ecom-DP Digital Pressure Measurement Instruction Manual
- June 6, 2024
- ECOM
Table of Contents
- ecom-DP Digital Pressure Measurement
- Design ecom-DP
- Initialization
- Numerical Indication
- Check Programs
- Clock Set
- Internal Instrument Settings
- Sensor Setup
- View Service Information
- Data Logger Program “ecom-DP Logger”
- Technical Data ecom-DP
- Maintenance Tip
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
ecom-DP Digital Pressure Measurement
Design ecom-DP
Initialization
Power Supply
The ecom-DP is powered by 3 batteries or 3 accumulators. The following types can be used to operate the instrument:
- Battery: 1.5 V AA (Mignon)
- Accumulator: 1,2 V AA NiCd (nickel cadmium)
- Accumulator: 1,2 V AA NiMh (metal hydride)
If you are using a loading unit, please adjust the function “Bat. Loading”
Open the battery compartment and insert the batteries. Be careful to ensure
proper polarity. If using the re-chargeable battery type, it will be necessary
to fully charge the batteries prior to initial use (approx. 4 hours). The
voltage of the batteries is monitored by the DP and will indicate a low
battery condition by approx. 3.3 V by beeping and flashing a signal on the LCD
display. If the voltage sinks (approx. 3 V) so the ecom-DP will automatically
switch off.
Accumulator Charging
The accumulator charging procedure is monitored by the ecom-DP. Once connected to AC power, the instrument switches on. During the charging procedure, a BATTERY indicator will be shown on the LCD display. Once the battery is fully charged, the BATTERY indica-tor on the display will blink. It is possible to operate the ecom-DP on AC power during the charging procedure by pressing the
Setting Clock
The internal clock and the internal RAM memory of the ecom-DP are buffered via a capacitor. If the instrument is without current supply during more than approx. 10 minutes, the recorded information gets lost. Before using the ecom- DP for the first time, adjust the internal clock and delete the internal RAM memory.
Press
-
Clock set
-
Delete RAM
-
Press OK to continue!
-
Press twice the
key and the measurement value menu will be displayed. -
Press
(wrench symbol) to call up the menu “In-strument settings”. -
With the cursor keys <UP/DOWN> scroll until the menu point “Set clock”.
-
Press
to activate.
Date and time can be set as follows:
- Select with the cursor keys <RIGHT/LEFT> the position to be changed (corresponding field blinks).
- Adjust the correct numbers with the cursor keys <UP/DOWN>.
- Press
to confirm correct date and time settings.
Numerical Indication
If “Pressure Auto -> O by start” is set on “YES”, so the internal pressure
sensors should be flown with pressure first after the switch-on of the
instrument, as they are zeroed during said switch-on procedure.
After switch-on, the specifications of the pressure sensor will be displayed:
denomination, measurement range and connection seizure (+/-). Press
- Before making measurements, the instrument must have adapted the ambient temperature!
- Make sure, that the ambient temperature remains constant during measurement!
Zeroing the Internal Pressure Sensor
If the instrument is attached to a metal surface with the magnetic holders,
the zero point may shift.
Please reset the zero point before measurement.
As the difference pressure sensor can drift because of its sensitivity in the zero point (also by attaching to met-al surfaces by magnetic holder), it is possible to reset the zero point.
- Press
(DP->0) to activate the function for zero point correction of the pressure sensors. - Press
(DP1) to zero the value of the pressure sensor 1. - Press
(DP2) to zero the value of the pressure sensor 2. - Press
(ESC) to quit the menu.
Different functions can be called up with the function key
Freezing of Measurement Value
- Press
(Hold) to “freeze” that measurement value displayed at the time the key is pressed (active function = Hold indication blinks). - Press again
(Hold) to quit the “freezing” and the current value will be displayed.
Minimal / Maximal Measurement Values
With
- Press one time
(Min/Max) to view the lowest determined values (Min/Max blinks). - Press
(CLR) to delete the minimal values (new scanning begins). - Press again
(Min) to view the highest values (Min/Max blinks). - Press
(CLR) to delete the highest values (new scanning begins). - Press one more time
(Max) to quit the menu.
Measurement Values Printout
Press
Condition hereto is a trouble free connection (direct sight) between the
sending diode at the front of the instrument and the receiving diode of the
printer. The distance between both instruments should not exceed 40 cm (1.3
foot), otherwise a trouble free transmission is not warranted at any time.
Contrast Adjustment and Lighting
Basing on the previous function seizure, the functions “LUMI” and “Light” are
displayed after pressing
Check Programs
- Use only air or inert gas for check !
- Respect the “Technical Rules for Gas installations!”
Call up the pre-programmed measurement routines “Preliminary Check”, “Main
Check” and “Usability Check” in the menu
Parameters to each measurement routine can be adjusted in the menu “Setup”.
Setup
Scroll with <UP/DOWN> until the menu “Setup”. Press
Pressure Check
- Stabilization time (1 – 10 min, default: 1 min)
- Measurement time (10 – 120 min, default: 5 min)
Preliminary Check
- Stabilization time (1 – 10 min, default: 1 min)
- Measurement time (10 – 120 min, default: 10 min)
- Test pressure (900 – 1200 hPa, default: 1000 hPa)
Main Check
- Stabilization time (1 – 10 min, default: 1 min)
- Measurement time (10 – 120 min, default: 10 min)
- Test pressure (90 – 160 hPa, default: 110 hPa)
Usability Check
- Stabilization time (1 – 270 min, default: 1 min)
- Measurement time (10 – 240 min, default: 10 min)
- Test pressure (20 – 160 hPa, default: 50 hPa)
- Max. operation pressure (10 – 100 hPa, default: 20 hPa)
- Air pressure (800 – 1200 hPa, default: 1013 hPa)
Each parameter can be adjusted the same way:
-
Select the parameter with <UP/DOWN>.
-
Press
to activate. -
Use the cursor keys <RIGHT/LEFT> to select the position to be changed (corresponding field blinks).
-
Adjust the correct figure using <UP/DOWN>.
-
Press
to confirm the corrected value.
Pressure Check
Proceed as follows:
-
Close the conduit with a suitable adapter (part of ecom-DPK).
-
Connect the components a described in ecom-DPK manual.
-
Scroll with <UP/DOWN> until the menu “Pressure check”. Activate with
-Taste. -
With the compressed-air pump create the pre-adjusted test pressure (the unit beeps as soon as the pressure level is achieved).
-
Interrupt the connection to the air pump (switch-off the ball valve).
-
Wait for stabilization time (if the pressure remains in the range “test pressure +/- 10% during stabilization time, so the measurement time will start).
-
Once the measurement time is over, the result is displayed and can be sent to the IR printer (option) thus pressing
(Print).
-
If the menu “Pressure Check” is selected again, so the result can be called up with
(as long as the ecom-DP is on) or a new measurement can be started with .
Preliminary Check
The preliminary check acc. to DVGW – TRGI Process Instructions G 600 by pipes (operation pressure up to 100 hPa) is registered as a measurement routine in the ecom-DP.
Proceed as follows:
-
Close the conduit with a suitable adapter (part of ecom-DPK).
-
Connect the components a described in ecom-DPK manual.
-
Scroll with <UP/DOWN> until the menu “Preliminary check”. Activate with
-Taste. -
With the compressed-air pump create the pre-adjusted test pressure (the unit beeps as soon as the pressure level is achieved).
-
Interrupt the connection to the air pump (switch-off the ball valve).
-
Wait for stabilization time (if the pressure remains in the range “test pressure +/- 10% during stabilization time, so the measurement time will start).
-
Once the measurement time is over, the result is displayed and can be sent to the IR printer (option) thus pressing
(Print). -
If the menu “Preliminary Check” is selected again, so the result can be called up with
(as long as the ecom-DP is on) or a new measurement can be started with .
Main Check
The main check acc. to DVGW – TRGI Process Instructions G 600 by conduits (operation pressure up to 100 hPa) is registered as a measurement routine in the ecom-DP. Proceed as follows:
-
Close the conduit with a suitable adapter (part of ecom-DPK).
-
Connect the components a described in ecom-DPK manual.
-
Scroll with <UP/DOWN> until the menu “Main Check”. Activate with
-Taste. -
With the compressed-air pump create the pre-adjusted test pressure (the unit beeps as soon as the pressure level is achieved).
-
Interrupt the connection to the air pump (switch-off the ball valve).
-
Wait for stabilization time (if the pressure remains in the range “test pressure +/- 10% during stabilization time, so the measurement time will start).
-
Once the measurement time is over, the result is displayed and can be sent to the IR printer (option) thus pressing
(Print) .
-
If the menu “Main Check” is selected again, so the result can be called up with
(as long as the ecom-DP is on) or a new measurement can be started with .
Usability Check
The usability check acc. to DVGW – TRGI Process form G 624 by conduits is
memorized as a measurement routine by the ecom-DP. The calculation of the leak
rate happens automatically according to the following equation and corresponds
herewith to the procedure of the DVGW-TRGI Process Form G 624:
*VB = V/TM ((PA + P1)/(PA + P2)-1) PB/P1 f**
with: VB = Gas leak volume in operation state (l/h)
V = Pipe content in litres
TM= Measurement duration in hours
PA = Barometer stand in hPa
P1 = Test pressure at meas. beginning in hPa
P2 = Test pressure at measurement end in hPa
PB = Maximal gas operation pressure in hPa
f = Factor for consideration of gas type
Proceed as follows:
-
Close the conduit with a suitable adapter (part of ecom-DPK).
-
Connect the components a described in ecom-DPK manual.
-
Scroll with <UP/DOWN> until the menu “Suitability/Leak rate”. Activate with
.
-
The pipe volume to be checked (V) is needed for the calculation of the leak rate. The ecom-DP offers two input possibilities:
-
Direct input of pipe volume after having pressed
(ADD) (by repeated input the value will be added).
-
The ecom-DP will calculate the pipe volume in sections basing on the diameter and the length for thread pipe acc. to DIN 2440 (middle heavy):
-
Press
(CAL). -
Use <UP/DOWN> to select the pipe diameter (1/2″ to 4″) and confirm with
. -
Select pipe length with the cursor keys and confirm with
. -
The volume for thread pipe acc. to DIN 2440 (medium heavy) will be calculated and added.
-
Repeat procedure until all pipes sections are registered.
-
-
Quit the volume input with
.
-
The maximal operation pressure adjusted in the set-up (PB) will be displayed. If need be, the value can be changed using the cursor keys. Confirm the value with
. -
The air pressure (PA) adjusted in the setup is dis-played. If need be, the value can be changed using the cursor keys. Confirm with
.
-
Use <UP/DOWN> to select the used gas type (f) and confirm with
. The following gas types are record-ed with their respective factors: - Natural gas
- Air
- City gas
- Propane
- Butane
- Hydrogen
-
With the compressed-air pump create the pre-adjusted test pressure (the unit beeps as soon as the pressure level is achieved).
-
Interrupt the connection to the air pump (switch-off the ball valve).
-
Wait for stabilization time (if the pressure remains in the range “test pressure +/- 10% during stabilization time, so the measurement time will start).
-
Once the measurement time is over, the result is displayed and can be sent to the IR printer (option) pressing
(Print). -
If the menu “Usability Check” is selected again, so the result can be called up with
(as long as the ecom-DP is on) or a new measurement can be started with .
Clock Set
-
Scroll with <UP/DOWN> until the menu point “Set clock”.
-
Press
to activate. Date and time can be adjust-ed as follows: -
Select with <RIGHT/LEFT> the position to be changed (corresponding field blinks).
-
Adjust the correct figures using <UP/DOWN>.
-
Confirm with
.
-
Internal Instrument Settings
Further instrument settings can be performed in the menu “Internal”. Press
Key Beep on/off Switch
- With the cursor keys <UP/DOWN> scroll until the menu point “Key beep”.
- Activate with
. - With the cursor keys <RIGHT/LEFT> modify the adjustment (YES or NO).
- Confirm with
.
Pressure Auto -> 0 by Start
The ecom-DP normally zeroes the internal pressure sensor during switch-on (adjustment “YES”). If the instrument is to remain connected to a pressure source, so this adjustment can be modified:
-
Scroll with the cursor keys <UP/DOWN> until the menu point “Pressure auto -> 0 by start”.
-
Press
to activate. -
Use the cursor keys <RIGHT/LEFT> to modify the adjustment (YES or NO).
-
Confirm the modification with
.
Choose Printer (ecom-P)
Please choose the used printer as follows:
- Scroll with the cursor keys <UP/DOWN> until the menu point “ecom-P”.
- Press
to activate. - Use the cursor keys <RIGHT/LEFT> to modify the adjustment (YES = printer ecom-P or NO = printer HP 82240B).
- Confirm the modification with
.
Battery Loading
The ecom-DP is able to work with accus and AC power (option):
- Scroll with the cursor keys <UP/DOWN> until the menu point “Battery loading”.
- Press
to activate. - Use the cursor keys <RIGHT/LEFT> to modify the adjustment (YES = using AC power for charging accus or NO = no AC power used).
- Confirm the modification with
.
Display Damping
The display damping can be adjusted to various measurement duties. 7 damping steps can be selected (1 = no damping; 7 = high damping).
Additionally the damping steps 3 – 7 enable the selection of a damping progression. This means:
not progressive:
by massive pressure changes the display does not react faster than by small
pressure changes.
low progressive:
by massive pressure changes the display reacts somewhat faster than by small
pressure changes.
high progressive:
by massive pressure changes the display reacts much faster than by small
pressure changes.
- To change the adjustment, scroll with the keys <UP/DOWN> until the menu point “Damping”.
- Press
to activate. - Adjust the desired damping step using the cursor keys <RIGHT/LEFT>.
- Press
to confirm.
Pressure Unit
The measuring unit of the internal pressure sensor can be changed. The following units are selectable:
- Hectopascal: hPa
- Centimetre water column: cmH2O
- Millimetre mercury column: mmHg
- Pound per square inch: PSI
- Inch water column: “H2O
- Inch mercury column: “Hg
For modification:
- Scroll with the cursor keys <UP/DOWN> until the menu point “Pressure unit”.
- Press
to activate. - Use the cursor keys <RIGHT/LEFT> to adjust the de-sired unit for the pressure sensor.
- Press
to confirm.
Auto OFF Function
The ecom-DP is able to switch off autonomously after 270 sec.:
- Scroll with the cursor keys <UP/DOWN> until the menu point “Auto OFF function”.
- Press
to activate. - Use the cursor keys <RIGHT/LEFT> to modify the adjustment (YES = activate the function or NO = disable the function
- Confirm the modification with
.
Contrast
The contrast of the display can be adjusted with the help of the cursor keys <RIGHT/LEFT>:
- Scroll with the cursor keys <UP/DOWN> until the menu point “Contrast”.
- Press
to activate. - To increase the contrast use cursor key
. - To decrease the contrast use cursor key
. - Confirm the modification with
.
Light
The lighting of the display can be switched ON and OFF.
- Scroll with the cursor keys <UP/DOWN> until the menu point “Light”.
- Press
to activate. - Use the cursor keys <RIGHT/LEFT> to modify the adjustment (YES = switch ON the light or NO = switch OFF the light
- Confirm the modification with
.
Press
Sensor Setup
The sensors in the instrument are given factory specified designations that
can be changed by the user. Using the <UP/DOWN> cursor keys, scroll to “Sensor
set-up” menu. Press
A maximum of 6 characters (letters, numbers and special characters) can be used in the sensor designation. Using the <UP/DOWN> cursor keys, select the position to be changed (sign blinks). Use the <UP/DOWN> cursor keys to select the character that should stand on this position.
Starting from the current one and using the
Blank sign
A – Z
a – z
Ä -Ü
ä – ß
0 – 9
Special characters
Once the renaming procedure is complete, press
View Service Information
- Use the cursor keys <UP/DOWN> to scroll thru until the menu point “Service”.
- Press
and an info window will be opened in which the address and phone number of the competent service Centre as well as the instrument version (e.g. ECOM DP V3.1) will be displayed.
Data Logger Program “ecom-DP Logger”
The measured value process of the ecom-DP can be logged with the help of the data logger program “ecom-DP Logger”. The program can be loaded free of charge from the download area of the ecom homepage www.ecom.de. To trans-fer the data a USB cable (kind becomes. – No.: 55818) and a driver (free down-load from www.ecom.de) is needed. The measured value rows can be stored in the .xls format (Excel).
The USB cable can be used as the power supply for the ecom DP. Therefore the function “Battery loading” must be adjusted on “no” when using 1,5 V AA (Mignon) batteries (see 6.4.)!
Technical Data ecom-DP
-
Power supply
- Battery: 3 x 1,5 V AA (Mignon)
- or 3 x 1,2 V AA NiCd (nickel cadmium)
- or 3 x 1,2 V AA NiMh (metal hydride)
- Internal loading unit
- Connection via 1.3 mm hollow plug
- Loading current approx. 500 mA
-
Power consumption
- With backlit approx. 90 mA
- Without backlit approx. 45 mA
- Operation time depending on used power supply
-
Display
- Graphic display
- Resolution 128 x 64 pixel
- Window approx. 43 x 28 mm
-
Operation:
(menu-monitored)- 4 cursor keys
- 1 confirmation key
- 4 function keys
- 1 ON/OFF key
-
Differential pressure measurement:
- Pressure range +/- 1500 hPa (mbar) or
- Pressure range +/- 70 hPa (mbar)
-
Resolution
- 0.1 hPa or
- 0.01 hPa
-
Accuracy
- < 3% of measurement value
-
Internal Clock
- Adjustment via menu and keyboard
- Buffer accu for battery change only (approx. 1 minute)
-
Dimensions (LxWxH)
- Approx. 170 x 75 x 35 mm
-
Weight
- Complete approx. 300 g
-
Ambient temperature
- 0 bis 50°C
Maintenance Tip
To secure the accuracy of your measuring instrument we recommend the annual check by an authorized ecom partner. In the case of strong demand (e.g. permanent several hours of measurement per day, rough conditions etc.) shorter intervals between checks should be selected – please contact your ecom partner. All ecom partners are listed under www.ecom.de.
Subject to technical changes
V3.2 02.2021
ecom GmbH
Am Großen Teich 2
D-58640 Iserlohn
Telefon: +49 (0) 23 71 – 9 45-5
Telefax: +49 (0) 23 71 – 4 03 05
Internet: http://www.ecom.de
Email: info@ecom.de
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>