Rotronic CP11 Humidity and Temperature Meter Instruction Manual

June 16, 2024
ROTRONIC

Rotronic CP11 Humidity and Temperature Meter

Specifications

  • Displays & records: relative humidity, temperature, CO2, dew point, and wet bulb
  • Built-in CO2 sensor with calibration cap
  • 5VDC adaptor port
  • PC connect port
  • Power supply: 4pcs AA alkaline batteries or 5VDC adaptor

To connect the CP11 to a PC or Laptop:

  1. Use a mini USB cable to connect the device to the computer.
  2. Download recorded data using the enclosed software.

Functions Key

The functions key allows you to:

  • Turn on or off the instrument.
  • Enter setup mode by pressing longer in normal mode.
  • Disable sleep mode by pressing with HOLD key.
  • Freeze current readings with HOLD/UP key.
  • Start automatic logging mode with START/ESC key.
  • Switch displayed parameters with MODE/RECALL key.

Operation :

To power on/off the device:

  • Press ON/OFF SET to turn on the meter.
  • To power off, press ON/OFF SET again.

FAQ:
Q: How can I change the settings on the CP11?
A: Most settings like memory clear, CO2 alarm, units, sampling rate, etc., can be changed using the function keys. Refer to the detailed instruction manual for specific instructions.

SHORT INSTRUCTION MANUAL

Congratulations on your purchase of the new portable humidity, temperature and CO₂ instrument. Please read these short instructions carefully before installing the device.

General description
The CP11 is an accurate hand-held datalogger that displays & records relative humidity, temperature, CO₂ and calculated parameters such as dew point and wet bulb. These short instructions are limited to a description of the main functions and installation of the device. The detailed instruction manual can be found on the internet at www.rotronic-humidity.com

Programming
Most of the settings, such as memory clear, CO₂ alarm, units(°C/°F), sampling rate, pressure compensation and real time clock can be changed using the function keys.

Power Supply
The CP11 uses 4pcs AA alkaline batteries or 5VDC adaptor (optional accessory). To insert (replace)the batteries, use the screw driver enclosed to loose 2 screws and put in the batteries in correct polarity with good contact.

Connection to PC or Laptop
Connect the CP11 via mini USB cable to a PC or Laptop. Recorded data download can be easily downloaded with the software enclosed.

Functions Key

Display

  • UpperLCD
  • Humidity/TEMP/CO2….reading
  • LowerLCD Real time clock display CO2
  • Carbon dioxide reading HOLD
  • Readings are hold unchanged MIN/MAX
  • Minimum/Maximum readings
  • AVG Average reading
  • Low battery indicator
  • DP TEMP WBT %rh °C/°F CAL REC
  • Dew point temperature
  • Temperature Wet bulb temperature
  • Unit of relative humidity Celsius/Fahrenheit of temperature In calibration status
  • In manual/automatic logging status
  • RECALL In manual records recall mode

 Operation

POWER ON/OFF
Press ON/OFE SET to turn the meter on. At power up, it emits a short beep and performs 28 sec. countdown (Fig. 1) for meter warm up, then enters normal mode with real-time clock displayed (Fig.2). Date and time are displayed in turns.
To power off, press ONIOFE SET again to turn off the meter.

DATE/TIME SETUP

When meter is powered on , hold down meter setup mode. Press key to enter. for more than 2 seconds until entering – key to select the program P80 (rtc) and press.
First, choose the time format as 12hour or 24hour by pressing and then press to confirm. Second, choose the date format as Month-Day-Year or Day- Month- Year by pressing to  and then, press key to confirm. Now, start to input the real time clock value from year, month, day, hour, minute to second. Press or to select and then press to confirm. Press to return to normal.

UNIT SETUP
When meter is powered on , hold down for more than 2 seconds until entering meter setup mode. Press the VEN key to select the program P50 (unit) and press key to enter. Press to normal to select °C or °F and then press to confirm.

Press  MEM to select °C or °F and then press to normal. to confirm. Press to return

TAKING MEASUREMENT
The meter starts measurement when power on and update readings every second. In the condition of operating environment change, it takes 30 seconds to respond for CO₂ sensor.

NOTE: Do not hold the meter close to faces in case exhalation affects CO2 to normal.

Humidity

Press MODE RECALY to switch to humidity display. (Fig.3).The lower display shows the real time clock.

TEMP, DP,WBT
Press MODEN RECALY to switch to temperatures display(Fig. 4,5,6) The lower display shows the real time clock.

S CO2 Carbon Dioxide)
Press to switch the mode to CO₂. User can get the CO2 reading in ppm unit on main display.(Fig.7). The lower display shows the real time clock.

DATA HOLD
In normal display mode, press to freeze the readings, “HOLD” icon is flashed on the left top of LCD (Fig.8). All current readings are kept unchanged. Press again to cancel the hold function.

NOTE: The hold function is disabled in Min/Max/Avg mode.

BACKLIGHT
The backlight will be activated for 10 seconds by pressing any key.

N.MAX.AVG. CURRENT REVIEW
This meter allows you to check the minimum, maximum, average, current value from the moment you press the key. Under normal mode, each press of displays MIN, MAX, AVG and current value in sequence. (Fig.9~11). To return to normal mode, press to escape(Fig. 12) In this max/min/avg/current modes, it shows the corresponding readings on main display and the accumulated time of how long you are in this mode. The maximum time interval for this function is 18 hours.
Press can switch the displayed parameters. (Fig. 13)

Distinguishing the normal display and current review by checking the time display format. In normal display, date & time are displayed in turns. In current review, the lower display is the accumulated time of how long you are in max/min/avg mode.

NOTE: “
HOLD” & “MANUALLY RECORD” & “RECALL” functions are disabled in Min/ Max Ava mode.

ALARM
The meter features audible alarm to give warnings when COz concentration exceeds the limit. (See “Meter Setup” section for setting alarm threshold). It emits beeps (Abt. 80dB) when CO₂ level goes over the set value and stops only when the readings fall below the set value. It beeps again when value goes over the limit.

AUTO POWER OFF
The meter turns off automatically after 20 minutes of inactivity To override the function, first, the meter is in power off status. Hold down ONIOFE SET and *ow for 2 seconds to turn on the meter until “n” appears. (Fig. 14)
NOTE: Auto sleep function will be disabled during calibration & auto recording mode. Auto sleep function WILL NOT be disabled automatically via using adaptor.

MANUALLY RECORDING

MEM The meter features a 99 point memory. In the normal or hold mode, press key to start recording, the REC icon and the main display will be flashed 3 times. The memory sequential number will show on the main display with 99 points in maximum. The data in the memory contains all parameter information and not only the displayed parameter. When all 99 points memory are occupied, a “FULL” message will be displayed on the screen. If you think the reading is changing too fast, you can press key to freeze current reading first then manually record the data. This recording function will be disabled in the Min/Max mode.

99 MEMORIES RECALL
In normal or hold mode, press RECALY key for more than 2 seconds until the Recall icon flashes. Press or to scroll the memories. The memory sequential number displayed on main LCD first and reading comes after. (Fig16&17). Press key to switch the display parameter.(Fig. 18). The time displayed in memory recall mode is the recording time of this memo.

DATA LOGGING
The meter can automatically record Humidity/TEMP./CO2 for long term environment monitor. The memory capacity is 6000 points for each parameter. Users can set up sampling rate from 1 second to 4hours 59 minutes and 59 seconds.
The factory preset rate is 30 second. To setup the sampling rate, when meter is powered 2 seconds until entering meter setup mode. Press P60 (rAtE) and press ( key to enter. hold down for more than key to select the program Press to select the sample rate from hour, minute to second and then press confirm. Press to return to normal. After sampling rate is selected, press SAy for 2 seconds under normal mode to start logging. The REG icon flashes to indicate the logging status and LCD main display shows the real time value. Lower displays are real time clock. (Fig. 19)

Repeat above to start another run of logging. While the logging space is full, it shows “FULL”. To terminate data logging, press for 2 seconds, REC icon stops flashing.
Press is key to switch to the mode you want to see. Except the mode switching function, auto power off, min/max, manual record, hold and recall functions are all disabled during logging.

Meter Setup

When meter is powered on, hold down meter setup mode. To exit setup, press SET for more than 2 seconds until entering STARI ESC to return to normal mode.
Press key to select the program and press The programmable setting flashes on display, press press key to confirm. To leave without saving, press key to select. or MEN key to select and key to return.

Battery low indicator
A battery low icon ” & ” appears when the supplied power is too low to maintain a normal operation. Please replace the batteries for accurate measuring. Suggest to use adaptor (optional part: AC1214) for long term measuring & datalogging.

Humidity Calibration
This meter can be calibrated either via 35% & 80% salt bottle or humidity generator. The ambient condition is recommended as stable 25°C. While the meter is powered off, plug the sensor probe into 35% salt. Press simultaneously for 3 seconds to turn on the meter and enter calibration mode. Press HOLD or H to select 35.0% calibration and press to start. “CAL” and calibrating value are blinking on the LCD.

CAUTION:
Do not calibrate the humidity without the default calibration salt. Otherwise, it will cause permanent damage. Contact ROTRONIC for calibration salt or services. Single point calibration will cause error code E11 or non accurate reading. Always do dual point calibration to complete a process. Suggest to calibrate low humidity first.

CO2 Calibration
The sensor has an automatic calibration function to secure long-term accuracy. If a calibration is desired, press  IN simultaneously for 3 seconds to turn on the meter and enter calibration mode. Press now or MEN) key to select 400 ppm CO₂ calibration (Outside Air calibration). Press to start and see “CAL” and CO₂ value blinking on LCD. Wait about 10 minutes until the blinking stops and the calibration is completed and back to normal mode. While choose 0 ppm CO₂ calibration, remove the cap from meter rear side and plug the CO₂ CALIBRATOR pipe into the sensor hole beneath. Wait about 10 min until the blinking stops and the calibration is completed and back to normal mode.

CAUTION:
Do not calibrate the meter in the air with unknown CO₂ concentration. Otherwise, it will be calibrated as 0 or 400ppm by default and leads to inaccurate measurements. Ensure the batteries are with full voltage during the calibration to prevent from interruption or failed calibration. When the accuracy becomes a concern after a long time usage, return to dealers for standard calibration.

Trouble Shooting

Error Messages Solution

  • E01 CO sensor is out of order Turn off meter and re-start again 2
  • E33 CO sensor is out of order Retry CO calibration 2 2
  • E02 measured value is under range Put meter in normal condition
  • E03 measured value is over range Put meter in normal condition
  • E04 Original data error results in this error (DP, WB)
  • E11 RH calibration error Retry humidity calibration
  • E31 Temp. sensor or AD damaged Return for repair
  • E32 Memory IC damaged Return for repair
  • E33 RH sensor or circuit damaged Return for repair

Technical Data

Optional Accessory

  • 5VDC universal adaptor: CO2

  • calibrator: Part Number

  • AC1214 CO2 CALIBRATOR ROTRONIC AG, CH-8303 Bassersdorf
    Tel. +41 44 838 11 44
    www.rotronic.com

  • ROTRONIC Messgeräte GmbH, D-76275 Ettlingen
    Tel. +49 7243 383 250
    www.rotronic.de

  • ROTRONIC SARL, 56, F – 77183 Croissy Beaubourg
    Tél. +33 1 60 95 07 10
    www.rotronic.fr

  • ROTRONIC Italia srl, I- 20157 Milano
    Tel. +39 2 39 00 71 90
    www.rotronic.it

  • ROTRONIC Instruments (UK) Ltd, West Sussex RH10 9 EE
    Phone +44 1293 571000
    www.rotronic.co.uk

  • ROTRONIC Instrument Corp, NY 11788, USA
    Phone +1 631 427-3898
    www.rotronic-usa.com

  • ROTRONIC Instruments Pte Ltd, Singapore 159836
    Phone +65 6376 2107
    www.rotronic.com

  • ROTRONIC Shanghai Rep. Office, Shanghai 200233, China
    Phone +86 40 08162018
    www.rotronic.cn

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

ROTRONIC User Manuals

Related Manuals