PRAXISDIENST 0009-023-000268-00 H3 How to connect external wifi and PCECG Owner’s Manual
- August 19, 2024
- PRAXISDIENST
Table of Contents
H3 How to connect external wifi and PCECG
We need to prepare the below parts, buy from COMEN.
H3 external module| |
---|---|---
0009-023-000268-00| external wireless network card| 1 50usd
023-000089-00| network card supporting network cable| 1
043-000665-01| TP-LINK wireless module holder| 1 100usd
POE switches and AP routers
023-00000055-00| TL-AP1202C-POE| 1 50usd
023-00000075-00| five port Poe Gigabit switch \tp-link\tl- sg1005p 1| 1
023-000032-00| COMEN PC-based ECG management software with soft dog| 1 100usd
115-003596-00
| |
Step:
first, ECG machine connected with AC power connect external wireless network card(wifi module)with ECG machine, like picture 2
POE switches adapter connect with power, Other side:
One port (a) of the POE switch adapter connects to the PC internet, and one port (b) connects with the TP-link switcher.
4. connect the WIFI module and TP-LINK switcher with the POE adapter. Like the below picture:
5. H3 ECG machine side NET set and port setting
Set IP address Make sure the IP section setting on your computer is the same
as the IP section on the wireless router
Network type: line
Transfer Type:ECG network
6. computer settings (Windows XP system)
1、 Set IP address Make sure the IP section setting on your computer is the
same as the IP section on the wireless router
“Run”, then input “CMD”)
then we use “Ping” tool to check the connection between ECG and PC, PC and TP-
LINK, and check if it is connected well.
INPUT: ping TP-LINK IP address. Ping WIFI module IP address. If the connection
succeeds, it will receive all data.
8. Install PCECG in PC.
After installation, set in PCECG soft side. insert the soft dog on the PC main host. And open the PCECG program.
9. Click “New” and fill in patient information and then Enter PCECG main
interface.
Finally, wait a few seconds to connect.
10. Then we set it on the PCECG side. Click Config. Enter system setup.
set server IP: 200.200.200.100
Server port: 5065, transfer type: SCP-ECG
connection port : 5065
after we see TCP already connected with PC-ECG machine below. Means we success.