ALAN Dental Equipment AR Heater With Display Instruction Manual

July 18, 2024
ALAN Dental Equipment

ALAN Dental Equipment AR Heater With Display

Product Description

AR heater is designed to heat the prepared inlay and veneer to improve the flowability during bonding, preventing premature curing of the composite material and heating the resin injection nozzle, anesthetic, etc. Using precision circuit to control its heating module, keeping the temperature tolerance within + 1°C. Depending on the material and personal preference, the temperature can be set to: 40°C, 45°C and 50°C.

Composition

  1. Filter cover
  2. Heating plate
  3. Honeycomb heating barrel
  4. Heating base
  5. Indicator light
  6. On/off
  7. TEMP button
  8.  DC
  9. Power adapter

Operation

  1. Place the heating base on a flat tabletop
  2. Insert the DC plug of the power adapter into the DC socket of the heating body, Then plug the power adapter into an electrical outlet. At this time, three temperature indicators are on at the same time (standby heating).
  3. Press the button on/off to turn on the heated body. Press the button again to turn off the heating body. The default temperature for heating base is 400 (When the heating body is off, hree temperature indicators are on at the same time).
  4. Press the button choose which temperature you need, see the detailed temperature list below
    Temperature| Indicator light color| Heating| Reach preset temperature
    ---|---|---|---
    40 [2l| White| White LED shining| White LED is always on
    45 •c| Green| Green LED shining| Green LED is always on
    50°c| Blue| Blue LED shining| Blue LED is always on
  5. When the indicator light changes from flashing to steady light, it indicates that the temperature has reached the preset temperature.
  6. The lights flash alternatively and slowly when cooling.
  7. The heating plate can be taken out from the heating base to heat. Also it can use for filtering without heating.

Note A

  • A. The composite material can not be agitated on the heating plate and heating base.
  • B. Avoid scratching the heating base and the bottom of heating plate.
  • C. To avoid cross-contamination, do not use again the used heating base and heating plate before disinfection.
  • D. The heating base and heating plate must be thoroughly cleaned if the composite material is stuck on it.

Cleaning and disinfection

All parts can be wiped clean with a sterilized water rag. Do not immerse any parts in any liquid. Can not be autoclave sterilization.

Fault promote

  • 3 lights together Temperature sensor failure

Technical parameters

  • Power adapter
  • Input: 100-240V AC 50/60Hz 0.8A Max
  • Output: 12V DC 2A
  • Power: 24W
  • Temperature setting: 40 ₫ 1 °C, 45 ÷ 1 °C and 50 ÷ 1°C
  • Filter range: (light <500nm) can’t pass through the filter cover

Warranty

All parts of this equipment are warranted for 1 year. Man-made damage is not within the free warranty Return to your dealer or manufacturer for repair if have any problems. Do not disassemble the equipment for maintenance. If you dismantle and repair yourself, you will no longer be free during the warranty period