Rotronic CP-11 Portable IAQ Meter Instruction Manual
- August 9, 2024
- ROTRONIC
Table of Contents
- INTRODUCTION
- MATERIAL SUPPLIED
- POWER SUPPLY
- LCD DISPLAY
- KEYPAD
- OPERATION
- AUTO POWER OFF
- SETUP
- CO2 CALIBRATION
- RH CALIBRATION
- TROUBLESHOOTING
- Can’t power on
- Fixed readings
- Slow response
- Error messages
- PC CONNECTION
- SPECIFICATION
- CO2 LEVELS AND GUIDELINES
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Rotronic CP-11 Portable IAQ Meter Instruction Manual
INTRODUCTION
Thank you for purchasing this portable IAQ meter. The meter measures CO 2 level, air temp., dew point, wet bulb temp. and humidity and is an ideal instrument for indoor air quality (IAQ) diagnosis. Poor indoor air quality is considered unhealthy because it causes tiredness, loss of ability to concentrate, and even illness(ex. Sick Building Syndrome). IAQ monitoring and survey, especially on CO 2 level and air ventilation become widely applied in public areas such as offices, classrooms, factories, hospitals and hotels. It is also suggested in regulations of industrial hygiene in some countries. (Appendix)
The portable CO meter uses NDIR (non- 2 dispersive infrared) technology to ensure the reliability and long term stability. It’s useful in verifying HVAC system performance and air ventilation control.
MATERIAL SUPPLIED
This package contains:
- Meter
- 4pcs AA batteries
- Operation manual
- This package contains:
- Mini USB cable and software CD
- Soft carrying case
- Big with blue backlight to use in dark area
- One touch to display CO2 /Dew point temperature/Wet bulb temperature/ Air temperature/Humidity in turns.
- Designed with NDIR (non-dispersive infrared) Waveguide technology CO2 sensor
- Programmable warning CO2 level
- Long time.
POWER SUPPLY
The meter is powered by either 4 AA batteries or a DC adaptor(5V/1A output).
Install the batteries into the battery compartment on the bottom and make sure they are in correct polarity and good contact. When an adaptor is used, it will cut off the power supply from batteries. The adaptor can’t be used as a battery charger.
When battery voltage gets low, will appear on the LCD Fig 1.
Fig.1
The CO sensor can’t work normally under low voltage, please replace with fresh batteries or connect with an adaptor.
LCD DISPLAY
Symbols
- UpperLCD CO2/Temp/RH…..reading
- LowerLCD Real time display
- CO2 Carbon dioxide reading
- HOLD Readings are freezed unchanged
- MIN/MAX Minimum/Maximun readings
- AVG Average reading
- Low battery indicator
- DP Dew point temperature
- TEMP Air temperature
- WBT Wet bulb temperature
- %rh Unit of relative humidity
- oC/oF Celsius/Fahrenheit of temp.
- CAL In calibration status
- REC In manual/automatic logging
- RECALL In manual records recall mode
KEYPAD
-
Turns on and off the meter.
Enters setup mode while meter is off.
Turn on with to become non-sleep mode -
Exits setup/recall page.
Start automatic logging. -
Press to switch displayed mode.
Long press to enter memory recall mode. -
Freezes the current readings.
Selects unit or increases value in setup.
Cancels data hold function. -
Press to manually record the reading.
Selects unit or decreases value in setup. -
Activates MIN,MAX,AVG function.
Saves and finishes settings.
OPERATION
POWER ON/OFF
Press to turn the meter on and off. At power up, it emits a short beep and performs 30 seconds countdown(Fig.2) for meter warm up, then enters normal mode with current CO2 and real time displayed (Fig.3). The real time display date and time in turns.
Fig.2
Fig.3
Suggest to wait for 30 seconds warm up time. For quick start, you may press for 2 seconds to end the warm up and entering normal display.
TAKING MEASUREMENT
The meter starts measurement when power on and update readings every second. In the condition of operating environment change (ex. from high to low temp.), it takes 30 sec to respond for CO2 sensor and 30 minutes for RH.
NOTE: Do not hold the meter close to faces in case exhalation affects CO2 levels.
CO2 (Carbon Dioxide)
Press to switch the mode to CO2. User can get the CO2 reading in ppm unit on main display.(Fig.4). The lower display shows the real time clock.
Fig.4
Fig.5
AIR,DP,WBT
Press to switch to temperatures display(Fig.5,6,7).The lower display shows the real time clock.
Fig.6
Fig.7
Humidity
Press to switch to humidity display. (Fig.8).The lower display shows the real time clock.
Fig.8
DATA HOLD
In normal display mode, press to freeze the readings, “HOLD” icon is flashed on the left top of LCD(Fig.9). All current readings are kept unchanged. Press again to cancel the hold function.
Fig.9
NOTE:
THE HOLD FUNCTION IS DISABLE IN MIN/MAX/ AVG MODE.
BACKLIGHT
The backlight will be activated for 10 seconds by pressing any key.
MIN,MAX,AVG,CURRENT REVIEW
This meter allows you to check the minimum, maximum, average, current value from the moment you press the key. Under normal mode, press to see the minimum, maximum, average and current value in turns. Each press of ,it displays MIN, MAX, AVG, current value in sequence. (Fig.10~12). To return to normal mode, press key to escape (Fig.13)
Fig.10
Fig.11
Fig.12
Fig.13
In this max/min/avg/current modes, it shows the corresponding readings on main display and the accumulated time of how long you are in this mode.
Parameter you choose to see on main display.
Fig.15
If the reading changes quickly, you can press to freeze the reading before manually recording the data.
The manually recording function is disabled in min/max mode.
99 MEMORIES RECALL
The meter features 99 point memories review function.
In
ALARM
The meter features audible alarm to give warnings when CO2 concentration exceeds the limit. (See P20 in setup for setting alarm threshold). It emits beeps(Abt.80dB)when CO2 level goes.
To terminate data logging, press for 2 seconds, REC icon stops flashing
While press to start the logging again, the previous logging records are overwrite.
Press key to switch to the mode you want to see. Except the mode switch function, min/max, manual record, hold and recall functions are all disabled during logging.
AUTO POWER OFF
The meter turns off automatically after 20 minutes of inactivity. To override the function, hold down and for 2 seconds to turn on the meter until “n” appears.(Fig.20)
NOTE:
AUTO SLEEP FUNCTION WILL BE DISABLED DURING CALIBRATION MODE.
Fig.20
Fig.21
SETUP
When the meter is off, hold down for more than 1 sec to enter setup mode. START To exit setup, press in P10 ~ P60 and it returns to normal mode. (Fig.21)
P10: 99 MEMORIES CLEAR
When entering setup mode, P10 and “CLr” (Fig.21) are displayed on the LCD. Press to go into P11 and decide to clear or keep all the manual records. The current set will be blinking on LCD (Fig.22).
Press or to choose NO or YES and press MIN M A X /A V G to confirm.
Fig.22
Press to escape and return to P10.
P20: CO2 ALARM
When entering setup mode, P20 and “ALAr” (Fig.23) are displayed on LCD. Press to go into P21 for setting CO2 alarm threshold. The current set value will be blinking on LCD(Fig.24). Press to increase the value or to decrease. The selectable alarm limits are 100~9900ppm,each press tunes 100ppm. When the preferred alarm value is set, press to save the setting or without saving and return to P20.
Fig.23
Fig.24
P30 TEMPERATURE UNIT
When entering setup mode, P30 and “unit” (Fig.25) are displayed on the LCD. Press to go into P31 and decide the temperature unit.
The current set will display on LCD (Fig.26).
Press or to choose C or F and press M A X /A V G to confirm.
Fig.25
Fig.26
Press to escape and return to P30.
P40:LOGGING SAMPLING RATE
Go into P40 for setting sampling rate of data logging(Fig.27). The range is from 1 second to 4 hours 59 minutes and 59 seconds.
Press and it goes into setting with blinking Hour digits on the lower display. To change the digit, press to increase and to decrease. Press to confirm and enter Minute setting . Press again to confirm and enter MIN Second setting (Fig.28). Press to confirm the rate setting or without saving and return to P40.
Fig.27
Fig.28
P50 PRESSURE COMPENSATION
When entering setup mode, P50 and “Pres” (Fig.29) are displayed on the LCD. MIN
Press to set pressure compensation value for CO2 measurement.
The current set will flash on LCD (Fig.30).The barometric pressure unit is kpa.
Press or to adjust the pressure value and press to confirm.
Fig.29
Fig.30
Press to escape and return to P50.
P60:REAL TIME CLOCK
Go into P60 for setting the real time clock of this meter.
Press and it goes into P61 to set the time format as 12hour or 24 hour. The current set will flash on LCD. Press or to change the format and press to confirm and entering real time clock setting.(Fig.31)
The blinking Year digits on the lower display. To change the digit, press or to adjust. Press to confirm and enter Month setting. Repeat above steps to complete the month/date/hour/ minute/second setting.(Fig.32)
Fig.31
Fig.32
CO2 CALIBRATION
The meter is calibrated at standard 400ppm CO2 concentration in factory. It’s suggested to do manual calibration regularly to maintain good accuracy.
Note:
When the accuracy becomes a concern after a long time usage, return to dealers for standard calibration.
CAUTION: Do not calibrate the meter in the air with unknown CO2 concentration. Otherwise, it will be calibrated as 400ppm by default and leads to inaccurate measurements.
The manual calibration is suggested to to be done in fresh outdoor air that is well ventilated and in sunny weather.
Place the meter in the calibration site. Press , simultaneously for 3 seconds to turn on the meter and enter calibration mode (Fig.33). Two calibration menu is available: Humidity /CO2.
Press key to select the CO2 calibration. 400ppm and “ CAL are blinking on the LCD. Press to start calibration. (Fig.34)
Wait about 10 minutes until the blinking stops and the calibration is completed automatically. To abort the calibration, turn off the meter at any time.
NOTE:
Ensure the batteries are with full voltage during the calibration to prevent from interruption or failed calibration.
Fig.33
Fig.34
RH CALIBRATION
The meter defaults to be calibrated the humidity with 33% and 75% salt solution. The ambient condition is recommended to be at 25oC and stable humidity(better to be close to the calibrating value). To abort calibration, just turn off the meter.
CAUTION:
Do not calibrate the humidity without the default calibration salt. Otherwise, it will cause permanent damage. Contact the dealer for calibration salt or services.
33% calibration
Plug the sensor probe into 33% salt bottle. Press , , simultaneously for 3 seconds to turn on the meter and enter calibration mode (Fig. 33.). Hold down to enter 33% calibration. ¡§CAL¡¨ and calibrating value (32.8% if at 25oC) are blinking on the LCD.
Meter is now calibrating, and will finish in about 60 minutes when ¡§CAL¡¨ and humidity stop blinking. To abort the calibration, turn off the meter at any time.
75% calibration
Plug sensor probe into 75% salt bottle and entering calibration mode. In calibration mode, press to select 75% calibration(Fig.35), press to start.
“CAL” and calibrating value (75.3% if at 25oC) are blinking on the LCD.
Meter is now calibrating, and will finish in about 60 minutes when “CAL” and humidity stop blinking. To abort the calibration, turn off the meter at any time.
Fig.35
TROUBLESHOOTING
Can’t power on
Press ON/OFF for more than 0.3 seconds and try again. Check whether batteries are in good contact and correct polarity, or the adaptor is well plugged. You may also remove the batteries for > 10 mins and then install the batteries again.
Fixed readings
Check whether data hold function was activated. (HOLD icon at the left top)
Slow response
Check whether the air flow channels on the rear were blocked.
Error messages
E01/E33: CO2 sensor is out of order. Retry CO2 calibration. E02: The value is under range. E03: The value is over range. E04: The original data error results in this error (DP, WB) E11: Retry humidity calibration. E31: Temperature sensor or AD damaged. E32: Memory IC damaged. E33: Humidity sensor or circuit damaged.
PC CONNECTION
Connect the logger to PC, users can do logging setting and data transmission with the included software.
Plug the USB cable into the socket at the right side of the meter and the other port to PC. Then install the software in your PC with the following procedure.
-
Insert the CD Rom and run installation. Select a preferred directory and click “Next” step by step and finish it.
-
After the software installation is completed, it will run USB driver installation next.
Software Operation
AutoConnect
Start the software and it will detect logger connection automatically and indicate the COM Port information at the bottom of the main screen as well as the COMPort setting column.
Logger Settings
To set up logging plan, click “Setting” icon and select “Logger”. The setting page is opened.
Users can set up the logger identifier number. Click OK for setting and Exit the screen to confirm the settings.
Data Transmission
To transmit auto recorded data from the meter, click “Download” icon on the left top side of the main screen. All auto recorded data in the logger will be transmitted. The raw data with time stamp will display at the lower part of the screen and the graph in the middle.
In the Graph display, the Y-axis indicates Temp, RH,DP, WBT, Co2 level in different line colors. And the X-axis can be switched to show Time or Recorded No. Stamp.
Data Review Function
The following functions help to view the GRAPH data in more detailed way.
Under View function, there are 4 tools to enlarge the Graph data for detailed data review.
Zoom in: Hold down “Ctrl” key on the PC keyboard and click any target point on the graph. It enlarges the point in each click.
Zoom window : Hold down “Ctrl” key and drag click the left button of the mouse to select an area on the Graph and the selected area will be enlarged .
Zoom Xaxis : Hold down “Ctrl” key and drag click the left button of the mouse to select an area on the Graph and the X-axis of the selected area will be enlarged.
Zoom Yaxis : Hold down “Ctrl” key and drag click the left button of the mouse to select an area on the Graph and the Y-axis of the selected area will be enlarged.
Three icons on the right top side of the main screen:
Redraw: After any review and zooming on the Graph, click “Redraw” to reset it to original format.
Clear: Click “Clear” to delete all data and Graph on the window. A warning box shows up for confirmation. Press “Yes” and all data will be cleared.
Cursor: Switch Cursor ON and it shows the raw data with the cursor. In any click on the graph, the digital data displayed.
Switch OFF to disable the function.
Data Processing
The downloaded data can be Saved or Printed via “File” processing functions.
Load: To retrieve saved files, just click “Load” and select a desired file and it will be loaded in the main screen with the file details at the left top side.
SPECIFICATION
- Measuring range: CO2
- Temperature: -20~60oC (-5~140oF)
- Accuracy CO2
- Temperature +0.3oC/+0.6oF at 5~40 oC
- CO2 Warm-up time: 30 seconds
- Response time
- CO2: <30 seconds(90% step change)
- Tair: <2 mins
- LCD / Meter size (mm)
- Operating condition (avoid condensation)
- Storage condition
- Power supply
- Weight :Standard package
CO2 LEVELS AND GUIDELINES
Non-Enforced Reference levels
NIOSH recommendations
- 250-350 ppm: normal outdoor ambient concentrations
- 600 ppm: minimal air quality complaints 600-
- 1000 ppm: less clearly interpreted 1000
- ppm: indicates inadequate ventilation; complaints such as headaches, fatigue, and eye/throat irritation will be more widespread. 1000 ppm should be used as an upper limit for indoor levels.
EPA Taiwan: 600ppm and 1000ppM
Type 1 indoor areas such as department stores, theaters, restaurants, libraries, the acceptable CO2 concentration of 8 hours avarge is 1000ppm.
Type 2 indoor areas with special requirements of good air quality such as schools, hospitals, day care centers, the suggested CO2 level is 600ppm.
Regulatory exposure limit
ASHRAE Standard 62-1989 : 1000ppm CO2 concentration in occupied building should not exceed 1000ppm.
Building bulletin 101 (BB101): 1500ppm UK standards for schools say that CO2 at averaged over the whole day(i.e. 9am to 3.30 pm) should not exceed 1500ppm.
OSHA: 5000ppm Time weighted average over five 8-hour work days should not exceed 5000ppm.
Germany, Japan, Australia, UK …: 5000ppm 8 hours weighted average in occupational exposure limit is 5000ppm.