VELP SCIENTIFICA F203A0510 MSL 8 Digital Magnetic Stirrer Instruction Manual

June 8, 2024
VELP SCIENTIFICA

VELP SCIENTIFICA F203A0510 MSL 8 Digital Magnetic Stirrer

VELP SCIENTIFICA F203A0510 MSL 8 Digital Magnetic
Stirrer

General Information

Before using the unit, please read the following instruction manual carefully

Do not dispose of this equipment as urban waste, in accordance with EEC directive 2002/96/CE.

This unit must be used for laboratory applications indoor only. The manufacturer declines all responsibility for any use of the unit that does not comply with these instructions. If the product is used in a not specified way by the manufacturer or with non specified accessories, product’s safety may be compromised.

This unit has been designed and manufactured in compliance with the following standards

Safety requirements for electrical equipment for measurement, control and for laboratory use

Electrical equipment for laboratory use
General requirement
VELP reserves the right to modify the characteristics of its products with the aim to constantly improving their quality.

Safety Regulations

The plug disconnects the instrument. Therefore, place the instrument where it can be quickly disconnected.

Place the power supply in the way to avoid liquid spills directly on it.

The stirred solution may release toxic, dangerous or poisonous gases. Adequate safety measures must be taken, in accordance with the safety regulations in force, including the presence of hood and personal protective equipment (masks, gloves, goggles, etc.).

Position the instrument on a flat surface, with a distance from the wall of 30 cm (at least).

Do not use with explosive and dangerous materials for which the equipment is not designed. The stirrer must not be used in explosive atmospheres, in bain- marie and to stir combustible liquids that have a low combustion temperature. The minimum fire point of flammable solution is 110 °C. Only small amounts (< 50 ml) of flammable liquid can be used with the device.

It is responsibility of the user appropriately decontaminate the instrument in case of dangerous substances fall on or in it. It is also responsibility of the user to use safety substances for cleaning or decontaminating, which do not react with internal parts of the instrument or with the material contained in it. In case of doubts on the compatibility of a cleaning solution, contact the manufacturer or local distributor.

In case of black-out, once the power is back the device will restart automatically.

Introduction

MSL DIGITAL instruments are able to stir up to 8, 25 and 50 L.
Thanks to the new brushless motor, the stirrer stays cold even after several days of continuous operation.
Speed Servo technology maintains constant speed as viscosity changes.
Epoxy painted metal structure, with stainless steel cover, is studied and tested in order to give to the instrument an excellent resistance to the attack of chemical agents and to corrosion in general.
The stirring speed can be set using the frontal knob within a range of 40 – 1500 rpm or 20 – 1000 rpm depending on the model.
The stirrer display shows the speed set point.
The instrument is equipped with advanced functions as auto-reverse of the stirring direction, intermittent mode programs and integrated timer to meet users’ workflow and safety lock feature to avoid possible alteration of the working conditions.
Introduction
Note: it’s important to choose the most suitable magnetic stirring bar related to the quantity and the liquid to be stirred as well as to the type of the beaker.

Installation

  • Unpacking
    • Check the integrity of the unit after unpacking
  • The box includes
    • MSL digital magnetic stirrer
    • Instruction manual
    • Plug for power supply
    • Switching power supply 100-240 V / 12 V
  • Installation
    • Place the unit on a non-flammable surface
    • Make sure that the technical specification of the unit and the power supply line are the same
    • Place the flask containing the sample and a suitable magnetic stirring bar on the stirring plate
    • Connect the instrument to an easily accessible socket (compliant with the current safety norms), using only the provided power supply and dedicated plug.

Operation

  • Push the knob to start stirring.
    Operation

  • Rotate the knob to set the stirring speed.
    Operation
    Note: push the knob to stop stirring.

Auto reverse, intermittent mode and timer functions

  • While the display shows OFF, push the knob for 3 seconds.
    Operation

  • Rotate the knob to set the auto reverse time (minutes/seconds).
    The rotation direction will change after each set interval time.
    Operation
    Note: select NNSS to disable the function.

  • Push the knob. Then rotate it to set the run time of the intermittent mode (minutes/seconds).
    Operation
    Note: select NNSS to disable the function.

  • Push the knob. Then rotate it to set the pause time of the intermittent mode (minutes/seconds).
    Operation
    The stirring function will be interrupted cyclically. The run time and the pause time should both be set.
    Note: select NNSS to disable the function.

  • Push the knob. Then rotate it to set the timer (hours/seconds).
    Operation
    After the timer has elapsed, the stirring will be stopped and the display will show End.
    Note: select HHnn to disable the function.

  • After the timer has elapsed, the stirring will be stopped and the display will show End.
    Note: select HHnn to disable the function.

Lock function

  • While the unit is working, push the knob for 3 seconds to enable the lock function.
    Operation

If the lock function is enabled, it is not possible to change the work conditions. The display shows Lock if the knob is pressed or rotated.
Note: push the knob for 3 seconds to disable the function.

Maintenance

No routine or extraordinary maintenance is necessary except periodically cleaning.

Cleaning
Disconnect the unit from the power supply and use a cloth dampened with a non- flammable non-aggressive detergent.
Repair
Repairs must be carried out by authorized VELP personnel only.

The instrument must be moved using the original shockproof packaging when freight forwarders, couriers or other are taking care of the unit transport. Follow the indications on the original packaging (e.g. place the instrument on a pallet).
It is the responsibility of the user, to properly decontaminate the unit in case of hazardous substances remaining on the surface or interior of the device. If in doubt about the compatibility of a cleaning or decontamination product, contact the manufacturer or distributor.

Technical data

 |  | MSL 8 D ig i t al| MSL 25 D ig i t al| MSL 50 Digitals
---|---|---|---|---

Ge n e r al fea t ure s

| Models| F203A0510| F203A0511| F203A0512
Voltage| 100 – 240 V / 50 – 60 Hz
Power input| 12VDC5W| 12 VDC 9 w| 12 VDC 10 w
Current consumption| 0.5A| 0.8A
Dimensions(WxHxD)| 168x60x183 mm| 253x71x269 mm| 354x77x369 mm
Weight| 0,95 kg| 1,8 kg| 4 kg
Construction material| Epoxy painted metal structure and stainless steel
Timer| 1 min … 99 h 59 min
Auto reverse / Intermittent mode| 5 s + 99 min 55s
Working in continuous| Admitted
Noisiness| << 60 dBa
Environmental temperature admitted| +5… +40 °C
Storage temperature admitted| -10…+60 °C
Max humidity| 80%
Overvoltage cateaory| II
Pollution degree CEI EN61010-1| 2
Max altitude| 2000 m

S ti r

| Stirring capacity| 8L H2O| 25 L H20| 50 L H20
Speed range| 40 + 1500 rpm| 20 + 1000 rpm
Motor type| BLOC
Speed selection| 10 rpm

Accessories & spare parts

A00001056| Stirring bar 6x35mm|
---|---|---
A00001061| Stirring bar 10x60mm
A00000387| Stirring bar 8x50mm
A00000356| Stirring bar 8x40mm
10007175| Knob 35D blue
10000231| Foot 11.5Dx4H| MSL 8

MSL 25

10000232| Foot 14Dx8H| MSL 50
10003081| Switching 100-240V/12V without plug|

MSL 8

10007986| Power supply 100-240V/12V 2.5A no plug| ****

MSL 25

MSL 50

Wiring diagram

Wiring diagram

  1. External power supply
  2. Socket of the instrument
  3. Electronic board
  4. Motor

Declaration of conformity

We, the manufacturer VELP Scientific, under our responsibility declare that the product is manufactured in conformity with the following standards:

EN 61010-1
EN 61326-1
2015/863/EU (RoHS III)
2012/19/EU (WEEE)
EN61010-2-051

and satisfies the essential requirements of the following directives:

  • Machinery directive 2006/42/EC /

 Declaration of conformity

Symbol

We, the manufacturer VELP Scientific, under our responsibility declare that the product is manufactured in conformity with the following regulations:

S.I. 2016/1101 Electrical Equipment (Safety) Regulations 2016
S.I. 2016/1091 Electromagnetic Compatibility Regulations 2016

according to the relevant designated standards:

EN 61010-1 Safety requirements for electrical equipment for measurement, control, and laboratory use
Part1: General requirements
EN 61010-2-051 Particular requirements for laboratory equipment for mixing and stirring
EN 61326-1 Electrical equipment for measurement, control and laboratory use – EMC requirements-
Part 1 : General requirements

and satisfies the essential requirements of regulations:

S.I. 2008/1597 Supply of Machinery (Safety) Regulations 2008
S.I. 2012/3032 Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic
Equipment Regulations 2012
S.I. 2013/3113 Waste Electrical and Electronic Equipment Regulations 2013 plus modifications.

Thank you for having chosen VELP!
Established in 1983, VELP is today one of the world’s leading manufacturer of analytical instruments and laboratory equipment that has made an impact on the world-wide market with Italian products renowned for innovation, design and premium connectivity. VELP works according to ISO 9001, ISO14001 and OHSAS 18001 Quality System Certification.
Our instruments are manufactured in Italy according to the IEC 1010-1 and CE regulation.

Customer Support

www.velp.com
VELP Scientifica Srl
20865 Usmate (MB) ITALY
Via Stazione, 16
Tel. +39 039 62 88 11
Fax. +39 039 62 88 120

VELP SCIENTIFICA Logo

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

VELP SCIENTIFICA User Manuals

Related Manuals