JAMECO P7100 1X/10X Oscilloscope Passive Probe User Guide

June 6, 2024
JAMECO

JAMECO P7100 1X/10X Oscilloscope Passive Probe

Frequency Compensation

Before taking any measurements using a probe, first check the compensation of the probe and adjust it to match the channel inputs. Most oscilloscopes have a square wave reference signal available at a terminal on the front panel used to compensate the probe. Connect the probe to the signal source on your oscilloscope.
Set the probe to 1 OX position. Adjust trimmer until seeing flat-top square wave on the display.

Voltage vs Frequency Rating Curve(RMS)

JAMECO-P7100 1X-10X-Oscilloscope-Passive-Probe - fig 2

  • CATII 11:c measurement category II is for measurements performed on circuits directly connected to low voltage installations. Examples are
    measurements on household appliances, portable tools and similar equipment.

  • Equipment fully protected by DOUBLE INSULATTION or REINFORCED INSULATION.

  • Review this user manual carefully to avoid any personal injury or damage to this product and any product connected to it. To avoid potential hazards, use this product only as specified.

  • The measurement category of a combination of a PROBE ASSEMBLY and an accessory (an auxiliary to the measurement) is the lower of the measurement categories of the PROBE ASSEMBLY and of the accessory.

  • Follow the instructions to use the PROBE ASSEMBLY will impair or ruin its self-owned protection function.

Accessories and Features

Probe is provided with several accessories designed to make probing and measurement simpler. Please take a moment to familiarize yourself with these accessories and their uses.

JAMECO-P7100 1X-10X-Oscilloscope-Passive-Probe - fig 4

Probe Characteristics

Probe Characteristics

Model| P7100       P7200      P7300
Attenuation Ratio| 1X&10X
Bandwidth| 100MHz               200MHz                300MHz
Rise time| 3.5ns                  1.75ns                  1.15ns
Input Resistance| 1MO/10MO±2%


Input Capacitance

| 1X:85pF~ 120pF
10X:18.5pF~22.5pF
Max. Input Voltage| 1X:150V RMS CAT II&10X:300V RMS CAT II
Compensation Range| 10pF-35pF                          8pF~25pF
Operation Temperature| -1o”C-+ss·c
Humidity| -40°C or below ..;go% +41·c to +50″C ..;so%
Altitude| Operating 3000m Nonoperating 15000m
Cable Length| 130±2cm
Weight| About55g
Safety| IEC61010-031

Accessory Kit

Accessory Kit

Item| Description| Quantity
1| Retractable Hook Tip| 1
2| Adjustment Tool| 1
3| Locating Sleeve| 2
4| Marker Rings| 8
5| Ground L ad| 1
6| Ground Spring(Above 100MHz)| 1
7| BNC Adapter(Above 200MHz)| 1

Note:

content of this document are subject to change without notice.