ZOLL DA-2006P Analyzer Test Equipment Instructions

August 6, 2024
ZOLL

DA-2006P Analyzer Test Equipment

“`html

Specifications:

  • Product: Zoll R Series Maintenance Test Supplemental
    Document

  • Manufacturer: BC Group Intl.

  • Analyzer Model: BC Biomedical DA-2006P Defibrillator
    Analyzer

Product Information:

The Zoll R Series Maintenance Test Supplemental Document is
designed to guide users when using the BC Biomedical DA-2006P
Defibrillator Analyzer for performing the Preventive Maintenance
procedure on a Zoll R-Series defibrillator. It provides
step-by-step instructions based on the Rev K of the R-Series
service manual.

Product Usage Instructions:

Performing Preventive Maintenance Procedure:

  1. 1. Follow physical inspection steps in section 1.0 of the service  

    manual.

  2. Step 2.3:

  * Press RANGE until the DA-2006P shows High Range 1000J Max under  

configuration.

  * Press ECG Waveforms.
  * Highlight ECG Group and press Choices.
  * Highlight Normal Sinus Rhythm and press Enter.
  * In the ECG Configuration Screen, highlight Heart Rate and press  

Choices.

  * Highlight 120 BPM using arrow keys and press Enter.
  * Press Exit.
3.  **Step 4.19:**
  * Use the Charge Timer as a substitute for a manual  

stopwatch.

  * Follow the steps for using the Charge Timer as described in the  

manual.

FAQ:

Q: Can I use the Charge Timer feature without a manual

stopwatch?

A: Yes, the DA-2006P is equipped with a Charge Timer that can be
used as a substitute for a manual stopwatch. Follow the steps
outlined in the manual for using this feature.

“`

Zoll R Series Maintenance Test Supplemental Document
For use with BC Biomedical DA-2006P Defibrillator Analyzer Prepared by BC Group Intl.
This document is to be used as a guide when using the DA-2006P to perform the Preventive Maintenance procedure for a Zoll R-Series defibrillator. This guide is based on the Rev K of the R-Series service manual. There are several steps in the PM procedure that do not require a defibrillator analyzer, such as the physical inspections in section 1.0 of the service manual. The heading of each step below corresponds to a specific test in the service manual.

2.3 – Press RANGE until the DA-2006P shows High Range 1000J Max under configuration – Press ECG Waveforms – With ECG Group highlighted, press Choices – Use the arrow keys to highlight Normal Sinus Rhythm. Press Enter. – In the ECG Configuration Screen, use the arrow keys to highlight Heart Rate. Press Choices – Use the arrow keys to highlight 120 BPM. Press Enter – Press Exit.

4.19

  • The DA-2006p is equipped with a Charge Timer as a substitute for a manual stopwatch. If you would like to use this option, follow these steps.
    – Press RANGE until the DA-2006P shows High Range 1000J Max under configuration
    – After the defibrillator is set to 200J, press the Start Charge Timer soft key on the DA-2006P. A 5 second countdown begins. At the end of the countdown, press the CHARGE button the defibrillator. As soon as the SHOCK button illuminates, discharge the defibrillator into the DA-2006P. The charge time will display under the Test Results section.

7.1 – Press RANGE until the DA-2006P shows High Range 1000J Max under configuration – Press ECG Waveforms – With ECG Group highlighted, press Choices – Use the arrow keys to highlight Normal Sinus Rhythm. Press Enter. – In the ECG Configuration Screen, use the arrow keys to highlight Heart Rate. Press Choices – Use the arrow keys to highlight 120 BPM. Press Enter – Press Exit. WI-7395111-10-00

1 of 5

9.2 – Press RANGE until the DA-2006P shows High Range 1000J Max under configuration – Press ECG Waveforms – With ECG Group highlighted, press Choices – Use the arrow keys to highlight Performance Waveforms. Press Enter. – In the ECG Configuration Screen, use the arrow keys to highlight Waveform. Press Choices – Use the arrow keys to highlight Sine Wave 60 Hz. Press Enter – In the ECG Configuration Screen, use the arrow keys to highlight Amplitude. Press Choices – Use the arrow keys to highlight Lead I 0.35mV Lead II 0.5mV. Press Enter – Press Exit.
10.1 – Press RANGE until the DA-2006P shows High Range 1000J Max under configuration – Press ECG Waveforms – With ECG Group highlighted, press Choices – Use the arrow keys to highlight Normal Sinus Rhythm. Press Enter. – In the ECG Configuration Screen, use the arrow keys to highlight Heart Rate. Press Choices – Use the arrow keys to highlight 120 BPM. Press Enter – Press Exit.
10.19 – Press ECG Waveforms – Use the arrow keys to highlight Heart Rate. Press Choices – Use the arrow keys to highlight 160 BPM. Press Enter – Press Exit.

WI-7395111-10-00

2 of 5

12.0 – Press RANGE until the DA-2006P shows High Range 1000J Max under configuration – Press ECG Waveforms – With ECG Group highlighted, press Choices – Use the arrow keys to highlight Normal Sinus Rhythm. Press Enter. – In the ECG Configuration Screen, use the arrow keys to highlight Heart Rate. Press Choices – Use the arrow keys to highlight 60 BPM. Press Enter – In the ECG Configuration Screen, use the arrow keys to highlight Amplitude. Press Choices – Use the arrow keys to highlight Lead I 0.70mV Lead II 1.0mV. Press Enter – Press Exit.
12.2 – Follow the directions outlined in 12.0 above. The DA-2006P is in Cardioversion Mode. Once a defibrillator is discharged into the DA-2006P, there will be a line item in the Test Results section of the DA-2006P that shows the Delay.
13.0 – Press RANGE until the DA-2006P shows High Range 1000J Max under configuration – Press ECG Waveforms – With ECG Group highlighted, press Choices – Use the arrow keys to highlight Normal Sinus Rhythm. Press Enter. – In the ECG Configuration Screen, use the arrow keys to highlight Heart Rate. Press Choices – Use the arrow keys to highlight a selection between 60 BPM and 120 BPM according to the R series Service Manual. Press Enter – In the ECG Configuration Screen, use the arrow keys to highlight Amplitude. Press Choices – Use the arrow keys to highlight Lead I 0.70mV Lead II 1.0mV. Press Enter – Press Exit.

WI-7395111-10-00

3 of 5

13.3 – Follow the directions outlined in 13.0 above. The DA-2006P is in Cardioversion Mode. Once a defibrillator is discharged into the DA-2006P, there will be a line item in the Test Results section of the DA-2006P that shows the Delay.
14.2 – Press RANGE until the DA-2006P shows Low Range 50J Max under configuration
14.8 – Press RANGE until the DA-2006P shows High Range 1000J Max under configuration
14.14 * The DA-2006p is equipped with a Charge Timer as a substitute for a manual stopwatch. If you would like to use this option, follow these steps. – After the defibrillator is set to 200J, press the Start Charge Timer soft key on the DA-2006P. A 5 second countdown begins. At the end of the countdown, press the CHARGE button the defibrillator. As soon as the SHOCK button illuminates, discharge the defibrillator into the DA-2006P. The charge time will display under the Test Results section.
16.2 – Press RANGE until the DA-2006P shows High Range 1000J Max under configuration – Press ECG Waveforms – With ECG Group highlighted, press Choices – Use the arrow keys to highlight Arrhythmias. Press Enter. – In the ECG Configuration Screen, use the arrow keys to highlight Waveform. Press Choices – Use the arrow keys to highlight Ventricular Fibrillation. Press Enter – Press Exit.

WI-7395111-10-00

4 of 5

16.4 – Press ECG Waveforms – With ECG Group highlighted, press Choices – Use the arrow keys to highlight Normal Sinus Rhythm. Press Enter.
17.0 – Press RANGE until the DA-2006P shows Pace: DO NOT DEFIB under configuration * There are two options for connecting the Pacer cables: (1) the Pacer Input Variable load connectors (2) the 50 fixed load at the Defibrillator plates. If option 1 is used:
– Press Pace Mode Setup soft key – Use the arrow keys to highlight Load. Press Choices – Use the arrow keys to scroll through the list of 50 ohm ­ 2300 ohm ­ open. Press Enter If option 2 is used: – Press Pace Mode Setup soft key – Use the arrow keys to highlight Load. Press Choices – Use the arrow keys to scroll to the top of the list and highlight Defib Paddles. Press Enter.
17.9 – Press RANGE until the DA-2006P shows High Range 1000J Max under configuration – Press ECG Waveforms – With ECG Group highlighted, press Choices – Use the arrow keys to highlight Normal Sinus Rhythm. Press Enter. – In the ECG Configuration Screen, use the arrow keys to highlight Heart Rate. Press Choices – Use the arrow keys to highlight 60 BPM. Press Enter – Press Exit.

WI-7395111-10-00

5 of 5