sensors that provide accurate measurements and can be configured to
set alarm thresholds to alert users if the levels exceed the
desired range.

Technical Specifications

  • Version: v4.3

  • Measurement Range: Temperature: -40°C to 85°C (-40°F to 185°F),
    Humidity: 0% to 100%

  • Accuracy: Temperature: ±0.5°C (±0.9°F), Humidity: ±2%

  • Resolution: Temperature: 0.1°C (0.1°F), Humidity: 0.1%

  • Logging Capacity: 16,000 readings

  • Logging Interval: Programmable from 1 second to 24 hours

  • Display: LCD with backlight

  • Power Supply: Lithium battery (replaceable) with up to 2 years
    of battery life

  • Dimensions: 92 mm x 62 mm x 22 mm (3.62 in x 2.44 in x 0.87
    in)

Product Usage Instructions

Logger Installation

Before installing the logger, ensure that the sensors are
securely connected. The logger can be attached using the provided
adhesive strips or placed in a protective case for added
durability.

First Installation

  1. Press the up arrow button to turn on the logger.

  2. Use the arrow buttons to select your preferred language and
    press the enter button.

  3. Enter the current date and time using the keypad and press
    enter.

Connecting the Measurement Sensors

  1. Connect the sensors to the logger by plugging them into the
    designated ports.

  2. Press the up arrow button to view the sensor readings.

Adding (Logging in) Sensors Loggicar C

  1. Press and hold the enter button until the display shows “Add
    Sensor”.

  2. Enter the sensor ID number and press enter.

Logger Setting (Configuration)

The logger can be configured for customized use by following
these steps:

  1. Select “Logger Setting” from the main menu.

  2. Select the desired setting to be changed (date and time,
    logging frequency, header, footer or system language).

  3. Adjust the setting and press enter to save changes.

Sensor Configuration

The sensor configuration allows users to add or delete sensors
as well as set alarm thresholds and sensor labels:

  1. Select “Sensor Configuration” from the main menu.

  2. Select “Add Sensor” to add sensors or “Delete Sensor” to remove
    sensors.

  3. Select “Logged in Sensors List” to view all sensors currently
    logged in.

  4. Select “Set Alarm Thresholds” to adjust the desired temperature
    and humidity ranges and set alarm thresholds.

  5. Select “Sensor Labels” to add or edit sensor labels for easier
    identification.

Printouts

Users can generate reports for data analysis:

  1. Select “Printouts” from the main menu.

  2. Select the type of report to be generated (latest delivery
    report, delivery report from list, date-to-date period report or
    settings report).

  3. Follow the prompts to generate and print the report.

Delivery Logging

The logger can be set to record data during transportation:

  1. Select “Delivery Logging” from the main menu.
  2. Select “Start Logging” to begin recording data.
  3. Select “Stop Logging” to end recording and save data.

Connecting the Loggicar-C Logger to a Computer

The logger can be connected to a computer for data transfer:

  1. Insert the provided USB cable into the computer and connect the
    other end to the logger.

  2. Select “USB Communication” from the main menu and follow
    prompts to transfer data.

Transport temperature and humidity logger.
v4.3

Operating Manual Loggicar ­ C v4.3

Contents.

Introduction. ……………………………………………………………………… 2

Operation description…………………………………………………………… 2

Technical specifications. ……………………………………………………….. 3

Loggicar ­ C package contents. ……………………………………………….. 4

Logger construction……………………………………………………………… 6

Logger installation……………………………………………………………….. 6

Keypad with display …………………………………………………………….. 7

Display. …………………………………………………………………………….. 8

First installation. …………………………………………………………………. 9

10. Connecting the measurement sensors. …………………………………… 12

11. Adding (logging in) sensors Loggicar C …………………………………….. 14

12. Logger setting (configuration). ……………………………………………… 17

12.1 Date and time…………………………………………………………………… 17

12.2 Logging frequency (SAMPLE) ………………………………………………… 18

12.3 Header……………………………………………………………………………. 20

12.4 Footer…………………………………………………………………………….. 22

12.5 System language ……………………………………………………………….. 23

12.6 Initialization …………………………………………………………………….. 25

13. Sensor configuration. …………………………………………………………. 26

13.1 Adding (logging in) sensors ………………………………………………….. 26

13.2 Deleting sensors ……………………………………………………………….. 27

13.3 Logged in sensors list………………………………………………………….. 28

13.4 Setting alarm thresholds ……………………………………………………… 30

13.5 Sensor labels (Sensor name) ………………………………………………… 33

14. Printouts …………………………………………………………………………. 34

14.1 Latest delivery report …………………………………………………………. 35

14.2 Delivery report from list………………………………………………………. 36

14.3 “Date to date” period report ………………………………………………… 38

14.4 Settings report………………………………………………………………….. 40

15. Delivery logging ………………………………………………………………… 41

16. Connecting the Loggicar-C logger to a computer ……………………….. 42

17. Printout description …………………………………………………………… 44

18. Reservations and notes ………………………………………………………. 45

19. Copyrights ……………………………………………………………………….. 45

20. EC Declaration of Conformity ……………………………………………….. 46

1

Operating Manual Loggicar ­ C v4.3
1. Introduction.
LOGGICAR-C was designed for the purposes of monitoring parameters, such as temperature and humidity of products transported in refrigerated trucks ­ according to the Commission Regulation (EC) No. 37/2005.
Reading this manual will increase your satisfaction with the reliable operation, low cost and safe use of the product.
2. Operation description.
The logger has 4 measurement channels to connect temperature or temperature and humidity sensors and 4 digital channels to connect door opening sensors. The device provides full control over the refrigerated truck.
The logger is equipped with a report printer in order to print out a delivery report for the customer or control authorities.
The memory of the logger (non-volatile) enables storing ca. 32000 measurements. When the memory is full, older results are deleted to make space for newer ones (overwriting).
The contents of the memory can be downloaded and archived using the Loggisoft LC software (optional).
Breakdown of memory capacity depending on logging frequency is included in par. 3 TECHNICAL SPECIFICATIONS below.
The built-in OLED screen provides for easy operation and local display of measurements.
The software of the logger enables setting alert thresholds (lower/upper threshold) ­ if these values are exceeded, the display will blink and the event will be shown on the printed report.
Printing paper is the only operational cost generated by the logger.
2

Operating Manual Loggicar ­ C v4.3
3. Technical specifications.

Overall dimensions Weight
Protection rating Power supply*

105 x 75 x 45 mm 0.5 kg IP 21 12-24 V

From the lighter port

Battery power supply Battery life
Logger memory** Logging frequency
Printer Printing paper
Measurement channels
Digital channels

7.4V; 1500mAh 24 hours
98000 samples 1-180 min. thermal Fi=30 mm L=56 mm 4
4

Li-Ion battery Without printing
For temperature and humidity sensors
For door opening sensors

  • – In the case of connecting the logger directly to the car battery, an internal power supply unit, cat. no. C228 made by MIKSTER is required. ** – Example of memory capacity depending on logging (sampling) frequency.

3

Operating Manual Loggicar ­ C v4.3

Breakdown of logged data, depending on sampling frequency.

Sampling frequency 5 min 10 min 15 min 30 min

Logging days 340 680 1020 2041

4. Loggicar ­ C package contents.
Loggicar-C logger.

USB/mini USB cable: For PC connection to download recorded data (using the LOGGISOFT LC software).

4

Operating Manual Loggicar ­ C v4.3 LCE-01 sensors connection cable: For connecting sensors to the logger. LLC-01 car charger: Input voltage – 12V-24V DC, output voltage – 9V DC Battery: Li-ion 7.4V / 1500mAh. Thermal paper: width 56mm, diameter 30 mm.
5

5. Logger construction. Thermal paper dispenser

Operating Manual Loggicar ­ C v4.3
Keypad with display 9V power supply

LCE-01 block connection port

Mini USB port COM service port

6. Logger installation
The Loggicar ­ C logger is a mobile device, therefore a specific mounting point is not specified. The user should install the logger in such a way as to enable safe operation of the vehicle and the logger during driving and parking. An industrial hook-and-loop fastener is provided with the logger. The fastener does not guarantee stable and permanent installation of the device on the vehicle, it only prevent movement. The strength holding down the logger depends on the cleanness of the surface, its type and installation accuracy. If the logger does not seem appropriately fixed, it should be installed in a different place. The logger can be installed in car holders for mobile devices.

6

Operating Manual Loggicar ­ C v4.3 Recommended location for the industrial fastener

NOTE!
The hook-and-loop fastener should not be stuck on the battery cover ­ this may result in the cover falling off and unwanted movement of the logger.

7. Keypad with display Battery charge status LED

“Delivery start” button

“On/Off” button

“Delivery start” button

Display

“Delivery end” button

7

Operating Manual Loggicar ­ C v4.3
Note!
The buttons have the following additional functions.
– Open system menu, back to previous function
– confirm selection – increase the value, move the cursor “up” – decrease the value, move the cursor “down”

8. Display.
Currently displayed measurement

Current sensor reading

Battery level or Loggicar ­ R operating
temperature (displayed alternatively)

Clock

Event icon
8

Operating Manual Loggicar ­ C v4.3 Even icons:
Icon

Description Out of paper
Delivery Alarm Battery status

9. First installation.
– Insert the battery into the logger

Note! First charging is recommended to last until the green LED lights up on the
device keypad. The red LED informs you that the battery is charging. In order to maintain the best battery life, do not completely discharge the
battery.
– Insert the thermal paper into the dispenser.

9

Operating Manual Loggicar ­ C v4.3
Note! Thermal paper can only be printed on one side. If the print is not visible, turn the
paper roll around.
– Connect the logger to power supply
The LCC-01 car charger connected to the car lighter port is typically use for power supply.

There is an option to use the vehicle electrical wiring for power supply by using the LCPS-01 power supply unit
or the LCWPS-01 wall plug charger.

LCPS-01 power supply unit

LCWPS-01 wall plug charger

The LCPS-01 and the LCC-02 charger can be purchased separately.

Switch on the logger by pressing

.

10

Operating Manual Loggicar ­ C v4.3
After the display lights up, set up the logger by selecting basic system parameters, such as language, date and time.
Follow the instructions and drawings below.

Use the

buttons to select system language.

Confirm by pressing

.

Use the the system clock.

buttons to set the day, month, year, hour and minute of

Confirm each selection by pressing

.

11

Operating Manual Loggicar ­ C v4.3 Note! After these settings, the logger is ready for connecting the sensors.
10. Connecting the measurement sensors.
Connect the temperature sensor cable to the port in the logger enclosure.
Make the following connections, depending on the type of the sensors. Note!
Connect leads marked with A to the port with a similar A marking ­ connect the B lead in a similar fashion.
12

Operating Manual Loggicar ­ C v4.3 LCT-01 sensor connection method.
LCT (H)-02 sensors connection method.
Note! In order for the device to recognise the sensors, add the sensors to the list of logged
in sensors ­ see par. 11. 13

Operating Manual Loggicar ­ C v4.3
Installation notes.
The measurement element of the sensors should be installed in the cargo section of the vehicle in such a way as to limit the unwanted effects of the external factors on the accuracy of the measurement. In case of a one sensor layout, install the sensor in the 2/3 of the length of the cargo section, from the driver’s cab. In the case of a two sensor layout, install the sensors in the 1/3 and 2/3 of the cargo section. Suggested mounting height form the floor: 1.5 m. Allow at least 50 cm distance from the internal light source.

11. Adding (logging in) sensors Loggicar C
In case only one temperature or temperature and humidity sensor is connected to the device, the logger automatically adds the sensor to the list.
In case more sensors are connected to the logger, add sensors manually by following the procedure and drawings below. Selection:

Press

.

Use the

buttons to select the SENSOR menu.

14

Operating Manual Loggicar ­ C v4.3

Confirm by pressing

.

Use the

buttons to select the number of the sensor to be logged in.

Confirm by pressing

.

Note! If you cannot find the number of the sensor you wish to log in on the list, check the
connection of the sensors ­ see par. 10.
Description of sensor markings.

15

Operating Manual Loggicar ­ C v4.3
Sensor number
Sensor type T – temperature H – humidity D – digital sensor for logging events such as:
– door opening/closing – unit on/off – thawing on/off
Note! Sensor number is provided on its enclosure.
After following this procedure, the sensor is ready for use.
16

Operating Manual Loggicar ­ C v4.3
12. Logger setting (configuration).

12.1 Date and time

Selection: Press

.

Use the

buttons to select the SETTINGS menu.

Confirm by pressing

.

Use the

buttons to select the TIME menu.

Confirm by pressing

.

Use the the system clock.

buttons to set the day, month, year, hour and minute of

17

Operating Manual Loggicar ­ C v4.3

Confirm each selection by pressing

.

12.2 Logging frequency (SAMPLE)
Data logging in the device is carried out by regular storing of measured values in the non-volatile memory.
The period between each of the measurements is called the sampling frequency ­ SAMPLE.
In practice, the sampling frequency depends on the transport time and is usually in the range of 5­60 minutes. Selection:

Press

.

Use the

buttons to select the SETTINGS menu.

18

Operating Manual Loggicar ­ C v4.3

Confirm by pressing

.

Use the

buttons to select SAMPLE.

Confirm by pressing

.

Use the

buttons to set the sampling frequency.

Confirm by pressing

.

19

12.3 Header

Operating Manual Loggicar ­ C v4.3

The header is the text on top of the printout, usually containing information about the user, such as company name and vehicle registration number. The maximum length of the header is two lines of 32 characters each.
Note! Due to the size of the display, the text is entered in four lines of 16 characters.

Header

Selection: Press Use the

buttons to select the SETTINGS menu.

20

Operating Manual Loggicar ­ C v4.3

Confirm by pressing

.

Use the

buttons to select HEADER.

Confirm by pressing

.

Use the

buttons to enter the header text.

Confirm each letter by pressing

.

21

Operating Manual Loggicar ­ C v4.3
IMPORTANT: After entering the whole header text, press and hold for 2 seconds. The header will be stored in the memory.
12.4 Footer
The footer is the text at the end on the printout. The maximum length of the header is two lines of 16 characters each.

footer

Selection: Press Use the

buttons to select the SETTINGS menu.

Confirm by pressing

.

22

Operating Manual Loggicar ­ C v4.3

Use the

buttons to select HEADER.

Confirm by pressing

.

Use the

buttons to enter the footer text.

Confirm each letter by pressing

.

IMPORTANT: After entering the whole footer text, press and hold for 2 seconds. The footer will be stored in the memory.

12.5 System language
The LOGGICAR-C logger has several menu languages. The user can select the system language setting. 23

Operating Manual Loggicar ­ C v4.3 Note! The logger is set to the default language ­ English. The number of languages may be different, depending on the firmware version.
Selection:

Press

.

Use the

buttons to select the SETTINGS menu.

Confirm by pressing

.

Use the

buttons to select the LANGUAGE menu.

Confirm by pressing

.

Use the

buttons to select the system language.

24

Operating Manual Loggicar ­ C v4.3

Confirm by pressing

.

12.6 Initialization

NOTE! The initialization of the logger deletes all recorded data and sets all configuration
values to default. Starting the initialization procedure is protected by code (default code: 02010).
It is recommended to change the code to prevent unauthorised users from deleting data from the logger. Initialization does not reset the system clock.
Selection:

Press

.

Use the

buttons to select the SETTINGS menu.

Confirm by pressing

.

25

Use the

Operating Manual Loggicar ­ C v4.3 buttons to select INITIALIZATION.

Confirm by pressing

.

Use the

buttons to enter the security code.

Confirm each of the numbers by pressing

.

After entering the code, press

and hold for 2 seconds.

If the security code is correct, the memory of the logger will be irreversibly wiped and default settings will be restored.

13. Sensor configuration.

13.1 Adding (logging in) sensors
See paragraph 11.
26

Operating Manual Loggicar ­ C v4.3
13.2 Deleting sensors
This function enables the user to delete a sensor from the logged in sensors list. Selection:

Press

.

Use the

buttons to select the SENSOR menu.

Confirm by pressing

.

Use the

buttons to select DELETE.

Confirm by pressing

.

Use the buttons to select the number of the sensor to be deleted (logged out).

27

Operating Manual Loggicar ­ C v4.3

Confirm by pressing

.

Use the

buttons to enter the security code.

Confirm each of the numbers by pressing

.

After entering the code, press

and hold for 2 seconds.

If the security code is correct, the sensor will be deleted from the list. Default code: 01234.

13.3 Logged in sensors list

This function displays the list of logged in sensors. Selection:

28

Operating Manual Loggicar ­ C v4.3

Press

.

Use the

buttons to select the SENSOR menu.

Confirm by pressing

.

Use the

buttons to select LIST.

Confirm by pressing logged in sensors.

.Use the

buttons to display the list of

29

Operating Manual Loggicar ­ C v4.3
13.4 Setting alarm thresholds
A minimum and maximum value can be set for each of the logged in sensors. Exceeding the value will result in flashing of the display (alarm). Newer logger models also emit an acoustic signal.
Exceeding the threshold values is logged in the memory and marked on the printout.

Minimum value exceeded
Maximum value exceeded

Selection:

Press

.

Use the

buttons to select the SENSOR menu.

30

Operating Manual Loggicar ­ C v4.3

Confirm by pressing

.

Use the

buttons to select ALARM.

Confirm by pressing

.

Use the defined.

buttons to select the sensor for which the alarm should be

Confirm by pressing

.

Use the

buttons to select the parameter. On/off, maximum alarm

value, minimum alarm value.

31

Operating Manual Loggicar ­ C v4.3

Confirm by pressing

.

Note! When editing the On/off field, each pressing of the the alarm on or off.

button will switch

When editing the Min and Max fields, the parameter setting screen is displayed.

Use the

buttons to enter the parameter value.

Confirm each of the numbers by pressing

.

After entering the value, press

and hold for 2 seconds.

Note!
In order to enter the minus sign, set the cursor in the first position from the left and

press

.

32

Operating Manual Loggicar ­ C v4.3
13.5 Sensor labels (Sensor name)
In order to make the identification of the logged in sensors easier, there is a possibility of naming the sensors with 16-character labels.
Selection:

Press

Use the

buttons to select the SENSOR menu.

Confirm by pressing

.

Use the

buttons to select NAME

Confirm by pressing

.

Use the defined.

buttons to select the sensor for which the name should be

33

Operating Manual Loggicar ­ C v4.3

Confirm by pressing

.

Use the

buttons to enter the name.

Confirm each letter by pressing

.

After entering the name, press stored in the memory.

and hold for 2 seconds. The name will be

14. Printouts
One of the basic features of the LOGGICAR-C logger is printing out reports with recorded data.
The user can select the following types of reports:
– delivery report
– “date to date” period report

34

Operating Manual Loggicar ­ C v4.3 – device configuration report
14.1 Latest delivery report
Printing out the report from the last delivery is a quick form of reporting during of after the delivery. The procedure of starting ad ending the deliveries is described in paragraph 10. If the delivery is completed, the report will include all data from the loading to the completion of delivery. If the printout is required during the delivery, it will include the data from the loading to the current time. Selection:
Press

Use the

buttons to select the DELIVERIES menu.

35

Operating Manual Loggicar ­ C v4.3

Confirm by pressing

.

Use the

buttons to select LATEST

Confirm by pressing

.

If there are any deliveries stored in the memory of the device, the report will be printed.

14.2 Delivery report from list
Printing the report from the list of deliveries enables the user to select any delivery recorded in the memory of the logger.
Selection:

Press

36

Operating Manual Loggicar ­ C v4.3

Use the

buttons to select the DELIVERIES menu.

Confirm by pressing

.

Use the

buttons to select LIST.

Confirm by pressing

.

Use the

buttons to select the delivery start date from the list.

37

Operating Manual Loggicar ­ C v4.3

Confirm by pressing

.

14.3 “Date to date” period report
Note! The LOGGICAR-C logger records data from the moment of logging in the sensors, regardless of starting the delivery.
The DATE printing option prints out a date to date report.

Selection: Press

Use the

buttons to select the DATE menu.

38

Operating Manual Loggicar ­ C v4.3

Confirm by pressing

.

Use the

buttons to enter the start date and time for the report.

Confirm each of the numbers by pressing

.

Confirm further by pressing

and holding for 2 minutes.

Do the same for the end date and time.

If the logger recorded any deliveries in the given period, it will print out the report.

39

Operating Manual Loggicar ­ C v4.3
14.4 Settings report
The Settings report option enables the user to print out the configuration of the logger. Selection:
Press

Use the

buttons to select the SETTINGS menu.

Confirm by pressing

.

40

Operating Manual Loggicar ­ C v4.3

15. Delivery logging
The “delivery” means the transport process from loading to unloading.

In order to mark the start of delivery, press

.

In order to mark the end of delivery, press

.

If the logger is in the delivery mode, the following icon will be shown on the display.

Delivery icon 41

Operating Manual Loggicar ­ C v4.3
16. Connecting the Loggicar-C logger to a computer

In order to connect the logger to a computer, use a USB/mini USB cable. Insert the

USB plug

into a free USB port of a PC computer and insert the miniUSB

plug

to the USB port in the logger.

Switch on the Loggicar-C logger after connecting the cable. 42

Operating Manual Loggicar ­ C v4.3 The computer will install drivers required for correct operation of the logger.
If the computer cannot find the drivers, download it from www.mikster.pl, section: software tools.
43

17. Printout description
Printout:
Sensor type and number

Operating Manual Loggicar ­ C v4.3
Header, no more than 4 lines.
Date and time of delivery start and end

Sampling frequency
Measurement

Door opened marker
Printing date and time

Footer, no more than 4
lines.
Deliveries: The delivery menu enables the user To print the last delivery or select it from the list (previous deliveries). Date: There is also a possibility to print by date ­ input the start and end time and select a delivery completed within the time frame.

44

Operating Manual Loggicar ­ C v4.3
18. Reservations and notes
– The LOGGICAR logger is a professional device and should be used according to its intended purpose – do not use a device with visible mechanical damage, such as broken enclosure, damaged wire insulation, broken ports, etc.
19. Copyrights
Copyright reservations All rights reserved. No part of this document can be copied, transmitted, processed, recorded in data recovery systems or translated to other languages or computer language in any form or using any electronic, mechanical, magnetic, optical, chemical, manual or other methods without prior written consent of MIKSTER Sp. z o.o.
Exclusion of liability MIKSTER Sp. z o.o. does not provide any indemnities or warranties, whether direct or implied, with regard to this document, an in particular ­ any implied guarantees or commercial value or usability for a specific purpose. Furthermore, MIKSTER Sp. z o.o. reserves the right to amend this document and introduce changes at any time without the requirement to notify anyone of the amendments or changes. All other logos, products or company names mentioned in this document may be registered trademarks or covered by copyright claims of other companies and are used in the document solely for informative purposes. MIKSTER and LOGGINET word and device are registered trademarks.
45

20. EC Declaration of Conformity

Operating Manual Loggicar ­ C v4.3

46

Operating Manual Loggicar ­ C v4.3 47

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals