Alan Dental Y-SMART Mini Dental Cordless Endodontic Handpiece User Manual
- July 15, 2024
- ALAN Dental
Table of Contents
- Alan Dental Y-SMART Mini Dental Cordless Endodontic Handpiece
- Product Introduction
- Product Installation
- Product basic accessories drawing
- Product Function and Use
- Operating Instructions
- Function Selection
- Operation Mode
- Option setting
- Built-in root canal tiles parameters
- Battery and charging
- Charging
- Shutdown
- Troubleshooting guide
- Maintenance
- Cleaning and Disinfection
- Storage and Transportation
- Symbol Description
- Environmental protection
- Packing list
- Product Warranty Certificate
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Alan Dental Y-SMART Mini Dental Cordless Endodontic Handpiece
Product Introduction
Product Description
The endo motor is mainly used in the root canal therapy process and is a
device for root canal formation and cleaning in the root canal preparation
stage. It helps the dentist complete the root canal treatment.
The features of this endo motor are as follows:
- a) HD OLED screen.
- b) Wireless handle and more convenient operation.
- c) Wireless charging to avoid poor contact of traditional contact type charging.
- d) 10 custom programs and 6 working modes.
- e) Large capacity battery and longer working hours.
Specification
Y-SMART
Please see the packing list for the machine configuration.
Perfrmance structure and composition
It consists of a charging base, a handle, a handpiece, and a power adapter. etc.
Scope of application
- This product is suitable for root canal therapy, root canal formation and cleaning in root canal preparation stage.
This product is only suitable for hospitals and clinics and must be used by qualified dentists.
Contraindications ,
if you are experiencing a patient with an implantable cardiac pacemaker (or other electrical equipment) and are warned not to use small appliances (such as electric shavers, hair dryers, etc.), it is recommended not to use this product.
Warning tips
-
Please read this manual carefully before you first use it.
-
The device must be used by a professional and qualified dentist at a suitable location.
-
Do not put the device directly or indirectly near the heat source.The device must be operated and maintained in a reliable environment.
-
This device requires special precautions regarding electromagnetic compatibility (EMC) and must be installed and used in strict accordance with EMC information. Do not use this device near fluorescent lights, wireless transmitters, remote controls, handheld and mobile HF communications equipment.
-
lt may cause the handle to overheat if use the reciprocating mode for a long time. In this case, it should be cooled and then used again. If the handle frequently overheats, please contact your local dealer.
-
This device must use the original contra-angle handpiece, otherwise it will not be used or cause adverse consequences.
-
Please do not make any modifications to the device. Any changes may violate the safety regulations and with internal power supply.cause harm to the patient. Any modifications will not be guaranteed.
-
Please use the original power adapter, other power adapters will cause damage to lithium batteries and control circuits.
-
The handle cannot be autoclaved and the surface can only be wiped with a neutral disinfectant or alcohol.
-
Do not press the pressing cover of the handpiece before the handpiece stops rotating, otherwise it will damage the hand piece.
-
The hand piece cannot be removed until the handle stops rotating, otherwise it will damage the hand piece and the inner gear of the handle.
-
Before starting the handle, make sure that the file is
installed and locked. -
The continuous rotating file can not be used in the reciprocating motion system program, and the reciprocating file can not be used in the continuous rotating system program.
-
Please set the torque and speed of the endomotor according to the parameters recommended by the files manufacturer.
-
The forward rotation mode is suitable for the continuous rotating file, the reciprocating motion mode is suitable for the reciprocating motion file (that is, one file), and when the continuous rotating file lock in the root canal,inversion can take out the files.
-
Do not start the handle motor before properly installing the handpiece, so as to avoid damage caused by the operator touching the gear.
Device Security classification
- Classified by running mode: continuous running device.
- Classified by type of electric shock: Class II equipment
- Classifed by degree of protection against electric shock: BF type application part.
- Degree of protection against liquid intake: ordinary equipment (IPX0).
- The degree of safety when used in the case of a flammable anesthetic gas mixed with air or a flammable anesthetic gas mixed with oxygen or nitrous oxide: a device that cannot be used in the presence of a flammable anesthetic gas mixed with air or a flammable anesthetic gas mixed with oxygen or nitrous oxide.
Main technical parameters
- Handle lithium battery: 3. 7V./800mAh
- Power Adapter
- Input: ~1 00V-240V.50Hz/60Hz .. 0.4A.Max
- Output: DC5V/1A
- Range of torque: 0.SNcm-4.0Ncm C5mNm~40mNm)
- Range of speed: 120rpm~800rpm
Operational environment
- Environmental temperature: -s·c ~+40°C
- Relative humidity: 30%-75%
- Atmospheric pressure: 70kPa~106kPa
Product Installation
Product basic accessories drawing
Handpiece instructions
-
The inside of this handpiece is driven by a precision gear with a transmission ratio of 1: 1.
-
Before the first use and after each use, please clean
and disinfect the contra-angle handpiece with the
commonly used neutral disinfectant, then use the special cleaning oil for oiling maintenance, and then carry on the high temperature and high pressure
sterilization with the temperature of 134°C and the pressure of 2.0bar / 2.3bar. (0.20MPa~0.23MPa). -
This contra-angle handpiece can only be used in conjunction with this device. It cannot be used in conjunction with another device. Otherwise, the contra-angle handpiece may be damaged.
-
The service life of the contra-angle hand piece is 1 year. According to the frequency and time of clinical use, and the degree of difficulty of different patient’s oral root canal during treatment, and its life span is also different
Installation and disassembly of hand piece
-
Installation
Point the positioning pin of the handpiece to the positioning hole of the handle, push it horizontally into the hand piece, and when you hear the sounds of “crack” , it indicates that it is installed in place. Three positioning pins on the handpiece are matched with 6 positioning holes on the handle.Handpiece can be installed at multiple angles. -
Disassembly
When the handle is not started, pull the handpiece horizontally outing
Installation and disassembly of files
Insert the files into the rotating shaft hole of the handpiece, gently rotate,
find the position corresponding to the semicircular part of the files, when
you find the position, insert a part of it first, do not press the button
above the handpiece, and then press the files into the handpiece with a little
greater force. When you hear the “crack”, the file is connected to the hand
piece.
If you need to remove the files, hold down the button above the hand piece and
pull the files out of the hand piece at the same time.
When installing the needle, please press the files several times, and make
sure the files is installed. Please keep the files clean, if dirt, etc. enter
into hand piece, will cause the vibration of the files.
Product Function and Use
Operating Instructions
The start and stop of the handle When the handle is off, press and hold the “Power On/Start” button to enter the standby state; then press the “Power On/Start” button to enter the working state; press the “Power On/Start” button again to return the standby state.
Function Selection
Press the function setting key “S” on the standby interface to set the working mode, the speed , torque and the automatic round-trip angle. After setting, press the “Power on/Start” button to return.
Operation Mode
In this interface, use the adjustment key to select the following six working modes:
-
Fwd: Forward 360° automatic movement, positive rotation plus torque reversal, remove the torque and resume normal rotation.
-
Rev: reverse automatic continuation: reverse 360 °automatic motion, reverse plus torque forward rotation, remove the torque and resume the reversal.
-
AF: Forward reciprocating motion, continuous cycle between forward rotation and reverses
-
AR: Reverse reciprocating motion, continuous cycle between reverses and forward rotation.
-
TCF: Automatic reverse close: forward 360 °,when encounter torque , will stop,it is used to remove the broken files with Anti-bearing knife.
-
TCR: Automatic forward rotation close: reverse 360°, when encounter torque, will stop, it is used to remove the broken files with positive knife.
-
Speed:
In this interface, use the adjustment key to select the speed, the range is 120-800rpm.The motor speed can also by changed by the adjusting key while motor spinning. -
Torque Limit:
In this interface, use the adjustment key to select the torque, the range is 0.5-4 Nern. When the torque is R.L, the speed and mode are not controlled by the torque, that is, there will be no resistance reversal or stop. Note that when the speed is greater than 600 rpm, the torque range is 0.5-2.0 Nern; when the speed is lower than 280, the torque range is 0.5-1.5 Nern. The torquer can also be changed by pressing the “S” key firstly then the adjusting key while motor spinning. -
Fwd/Rev Angle:
This option interface appears only in AF and AR modes. Use the adjustment key to select the round-trip angle from 30°to 370 °
Option setting
In the standby interface, press and hold the adjustment key to enter the setting interface, then press the “S” key and the adjustment key to set the auto power-off time, buzzer volume, left and right hand mode, Opticlight time after motor stop can be selected from 3S-1 OS by the adjusting key , power-on default program and reset. After setting, press the “Power On/Start” button to return.
7. Reset
After reset and in default mode, the parameters of M 1-M 10 are as follows:
Function | Ml | M2 | M3 | M4 | M5 | M6 | M7 | M8 | M9 | Ml0 |
---|---|---|---|---|---|---|---|---|---|---|
Mode | Fwd | Fwd | AF | TCF | Rev | Rev | AR | TCR | Fwd | Rev |
Speed | 350 | 500 | 300 | 350 | 350 | 500 | 300 | 350 | 800 | 1000 |
TQ
|
3
|
2
| 90° I 30°|
2
|
3
|
2
| 90° I 30°|
2
|
1. 5
|
1
Built-in root canal tiles parameters
In the standby interface, long-press the adjustment key to enter the built-in parameter interface. The software has built-in our DS model root canal tiles parameters and some common root canal tiles parameters at home and abroad and press the key to select. After selecting, press the “Power on/Start” button to confirm and return. At this time, the interface parameter lock and cannot be adjusted. You can switch the root canal model by adjusting the key. If you want to exit, you need to press and hold the key again. The details of built- in root parameters are shown in the following table:
Files size | Speed | Torque | Mode |
---|---|---|---|
DS(Pro)-Dx | 300 | 3 |
Fwd
DS(Pro)-Path| 350| 1. 5
DS-D1| 350| 2
DS-D2/D3| 350| 1. 5
DS-Pro-D1/D3| 350| 2
DS-Pro-D2/D4| 350| 1. 5|
Fwd
---|---|---|---
DS-M-Path| 300| 1. 5
DS-M-Dl/D2/D3| 350| 1. 5
DS-R| 400| 2
DS-Path-DPl/2/3| 350| 1. 5
M3-0pen| 300| 3
M3-Path| 350| 1. 5
SANI-S3| 500| 2.5
Protaper-SX/S1|
250
| 3. 0
Protaper-S2| 1.0
Protaper-Fl| 1. 5
Protaper-F2/F3| 2.0
Protaper NEXT| 300| 2.0
TF| 450| 2.5
Mtwo| 280| 1. 5
Hyflex. CM| 500| 2.5
Hyflex. NT| 400| 2.5
Wave0ne| 150| 30. 0|
AF
Reciproc| 150| 30. 0
Battery and charging
Battery indication
There is a real-time power indication in the lower left corner of the screen interface:
Note: It is normal for the battery to change when the host adds load. For safety, the motor automatically turns off when charging.
Charging
The endomotor has a built-in 800mA large-capacity battery and is wirelessly
charged. The charging time is about 4 hours.
Insert the charging base into the adapter, and make sure that there is no
foreign matter in the hole in the main body of the charging base. After
placing it correctly, put the host handle into the charging base, then the
buzzer “drip” two times, and the “CHG”word will appear on the LCD screen and
indicates that the charging is in progress; at the same time, the screen power
is animated.
When charging for a period of time, two buzzer “drips” appear, and the word
“FUL”is displayed on the LCD screen, indicating that the battery is fully
charged.
Shutdown
Two ways to shut down
Automatic shutdown without operation, and the shutdown
time can be selected for 3-15 minutes in the setting.
Press and hold the function key “S”, and press the “Power On/Start” button to turn off the host; after shutdown, the host will store the parameter settings before shutdown for the next use.
Notes
-
The forward rotation function and the reciprocating function mentioned in this instrument use different root canal files.
-
Do not use curved cracked files that do not conform to specification. If such files are used, the files will be suddenly interrupted during the rotation, causing injury
-
When using, please follow the speed and torque of the files specified by the files manufacturer.
-
Do not use or place this product in high temperature places such as strong sunlight, fireside or stove.
-
When placing the main unit into the charging base, please pay attention to the direction of the loading {the operation panel is outward). If the direction is incorrectly placed, it may cause damage or cannot be
-
Do not lubricate the main unit with oil injection and sterilize the main unit with high temperature and high
-
This product is an electronic product and does not have a waterproof function.
-
Do not drop the main unit and charger base, and place the charging base in a stable place.
-
Please do not turn it on and use while charging. j))Please use the handpiece provided or approved by our
-
company, otherwise it may lead to misoperation of the machine.
-
This instrument will not cause electromagnetic interference to equipment in the same environment, and will be subject to electromagnetic interference ingeneral use environment. Please avoid using it in the electromagnetic environment and strong electric field.
-
The charger plug shall be removed from the power socket after charging.
-
This device should not be used in the presence of flammable anesthetic gas mixed with air or flammable anesthetic gas mixed with oxygen or nitrous oxide.
-
This endomotor cannot be repaired on
Contraindication: -
please do not use with high frequency surgical equipment and heart patients should be used with caution!
-
patients who use implantable electronic devices such as pacemakers should not be advised to use this product if they do not have a doctor’s advice!
Troubleshooting guide
Common faults and treatment methods
NO | Troubles | Possible problems and solutions |
---|
1
|
Can’t turn on the device
| Lack of power. Please charge
Press the power button for short time. Please long press it over 1s
2
|
No display when charging
| Poor contact between the adapter and the charging base. Propery rotate the
plug that connects the adapter to the charging base
There is a foreign object on the bottom of the main unit and the charging
base. Please clean up.
The main unit is not properly placed on the charger base
Maintenance
- Products that are not used for a while, be sure to check that the product is operating properly and safely before reusing it.
- when the battery is not used for a long time, it should be charged once a month to ensure the service life of the lithium battery.
- The working life is approximately 5 years from the date of
- when there is a stain on the surface of the main body, apply some neutral detergent to the soft cloth and wipe it with a wet cloth.
- Lubricate the machine with a lubricant for more than once a day after treatment or before autoclaving. Rotate the E-shaped nozzle into the outlet of the cleaning lubricant, remove the burs, and then insert it into the outer sheath of the handpiece and spray at least 2 seconds.
- When the handpiece is repeatedly steam-sterilized, it is recommended to perform regular maintenance to ensure that the handpiece is in good working condition and weekly maintenance is recommended.
Notes:
- 1. It is strictly forbidden to wipe with ethanol (alcohol), banana water, light grease (light gasoline) and so on.
- It is strictly forbidden to place the machine under the sun or beside the fire to avoid battery
- Do not disassemble the endomotor if there is any quality problem.
Cleaning and Disinfection
Please sterilize the handpiece, root canal file and LED light only.
- Please brush away the dirt attached to the surface with a brush (metal is strictly prohibited) and wipe it clean with a soft cloth stained with disinfection alcohol.
- During disinfection, the temperature shall not exceed 135°C(134°, 3.5mins recommended).
- When not in use, please seal it in a sterile bag and place it in a dry, well-ventilated
- Please sterilize the handpiece, root canal file and LED light only.
- Because the bottom layer of the pyrolysis rack is close to the heat source, the local temperature will be high, so the vessel should be placed in the middle or upper layer as much as possible.
- It is strictly forbidden to use chlorinated water (strongly acidic water, super acid water) or sterilizing liquid to wash, impregnate and wipe the vessel.
- Wrap it with a sterilized bag and mark it.
Storage and Transportation
Storage conditions
storage temperature:- 40° C~ +55° C
Relative humidity
95% Airpressure: 50kpa~-106kpa Environmental conditions:
It should be stored in places without light, rain, dust, corrosive gases, chemicals and good ventilation.
Transportation
- Prevent excessive shock and vibration during transportation, handle with care and avoid inversion.
- It should not be mixed with dangerous goods during
- The sun or rain and snow should be avoided when transporting.
Symbol Description
Environmental protection
The metal part of the product is treated as scrap metal. Synthetic materials, electrical components and printed circuit boards are treated as waste electrical appliances. The handling of related materials must be handled in accordance with local regulations and should be negotiated with the local company that specializes in the disposal of waste products.
Packing list
- Main body······················· 1
- Charging base················ 1
- Adapter····························· 1
- Handpiece························ 1
- Operating manual··········· 1
- product certification·········· 1
- After-sales service
The product can be replaced due to quality problems within seven days and free maintenance within one year. If you buy this product over one year, you need to pay the charge of maintenance accessories. About accessories, we provide free maintenance for six months. If accessories(such as endomotor shell. LED light.etc.) damaged or run out, you can purchase from our company.
Product Warranty Certificate
Client information
Client information|
Contact number|
Address|
Postcode|
Sales information|
Name of marketing unit|
Contact number|
Address|
Factory number|
Invoice number|
Sales date|
Seal of marketing unit|
Maintenance Records
Date
| The diagnosis| Fault cause| Examination Date| Maintenance Engineer
---|---|---|---|---
| | | |
Treatment
|
Date
| The diagnosis| Fault cause| Examination Date| Maintenance Engineer
---|---|---|---|---
| | | |
Treatment
|
Date
| The diagnosis| Fault cause| Examination Date| Maintenance Engineer
---|---|---|---|---
| | | |
Treatment
|
Special Note
The company reserves the right to modify the machine design, product technology or accessories, instruction manual and machine packaging contents at any time without further notice. The product is subject to the actual product, and our company reserve all the rights for the final explanation.