ZOLL IC-4593 AE Quattro Intravascular Heat Exchange Catheter Kit Instruction Manual

June 3, 2024
ZOLL

ZOLL IC-4593 AE Quattro Intravascular Heat Exchange Catheter Kit

Quattro® Intravascular Heat Exchange Catheter Kit includes

The Quattro intravascular heat exchange catheter (“Quattro catheter” or “catheter”) is a sterile, single-use flexible 9.3 F catheter designed for placement in the inferior vena cava from an insertion site in the femoral vein. The Quattro catheter is to be connected to a single-use disposable ZOLL® Start-Up Kit and the Coolgard®/ Thermogard® family of consoles (“console”) (all supplied separately). A dilator and guidewire are required for the percutaneous insertion of the Quattro catheter. Three Luers are available for infu-sion and sampling.

Ethylene oxide sterilized. The catheter is supplied sterile for single use only and should not be resterilized. The package should be inspected prior to use to ensure that the sterility barrier has not been compromised. Store in a cool, dry place. Avoid freezing and excessive heat.

Intend used

The Quattro catheter in combination with the Coolgardie or Thermogard System permits circulation of temperature-controlled saline through a heat exchanger to cool/rewarm the patient’s blood in patients for whom the risks of a central line are warranted. Maximum use period: 4 days.

  1.  The risks of the catheter are essentially those of a central line. The catheter should not be used in patients for whom central line placement is not indicated.
  2.  Bleeding diathesis.
  3.  Active sepsis.
  4.  Infection or active bleeding at the site of catheter insertion.
  5.  Patients with no vascular access, or a vascular system that will not accommodate a catheter, including patients with vena cava filters or other implanted impediments to passage of the catheter.
  6.  Patients for whom the required temperature monitoring cannot be established.
  7.  Hypothermia is contraindicated in patients who have hematological diseases that will be made worse with hypothermia (e.g., any disease that produces cryoglobulinemia, any hemoglobinopathy in which hemolytic anemia can be precipitated by cold including Sickle Cell Disease or Thalassemia).
  8.  Not intended for pediatric or neonatal use.

warning
Central venous catheterization should only be performed by well-trained personnel well-versed in anatomical landmarks and safe techniques. Personnel should also have knowledge of potential complications.

warning
The catheter should be placed via a femoral vein approach only. Do not allow the catheter to be placed into the right atrium or right ventricle. Placement in the right atrium or right ventricle can result in severe patient injury or death. The Quattro catheter and Start-Up Kit could potentially misconnect with small bore connectors from other medical devices. Such connection errors could result in patient injury or death. The custom Luers contained on the catheter and Start-Up Kit may reduce the risk of misconnections, but still have the potential for misconnections with these specific medical device applications: Breathing Systems & Driving Gases applications, Enteral & Gastric applications, Urethral & Urinary applications, Limb Cuff Inflation applications, Neuraxial applications, and Intra-vascular or Hypodermic applications. Always use caution when connecting catheters and Start-Up Kits to these and other medical device applications. Ensure that the catheter and/or Start-Up Kit are not disconnected to an IV or other medical devices.

The product is designed for single use only. Do not resterilize or reuse. Do not reinsert, once removed from the patient. Do not alter the catheter in any way. Potential risks with the re-use of a single-use device include but are not limited to:

  •  Potentially life-threatening infection
  •  Toxic shock due to the degradation of materials
  •  Increased risk of thrombosis
  •  Reduced heat exchange power
  •  Device failures
    1.  The catheter should be positioned so that the distal tip of the catheter is in the inferior vena cava, below its junction with the right atrium and parallel to the vessel wall. X-ray examination should be used to ensure that the catheter is not in the right atrium or ventricle.
    2.  Possible complications with central venous catheters include: atrial or ventricular perforation, cardiac tamponade, air embolism, catheter embolism, thoracic duct laceration, bacteremia, septicemia, thrombosis, inadvertent arterial puncture, hematoma formation, hemorrhage, nerve damage and dysrhythmias.
    3.  Caution. If blood is observed within the saline circuit, stop the procedure and perform a check for a catheter leak.
    4.  All Luer-Lock connections and covers must be securely tightened to prevent air embolism or fluid or blood loss.
    5.  Never use excessive force in moving the catheter or guidewire. If resistance is encountered, an x-ray should be performed to identify the reason for the resistance.
    6.  Passage of the guidewire into the right heart can cause dysrhythmias, right bundle branch block, vessel wall, atrial or ventricular perforation.
    7.  Use only sterile normal saline for catheter priming. It is the circulating fluid in the catheter.
    8.  The catheter should be routinely inspected for flow rate, security of dressing, correct catheter position, and for secure Luer-Lock connections. Use the centimeter markings to identify if the catheter position has changed. A potential risk associated with multi-lumen balloon catheters is a lack of therapy, including failure to infuse drug therapy through the infusion lumens.
    9.  Only x-ray examination can ensure that the catheter tip has not entered the heart or no longer lies parallel to the vessel wall. If the catheter position has changed, perform an x-ray examination to confirm the catheter tip position.
    10.  For blood sampling, temporarily shut off the remaining infusion ports through which solutions are being infused.
    11.  Use only a 30 cc or smaller syringe for blood sampling.
    12.  Alcohol and acetone can weaken the structure of the shaft material. Care should therefore be taken when infusing drugs containing alcohol or when using alcohol or acetone when performing routine catheter care and maintenance. Alcohol should not be used to declot the catheter.
    13.  Use only the ZOLL suture tab and clip provided in the kit to prevent catheter damage.
    14.  Use of a syringe smaller than 10 ml to irrigate or declot an occluded catheter may cause intraluminal leakage or catheter rupture.
    15.  Do not infuse into the orange IN and OUT Luer-Lock connections, as this will result in lack of therapy.
    16. Fever may have infectious and/or non-infectious causes in patients. Mitigation of fever as a sign of possible infection necessitates daily, meticulous assessment for other signs of infection.
    17. Use care when infusing drugs that may be affected by cool temperatures (as low as 4ºC). Solutions containing mannitol are temperature-sensitive and must not be delivered through the catheter except for a rapid push of a solution of up to 20 %mannitol, followed by a saline flush. Higher than a 20%
    18. Cardiac Tamponade: The placement of indwelling catheters in the right atrium is a practice that may lead to cardiac perforation and tamponade. Practitioners placing central venous catheters must be aware of this potentially fatal complication before advancing the catheter too far relative to patient size. The actual position of the tip of the indwelling catheter should be confirmed by x-ray after insertion. Central venous catheters should not be placed in the right atrium unless specifically required for special relatively short-term procedures, such as aspiration of air emboli during neurosurgery. Such procedures are nevertheless risk-prone and should be closely monitored and controlled.
    19.  WARNING. When connecting infusion sets/injection systems to the catheter, do not exceed 100 psi/689 kPa.
    20.  For patients being made hypothermic, the hypothermia itself may exacerbate some disease states. Care should be taken to properly monitor patient homeostasis during hypothermia.
    •  Cardiac rhythm disturbances–both bradycardia and ventricular tachyarrhythmia.• Clotting and coagulation function. Patients at risk for disturbances of their clotting or coagulation function should be closely monitored during hypothermia.
    •  Blood gas and pH analysis. Hypothermia modifies resting pH and PaCO2. Physicians should be aware of the effect of temperature on the result.
    •  Prolonged hypothermia depresses the immune response and lung function.

Intraluminal leakage (between the saline lumen and infusion lumens) or balloon leakage is a potential catheter failure mode. In the event of such a failure, sterile saline from the cooling circuit is introduced into the patient. Intraluminal leakage or balloon leakage is typically associated with a fluid loss alarm once the saline bag has been depleted and stops the system.

The cooling circuit is a closed loop system – usually, fluid loss alarms indicate a breach somewhere in this closed loop. With any fluid loss alarm, check both the integrity of the catheter and the Start-Up Kit (see below). you notice a depleted saline bag or an air trap alarm, do not replace the saline bag prior to identifying the location of the leak and taking the appropriate mitigation. Check for system leaks according to the instructions in the Check for a Start-Up Kit leak and Check for a catheter leak sections below. (Note that a leak could be external or internal.)

Replacing the saline bag repeatedly without investigating the leak or loss of saline may result in an unintended infusion of saline into the patient. The saline infusion may lead to the following adverse effects: local swelling that can cause subsequent local tissue damage; systemic fluid overload that can lead to dependent edema and subsequent skin breakdown; internal organ fluid overload, with subsequent overloading of the brain, lungs or heart. In some cases, this fluid overload may lead to life-threatening events. The console emits an alarm when the saline bag is empty. The bag must be completely empty and additional saline must have drained between the saline spike and the air trap

  1.  Check the air trap for condensation. If the air trap shows signs of condensation, wipe the air trap and reinstall it in the console. In the case of an air trap alarm, verify that the air trap alarm is cleared after this step.
  2.  Carefully check the saline path from the saline bag to the console for any leaks. Check if there is saline on the floor, console, or the patient’s bed.
  3.  If there is any saline on the floor, console, or the patient’s bed, check that the Luers on the catheter and Start-Up Kit are not cracked or damaged and that the connections are tight enough to prevent leaks.
  4. If you find a leak in the Start-Up Kit, replace the Start-Up Kit and see if there is also a leak in the catheter.
  5. If you do not find a leak in the Start-Up Kit, there is likely a leak in the catheter. Investigate further.

check for a Cather leak

  1.  Disconnect the Start-Up Kit from the catheter. Properly cap both the catheter and Start-Up Kit using an aseptic technique.
  2.  Fill a sterile 10 mL slip-tip syringe with sterile saline.
  3.  Connect the syringe to the IN Luer of the catheter and disconnect the OUT cap. Infuse 10 mL of saline – the saline should flow out the OUT Luer. If the saline does not flow out of the OUT Luer, a catheter leak is indicated.
  4.  Cap the OUT Luer and pull 5 cc of vacuum. Sustain for at least 10 seconds. Up to 4 mL of saline (not blood) should enter the syringe and you should be able to maintain the vacuum. If traces of blood are seen in the syringe or vacuum cannot be maintained, it indicates a catheter leak.
  5.  If you find a leak in the catheter, replace the catheter.
  6.  Replace the saline bag and re-prime the Start-Up Kit.
  7.  Ensure leak-tight Luer connections to the Start-Up Kit and continue the therapy.

Material Required

  • 1 Quattro Catheter Kit for percutaneous introduction
  • 1 500 cc bag of sterile normal saline (not provided)
  • 1 Start-Up Kit (provided separately)
  •  6 ft (183 cm) standard tubing or
  • 9 ft (274 cm) extended tubing
  • 1 Coolgard or Thermogard console (provided separately)
  • 1 YSI-400 temperature probe (not provided)

The catheter has a radiopaque marker band to assist in the identification of the catheter during and after insertion when viewed using x-ray equipment. The proximal end of the proximal balloon has one marker band. The tip of the catheter contains barium sulfate to make it radiopaque. Use a sterile technique. The IN and OUT Luer-Locks on this catheter are custom-manufactured and are intended to connect only with the Start-Up Kit listed in Materials required.

  1.  Obtain femoral venous access using standard percutaneous techniques. Access should be maintained with a 0.032″ (0.81 mm) guidewire. See Guidewire Instructions for Use.

  2. Do not attempt to reinsert a partially or completely withdrawn OTN (over the needle) introducer needle from its catheter.

  3. o not use a guidewire larger than 0.032″ (0.81 mm) with the catheter.

  4.  Holding the guidewire in place, remove the introducer catheter. `~ìíáçåK=Maintain a firm grip on the guidewire at all times.

  5. Enlarge the cutaneous puncture site with the cutting edge of the scalpel positioned away from the guidewire.

  6. Do not cut the guidewire. Use the vessel dilator to enlarge the site as required. Do not leave the vessel dilator in place as an indwelling catheter, to minimize the risk of a possible vessel wall perforation.

  7. Thread the tip of the catheter over the guidewire. Maintain a sufficiently firm grip on the guidewire during catheter insertion. Grasping the catheter tip near the skin, advance the catheter into the vein. Continue to advance the catheter over the guidewire, placing your fingers just proximal to the balloon.

  8. Using centimeter marks on the catheter as positioning reference points, advance the catheter to at least the minimum marker number to ensure the proximal infusion port is in the vessel.

  9. Hold the catheter at the desired depth and remove the guidewire. If resistance is encountered when attempting to remove the guidewire after catheter placement, the guidewire may be kinked at the tip of the catheter. If resistance is encountered, withdraw the catheter relative to the guidewire about 2–3 cm and attempt to remove the guidewire. If resistance is encountered again, remove the guidewire and catheter simultaneously.

  10. Do not apply undue force to the guidewire.

  11. verify that the guidewire is intact upon removal.

  12. Check catheter placement by attaching a syringe to the distal infusion Luer and aspirate until a free flow of venous blood is observed. Connect the infusion Luer to the appropriate Luer-Lock line as required. The unused infusion port may be locked through the injection cap using standard hospital protocol. A slide clamp is provided on the tubing to occlude flow through the infusion Luer during line and injection cap changes.

  13. 9. Secure and dress the insertion site and catheter temporarily.
    10. Verify the catheter tip position by chest x-ray immediately after placement. X-ray examination must show the catheter located in the IVC with the distal end of the catheter parallel to the vena cava wall. If the catheter tip is malpositioned, reposition and reverify.
    11. The proximal radiopaque marker indicates the proximal end of the balloon. Ensure that the balloon resides completely in the vessel. If the catheter is malpositioned, reposition and reverify.
    12. Secure the catheter to the patient. Use the suture tab side wings as the primary suture site to minimize the risk of catheter migration.
    13. The ZOLL suture tab and clip can also be used as an additional attachment point. Ensure that the catheter body is secure and does not slide.

  14. Use only the ZOLL suture tab and clip provided in the kit. Catheter damage may result if other tabs or clips are used.

  15. Do not suture directly to the outside diameter of the catheter, to minimize the risk of cutting or damaging the catheter or impeding catheter flow.

  16. Dress the puncture site per hospital protocol. Maintain the insertion site with regular meticulous redressing using an aseptic technique.

  17.  Record on the patient’s chart the indwelling catheter length, using the centimeter marks on the catheter shaft as reference. Frequent visual reassessments should be made to ensure that the catheter has not moved.

  18. Attach a primed Start-Up Kit to the catheter: Connect the male Luer of the Start-Up Kit to the female IN Luer of the catheter (labeled “IN”). Connect the female Luer of the Start-Up Kit to the male OUT Luer of the catheter (labeled “OUT”). White “ZOLL” tags are fitted loosely to the IN and OUT extension tubes to help identify them.

  19. The Start-Up Kit IN and OUT Luers are intended to connect to the catheter IN and OUT Luers only and are not intended to connect to standard Luer Lock syringes. They have ZOLL custom fittings and are orange in color for easy identification.

  20. Ensure that a sufficient amount of sterile saline is present at the ends of the Luers to make an air-free connection. Refer to the operation manual.

  21. Failure to connect the Start-Up Kit correctly to the catheter could result in catheter failure. Do not attach the Start-Up Kit (orange) Luers to the blue, white, or brown Luers.

  22. Do not attach the Start-Up Kit to the distal port.

  23. Do not place any stopcocks in line that may be inadvertently shut off. This can cause line blockage and possible failure.

  24. Pump saline through the Start-Up Kit and catheter to ensure that all connections are secure and that there is no leaking. Allow any remaining air in the system to be purged out as described in the operation manual.

Removing the Catheter

  1.  Stop pumping saline through the catheter.
  2.  Disconnect the Start-Up Kit from the catheter. råÅ~é=çê=äÉ~îÉ= ìåÅ~ééÉÇ=íÜÉ=fk=~åÇ=lrq=iìÉêë=çÑ=íÜÉ=Å~íÜÉíÉêK This allows residual saline within the circuit to be expressed. As the catheter is withdrawn, the balloons are compressed. Saline within the balloons must be free to pass out of the balloon or the balloon will not deflate, making the catheter difficult to remove.
  3.  Optionally, attach a 20 or 25 cc syringe to the catheter IN Luer. Pull and hold a vacuum for 15 seconds to allow residual saline to be removed from the catheter balloon section prior to removing the catheter. For convenience, a 20 or 25 cc syringe is included with the Start-Up Kit package. Hang it on the saline hook on the console until ready for use. Discard after each patient.
  4. Place the patient in a supine position. Remove the dressing. Remove the sutures from the suture site.
  5. Do not place a Luer cap on the OUT Luer.
  6. Slowly remove the catheter from the patient. As the catheter exits the site, apply pressure with a dressing impermeable to air (e.g. Vaseline gauze).
  7. Do not move the catheter if resistance is felt. Check to ensure that the IN and OUT Luers of the cooling circuit are NOT capped. If they are capped, uncap them, deflate the balloon, and try removing the catheter again. If resistance is still encountered, an x-ray should be performed to identify the reason for the resistance.

MRI SAFETY INFORMATION

The Quattro Intravascular Heat Exchange Catheter is MR Conditional. A patient with this catheter may be safely scanned under the following conditions. Failure to fol-low these conditions may result in injury to the patient.

Guidewire Instructions for Use

This information applies only to the use of guidewires in the Seldinger technique of catheter placement in the vasculature. The supplied guidewire is designed for single use only. Do not resterilize or reuse. Do not reinsert, once removed from the patient. Should resistance occur during insertion or withdrawal, do not continue to move the guidewire. Determine the cause under fluoroscopy and take action as needed. Use extreme caution when moving a guidewire through a stent. Use of a guidewire in stented vessels creates additional patient risk. Avoid withdrawing the guidewire through metal needles; the guidewire may shear. Because of the delicate and fragile nature of guidewires, extra care in handling must be taken. Avoid bending or kinking. Do not use dam-aged guidewires. During storage or procedures, avoid coiling the guidewire in less than 8 inch (20.32 cm) diameters, as smaller diameters place unnecessary stress on the guidewire and may cause it to kink. The provided dispenser is the best means of storage and handling of the guidewire.
Sufficient guidewire length must remain exposed to maintain a firm grip on the guidewire at all times. Every guidewire is provided in a dispenser package. Remove the guidewire anti-migration clip before dispensing the guidewire. Remove the guidewire protective cap immediately prior to guidewire use. Prepare the guidewire prior to insertion. It is recommended that the dispenser be filled with heparinized solutions (e.g. saline or dextrose) to bathe the guidewire during insertion. The preformed “J” guidewire will resume its shape when removed from the product dispenser.

  1.  Guidewire protector cap
  2.  Guidewire anti-migration clip

Inspection
Inspect the guidewire prior to use and discard if any deformities are present in the guidewire. Guidewire placement should be routinely monitored by x-ray or fluoroscopic procedure.

  1.  Puncture the vessel.
  2.  Insert the guidewire into the needle hub and gently advance 5- 10 cm of the guidewire into the punctured vessel. Navigate the guidewire to the desired position. Avoid rough or overly vigorous manipulation of the guidewire to prevent damage to the guide or the vessel.
  3.  Remove the needle from the guidewire.
  4.  Dilate the tissue and vessel with the dilator using a slight rotary motion.
  5.  Remove the dilator (the vessel dilator is intended for vascular dilation only).
  6.  Introduce the catheter by sliding it over the guidewire.
  7. Remove the guidewire.

QUATTRO CATHETRE

  1.  IN Luer
  2.  OUT Luer
  3.  Infusion ports
  4.  Distal Luer

Medical Corporation. All rights reserved. ZOLL and Quattro are trademarks or registered trademarks of ZOLL Medical Corporation and/or ZOLL Circulation Inc. in the United States and/or other countries. All trademarks are the property of their respective owners. Patent: www.zoll.com/patents

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

ZOLL User Manuals

Related Manuals