ADAM Advanced Label Printing Scales User Manual
- June 7, 2024
- ADAM
Table of Contents
Advanced Label Printing Scales
This document hosted by: www.oldwillknottscales.com
Adam Equipment
BCT
Software rev: V 1.00 & above
© Adam Equipment Company 2019
Easy Reference: Model name of the indicator/scale: Serial number of the unit:
Software revision number (Displayed when power is first turned on): Date of
Purchase: Name of the supplier and place:
© Adam Equipment Company 2019
CONTENTS
PN 3.11.6.6.15391, Rev B, August 2019
1.0 INTRODUCTION ………………………………………………………………………………………….2 2.0
SPECIFICATIONS………………………………………………………………………………………..3 3.0
INSTALLATION……………………………………………………………………………………………5
3.1 UNPACKING …………………………………………………………………………………………….5 3.2 LOCATING
……………………………………………………………………………………………….5 4.0 KEY
DESCRIPTIONS……………………………………………………………………………………6 5.0 DISPLAYS
…………………………………………………………………………………………………..7 5.1 WEIGHT DISPLAY
…………………………………………………………………………………….7 5.2 UNIT WEIGHT DISPLAY
…………………………………………………………………………….8 5.3 COUNT
DISPLAY………………………………………………………………………………………8 5.4.1 SYMBOL ICONS
…………………………………………………………………………………….8 6.0
OPERATION………………………………………………………………………………………………..9 6.1 ZEROING THE DISPLAY
……………………………………………………………………………9 6.2 TARING
……………………………………………………………………………………………………9 6.3 CHECKWEIGHING
…………………………………………………………………………………10 6.4 PARTS
COUNTING…………………………………………………………………………………11 6.5 SETTING THE PLU DATA
……………………………………………………………………….12 7.0 USER PARAMETERS
…………………………………………………………………………………15 7.1 RS 232 PARAMETERS
…………………………………………………………………………….15 7.2 SYSTEM PARAMETERS
………………………………………………………………………….16 8.0 CALIBRATION
…………………………………………………………………………………………..19 8.1 SETTING CAPACITY AND DIVISION
…………………………………………………………19 8.2 CALIBRATION ……………………………………………………………………………………..21
9.0 LABEL PRINTING FUNCTION …………………………………………………………………….22 9.1 LABEL EDITING
COMPUTER SOFTWARE ………………………………………………..22 9.2 LABEL DOWNLOAD PROCEDURE VIA RS
232 CABLE ………………………………28 10.0 RS-232 INTERFACE
………………………………………………………………………………..29 10.1 INPUT COMMANDS FORMAT
……………………………………………………………….31 10.2 REAL TIME CLOCK SETUP
…………………………………………………………………..32 10.3 AUTO SLEEP FUNCTION
……………………………………………………………………..32 11.0 BATTERY AND BACKLIGHT
OPERATION………………………………………………..32 11.1
BATTERY…………………………………………………………………………………………….32 11.2 CHARGING
………………………………………………………………………………………….33 11.3 BACKLIGHT FOR LCD
………………………………………………………………………….33 12.0 ERROR CODES
………………………………………………………………………………………34 13.0 REPLACEMENT PARTS AND
ACCESSORIES…………………………………………..35 14.0 SERVICE
INFORMATION…………………………………………………………………………35 WARRANTY
INFORMATION………………………………………………………………………………..36
© Adam Equipment Company 2019
1.0 INTRODUCTION
The BCT series provides a versatile check counting and weighing platform scale
perfect for general warehouse and production.
The BCT can be set to verify counts up to 60000 divisions All indicators are
supplied with a RS-232 bi-directional interface and real time clock
(RTC). The BCT indicator has a sealed keypad with colour coded membrane
switches and
there is a large, easy to read liquid crystal type display (LCD). The LCD is
supplied with a backlight. The BCT includes parts counting, check counting and
averaging applications.
© Adam Equipment Company 2019
2.0 SPECIFICATIONS
BCT Models
Model name:
BCT 16
Maximum Capacity
16kg
Readability
0.5g
Resolution
1:32000
Repeatability (sd) 0.5g
Linearity
1g
Eccentric Loading 1g
Hysteresis
1g
BCT 32
32 kg
1g 1:32000 1g 2g 2g 2g
BCT 60
60kg
2g 1:30000 2g 4g 4g 4g
BCT 75
75kg
0.005kg 1: 15000 0.005kg 0.01kg 0.01kg 0.01kg
BCT 150
150kg
0.01kg 1: 15000 0.01kg 0.02kg 0.02kg 0.02kg
BCT 300
300kg
0.02kg 1: 15000 0.02kg 0.04kg 0.04kg 0.04kg
BCT a Models Model name:
BCT 35a BCT 65a BCT 130a BCT 165a BCT 330a BCT 660a
Maximum Capacity 35lb 16kg
/ 65lb 30kg
/ 130lb 60kg
/ 165lb 75kg
/ 330lb / 660lb /
150kg
300kg
Readability
0.001lb / 0.002lb / 0.005lb / 0.01lb / 0.02lb / 0.05lb /
0.5g
1g
2g
0.005kg
0.01kg
0.02kg
Resolution
1:32000 1:30000 1:30000 1:15000
1: 15000 1:15000
Repeatability (sd) 0.001lb / 0.002lb / 0.005lb / 0.01lb / 0.02lb / 0.05lb /
0.5g
1g
2g
0.005kg
0.01kg
0.02kg
Linearity
0.002lb / 0.005lb / 0.01lb / 0.02lb / 0.05lb / 0.1lb
/
1g
2g
5g
0.01kg
0.02kg
0.05kg
Eccentric Loading 0.002lb / 0.005lb / 0.01lb / 0.02lb / 0.05lb / 0.1lb
/
1g
2g
5g
0.01kg
0.02kg
0.05kg
Hysteresis
0.002lb / 0.005lb / 0.01lb / 0.02lb / 0.05lb / 0.1lb
/
1g
2g
5g
0.01kg
0.02kg
0.05kg
© Adam Equipment Company 2019
Other specification
Pan size
BCT 16, 32, 60: 300 x 400mm
BCT 75, 150, 300: 400 x 500mm
BCT 35a, 65a, 130a: 300 x 400mm
BCT 165a, 330a, 660a: 400 x 500mm
Operating temperature Power supply
0C-40C
6V 10Ah Rechargeable battery and 12V DC 3A external adapter
Battery life
90 hours typical. Battery Life is less when the backlight is used.
Interface Display
Scale housing
RS-232 Bi-directional
3 displays showing 6, 6 & 7 digits
Indicator: ABS plastic Platform: Stainless steel
Overall dimension
Net weight Functions Weighing units Calibration
BCT 16, 32, 60, 35a, 65a, 130a: BCT 75, 150, 300, 165a, 330a, 660a:
805x425x170mm
870x475x205mm
8.5 kg
13.5 kg
Weighing, parts counting, check counting
Grams, Kilogram, pound, ounce
Automatic external calibration
© Adam Equipment Company 2019
3.0 INSTALLATION
3.1 UNPACKING This scale must be connected to a load cell platform and
calibrated as necessary to match the platform and user requirements. The
user’s application and the technical specifications of the platform or load
cell will determine the necessary configuration. 3.2 LOCATING
The scale should not be placed in a location that will reduce the accuracy.
Avoid extremes of temperature. Do not place in direct sunlight or near air
conditioning vents.
Avoid unsuitable tables. The table or floor must be rigid and not vibrate.
Avoid unstable power sources. Do not use near large users of electricity such
as welding equipment or large motors.
Do not place near vibrating machinery. Avoid high humidity that might cause
condensation.
Avoid direct contact with water. Do not spray or immerse the indicator in
water. Avoid air movement such as from fans or opening doors. Do not place
near open windows or airconditioning vents. Keep the scale clean. Do not stack
material on the indicator when they are not in use.
© Adam Equipment Company 2019
4.0 KEY DESCRIPTIONS
Keys [0-9]
Functions
Numeric entry keys, used to manually enter a value for tare weights, unit
weight, and sample size.
Secondary functions: Used to type letters where necessary; used to add a decimal point.
[PST/ CE]
Clear the preinstall value.
[M+] [AVG.]
Add all the quantities and store, the max Total Number of Additions is 99.
On/Off the automatic average function.
[QTY/ PST] Preinstall the High/low Limit values. [ID]
Press this key to enter the ID setting.
[UNITS]
Choose the units
[PLU]
Used to access any stored PLU weight values
[CE/ BACKLIGHT]
Used to clear numeric totals on the display.
Secondary function: Allows the backlight to be configured to ON/ OFF/ AUTO
© Adam Equipment Company 2019
[U 1-6]
Fetch a stored PLU value directly Used to add a decimal point when entering a number on the display
[PRINT/ SAVE]
Print the total weight, unit weight, quantity of percent
[SMPL]
Used to input the number of items in a sample.
[UNIT W.T.] Used to enter the weight of a sample manually. [Tare ]
Tares the scale. Stores the current weight in memory as a tare value, subtracts the tare value from the weight and shows the results. This is the net weight. Entering a value using the keypad will store that as the tare value.
[0/ Zero] Sets the zero point for all subsequent weighing to show zero.
5.0 DISPLAYS
The scale includes one LED display separated into three. These are “Weight”,
“Unit Weight” and “Count”.
5.1 WEIGHT DISPLAY : Stability indicator
0 : Zero indicator Weight: The weight of goods. Unit weight: The unit weight
of goods. Total Count: The total count of goods.
Tare: Used as tare key in pricing mode Lo-bat: When battery power lower than
5.6V, it will show. Kg / Lb / g / oz /: Symbols shown for the units
Net: Net weight- The scale has been tared
© Adam Equipment Company 2019
5.2 UNIT WEIGHT DISPLAY
This display will show the unit weight of a sample. This value is either input
by the user or computed by the scale. The unit of measurement is grams if
grams or kilograms is selected for the weighing unit, or pounds if pounds is
selected.
5.3 COUNT DISPLAY
This display will show the number of items on the scale or the value of the
accumulated count.
5.4.1 SYMBOL ICONS
“Z” or Zero: “Zero” indication.
Tare:
“Tare” displays the deduction of the container weight.
M+: “M+”
Totaliser.
If the piece weight is smaller than “Minimum Sampling Weight”, ” ” will be displayed. In order to clear the ” ” symbol, increase the size of the sample.
If the piece weight is smaller than “Minimum Unit Weight”, ” ” will be
displayed. The scale will still weigh even though the piece weight is too
small, however this may affect the count accuracy.
Press [Q’TY PST] key to enter the H/L Limit value for Quantity Preset in the
weighing mode. “LIMOF” is used to configure the alarm settings, quantity or
weight. In the unit weight Window, it will show “weight” or “count”, press
[M+] key to change the alarm settings.
© Adam Equipment Company 2019
6.0 OPERATION
NOTE: BCT scales only SETTING THE WEIGHING UNIT, lb or kg
The BCT scale will turn on displaying the last weighing unit selected, either
kilograms or pounds. To change the weighing unit press the [Units] key when
the “Unit Weight” display shows zero. If necessary press the [CE] key to clear
the unit weight before changing.
6.1 ZEROING THE DISPLAY Press the Zero key to re-zero the display with no load
on the weigh pan. If the balance was damaged or the weight is bigger than the
range of the zero, this function will be inactive. It may be necessary to re-
zero the scale over a period of time to account for minor drifting.
6.2 TARING Zero the scale by pressing the [Zero/0] key if necessary. The
indicator “0” will be ON.
Place a container on the platform and its weight will be displayed.
Press [Tare] to tare the scale. Once the scale stabilises, it will display the
net weight and the NET sign will be displayed.
As a product is added, only the weight of the product will be shown. The scale
could be tared a second time if another type of product was to be added to the
first one. Again only the weight that is added after taring will be displayed.
When the container is removed a negative value will be shown. If the scale was
tared just before removing the container, this value is the gross weight of
the container plus all products those were removed. The indicator “0” will
also be ON because the platform is back to the same condition as it was when
the [Zero/ 0] key was pressed last.
TARE WEIGHT IS KNOWN (PRE-TARE)
When the pan is empty, press [Tare] key, the display will show “P- TARE”.
Input the weight of the container, press [Tare] key, Pre-tare complete. User
can set follow function ON or OFF in F5 PTE:
In weighing mode, when tare weight is known, input tare weight, then press
[Tare] key to Pre-tare.
© Adam Equipment Company 2019
Specific operation as follows:
Put the container and weights on the pan, input the container’s weight and
press [Tare] key, the weight window will show the net weight of the weights.
Take away the container and weights, the weight windows will show a negative
value. At this time, press [Tare] key to remove the tare weight value. Now,
the weight window back to zero.
6.3 CHECKWEIGHING 6.3.1 SETTING THE CHECKWEIGHING LIMITS
To set the checkweighing limits ensure there is nothing on the scale/ platform
and that the display shows zero.
Press the [Q’TY/PST] key to bring up the checkweighing display. You should now
see “LIMOF” then “Weight” or “Count” on the display, press the [M+] key in
order to alternate between the two.
Once selected, press the [Q’TY/PST] key to move to the next setting. “CHLn0d H
L” should now appear on the display.
Press the [M+] key to set the check alarm configuration, this will position a
small dash on either side of the “H” and “L” or in between. (i.e. H-L, -H L, H
L- etc).
There are 6 various customisable alarm options:
-H L
H – L H L-H L- H – L H – L –
The beep is on when the weighing mass is more than the set high limit. The
beep is on when the weighing mass is between the set high and low limit. The
beep is on when the weighing mass is below the set low limit.
The beep is on when the weighing mass is both more than and below the set
limits. The beep is on when the weighing mass is in between the set limits and
more than the set high limit. The beep is on when the weighing mass is in
between the set limits and below the set low limit.
© Adam Equipment Company 2019
Press [Q’TY/PST] to confirm and move to the next setting. “LIM-Hi” should now
appear on the display with zero digits on the right of the display. Use the
numeric keys [1-9] to manually enter the high check limit. Press [Q’TY/PST] to
confirm.
“Lim-Lo” should now appear on the display with the same zero digits on the
right of the display. Use the numeric keys to enter to low limit and press the
[Q’TY/PST] key to confirm.
6.4 PARTS COUNTING
6.4.1 THE UNIT WEIGHT OF OBJECT IS UNKNOWN
Place a sample quantity of items on the scale/ platform, Enter the quantity of
items by using the numeric keys on the indicator, then press the [SMPL] key.
Once the scale stabilises, the quantity will show on the total quantity
display and the balance will enter the counting mode automatically.
6.4.2 THE UNIT WEIGHT OF OBJECT IS KNOWN
Enter the known weight of the item using the numeric keys on the indicator,
Press the [UNIT W.T.] key, the quantity is shown on the quantity display and
enter the counting mode automatically.
The larger the sample weight is, the more accurate the quantity will be. With
both the “Unit weight” and “Quantity” displays showing zero, pressing either
the [UNIT W.T]. Key or the [SMPL] key will recall the latest piece weight
value. When Press the [UNIT W.T]. key to get the unit weight, if the value
which you place
next time is more than 10% of the last sampling value and less than the last
sampling value of present, the balance will automatically re-calibration the
unit weight. User can set this function on /off in “F2 AVG”. If the unit
weight and quantity all known, the balance will recalculate the unit weight
according the quantity shows in the total quantity display when you press the
[UNIT W.T].
6.4.3 AUTOMATIC UPDATE OF UNIT WEIGHT
The scales will automatically update the unit weight when a sample less than
the initial sample count is added. A beep will be heard when the value is
updated. It is wise to check the quantity is correct when the unit weight has
been updated automatically.
The user can turn this function ON/OFF by pressing the AVG key.
© Adam Equipment Company 2019
6.4.4 ACCUMULATION/ TOTALISING
The totalising function can be used up to a maximum of 99 times before it must
be reset. The totalising display is limited to seven digits maximum. When the
quantity is shown on the quantity display, Press [M+] Key, and then enters the
totalising function. The quantity display shows “Add XX “, XX means the Total
Number of Additions. After 3 seconds, the scale will return to the counting
mode.
VIEWING ACCUMULATED DATA When the weight window shows zero, press [M+] key to
show the accumulative data. First, the scale will show total accumulative data
as indicated by t XX in the weighing window. Press [M+] again, the scale
will show all individual stored values as indicated by Add XX in the weighing
window, followed by the stored amount. Press [M+] to view each stored value.
When the scale shows one of the stored values, press [PST CE] to delete the
selected stored value. When the scale shows all accumulated data, press [PST
CE], all accumulated data will be deleted. In weighing mode, press [PST CE],
all accumulated will be deleted.
6.5 SETTING THE PLU DATA
In normal weighing mode, use number keypad to input PLU serial number, then
press [PLU], can get the weights name and unit weight of the corresponding
weights. The display will show the item name once and back to normal weighing
mode.
6.5.1 EDIT THE PLU DATA
In normal weighing mode, press and hold [PLU] key for 3s to enter PLU date
edit mode.
The weight window shows “set”, user input the PLU number, then press [PLU] (or
press “Ux” key directly), the weight window will show the input PLU number.
Now, scale enter into unit weight edit mode.
If this PLU number is already stored, the weight window will show the stored
unit weight. If it is a new one, the weight window will show the unit weight
which get in normal weighing mode. The user can edit the unit weight by use
keypad.
© Adam Equipment Company 2019
Press [PLU] again to store the unit weight and enter SKU ID edit. User can use
number and letter keypad to input. SKU ID can include letters, please check
the step 6.5.2 for details on how to input. After inputting ID, press [PLU] to
store. When finished, press [CE] to exit.
6.5.2 INPUT LETTER Press [.ABC] once, there will have an arrow sign appear on
the display point to “ABC”, or show an “ABC” sign on the display. This means
enter to letter input mode. Now, the user can input a letter by using the
numeric keypad. Press [2abc] once, the display will now show “A” and blink,
press once again, the display will show “B”, and so on. After the letter stops
flashing, it means this letter has been inputted. To return to number input
mode, press [.ABC], the letter input sign will also disappear.
6.5.3 SETTING BY PC (RECOMMEND METHOD) Connect the scale with PC by RS232,
open ScaleSetPlu.exe, and choose the right port.
SINGLE PLU SETTING Press , software will show Input the serial number, unit
weight, SKU ID, name as follow picture. if need scale increase the serial
number automatically, please choose “Auto inc”.
© Adam Equipment Company 2019
Choose the right com port, Click
, if this succeeds, this will show “Success”.
The user can then set the next PLU. DOWNLOAD MULTIPLE PLU BY TABLE (RECOMMEND)
Use table edit software (such as Excel, WPS, Numbers) to new built a file,
then input PLU data by follow format:
Then save as xls file.
1. In ScaleSetPlu.exe, click “Open”
© Adam Equipment Company 2019
2. Choose the right com port, Click “download file”
if succeed, will show “Success”.
7.0 USER PARAMETERS
To access the user parameters menu, press the [PLU] key and the [CE/
BACKLIGHT] key at the same time to enter the user parameter setting, the
display will show the first parameter “P1 TRA”, this is RS232 communication
setting.
In this menu. Pressing the [PRINT] key will enter the next parameter. Pressing
the [PLU] key will enter the sub parameter setting. Alternatively, the user
can press the number key that links to a corresponding sub parameter setting.
For example, in “T1 FC”, press [3], the display will jump to “T3 bps”. Press
the [M+] key use to choose the right parameter options.
7.1 RS 232 PARAMETERS The RS232 parameters are divided into two categories: P1 TRA and P2 FUN.
P1 TRA Communication settings
Parameter
Definition
Options
T1 FC
Communication mode: Continuous transfer 1-6 mode, stable transfer mode and response.
T2 Add
Communication address, effective in multi- 1-20 machine communication.
T3 Bps
Communication baud rate
2400, 4800, 9600, 19200, 38400, 56000, 115200
T4 Fmt
This parameter can be set to: n81 check bit/o71 N81
data bit /e71 stop bit.
O71
E71
T5 PFM
Label format. Set to 0, can print out the format 0-19 which named FORMCOIN LP-50. For LP-50, please choose 0 or 1.
© Adam Equipment Company 2019
T6 Chk
Turn XOR check on or off.
T7 FC2
Communication mode:
1-6
T8 PF2
Print functions
1-9
P2 FUN Other function settings
Parameter
Definition
Options
F1 BL F1C SL
For setting the backlight
ON OFF Auto
Setting the seconds.
sleep
time
of
the
backlight
in
1-45 seconds
F2 ARG F3 CST F4 MAE F5 PTE
For setting the average unit weight function automatically.
Tare and zero stable waiting time.
Turn ON/OFF next add up when not returning to zero after the previous add up.
Set pre-taring ON/OFF when pre-inputting the tare value.
ON OFF
ON OFF
ON OFF
ON OFF
F6 UNT
Selecting the Weighing Unit
Kg, lb, oz, g
F7 FIL F8 ZPE
Filter speed settings To set if the scale can alarm when weight is 0.
ON OFF
ON OFF
F9 FAL
On means full alarm mode. OFF means simple alarm mode.
ON OFF
F10 BC
To set if the Hi-Lo relay runs after stable.
ON OFF
Press `’ID” to escape and return to normal weighing mode.
7.2 SYSTEM PARAMETERS
To access system parameters, turn the indicator off, then press and hold down
the CE/Backlight & PLU buttons, and turn on the indicator using the on/off
rocker switch. The indicator will display CA:C3:-3 —— Using the keypad key in
123456 & then press the PRINT key, repeated pressing of the PRINT key will
cycle through the available options listed below:
© Adam Equipment Company 2019
Main- setting
Sub- setting
setting
CODE
Input new password. If user does not want to modify password, press [PRINT] to enter the next parameter.
F1 FIL
Filtering. Pressing the M+ key increases the intensity of the filter, the higher the filter value the more stable the weight reading is.
P1 FIL { Filter }
F2 ZEO
Zero track parameter. If the zero point often moves, user can set this parameter. Pressing the M+ key increases the zero tracking filtering, however increasing this value may make it difficult to weigh small items.
F3 ZTR
Creep tracking parameter. If the connected load cell suffers from creep, user can set the parameter. Pressing the M+ key increases the creep tracking filtering, however increasing this value may make it difficult to weigh small items.
P2 C-D
{Capacity and division}
The unit weight window : CXXXX
The total quantity window : DXXXX
dp xxx
XXXX is the capacity. If weigh is more than capacity+9 d, display will shows
over load
Division value.
Open or close the dual division.
CU xx
The unit of capacity and calibration.
P3 LIN X
{ Multi-range calibration }
User can set multi-range calibration on/off here. NoteIf user does not want to use the multi-range calibration, please set X off. If choose on, please remove all the weights from the pan before doing multi-range calibration.
LINEX
The state of multi-range calibration, X can be 1, 2, 3, it means setting step.
Input the weights value, the put the right weights to calibrate.
Remark: In the second step, value of weights must bigger than the first step,
and in the third step, value of weights must bigger than the second step.
P4 CAL
The total quantity window : WEIXXXX
The unit weight window : LOAD The total quantity window : XXXXX
Use number key to input the weight value, XXXX is the value of the weights for
calibration. Press [PLU] to enter the calibration. The display will show the
A/D value, with no weight on the top pan press the PLU key when the stable
indicator is showing
In this step, put on the correct weight, after stable indicator appear, press
[PLU] can complete calibration.
© Adam Equipment Company 2019
P5 EC
Calibration parameter record. User can review or use it for setting. When calibration parameters are lost, user can use this record to calibrate without weight. Note we would recommend using P4 Cal setting and doing a fresh calibration as this is the best method
P1 Rul
Calibration parameter the number is in the unit
weight window, this value affects the weighing accuracy directly. Please note
down this parameter after calibration. When this parameter lost, user can
input the value by number key which been noted to calibration without weight.
Press [M+] key, the digit of total weight window will keep blinking. Then
input the value and press [PLU] to store and enter to another parameter.
P2 Ld0
This is the calibration parameter is for multi-range calibration.If user did
not do multi-range calibration;
do not need note this parameter. But need set P3 to OFF.how to use and set
this parameter, it is the
same as standard calibration.
P2 Ld1
The same as Ld0.
P2 Lr0
The same as Ld0.
P2 Lr1
The same as Ld0.
© Adam Equipment Company 2019
8.0 CALIBRATION
8.1 SETTING CAPACITY AND DIVISION To set up the capacity and division follow
the procedure below:
With the indicator turned off, press and hold down the [PLU] and
[CE/Backlight] keys and then turn the scale on using the rocker switch at the
right back hand side of the indicator. The display will show:
Enter 123456 which is the default passcode and you should see the following
screen:
Press the [Print/Save] button, this will take you into the system parameters
(P1 FIL is to do with filtering) , press the [Print/Save] button again, and
you should see the system parameter for settings capacity & divisions (P2
C-d), screen should look similar to as follows:
© Adam Equipment Company 2019
In the above example you can see the capacity is set for 20kg x 1g. To
increase the capacity by 10 press the [M+] key; the capacity reaches a maximum
by continuing pressing the [M+] key and then returns to the lowest point.
Pressing the [Zero/0] key increases the flashing capacity value to various
preset values, pressing the [Tare] key decreases the flashing capacity value.
Once you are happy with the capacity value being correct, press the PLU
key.The indicator has fixed divisions that can be set, but these are all
normal multiples.To change the flashing division value press the [M+] key,
repeated pressing of the M+ key cycles the values round again. Once you are
happy with the division, press the PLU key, you will then see the following
screen:
The flashing dp, refers to dual division; leave this set to off and press the
PLU button to move to setting calibration unit which is shown as CU kg this
is the weighing unit for calibration. To change the weighing unit form kg to
lb press the [M+] key.
To move to the next parameter group setting press [Print/Save] key. The
parameter shopwn on the display is P3 LIN (muli range calibration) leave
this set to OFF; if it is set to ON pressing the [M+] key toggles the setting.
Press the [Print/Save] key to go to calibration -P4 CAL
© Adam Equipment Company 2019
8.2 CALIBRATION The calibration display should look as follows:
To calibrate the base, make sure the load cell cable is plugged into the back
of the indicator, where it is labelled up load cell, the scale base is level,
and there is no weight on the top pan.
Enter the weight you wish to calibrate the scale using the Green keys; you can
use the abc button to add a decimal point to the calibration value. Pressing
the [CE/backlight] button clears any entered value. Press the PLU key to show
the zero A/D counts; when the stable indicator is showing press the PLU key to
accept the zero point.
The screen should look similar as follows:
Load the weight you wish to calibrate with on the top pan, and press the PLU
key. You should see Pass on the display, and then the indicator will return to
normal weighing.
NOTE: BCT scales will have the lb or kg indicator on, to show the unit of the
weight requested. If the scale was in pounds before starting the calibration,
the weights requested will be in pound values or if the scale was weighing in
kilograms then metric weights will be requested.
© Adam Equipment Company 2019
9.0 LABEL PRINTING FUNCTION
In normal weighing mode, press [Print/Save] to print through built in label
printer. Printing can also be done through RS 232 connectivity.
Labels can also be printed via the built-in printer when in the accumulation
recall function and in parts counting.
9.1 LABEL EDITING COMPUTER SOFTWARE The BCT is supplied with RS-232 interface
which allows for connection between computer and indicator to edit labels for
the inbuilt label printer and set printing parameters.
© Adam Equipment Company 2019
9.1.1 LABEL EDITOR SOFTWARE INTERFACE
9.1.2 OPERATION METHOD (1) Connect the scale to the PC using the RS 232 cable
provided, this is so the
customised label can be downloaded straight to the indicator after completion
(2) Start up the label editing software
© Adam Equipment Company 2019
(3) Press the barcode icon in the top left-hand corner, the following menu will be displayed:
From here select to start a new label design, open a previously saved template or save a template to your PC.
(4) Set the parameters.
Printer: Chose the type of printing method (Default: built in printer) Width: Length of label (unit is dot, 8 dot = 1mm, 400 dot = 50mm) Height: 400 dot = 50mm Print speed: The default value is 4 Direction: Print direction, 0 is positive, 1 is negative Prints: How many labels will be printed when the print button is pressed For madd: For the accumulative function, will print when adding up.
To change the parameters, left click the desired parameter and a drop-down arrow will appear with different options to select from.
9.1.3 LABEL CUSTOMISATION
Once parameters have been set, the label can be customised and edited to the
desired preference.
© Adam Equipment Company 2019
9.1.4 ADDING TEXT
Various data can be added onto the label in which the text can be customised.
Left click the TEXT icon and the drop down will appear
Left click on any variable to add to the label. For example,
net weight
Net weight will appear on the label as:
The added information can be moved and place anywhere on the label by clicking and holding down.
The text can also be further customised in the parameter section on the left- hand side
Taking the net weight example:
To add text, click the `’TEXT” parameter on the left hand
side of the display.
Add the desired text into the box.
© Adam Equipment Company 2019
9.1.5 ADDING PHOTOS
Pictures can be added to a label design, the pictures will show as a black and
white picture. To add a picture, left click the picture icon in the editing
tools section.
The picture will appear as a small icon before uploading the image file via
the parameter settings section
X: 8 = 1mm, 500=50 mm Y: 8 = 1mm, 400 = 50mm Rename: Rename picture, if there
are pictures with the same name from previous labels they may be replaced
File: left click on the 3 dots that appear to select the picture to put on the
label Editable: whether the part of the template editing mode can be edited
Text images can also be added to the label design, follow the same procedure
as above by pressing the Text picture’ icon in the editing tools section and add text in the
TEXT’ section.
9.1.6 ADDING BARCODES Bar codes can be added to the label which are capable of
holding various weighing data.
Left click the arrow next to the barcode icon and select the information for
the barcode to hold, the barcode will then be added to the label.
© Adam Equipment Company 2019
The barcode also has customisable parameters:
Height: The height of the barcode Thickness: Minimum line width Barcode type:
Type of barcode (food retailers usually use EAN13) Readable: Enable or Disable
the Print out of the text below the barcode Rotation: Rotation angle (0 =
clockwise) Text: Can insert this variable (EAN13 barcodes must be numeric and
12 digits long. SKU must be 5 digits long, for price no dot needs to be
chosen)
NOTE: if the text for the barcode is entered incorrectly then an error message will
appear indicating a wrong
entry.
9.1.7 ADDING QR CODES
Left click the QR code icon located in the editing tools section and click on
the label area to place the QR code picture. The parameters for the QR code
are as follows:
X: 8=1mm, 400=50mm Y: 8=1mm, 400=50mm Thickness: The size of the QR code will
affect the quality of the QR code. The bigger in size, the easier to scan
Error Correction Level: The bigger the percentage, the easier it is to recover
lost print data Text: The text entered can be a web link or common words so
when the QR code is scanned with a smart device it will take the user to the
address entered in this parameter.
© Adam Equipment Company 2019
9.1.8 DELETING ITEMS Select the item you want to delete by left clicking it
and press the delete button on
the keypad.
A message will appear asking the user to confirm the deletion. Press OK’ to confirm and the item will then be deleted from the label. Alternatively, right click the item and click
delete’. The same message of
confirmation will appear, click `OK’ to confirm.
9.2 LABEL DOWNLOAD PROCEDURE VIA RS 232 CABLE
Save your final label design in preparation for download via RS 232 to the BCT
by pressing the save icon located in the top left hand corner of the software
Export the image by selecting the export icon from the barcode icon in the top
lefthand corner
Press the download icon Select the necessary serial port and baud rate and
click “download”
© Adam Equipment Company 2019
10.0 RS-232 INTERFACE
BCT scales are supplied with a RS-232 bi-directional interface. The scale when
connected to a printer or computer through the RS-232 interface, outputs the
weight, unit weight and count. Specifications:
RS-232 output of weighing data ASCII code Adjustable Baud rate, 600, 1200,
2400, 4800, 9600 and 19200 baud 8 data bits No Parity
Connector:
9 pin D-subminiature socket Pin 3 Output Pin 2 Input Pin 5 Signal Ground
The scale can be set to print text in English, French, German or Spanish. See
the RS-232 parameters section for details. All lines end with line feed
© Adam Equipment Company 2019
Data Format-Normal Output:
ID
QTY
0 pcs
U. W 123 g
Kg or g for metric weights and lb for pounds.
G. W 1.234 Kg Net Wt. if net weight is displayed
Item:
[barcode]
Includes 2 line feeds with carriage return
at beginning and end of the form
DATE 12/09/2006
TIME 14:56
Data Format- Memory Recall Print:
——————
TOTAL
No.
5
Wt. 1.234 Kg
Pcs
10 pcs
——————
DATE 12/09/2006
Includes 1 line feed 2 line feeds, carriage return TIME 14:56
© Adam Equipment Company 2019
In other languages the format is the same but the text will be in the language selected.
Description
ENGLISH
FRENCH
GERMAN
SPANISH
Item barcode
Item
Article
Artikel
ít
Assigned
product/ ID
transaction ID
Product name/ item Name label
Print gross weight
G. W
Net weight
N. W
Weight per unit counted U. W
QTY / Number of items Pcs counted
Number of weighings No. added to subtotals
Total weight and count Total printed
Print date
Date
Print time
Time
–
Nom
Pds Brut Pds Net Pds unit Pcs
Nb.
Total
Date Heure
–
Name
Brut-Gew Net-Gew Gew/Einh Stck.
Anzhl
Gesamt
Datum Zeit
–
Nombre
Pso Brut Pso Net Pso/Unid Piezas
Num.
Total
Fecha Hora
10.1 INPUT COMMANDS FORMAT
The scale can be controlled with the following commands. The commands must be sent in upper case letters, i.e. “T” not “t”. Press the Enter key of the PC after each command.
T
Tares the scale to display the net weight. This is the same as pressing [Tare] key.
Z
Sets the zero point for all subsequent weighing. The display shows zero. Same as pressing the [Zero 0] key.
P
Prints the weight, unit weight and totals same as pressing the [PRINT/ SAVE] key.
© Adam Equipment Company 2019
10.2 REAL TIME CLOCK SETUP
The Real Time Clock (RTC) is used only for the RS-232 output. The Date and
Time can be set as required. The scale will keep the clock running even when
the power is off.
In weighing mode, press and hold [PLU] key, then press [CE] once to enter
setting mode. The display will show “P1 TrA”, press [PRINT] key once, the
display will show “P2 FUN”.
Press [PLU] key again and again to the unit weight display shows “F9 FAL”. At
this time, press and hold the [PLU] key, press [PRINT] once, then enter into
time setting. The weight window will show “Date”, use number keys to input the
data information. For example, 11/10/2013, input “130111”. Then press [PLU]
key to enter time setting. The weight window shows “Time”, use the numeric
keys to input the time information. For example, 10:10, input “1010”. Then
press [PLU] to complete the settings.
Press `’ID” to escape and return to normal weighing mode.
10.3 AUTO SLEEP FUNCTION
To set the auto sleep function you will need to enter the Other function’ parameter setting in the User parameters (see section 7). Press the [PLU] key and the [CE/ BACKLIGHT] key at the same time to enter the user parameter setting, the display will show the first parameter “P1 TRA” Press the [PRINT/SAVE] key to move to the other function parameters and press the numeric button [2] on the keypad to access the SL parameter. Press the [M+] key to increase the sleep time in seconds and press the [Q’TY/PST] key to confirm. Press
’ID” to escape and return to normal weighing mode.
11.0 BATTERY AND BACKLIGHT OPERATION
11.1 BATTERY
The scales can be operated from the battery, if desired. The battery life is
approximately 70 hours.
When the battery needs charging the battery symbol will appear .
The battery
should be charged as soon as possible. Once the “LO bAT” message is shown the scale
will still operate for about 10 minutes after which it will automatically switch off to protect
the battery.
© Adam Equipment Company 2019
11.2 CHARGING
When charging, the indicator charge light will appear red. It will turn green
when charging finished (the charging time should be 8-10h).
Attention: do not use up the battery, otherwise the battery may break. If the
indicator is not in use for a long time, the battery should recharge in 40
days.
After charge, please take off the power supply. We recommend that you do not
always connect to the power supply.
11.3 BACKLIGHT FOR LCD The backlight of the LCD can be set to be”1″: ON at all the time, “2”: ON only when a weight is placed on the scale or “3”: Turned off. To set the backlight press and hold [CE/BACKLIGHT] key for 4 seconds. The weight display will show “ON/ OFF/ AUTO” Press and hold the [CE/BACKLIGHT] key to alternate between these options:
“BL on” “BL Au” “BL OFF”
Sets the backlight to be on at all times.
Sets the backlight to operate automatically when a weight is placed on the
scale or a key is pressed.
Sets the backlight to be off.
© Adam Equipment Company 2019
12.0 ERROR CODES
During the initial power-on testing or during operation, the scale may show an
error message. The meaning of the error messages is described below.
If an error message is shown, repeat the step that caused the message, turning
the balance on, carry out the calibration or other functions. If the error
message is still shown contact your dealer for further support.
ERROR CODE DESCRIPTION
Solution
ADO—
Over A/D resolution Remove the weights of the pan or send
range
back to the agent.
ADL–OVR–ERR-Z
Battery flash
symbol
Over A/D resolution range
Over load (max: capacity +9e) AD value exceed the pow er on zero range when po
wer on
Low power alarm
Remove the weights of the pan or send back to the agent. Check the weight of
the weights, cannot exceed Max Cap.+9e Power on again after remove the weights
or re-calibrate
Charge battery
Lo-Bat
Low power alarm
Charge battery
© Adam Equipment Company 2019
13.0 REPLACEMENT PARTS AND ACCESSORIES
If you need to order any spare parts and accessories, contact your supplier or
Adam Equipment. A partial list of such items is as follows-
Main Power cord Power supply Replacement Battery Printer
14.0 SERVICE INFORMATION
This manual covers the details of operation. If you have a problem with the
scale that is not directly addressed by this manual then contact your supplier
for assistance. In order to provide further assistance, the supplier will need
the following information which should be kept ready:
A. Details of your company
-Name of your company:
-Contact person’s name: -Contact telephone, e-mail,
fax or any other methods:
B. Details of the unit purchased
(This part of information should always be available for any future
correspondence.
We suggest you to fill in this form as soon as the unit is received and keep a
print-out in
your record for ready reference.)
Model name of the scale: Serial number of the unit:
BCT _____
Software revision number
(Displayed when power is first turned on):
Date of Purchase:
Name of the supplier and place:
C. Brief description of the problem
Include any recent history of the unit. For example:
-Has it been working since it’s delivered -Has it been in contact with water
-Damaged from a fire
-Electrical Storms in the area -Dropped on the floor, etc. =
© Adam Equipment Company 2019
WARRANTY INFORMATION
Adam Equipment offers Limited Warranty (Parts and Labour) for any components
that fail due to defects in materials or workmanship. Warranty starts from the
date of delivery.
During the warranty period, should any repairs be necessary, the purchaser
must inform its supplier or Adam Equipment Company. The company or its
authorised Technician reserves the right to repair or replace the components
at any of its workshops at no additional cost, depending on the severity of
the problems. However, any freight involved in sending the faulty units or
parts to the Service Centre should be borne by the purchaser.
The warranty will cease to operate if the equipment is not returned in the
original packaging and with correct documentation for a claim to be processed.
All claims are at the sole discretion of Adam Equipment.
This warranty does not cover equipment where defects or poor performance is
due to misuse, accidental damage, exposure to radioactive or corrosive
materials, negligence, faulty installation, unauthorised modifications or
attempted repair, or failure to observe the requirements and recommendations
as given in this User Manual.
This product may include a rechargeable battery that is designed to be removed
and replaced by the user. Adam Equipment warrants that it will provide a
replacement battery if the battery manifests a defect in materials or
workmanship during the initial period of use of the product in which the
battery is installed.
As with all batteries, the maximum capacity of any battery included in the
product will decrease with time or use, and battery cycle life will vary
depending on product model, configuration, features, use, and power management
settings. A decrease in maximum battery capacity or battery cycle life is not
a defect in materials or workmanship, and is not covered by this Limited
Warranty.
Repairs carried out under the warranty do not extend the warranty period.
Components removed during warranty repairs become company property.
The statutory rights of the purchaser are not affected by this warranty. The
terms of this warranty is governed by the UK law. For complete details on
Warranty Information, see the terms and conditions of sale available on our
web-site.
© Adam Equipment Company 2019
WEEE 2012/19/EU
TEaCêDDDnhUteiiiressdit,eppsspaeroodeepesssefsiiprfGvttteaiiritvveccihrcootreeeuätiinnimlétorooennnasneisppycc.oupehpnnceuteoifdòmfutoiectebritmsseêresdetéedrreqmeridmusepeeiérsosHnlemiestmamceaauhidelnustanméxiodttoflsüoao.ilvnnijlDsueeedcnieinsottrptloimesrofeosicsuesroadtstglintrétidio.cccloofhtwimseboatranessetsststmteeliiod.crécuiiT.enaohsalsiegs(sdeia.forlfssmi.ottéLea’sépdtlip)icmlomiesinusastttoicocononudfnoetrrmliaesbtoaotultotescriaideleldatwohiest
FCC / IC CLASS A DIGITAL DEVICE EMC VERIFICATION STATEMENT
NOTE: This equipment has been tested and found to comply with the limits for a
Class A digital device, pursuant to Part 15 of the FCC rules and Canadian
ICES-003/NMB-003 regulation. These limits are designed to provide reasonable
protection against harmful interference when the equipment is operated in a
commercial environment. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio communications.
Operation of this equipment in a residential area is likely to cause harmful
interference in which case the user will be required to correct the
interference at his own expense.
CALIFORNIA PROPOSITION 65 – MANDATORY STATEMENT WARNING: This product includes
a sealed lead-acid battery which contains chemicals known to the State of
California to cause cancer and birth defects or other reproductive harm.
Adam Equipment products have been tested with, and are always supplied with
mains power adaptors which meet all legal requirements for the intended
country or region of operation, including electrical safety, interference and
energy efficiency. As we often update adaptor products to meet changing
legislation it is not possible to refer to the exact model in this manual.
Please contact us if you need specifications or safety information for your
particular item. Do not attempt to connect or use an adaptor not supplied by
us.
© Adam Equipment Company 2019
ADAM EQUIPMENT is an ISO 9001:2015 certified global company with more than 40
years’ experience in the production and sale of electronic weighing equipment.
Adam products are predominantly designed for the Laboratory, Educational,
Health and Fitness, Retail and Industrial Segments. The product range can be
described as follows:
-Analytical and Precision Laboratory Balances -Compact and Portable Balances -High Capacity Balances -Moisture analysers / balances -Mechanical Scales -Counting Scales -Digital Weighing/Check-weighing Scales -High performance Platform Scales -Crane scales -Mechanical and Digital Electronic Health and Fitness Scales -Retail Scales for Price computing
For a complete listing of all Adam products visit our website at
www.adamequipment.com
Adam Equipment Co. Ltd. Maidstone Road, Kingston
Milton Keynes MK10 0BD UK
Phone:+44 (0)1908 274545 Fax: +44 (0)1908 641339
e-mail: sales@adamequipment.co.uk
Adam Equipment Inc. 1, Fox Hollow Rd. 06478
USA Phone: +1 203 790 4774 Fax: +1 203 792 3406
e-mail: sales@adamequipment.com
AE Adam GmbH. Instenkamp 4 D-24242 Felde
Germany Phone +49 (0)4340 40300 0 Fax: +49 (0)4340 40300 20
e-mail: sales@adamequipment.co.za
Adam Equipment S.A. (Pty) Ltd. 7 Megawatt Road, Spartan EXT 22 Kempton Park,
Johannesburg,
Republic of South Africa
Phone +27 (0)11 974 9745 Fax: +27 (0)11 392 2587
e-mail: sales@adamequipment.co.za
Adam Equipment (S.E. ASIA) PTY Ltd
70 Miguel Road Bibra Lake Perth WA 6163 Australia
Phone: +61 (0) 8 6461 6236 Fax +61 (0) 8 9456 4462 e-mail:
sales@adamequipment.com.au
Adam Equipment (Wuhan) Co. Ltd. A Building East Jianhua Private Industrial
Park Zhuanyang Avenue
Wuhan Economic & Technological Development Zone 430056 Wuhan P.R.China
Phone: + 86 (27) 59420391 Fax + 86 (27) 59420388 e-mail:
info@adamequipment.com.cn
© Copyright by Adam Equipment Co. All rights reserved. No part of this
publication may be reprinted or translated in any form or by any means without
the prior permission of Adam Equipment. Adam Equipment reserves the right to
make changes to the technology, features, specifications and design of the
equipment without notice. All information contained within this publication is
to the best of our knowledge timely, complete and accurate when issued.
However, we are not responsible for misinterpretations which may result from
the reading of this material.
The latest version of this publication can be found on our Website.