VELP SCIENTIFICA F202A0270 TX4 Digital IR Vortex Mixer Instruction Manual
- June 8, 2024
- VELP SCIENTIFICA
Table of Contents
VELP SCIENTIFICA F202A0270 TX4 Digital IR Vortex Mixer
General Information
Before using the unit, please read the following instruction manual carefully.
Do not dispose of this equipment as urban waste, in accordance with EEC directive 2002/96/CE.
This unit must be used for laboratory applications only.
The manufacturer declines all responsibility for any use of the unit that does
not comply with these instructions. If the product is used in a not specified
way by the manufacturer or with non specified accessories, product’s safety
may be compromised.
This unit has been designed and manufactured in compliance with the following standards:
Safety requirements for electrical equipment for measurement, control and for laboratory use.
Electrical equipment for laboratory use
General requirement
VELP reserves the right to modify the characteristics of its products with the aim to constantly improving their quality.
Safety Regulations
Connect the instrument using only the supplied external power supply. The plug disconnects the instrument. Therefore, place the instrument where it can be quickly disconnected.
Position the instrument on a flat surface, with a distance from the wall of 30 cm (at least).
Before placing the tube on the stirring cup, make sure that its cap is tightly closed to prevent sample leakage.
The solution may release toxic, dangerous or poisonous gases. Adequate safety measures must be taken, in accordance with the safety regulations in force, including the presence of hood and personal protective equipment (masks, gloves, goggles, etc.).
Do not use with explosive and dangerous materials for which the equipment is not designed. The stirrer must not be used in explosive atmospheres.
Keep hand/arm vibration to a minimum (Directive 2002/44/EC) when operated manually:
- Keep the sample container as vertical as possible.
- Only exert the minimum pressure necessary for the mixing process.
- Use attachments for multiple samples when there is a large number of samples.
Introduction
Mixing by spinning can be carried out quickly and safely by placing the tube
containing the sample on the mixing cup head. The Vortex Mixer with LCD
display can be used in automatic (SENSOR) or continuous mode (TIMER): in both
cases the mixing speed can be adjusted using the speed selector knob (1).
When the instrument is switch on, it is always set in SENSOR mode. Mixing
starts when the test-tube crosses the infrared The instrument is factory set
to continuous mode (Timer), mixing starts automatically when the speed
selector / settings knob is pressed. To select Sensor mode, rotate the knob to
highlight “Sensor” on the display, press the knob to activate the operating
mode selected. Mixing starts when the test-tube crosses the infrared sensor
field (5) near the mixing cup head (4). The press-on fitting means that the
mixing cup head can easily be replaced by any of the various accessories
available for test-tubes of different sizes.
Assembly and installation
Check the integrity of the unit after unpacking. The box includes:
- TX4 Digital IR Vortex Mixer with mixing cup head.
- Power supply 100-240V/12V no plug.
- EU adapter plug.
- Instruction manual.
Electrical connections
After having unpacked the instrument, place the unit on the laboratory bench.
Before connecting the instrument to the power supply, make sure that the
values on the rating plate correspond to those of the power supply. Connect
the unit to the power supply using the transformer supplied.
Ensure that the socket and the relative cut-off device conform to current
safety norms and are easy to reach.
Start-up
Check that the mixing cup head is correctly positioned. Also check that the speed regulation knob is turned to “0 rpm”. Use the on-off switch on the front of the unit to turn the unit on (position “ON”). The start-up and the control of the speed are effected with the knob. When switched on the unit will automatically start in the last operating mode set (Timer / Sensor).
Information about construction materials
Housing
Paint
Frontal control panel
Cup head| Zinc alloy/Polymer
PET
PET
SEBS| Feet
Platform
Foamed part for platforms| PVC
PP
PE
---|---|---|---
Operating controls
Connect the unit to mains and turn it on using the on-off switch. Select the operating mode using the mode select button on the front of the unit.
When Timer mode is selected, the timer can be set and the mixing cup head runs
continuously at the set speed.
When the Touch mode is selected, mixing starts automatically when the test-
tube crosses the sensor field (infrared optical system) shown in Figure 2.
In both operating modes there is no need for any pressure on the mixing cup
head. The sensor field is purpose-designed to detect the presence of any kind
of test-tube.
NOTE: in case of black-out, the device will not automatically restart
after the break and reconnect the power supply.
SETTINGS KNOB
With the instrument in stand-by, turn the knob to select and modify the operating parameters (rpm, timer, Start and Sensor) (Fig. 3). With the instrument running, turn the knob to increase or decrease the speed from 0 to 3000rpm or from 0 to 800rpm depending on the full-scale setting. Press the speed selector / settings knob to start or stop mixing in Timer mode (Start or Stop highlighted – Fig 4), to select the operating mode (Sensor or Timer highlighted – Fig 5) and/or to modify speed and time parameters (rpm or timer highlighted – Fig 6 and 7). With the timer highlighted (Fig. 7), turn the knob to set the time from 1 second to 999min. and 59 seconds in Timer mode. During use the display will show the countdown and then END (Fig. 9). To disactivate the time keep the speed selector / settings knob pressed for 3 seconds. The display will show — : –: mixing will be active for an infinite time. When a parameter is highlighted (Fig. 3), keep the knob pressed for 3 seconds to set the max speed (Fig. 8). This function can be used in Timer mode only. Use the 800rpm full scale setting if you are using a platform. When Timer is highlighted (Fig. 5), keep the knob pressed for 3 seconds to set the sensor time from 1 second to 999min. and 59 seconds (Fig. 10), this function can be used in Sensor mode only. The display shows the countdown and mixing stops automatically.
The on-off switch turns the unit on and off. If the switch is in the “OFF”
position the unit is off; if the switch is in the “ON” position the unit is
on.
Always turn the unit off after use.
Maintenance
No routine or extraordinary maintenance is necessary apart from periodically
cleaning the unit as described in this manual. In compliance with the product
guarantee law, repairs to our units must be carried out in our factory, unless
previously agreed otherwise with local distributors.
To replace the cup head, pull it upwards and off. The other push-fit
accessories can then be inserted.
The instrument must be transported in its original packaging and any
indications present on the original packaging must be followed (e.g.
palletized).
It is the responsibility of the user, to properly decontaminate the unit in
case of hazardous substances remaining on the surface or interior of the
device. If in doubt about the compatibility of a cleaning or decontamination
product, contact the manufacturer or distributor.
Cleaning
Disconnect the unit from the power supply and use a cloth dampened with an non-inflammable non aggressive detergent.
Technical data
Voltage | AC 100-240V (±10%); 50-60 Hz |
---|---|
Power input | 12 VDC ; 9.2W |
Current consumption | 0.77A |
Dimensions (WxHxD) | 150x130x165 mm (5.9×5.1×6.5 in) |
Weight | 2.7 Kg (6.2 lb) |
Speed range | 0-3000 rpm or 0-800 rpm |
Timer range | 1 sec. – 999min. 59sec. |
Spin diameter | 4.5 mm (0.177 in) |
Max. weight on the mixing cup head / Accessories | 0.5 kg (1.1 lb) |
Permitted operation | Continuous |
Operation mode | Timer or Sensor |
Level of electrical protection IEC/EN60529+A1 | IP 42 |
Noise level | << 80 dBa |
Admitted temperature | Storage: -10…+60 °C (+14…+140 F°) Working: 5°…+40 °C |
(+41…+104 °F)
Admitted humidity| Max. 80%
Overvoltage category| II
Pollution degree IEC/EN61010-1| 2
Max. altitude| 2000 m
Accessories / Spare parts
-
A00000012 Foam stand for 19 microvials 1.5 ml – Eppendorf®
-
A00000013 Customizable soft foam top
-
A00000014 Foam stand 5 test tubes Ø16mm
-
A00000015 Foam stand for microtiter
-
A00000016 Small rubber supporting plate Ø 50mm
-
A00000019 Foam stand 4 test tubes Ø29mm
Do not exceed 800 rpm when using one of the above accessories. High speeds may cause the sample to spill.
Thank you for having chosen VELP!
Established in 1983, VELP is today one of the world’s leading manufacturer of
analytical instruments and laboratory equipment that has made an impact on the
world-wide market with Italian products renowned for innovation, design and
premium connectivity. VELP works according to ISO 9001, ISO14001 and OHSAS
18001 Quality System Certification.
Our instruments are manufactured in Italy according to the IEC 1010-1 and CE
regulation.
Our product lines:
Analytical instruments
Elemental Analyzers
Digestion Units
Distillation Units
Solvent Extractors
Fiber Analyzers
Dietary Fiber Analyzers
Oxidation Stability Reactor
Consumables
Laboratory Equipment
Magnetic Stirrers
Heating Magnetic Stirrers
Heating Plates
Overhead stirrers
Vortex mixers
Homogenizers
COD Thermoreactors
BOD and Respirometers
Cooled Incubators
Flocculators
Overhead Shakers
Turbidimeter
Radiation Detector
Open Circulating Baths
Pumps
Declaration of conformity
We, the manufacturer VELP Scientifica, under our responsibility declare that the product is manufactured in conformity with the following standards:
EN 61010-1 EN 61010-2-051 EN 61326-1
and satisfies the essential requirements of the following directives: e soddisfa i requisiti essenziali delle direttive:
2006/42/EC
2014/30/EU
2015/863/EU (RoHS III)
2012/19/EU (WEEE)
plus modifications
Declaration of conformity
We, the manufacturer VELP Scientifica, under our responsibility declare that the product is manufactured in conformity with the following regulations:
S.I. 2016/1101 Electrical Equipment (Safety) Regulations 2016
S.I. 2016/1091 Electromagnetic Compatibility Regulations 2016
according to the relevant designated standards:
EN 61010-1: Safety requirements for electrical equipment for measurement, control, and laboratory use Part 1: General requirements.
EN 61010-2-051: Particular requirements for laboratory equipment for mixing and stirring
EN 61326-1 Electrical equipment for measurement, control and laboratory use – EMC requirements – Part 1: General requirements
and satisfies the essential requirements of regulations:
S.I. 2008/1597 Supply of Machinery (Safety) Regulations 2008
S.I. 2012/3032 Restriction of the Use of Certain Hazardous Substances in
Electrical and Electronic
Equipment Regulations 2012
S.I. 2013/3113 Waste Electrical and Electronic Equipment Regulations 2013
plus modifications.
We respect the environment by printing our manuals on recycled paper.
Customer Support
www.velp.com
VELP Scientifica Srl
20865 Usmate (MB) ITALY
Via Stazione, 16
Tel. +39 039 62 88 11
Fax. +39 039 62 88 120