iVOOMi YT-EWQ-04 Non Contact Infrared Thermometer User Manual
- June 6, 2024
- iVOOMi
Table of Contents
iVOOMi YT-EWQ-04 Non Contact Infrared Thermometer User Manual
1. Product Performance, Main Structure Composition & Scope of
Application
Product Performance & Main Structure Composition:
Technical Parameters
Measurement Range | 3cm—Scm |
---|---|
Measuring Range | The Human Body Temperature |
(89.6° F— 109.2°F)
The Object Temperature| 0°C-100°C (32° F— 212°F)
Accuracy of Measurement| The Human Body Temperature| (35-42)°C within (±0.2°C
—0.4° F)
(35-42)°C Outside (±0.3°C —0.54° F)
The Object Temperature| ±1.0°C/1.8° F
The Operating Environment| 16°C-35°C (60.8° F— 95°F RH 5 85% (Noncodensing)
Storage Environment| 10°C 55°C (14° F— 131°F) RH 5 85% (Noncodensing)
Power supply| DC 1.5V*2 (2AAA Batteries)
Alert| Low Battery Indicator
Memory| Auto Store Measurement
Automatic Shutdown| Automatic shutdown without any operation within 30 seconds
Product Size| 149 mm x 77 mm x 43 mm
The Product Weight| 100g± 2g
The Infrared Thermometer measures the body temperature by collecting infrared thermal radiation from the forehead of the humanbody. Its operation is simple and hygienic, and the measurement is fast and accurate. Theuser only needs to point the probe head at the forehead and press the measurement button,and the body temperature can be quickly and accurately measured in 1 second, which canbe used by the family and medical departments to measure the body temperature.This product is a type 2 medical device, which belongs to the internal power supplyequipment, type B application part, and the protection class is ordinary equipment (IP20). Donot use it in the presence of flammable anesthetic gas and air, nitrous oxide gas, for continuous operation of the equipment. The EU classification is Ilb.
Scope of Application
Used by families and medical departments to measure Body/Surface temperature.
Product Features
- Top quality infrared sensor ensures very little variations.
- Great Environment adaptability. Works in most environments.
- New control probe structure ensures more accurate measurement.
- Conveniently switch between human and object modes with just one button press.
- High Temperature alarm with red backlight color.
- Automatically saves 32 temperature measures.
- Choice of two temperature measuring units – Celsius and Fahrenheit .
- Automatic shutdown for energy saving.
Contraindications, Precautions, Warnings and Contents of Reminders
Matters Needing Attention
- Please read this manual carefully before using and make sure the battery is installed.
- Do not immerse the forehead of the thermometer in any liquid.
- Do not use it in high or low temperature environment for a long time.
- Do not crash, drop or mix with sharp objects.
- Do not disassemble.
- Do not use in the environment of strong electromagnetic interference.
- Keep the thermometer out of reach from children.
- Thermometer Measurements are not a substitute for a physician’s diagnosis.
Warning and prompt contents
This product is only for detection and NOT for treatment
Instruction for Using the Product (Drawings)
Display the Icon Definition
Icon to Define | Icon | State description | Fault Code |
---|
Low Battery
Indicator| | Show| Replace battery immediately
Don’t show| Enough Power available for usage| Err1| Err2
Prompt| | Show| Prompt sound is on
Don’t show| Prompt sound is off| Operating
environment
temperature is
too low,
Measure
under
normal
operating
conditions| Operating
environment
temperature is
too high,
Measure
under
norms
operating
conditions
Surface
Selection| | Human Mode. Select when measuring human body. Very Important to
select the correct mode.
–
| Object Mode. Select when measuring object.
Temperature
Measuring Unit|
˚C
| Celsius
˚F
| Fahrenheit
Display| | Displays the temperature.
Memory
Symbol| M| Displays the Memory. Can store upto 16 measurements.
16 Groups of Memory Data Queries
When the thermometer is turned on, press the “+” or “2 button to display the first or last memory on the LCD Value.To toggle between the memory data, press ‘+’ to move up and ‘-‘ to move down. You can store up to 32 data points.
Function Key Description
-
Liquid Crystal Display (LCD)
-
Function setting key
The Function
Keys| Functional description
---|---
Set| Enter/exit the Settings menu and switch the Settings key
+| Move up the Selection
–| Move Down the selection
Setting Operation
Thermosafe comes with three settings: Surface selection, prompt tone on/off, and selection of measuring unit(°C and °F)
The comparison table of setting menu is as follows:
The Menu| Function| Set| +| —| The
Initial Value
---|---|---|---|---|---
Fl| Surface Selection| Press
for F2 Setting| Prompt| No
Prompt| Prompt
F3| Measuring Unit| Press
Again to Finish| °C| °F| °C
Setting of Forehead Temperature/Object Mode Conversion
At start-up, press and hold “SET” till the screen displays Ft. Press “+” or “-” button to toggle the surface between human and object. Choose appropriately.
Prompt Tone Switch Setting
At start-up, press and hold “SET” till the screen displays A on the screen. Press “SET” once to use F2 setting. Press the “+” or “2 button to switch the prompt tone on and off.
Celsius/ Fahrenheit Setting
At start-up, press and hold “SET” till the screen displays Fl on the screen, & press “SET” twice to use F3.Press the “+” or “2 button to toggle between Celsius/Fahrenheit.
SAFETY INSTRUCTIONS
Measurement of Human Body Temperature
- After loading the battery, press the trigger button for 2 seconds.
- © Make sure the measuring mode at the top of the screen is body mode.
- Aim the thermometer probe at the bare body- head or arms, whilst maintaining a distance of 3-5cm. Then press the trigger to record the temperature. The target’s temperature will display on the screen 1 second after pressing the trigger.
- The thermometer will automatically shutdown after 10 seconds if left idle.
- 32°C – 37.3°=> Green Light; 37.40C- 37.90C => Yellow Light: Temperature >38°C above, Red Light.
Temperature <32 °C => Red Light and shows “LO”.
Temperature > 42.9 CC => Red Light and shows “HI”.
Warm Prompt
- Keep the inner cavity of the sensor and probe clean before and after use.
- Place the thermometer in a stable temperature environment. When the ambient temperature varies greatly (e.g. from indoor to outdoor), Please leave it for about ten minutes and then measure it.
- Do not measure the temperature of an extremely high or low temperature object. Start taking your temperature immediately. Leave it for ten minutes.
- When there is a big difference in temperature of surface relative to the environment let thermometer adjust to the environment for 5 minutes and attempt again.
- For accurate readings, wait 30 minutes after exercising, showering or eating before measuring temperature
Object Temperature Measurement
- Toggle the measurement mode to object ****using the function key.
- Point the thermometer probe at the measuring object about 3-5cnn away. Then press the trigger to record the temperature. The target’s temperature will display on the screen 1 second after pressing the trigger.
Read User Manual Online (PDF format)