INVACARE Softform Active 2 Mattress with RX pump User Guide

June 6, 2024
INVACARE

Softform Active 2 Mattress with RX pump

User Guide

Control unit menu display

Control unit menu display

Control unit menu display

Operation

1. The power indicator will flash and control unit will enter static mode initially until pressure reaches 15 mmHg.

Then control unit will enter alternate mode and power indicator will remain on. The cycle time is set at 10 min, and the pressure is set at 60 ± 3 mmHg.

The first inflation of new mattress: the indicator function (visible and audible indicators) will activate if the mattress does not inflate completely within 15 minutes.

Operation failures

• During control unit operation, low pressure indicator will activate within 1 minute if air hoses are disconnected or air cells are broken.
• In the case of abnormal alternate or no alternate, the audible signal will be activated and alternate failure indicator will flash.

Installing the system

Installing the system

1. Hang the control unit by means of the 2 built in hangers (A) at the end of the bed or place it on a horizontal surface. (Placing the control unit on the floor will not affect the performance, but may expose it to accidental damage.)

2. Connect the CPR hose E to air outlets on the mattress and control unit B, ensuring that the hose is not kinked or twisted. Allow space for bed rails to drop freely.

3. Connect the mains power cable to the power supply socket C of the control unit and a suitable outlet.

4. Switch on mains power of the outlet if present.

5. Move the mechanical switch D on the left side of the Control unit to the On position. A faint single beep sounds and the system in cycle light flashes while the system powers up.

CPR procedure

The Invacare® Softform Active® 2 Rx mattress has been fully tested to comply with the current CPR standard of 5 – 6 cm compression depth. This was achieved at all stages of inflation/deflation.

1. Hold down Red CPR button A.

2. Pull hose connector firmly away from the control unit B.

3. Switch off the control unit.

Mattress will start to deflate. The deflation time is 20 seconds.

4. When CPR is complete reactivate the system following section Installing the system.

Identifying and repairing faults

There are audio and visual alarms present on the control unit.

Identifying and repairing faults

Invacare Limited
Pencoed Technology Park, Pencoed
Bridgend CF35 5AQ
UK

Invacare Portugal, Lda
Rua Estrada Velha 949
4465-784 Leça do Balio
Portugal

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals