VELP SCIENTIFICA F203A0450 MST Digital Magnetic Stirrer Instruction Manual

June 8, 2024
VELP SCIENTIFICA

**HOOVER HHP30C011 Air Cleaner User Manual

**

General Information

Before using the unit, please read the following instruction manual carefully.

Do not dispose of this equipment as urban waste, in accordance with EEC directive 2002/96/CE.

This unit must be used for laboratory applications indoor only. The manufacturer declines all responsibility for any use of the unit that does not comply with these instructions. If the product is used in a not specified way by the manufacturer or with non specified accessories, product’s safety may be compromised.

This unit has been designed and manufactured in compliance with the following standards:

Safety requirements for electrical equipment for measurement, control and for laboratory use:
IEC/EN 61010-1
IEC/EN 61010-2-051

Electrical equipment for laboratory use: UL 61010-1
General requirement – Canadian electrical code: CAN/CSA-C222 No.61010-1
VELP reserves the right to modify the characteristics of its products with the aim to constantly improving their quality.

Safety Regulations

The plug disconnects the instrument. Therefore, place the instrument where it can be quickly disconnected.

Place the power supply in the way to avoid liquid spills directly on it.

The stirred solution may release toxic, dangerous or poisonous gases. Adequate safety measures must be taken, in accordance with the safety regulations in force, including the presence of hood and personal protective equipment (masks, gloves, goggles, etc.).

Position the instrument on a flat surface, with a distance from the wall of 30 cm (at least).

Do not use with explosive and dangerous materials for which the equipment is not designed. The stirrer must not be used in explosive atmospheres.

It is responsibility of the user appropriately decontaminate the instrument in case of dangerous substances fall on or in it. It is also responsibility of the user to use safety substances for cleaning or decontaminating, which do not react with internal parts of the instrument or with the material contained in it. In case of doubts on the compatibility of a cleaning solution, contact the manufacturer or local distributor.

Do not use with explosive and dangerous materials for which the equipment is not designed. The stirrer must not be used in explosive atmospheres, in bain- marie and to stir combustible liquids that have a low combustion temperature. The minimum fire point of flammable solution is 200 °C. Only small amounts (< 50 ml) of flammable liquid can be used with the device.

In case of black-out, once the power is back the device will restart automatically.

Introduction

The MST magnetic stirrer is a small, simple and effective stirrer.
Thanks to the new brushless motor, the stirrer stays cold even after several days of continuous operation.
Thanks to the white cover surface it is specially designed for microtitrations, and it is extremely useful where reliable and small instrument is needed.
The stirring speed can be set using the frontal knob within a range of 80 – 1500 rpm
The stirrer display shows the speed set point.
The instrument is equipped with advanced functions as auto-reverse of the stirring direction, intermittent mode programs and integrated timer to meet users’ workflow and safety lock feature to avoid possible alteration of the working conditions

NOTE: it’s important to choose the most suitable magnetic stirring bar related to the quantity and the liquid to be stirred as well as to the type of the beaker. The magnetic stirring bar that satisfies most of the application is A00001056 (6x35mm).

Installation

  • Unpacking
    • Check the integrity of the unit after unpacking
  • The box includes
    • MST Digital magnetic stirrer
    • Instruction manual
    • Plug for power supply
    • Switching power supply 100-240 V / 12 V
  • Installation
    • Place the unit on a non-flammable surface
    • Make sure that the technical specification of the unit and the power supply line are the same
    • Place the flasks containing the sample and a suitable magnetic stirring bar on the stirring plate
    • Connect the instrument to an easily accessible socket (compliant with the current safety norms), using only the provided power supply and dedicated plug.

Operation

  • Push the knob to start stirring
    Operating Instruction

  • Rotate the knob to set the stirring speed.Operating Instruction

Note: push the knob to stop stirring.

Auto reverse, intermittent mode and timer functions

  • While the display shows OFF, push the knob for 3 seconds.
    Operating Instruction

  • Rotate the knob to set the autoreverse time (minutes/seconds).Operating Instruction
    The rotation direction will change after each set interval time.
    Note: select NNSS to disable the function.

  • Push the knob. Then rotate it to set the run time of the intermittent mode (minutes/seconds).
    Operating Instruction
    Note: select NNSS to disable the function.

  • Push the knob. Then rotate it to set the pause time of the intermittent mode (minutes/seconds).
    Operating Instruction
    The stirring function will be interrupted cyclically. The run time and the pause time should both be set.
    Note: select NNSS to disable the function.

  • Push the knob. Then rotate it to set the timer (hours/seconds).Operating Instruction
    After the timer has elapsed, the stirring will be stopped and the display will show End.
    Note: select HHnn to disable the function.

  • Push the knob or wait 10 seconds to confirm. Dedicated blinking leds show the active functions.
    Note: each function can be set individually or in combination with the others.

Lock function

  • While the unit is working, push the knob for 3 seconds to enable the lock function.
    Operating Instruction

If the lock function is enabled, it is not possible to change the work conditions. The display shows Lock if the knob is pressed or rotated.
Note: push the knob for 3 seconds to disable the function.

Maintenance

No routine or extraordinary maintenance is necessary except periodically cleaning.

Cleaning
Disconnect the unit from the power supply and use a cloth dampened with a non- flammable non-aggressive detergent.

Repair
Repairs must be carried out by authorized Velp personnel only.

The instrument must be moved using the original shockproof packaging when freight forwarders, couriers or other are taking care of the unit transport. Follow the indications on the original packaging (e.g. place the instrument on a pallet). It is the responsibility of the user, to properly decontaminate the unit in case of hazardous substances remaining on the surface or interior of the device. If in doubt about the compatibility of a cleaning or decontamination product, contact the manufacturer or distributor.

Technical data

| |
---|---|---
General features| Model| F203A0450
Voltage| 100 – 240 V / 50 – 60 Hz
Power input| 12 VDC 2.5 W
Current consumption| 0.2 A
Dimensions (WxHxD)| 130x50x150 mm
Weight| 0.4 kg
Construction material| PA
Timer| 1 min ÷ 99 h 59 min
Autoreverse / Intermittent mode| 5 s ÷ 99 min 55s
Working in continuous| Admitted
Noisiness| < 35 dBa
Environmental temperature admitted| +5…+40 °C
Storage temperature admitted| -10…+60 °C
Max humidity| 80%
Overvoltage category| II
Pollution degree CEI EN61010-1| 2
Max altitude| 2000 m
Stir| Stirring capacity| 5 L H2O
Speed range| 0 ÷ 1500 rpm
Motors type| BLDC
Speed selection| 10 rpm

Declaration of conformity

We, the manufacturer VELP Scientifica, under our responsibility declare that the product is manufactured in conformity with the following regulations:

S.I. 2016/1101 Electrical Equipment (Safety) Regulations 2016 S.I. 2016/1091 Electromagnetic Compatibility Regulations 2016 according to the relevant designated standards: EN 61010-1 Safety requirements for electrical equipment for measurement, control, and laboratory use Part 1: General requirements EN 61010-2-051 Particular requirements for laboratory equipment for mixing and stirring EN 61326-1 Electrical equipment for measurement, control and laboratory use –  EMC  requirements – Part 1: General requirements and satisfies the essential requirements of regulations: S.I. 2008/1597 Supply of Machinery (Safety) Regulations 2008 S.I. 2012/3032 Restriction of the Use of Certain Hazardous Substances in Electrical and Electronic Equipment Regulations 2012
S.I. 2013/3113 Waste Electrical and Electronic Equipment Regulations 2013 plus modifications.

Thank you for having

Established in 1983, VELP is today one of the world’s leading manufacturer of analytical instruments and laboratory equipment that has made an impact on the world-wide market with Italian products renowned for innovation, design and premium connectivity. VELP works according to ISO 9001, ISO14001 and OHSAS 18001 Quality System Certification. Our instruments are manufactured in Italy according to the IEC 1010-1 and CE regulation.

Our product lines:

Analytical instruments

Elemental Analyzers  Digestion Units  Distillation Units Solvent ExtractorsFiber Analyzers Dietary Fiber Analyzers Oxidation Stability Reactor Consumables

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

VELP SCIENTIFICA User Manuals

Related Manuals