UEi TEST C163 Rseidential Commercial Combustion Analyzer User Guide
- June 6, 2024
- UEi TEST
Table of Contents
UEi TEST C163 Rseidential Commercial Combustion Analyzer
OVERVIEW
MENU SCREEN
- Press ”Power” Button for 2 seconds. The analyzer will perform a 60 second purge. This should be done in fresh air.
- Rotate Selector Dial to MENU to set up or customize your settings.
- Selected parameter are centered, highlighted with arrow icons on either side.
- Use and buttons to scroll menu options, Press to select and edit a parameter.
- Press and buttons to change contents of a field.
- Press button to scroll fields.
- Press to enter content and move to next field.
STATUS SCREEN
BEFORE TESTING CHECKLIST
- Clean particle filter
- Water trap and probe line are free of condensate
- All hose and thermocouple connections are properly secured
- Water trap is fitted correctly
- Power on and zero
- Flue gas probe is sampling ambient FRESH air
- Remove the rubber boot and red drain plug. Allow the water to drain out. Re-insert the rubber plug and replace boot cover.
- Remove protective rubber boot. Remove water trap from analyzer. Remove particle filter from water trap and replace with new one. Reattach trap and boot.
COMBUSTION ANALYSIS
- Insert flue probe into combustion input. Insert K-type temperature probe into input T1.
- Insert flue probe into stack. Adjust cone so the probe tip is approximately at the center of the stack.
- Rotate function rotary dial to O2/Eff for Test Screen 1 parameters.
NOTE: If there is no hole, check with service and installation instructions of appliance for best testing method or check with your local authority on best practice
- O2/Eff displays CO2, O2, Efficiency, Flue temperature, Inlet temperature.
- Rotate the function rotary dial to CO/CO2 for Test Screen 2 parameters.
- CO/CO2 displays CO (ppm), CO2, Losses calculated, Excess air %. Make any adjustments needed for proper Combustion and wait for analyzer to display change in readings. (Repeat as necessary).
CHANGE FUEL TYPE
- Rotate the dial to either STATUS, AUX, O2/Eff, CO/CO2 or Temp setting.
- Press button until the Fuel Type displays on the Status Bar. Press and Hold until two arrow icons display.
- Press and buttons to scroll fuel selections. Press to select fuel
AUXILIARY SCREEN
- When performing a combustion Test, Rotate function rotary dial to AUX to view the Auxiliary Test Screen’s custom param-eters.
- AUX displays 5 lines of editable parameters. To edit line values Press button until Edit? displays on the Status Bar. Press and Hold until the Status Bar reads LINE 1
- Press and buttons to scroll parameter selections. Press to select a parameter and advance to the next Line. Continue until cursor is back to Status Bar to exit.
The AUX screen allows you to assign any of the following parameters on any
display line.
O2, CO, Display Fuel Type, CAL, XAIR, LOSS, Efgc, ATM, Ti, T, T2, T1, CO2,
COa, or COn. Pressure(C163 and !64 only). NOX
and No (C164 only)
TEMPERATURE TEST
- Insert K-type temperature probes into input T1 and T2.
- Rotate function rotary dial to TEMP for Temperature Test Screen.
- TEMP displays T1, T2 and Differential Temperature. (Pressure C163 C164 only)
PRINTING & LOGGING
- To Print: Align printer (sold separately) above Analyzer.
- During testing, Press to startprinting. Press again to abort.
- To Log: During testing, Press and Hold to log a report. To view these Logs, see View Saved Logs section.
VIEW SAVED LOGS
- Rotate the dial to MENU. Press button until the LOGS is selected between two arrow icons. Press to select.
- Screen displays all saved LOGS. With VIEW selected in the Status Bar, Press to select
- Press and buttons to scroll Test Report categories. Press to select a category.
- Line 1 displays the LOG number on screen. To view another LOG, Press and buttons to NEXT. Press to advance to next log. Repeat until you reach desired LOG.
- To print the LOG, Press and buttons until PRINT is selected between two arrow icons. Press to select to start printing.
- To EXIT, Press and buttons until EXIT is selected between two arrow icons. Press to EXIT screen. Repeat until return to MENU screen.
POST TESTING
Remove the probe from the flue and allow analyzer to purge with fresh air until
readings return to zero. O2 to 20.9% and CO below 20 ppm
CAUTION: probe tip will be HOT.
POWERING OFF
When you power off the Analyzer, there is a 10 second purge.
CAUTION
Make sure you do not exceed the analyzer’s operating specifications.
- Do not exceed the flue probes maximum temperature (1112˚F)
- Do not exceed the analyzer’s internal temperature range 112°F (50°C)
- Do not put the analyzer on a hot surface
- Do not exceed the water trap’s level
- Do not let the particle filter become dirty and blocked
The analyzer is fitted with 3 (AA) size rechargeable batter-ies. Observe
correct polarity when replacing.
Always check operation immediately after battery replacement.
Do not leave batteries in unit for extended periods of non-use.
WARNING
Do not recharge with Alkaline batteries fitted. Alkaline batteries are not
rechargeable. Attempting to recharge alkaline batteries may result in product
damage and create a fire risk. Time and Date will need to be reset after
changing batteries. Always dispose batteries using approved disposal methods
that protect the environment.
WARRANTY
The C162/C163/C164 combustion analyzer is warranted for one year including sensors. If within the warranty period your instrument should become inoperative from such defects, the unit will be repaired or replaced at UEi’s option. This warranty covers normal use and does not cover damage which occurs in shipment or failure which results from alteration, tampering, accident, misuse, abuse, neglect or improper maintenance (calibration). Batteries and consequential damage resulting from failed batteries are not covered by warranty. Any implied warranties, including but not limited to implied warranties
of merchantability and fitness for a particular purpose, are limited to the
express warranty. UEi shall not be liable for loss of use of the instrument or
other incidental or consequential damages, expenses, or economic loss, or for
any claim or claims for such damage, expenses or economic loss. A purchase
receipt or other proof of original purchase date will be required before
warranty repairs will be rendered. Instruments out of warranty will be
repaired (when repairable) for a service charge. Warranty only covers hardware
and does not extend to software applications or accessories. Contact UEi for
specific warranty and service information.
This warranty gives you specific legal rights. You may also have other rights
which vary from state to state.
Complete care for your C160 series combustion analyzer.
When you:
Request Recertification or Service Online Within 1 Year of Purchase or
Service
UEi will:
- 6-Year Warranty: All UEi combustion analyzers have a standard
- 1-year warranty. Each recertification includes a 30-point inspection and extends the warranty for 1 more year for up to 6 years from the date of warranty.
- 24-hour Service: Analyzers received for standard recertification and service through UEi Service+ are shipped back on the next business day.
- Free Shipping: UEi Service+ offers free shipping both to and from our service center. When customers book their recertification, they receive a prepaid UPS Ground shipping label.
Register Online
Registering you analyzer online is quick and easy. Just log in or setup an
account, it only takes a couple of minutes. Once logged in you can register
you analyzer by providing some product information and uploading a proof-of-
purchase. When it’s time to request recertifcation, just log into your
account, select the analyzer, select the service and place your order.