Microtech Depth Gauge EE Instructions
- August 17, 2024
- microtech
Table of Contents
- Microtech Depth Gauge EE
- Product Information
- Product Usage Instructions
- Description
- Operating features of the instrument
- Basic functions
- Advanced functions
- Automatic references
- Bluetooth® configuration
- Favourite key
- Switching off
- List of the main command
- Specifications
- FCC
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Microtech Depth Gauge EE
Product Information
Specifications:
- Battery: lithium 3V, type CR2032
- Frequency Band Modulation: 2.4GHz (2.402 – 2.480GHz) GFSK (Gaussian Frequency Shift Keying)
- Max Output Power: Class 3: 1mW (0dBm)
- Range: Open space: up to 15m, Industrial environment: 1-5m
- Battery Life:
- Continuous: up to 2 months – Always connected with 4 values/sec.
- Saver: up to 5 months – The instrument sends value only when the position has changed.
- Blind/Push: up to 7 months – Value is sent from the instrument (button) or requested from the computer.
Product Usage Instructions
Operating Features of the Instrument
The instrument has two operating modes: basic functions and advanced
functions. You can select references, work in Automatic Reference mode, and
enter a multiplication factor.
Start
Press the MODE button to start the instrument.
Basic Functions
Short press on MODE gives direct access to basic functions like selecting
references and inputting preset values.
Advanced Functions
Prolonged press on MODE accesses advanced functions such as unit selection,
choice of measurement direction, and multiplication factor input.
FAQ:
-
Q: How do I change the measuring direction?
A: To change the measuring direction, a displacement of more than 0.2mm in the opposite direction is required. -
Q: How can I clear pairing information?
A: To clear pairing information, navigate to the rESEt menu and select the option to clear pairing information.
Description
- Support
- Perche
- Movable cursor
- MODE button
- Favourite button
- SET button
- Base
- Measuring button (interchangeable)
- Battery compartment or power cable
- Clamping screw
- Unit of measurement (mm/INCH)
- +/- indicator
- Low battery
- Freezing the measured value
- Preset mode
- Active reference
- Locking the buttons
- Sending data
- Bluetooth® connection
- Display – 6 digits
- Multiplication factor /Ref Auto
Operating features of the instrument
-
The instrument has two operating modes: basic functions (direct access) and advanced functions. In addition to the configuration functions, you can select 2 references, or work in Automatic Reference mode (details see chapter 5). You can also enter a multiplication factor (see chapters 3 and 4).
-
The «favourite» key gives direct access to the function used most often (see chap. 7).
-
Sets a Preset value, verifies a selection, and controls switching off the instrument. By default, SIS mode enables automatic switch-off with no loss of origin (see chap. 8)
-
Personalising the functions
It is possible to activate or de-activate certain functions of the instrument via Power RS/USB cable, or Bluetooth® (see chap. 10). -
Data transmission parameters 4800Bds, 7 bits, even parity, 2 stop bits.
Start
Press a button.
For a Bluetooth® connection (see chap. 6).
Basic functions
Each short press on gives direct access to the basic functions:
- rEF Selection of the reference (1 to 2), or Automatic references (see chap. 5)
- PrE Inputting a Preset value next digit 0…9 save PRESET
- bt Bluetooth® Enable / disable, reset Bluetooth® module or display its MAC address.
Advanced functions
Prolonged pressure (>2s) on gives access to the advanced functions.
Then, each short press on accesses the required function:
- Unit Unit selection (mm or inch)
- dir Choice of measurement direction (positive or negative direction)
- Mult Multiplication factor, enable or disable multiplication factor (value can be modified if On is confirmed by button)
- Input of multiplication factor, next digit 0….9 save MULT
- CSt Introduction of a constant value (see chapter 5)
- OFF Automatic switch-off mode / MAn = disabled, Auto = active (after 10 min. by default).
- bt.CFG Bluetooth® profile selection. (see chap. 6 for details) The + sign indicates the currently active profile.
- Loc Keypad lock Only the favourite key remains active.(to unlock the keypad, press for 5 sec)
Automatic references
Depending on the application, when reversing the measuring direction, an offset value can be managed to compensate for the dimensions of the measuring keys.
To use this operating mode, select the rEF menu to Auto.
The value of the measuring key constant must first be entered in the CSt menu.
Note:
- In Auto Reference mode, the preset value entry is assigned to the active reference of the measuring direction:
- For the change of measuring direction to take place, a displacement >0.2mm in the opposite direction is required.
Bluetooth® configuration
The connection procedure has been designed to be simple and is signalled by the following three states:
- Symbol off ………….. disconnected mode
- Symbol blinking …… advertising mode
- Symbol on ………….. connected mode
The following options can be selected to control the Bluetooth® module.
- On Enable Bluetooth® module (start advertising mode).
- OFF Disable Bluetooth® module (terminate active connection).
- rESEt Clear pairing information.
- MAC Display the MAC (Media Access Control) address.
Three Bluetooth® profiles are available.
- SIMPLE Profile without pairing (default).
- PAIr Paired and secured profile.
- HId Virtual keyboard mode (compatible with recent equipment without driver installation).
Note: Bluetooth® pairing information is cleared when the profile is changed.
Connection:
- Activate Bluetooth® compatible software and hardware (Master: PC, Display Unit).
- Start the instrument. By default the Bluetooth® module is active and the instrument is available for connection (advertising mode).
- If no connnection is established during the advertisement period reactivate the Bluetooth® module using the bt / On menu.
- Instrument is ready to communicate (connected mode.)
Only with paired profile:
Pairing with master is automatically done at first connection. To connect the
instrument to a new master (new pairing), pairing information on the
instrument must be cleared using the bt / rESEt menu.
Bluetooth® specifications
Frequency Band | 2.4GHz (2.402 – 2.480GHz) |
---|---|
Modulation | GFSK (Gaussian Frequency Shift Keying) |
Max Output Power | Class 3: 1mW (0dBm) |
Range | Open space: up to 15m Industrial environment: 1-5m |
Battery life | Continuous: up to 2 months – Always connected with 4 values / |
sec.
Saver: up to 5 months – The instrument sends value only when the position has changed.
Blind/Push: up to 7 months – Value is sent from the instrument (button) or requested from the computer.
Other specifications on the manufacturer’s website.
Favourite key
The «favourite» key gives direct access to a predefined function, and can be configured according to the needs of the user. In order to assign a function to the «favourite» key, give a prolonged press on , and then select the required function:
Validation of selection: By a prolonged press on or a short press on or .
Note:
- A function can also be assigned via RS232 using the command <FCT + Function No.> (FCT 0..9 A..F)
Example: Unit change=, no function = .
Switching off
The dial gauge goes automatically into stand-by if not used for 10 minutes,
unless automatic switch-off mode has been turned off (see Chap. 4, advanced
functions).
Stand-by mode can be forced by a prolonged press (> 2 sec) on :
In stand-by mode, the value of the origin is retained by the sensor (SIS mode), and the instrument automatically restarts with any movement of the measurement probe, RS command, Bluetooth® request or press on button. The instrument can be switched off completely for a long period of nonuse, but this will necessitate a zero reset on restart (the origin will be lost):
Prolonged press (>4 sec) on :
Re-initialising the instrument
The initial instrument settings can be restored at any time by a prolonged
press (>4 sec) simultaneously on and until the message rESEt is displayed.
Personalising the instrument
Access to the functions of your instrument can be personalised, for more
information see manufacturer’s website (requires you to connect your
instrument via Power RS / USB cable, or Bluetooth®).
- Possibilities:
- Enable or disable the required functions.
- Modify access to the advanced functions (direct access).
Connecting the instrument
The instrument can be connected to a peripheral via a Power (RS or USB)
cable or Bluetooth®. See page 4 for connecting the Power cable. Measured
values can be transmitted and the instrument driven using predefined commands
(see chap. 12 for a list of the main commands).
List of the main command
Selection and configuration
- CHA+ / CHA- Change measurement direction
- FCT0 …9…A…F Assign «favourite» function
- MM / IN Change measurement unit
- KEY0 / KEY1 Lock / unlock keypad
- MUL [+/-]xxx.xxxx Modify multiplication factor
- PRE [+/-]xxx.xxx Modify preset value
- STO1 / STO 0 Activate / de-activate HOLD
- ECO1 / ECO 0 Activate / de-activate economic mode
- LCAL dd.mm.yy Modify last calibration date
- NCAL dd.mm.yy Modify next calibration date
- NUM x…x (up to 20chars) Modify the number of the instrument
- UNI1 / UNI0 Activate / de-activate change of units
- OUT1 /OUT0 Activate / de-activate contin. data transmission
- PRE ON / PRE OFF Activate / de-activate Preset function battery
- PRE Recall Preset
- SET Zero reset
- REF1/REF2 Change of active reference
- CST [+/-]xxx.xxx Introduction of the constant value
- REFAUTO1 / REFAUTO0 Activate / de-activate automatic reference
- SBY xx xx number of minutes before stand-by
- BT0/BT1 Activate / de-activate Bluetooth® module
- BTRST Clear pairing information
Interrogation
- ? Current value?
- CHA? Measurement direction?
- FCT? «favourite» function active?
- UNI? Measurement unit active?
- KEY? Keypad locked?
- MUL? Multiplication factor?
- PRE? Preset value?
- STO? Status of HOLD function?
- ECO? Current economic mode
- LCAL? Date of last calibration?
- NCAL? Date of next calibration?
- NUM? Instrument number?
- SET? Main instrument parameters?
- ID? Instrument identification code?
- CST? Valeur de constante ?
- REFAUTO? Référence automatique ?
Maintenance functions
- BAT? Battery status (BAT1 = OK, BAT0 = low battery)
- OFF Switch-off (wake up using a button or RS)
- RST Re-initialisation of the instrument
- REF? Active reference ?
- SBY Put instrument in stand-by (SIS)
- VER? Version No. and date of firmware
- MAC? Bluetooth® MAC address ?
Specifications
Measuring range | 300 mm / 12’’ | 600 mm / 24’’ |
---|---|---|
Total measuring range | 335 mm / 13.2’’ | 625 mm / 24.6’’ |
Resolution | 0.01 mm / .0005’’ | |
Accuracy | 30 µm / .0012’’ | 40 µm / .0015’’ |
Repeatability | 10 µm / .0004’’ (±1 digit) | |
Max. travel speed | >2 m/s / > 80’’/s | |
No. of measurements per second | Up to 10 mes/s | |
Units of measurement | Metric (mm) / English (Inch) (direct conversion) | |
Maximum preset | ±999.99mm / ±39.9995 IN | |
Measuring system | Sylvac inductive system (patented) | |
Power supply | 1 lithium battery 3V, type CR 2032, capacity 220mAh | |
Average autonomy | 8’000 hours (with Bluetooth® switched on, see chapter 6.1) | |
Data output | RS232 / Bluetooth® 4.0 compatible (see chapter 6) | |
Working temperature (storage) | +5 à + 40°C (-10 à +60°C) | |
Electromagnetic compatibility | According to EN 61326-1 | |
IP specification (electronic unit) | IP 54 (according to IEC60529) | |
Weight | 440g | 550g |
CERTIFICATE OF CONFORMITY
We certify that this instrument has been manufactured in accordance with our
Quality Standard and tested with reference to masters of certified
traceability by the Federal Institute of metrology.
Calibration certificate
Because we make our instruments in batches, you may find that the date on your
calibration certificate is not current. Please be assured that your
instruments are certified at point of production and then held in stock in our
warehouse in accordance with our Quality Management System ISO 9001. Re-
calibration cycle should start from date of receipt.
The Bluetooth® word mark and logos are registered trademarks owned by the Bluetooth® SIG, Inc. and any use of such marks by Sylvac is under license. Other trademarks and trade names are those of their respective owners.
U.S./Canada Certification
NOTICE: Changes or modifications made to this equipment not expressly approved by Sylvac may void the FCC aut horization to operate this equipment.
FCC
NOTICE: This device complies with Part 15 of the FCC Rules and with
RSS-210 of Industry Canada. Operation is subject to the following two
conditions.
(1) this device may not cause harmful interference, and
(2) this device must accept any interference received, including interference
that may cause undesired operation.
NOTE: This equipment has been tested and found to comply with the limits
for a Class A digital device, pur suant to Part 15 of the FCC Rules. These
limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environment. This
equipment generates, uses, and can radiate radio frequency energy and, if not
installed and used in accordance with the instruction manual, may cause
harmful interference to radio communications. Operation of this equipment in a
residential area is likely to cause harmful interference in which case the
user will be required to correct the interference at his own expense.
Radiofrequency radiation exposure Information:
This equipment complies with FCC radiation exposure limits set forth for an
uncontrolled environment. This equipment should be installed and operated with
minimum distance of 20 cm between the radiator and your body.
This transmitter must not be co-located or operating in conjunction with any
other antenna or transmitter.
Brazil Certification
Description:
This module is based on Nordic Semiconductor nRF8001 μBlue Bluetooth® Low
Energy Platform. The nRF8001 is a single chip transceiver with an embedded
baseband protocol engine, suitable for ultra-low power wireless applications
conforming to the Bluetooth® Low Energy Specification contained within v4.0 of
the overall Bluetooth® specification. The nRF8001, used in the current
revision of ISP091201, is a production product using a RoM for the baseband
protocol engine.
Changes without prior notice:
Edition: 2020.11 / 681-273-07
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>