oticon 232298 Smart Charger miniRITE R Hearing Aids Instruction Manual
Introduction to this booklet
This booklet guides you on how to use and maintain your new Oticon SmartCharger miniRITE R. Ensure you read this booklet carefully, including the Warnings section. This will help you get the most benefit from your charger. If you have additional questions about the use of your charger, contact your hearing care professional.
IMPORTANT NOTICE
For details on the functionality of your specific hearing aid solution, refer
to your hearing aid’s instructions for use.
Intended use
Intended use| The charger is intended to recharge the hearing aids and CROS transmitter that contain rechargeable batteries.
The product is intended to be used together with the given hearing solution.
---|---
Indications for use| There are no indications for use (diagnoses) for the
charger itself.
Intended user| Person with hearing loss using a hearing aid and their
caregivers. Hearing care professional responsible for adjusting the accessory.
Intended user group| Adults and children older than 36 months.
Use environment| Indoor and outdoor.
Contraindications| No contraindications.
Clinical benefits| See clinical benefits of the hearing aid.
Package contents
The packaging box contains the following elements:
Charger
What it is and does USB cable
Temperature conditions
Temperature | Charging condition |
---|---|
Below +5°C (+41°F) | No charging due to low temperature. |
+10°C to +35°C (+50°F to +95°F) | Best charging condition. The charging of |
hearing aids can take up to 3 hours.
+5°C to +10°C and
+35°C to +38°C
(+41°F to 50°F and
+95°F to +100°F)
| Extended charging. The charging of hearing aids can take up to 4 hours.
Above +38°C (+100°F)| No charging due to high temperature.
Prepare charger
Connect to power source
Charger is turned ON
When the charger is connected to power, the charger’s power bank charging
status is shown.
First use – Charge the power bank’s battery When you use your charger for
the first time, it is important that you fully charge the charger’s built-in
battery.
Charge the hearing aids
Place hearing aids in charger Place the hearing aids in the charging
ports as shown in the picture and close the lid. Make sure that your hearing
aids match the right and left color indicators on the charger’s charging
ports.
The RED color indicator is for the hearing aid on the right-hand side. The
BLUE color indicator is for the hearing aid on the left-hand side.
Charging status
The hearing aid LED lights turn ON when the hearing aids are correctly placed in the charger.
Hearing aid LED light charging status:
ORANGE = Charging
GREEN = Fully charged
Charging your hearing aids
Ensure you fully charge your hearing aids before first use and charge them
every night. That ensures that you start your day with fully charged hearing
aids.
If your hearing aid’s battery is completely drained, the normal charging time
is:
3 hours | 1 hour | 0.5 hour |
---|---|---|
Fully charged | 50% charged | 25% charged |
When the battery in your hearing aids is fully charged, the charging process
stops automatically.
Charging time may vary depending on the remaining capacity of the battery and
between the left/right-hand side hearing aid.
The charger’s lid serves as protection and storage for your hearing aids.
During charging, when you open the lid, the LEDs in your hearing aids light up
to indicate their charging status.
Note
If you have not used your hearing aid in a long time and you then try to
recharge it, it may take up to 30 seconds before the hearing aid begins
charging. For more information, see your hearing aid’s instructions for use.
IMPORTANT NOTICE
Ensure that your charger is powered or that the charger’s built-in battery is
charged when the hearing aid or transmitter is seated in the charging port.
Otherwise your hearing aids or transmitter will remain turned ON and continue
using battery power.
Charging the charger
Your SmartCharger contains a built-in lithium-ion rechargeable battery that
doubles as a power bank. This lets you fully charge a pair of hearing aids a
minimum of three times on a fully-charged power bank while you are on-the-go.
When the charger is plugged into a power source
-
Your hearing aids and the charger’s built-in battery are being charged.
When the charger is unplugged from a power source -
The charger enters Power bank mode, letting you charge your hearing aids when you do not have access to a power socket.
Charging times
Power bank battery | Power bank + hearing aids |
---|---|
Up to 4 hours depending on charge level | Up to 8 hours depending on charge |
level
Note
It is recommended that you keep the lid closed at all times while you charge
your hearing aids. When the charger is connected to a power source, the LEDs
on your charger indicate the charging status of the power bank. The LEDs turn
OFF shortly after the power bank battery is fully charged.
IMPORTANT NOTICE
When charging, the light indicators on your charger should be lit as described
in this booklet. If the light indicators are not lit as described, disconnect
the charger and consult your hearing care professional.
Power bank function
When the charger is disconnected from power, the LEDs on your charger indicate the charger’s power bank battery level. You can check the remaining battery level by opening the charger’s lid.
Long-term storage
Before you put away or store the charger for a prolonged period of time (more
than 14 days), ensure you first fully charge the charger’s power bank battery.
This way the battery can be charged again.
Note
To protect the rechargeable battery, it is necessary that you fully charge the
charger’s power bank battery every 12 months, otherwise the rechargeable
battery may need to be replaced.
Power and charge using other sources
You can power and charge the charger’s battery from other sources than a power
socket. This can be done by using the charger’s USB cable.
Ensure that the power source is USB 2.0 compliant, with a minimum of 500mA
output.
Examples of power sources:
- Computer
- Car
IMPORTANT NOTICE
When using a power socket, use the supplied original power supply (5.0V/1A)
for the charger.
Drying your hearing aids
The SmartCharger uses heat from inductive charging to dry your hearing aids when you place them in the charger ports. The drying of your hearing aids is activated when you place the hearing aids in the charger ports and charging begins.
IMPORTANT NOTICE
Ensure you remove the USB cable when you transport the charger.
Caring for your charger
Always ensure that the charger is dry and clean before use. Keep the charging ports clean from dust by using a dry cloth. Do not use water or cleaning fluids to clean the charging ports. Store your charger in a dry location, e.g. in the bedroom. Ensure you keep the charger in an environment with a temperature between +5°C and +38°C (+41°F and +100°F).
IMPORTANT NOTICE
To clean the charger, use a soft, dry cloth. The charger must never be washed
or immersed in water or other liquids.
General warnings
For your personal safety and to ensure correct usage, you should familiarize
yourself fully with the following general warnings before using your charger.
Consult your hearing care professional if you experience unexpected operations
or serious incidents with your charger during use or because of its use. Your
hearing care professional will support you with issue handling and, if
relevant, reporting to the manufacturer and/or the national authorities.
Usage of charger
The charger should be used only as directed and adjusted by your hearing care
professional.
Risk of burns
Do not place anything but the hearing aids in the charger ports.
Choking hazards
Charger, hearing aids and their parts should be kept out of reach of children
and anyone who might swallow these items or otherwise cause injury to
themselves.
If a battery or hearing aid is swallowed, see a doctor immediately and contact
the National Poison Center at
1-800-222-1222 or National Battery
Ingestion Hotline at 1-800-498-8666.
Heat
The charger must never be exposed to extreme heat, e.g. left inside a parked
car in the sun. The charger must not be dried in microwave ovens or other
ovens.
Explosives
The charger is safe to use under normal usage conditions. The charger has not
been tested for compliance with international standards concerning explosive
environments.
Therefore, do not use the charger in environments with danger of explosions
e.g. mines, oxygen rich environments or areas where flammable anaesthetics are
handled.
Rechargeable battery
Do not attempt to open the charger, as it may damage the battery. Never
attempt to replace the battery. If battery replacement is needed, return your
device to the supplier. The service guarantee is void if there are signs of
tampering. In case of battery leakage, do not wear or carry your charger, as
it may cause skin irritation due to acids leaking from the battery. If your
skin has been in contact with the leaked battery acids, use a wet cloth to
wipe it off and ensure no acid is left on your skin. If you experience skin
irritation, consult your doctor. For further handling instructions of your
charger consult your hearing care professional.
The safety of recharging batteries using a USB connector is determined by the
external signal source. When connected to external equipment plugged into a
power socket, this equipment must comply with IEC 62368-1 or equivalent safety
standards.
Connection to external equipment
The safety of the charger, when connected to external equipment with USB
cable, is determined by the external signal source. When the charger is
connected to external equipment which is plugged into a power socket, this
equipment must comply with IEC 62368-1 or equivalent safety standards.
Fire hazards
The power source in your charger has insufficient energy to cause fire in
normal usage conditions.
Be aware of your surroundings
Avoid interacting with your charger when your attention is needed elsewhere.
This may take away your attention from your surroundings, for example while in
traffic. Refer to national traffic laws regarding the use of communication
devices while in traffic.
Possible side effects
The otherwise non-allergenic materials used in your charger may, in rare
cases, cause a skin irritation or other side effects. If these conditions
occur, seek consultation with a physician.
Transport on aircraft
Consult the airline for details regarding transportation of your hearing aid
charger on aircraft.
Caution must be taken with active implants
It is recommended to keep the charger at least 15 centimeters away from the
implant, e.g. do not carry it in a breast pocket as it contains magnets. In
general, please follow the guidelines recommended by the manufacturers of
implantable defibrillators and pacemakers on use with portable electronics and
magnets.
If you have an active implant, contact the manufacturer of your implantable
device for information about the risk of interference.
Active Implants
The charger has been thoroughly tested and characterized for human health
according to international standards for human exposure (Specific Absorption
Ratio — SAR), induced electromagnetic power and voltages into the human body.
The exposure values are well below international accepted safety limits for
SAR, induced electromagnetic power and voltages into the human body defined in
the standards for human health.
Charging
To prevent potential battery damage during storage charge your charger every
12 months.
Use of third-party accessories
Only use accessories, transducers, or cables supplied by the manufacturer.
Non-original accessories may result in reduced electromagnetic compatibility
(EMC) of your charger.
Troubleshooting
Symptom | Possible causes | Solutions |
---|---|---|
The hearing aids do not turn ON after being removed from the charger |
The charger did not have enough time to turn the hearing aids ON before they were removed from the charger
|
Place the hearing aids in the charger ports and try again, or use the push button to turn the hearing aids ON manually
Two charger LEDs show ORANGE blinks| The hearing aid or charger’s temperature
is too cold or too warm| Move the charger and hearing aid to a location with a
temperature between +5°C and +38°C (+41°F and +100°F)
Three charger LEDs show ORANGE blinks| System error| Contact your hearing care
professional
The charger LEDs remain turned OFF when the charger is connected to power|
The charger is not charging
| Verify that the power plug is correctly connected
Ensure that the power source delivers sufficient output for your charger
Symptom | Possible causes | Solutions |
---|
The hearing aid LED blinks GREEN when the hearing aid is placed in the charger| ****
The hearing aid has not been in use for an extended period of time
| Depending on how depleted the battery in the hearing aid is, the hearing aid automatically resumes normal charging mode after a timeframe of up to 10 minutes.
Ensure you leave the hearing aid in the charger during this process.
| The charger is not turned ON| Verify that the charger power plug is
correctly connected or the power bank has enough battery
| The hearing aid or charger’s temperature is either too warm or| Move the
charger and hearing aid to a location with a temperature between +5°C and
+38°C
The hearing aid LED| too cold| (+41°F and +100°F)
remains turned OFF| |
|
when the hearing aid is placed in the charger| Charging is incomplete. The
room temperature exceeds
+35°C (+95°F), which extends the charging time. The charger has stopped charging to protect the battery.
| Reinsert the hearing aid into the charger. This completes the charging
within approximately 15 minutes.
| The hearing aid is not correctly seated in the charger| Check the charging
ports for foreign objects
The hearing aid LED blinks ORANGE when the hearing aid is placed in the
charger| ****
System error
| ****
Contact your hearing care professional
Conditions of use
Operating conditions| Temperature: +5°C to +38°C (+41°F to +100°F) Humidity:
5% to 93%, non-condensing Atmospheric pressure: 700 hPa to 1060 hPa
---|---
Transportation and storage conditions| Temperature and humidity shall not
exceed the below limits for extended periods during transportation and
storage:
Temperature: -20°C to +60°C (-4°F to +140°F) Humidity: 5% to 93%, non- condensing
For information about your hearing aid’s conditions of use, see your hearing aid’s instructions for use.
International warranty
Your device is covered by an international warranty issued by the
manufacturer. This international warranty covers manufacturing and material
defects in the device itself, but not in accessories such as batteries,
tubing, speakers, earpieces and filters, etc. Problems arising from
improper/incorrect handling or care, excessive use, accidents, repairs made by
an unauthorized party, exposure to corrosive conditions, physical changes in
your ear, damage due to foreign objects entering the device or incorrect
adjustments are NOT covered by the international warranty and may void it. The
above international warranty does not affect any legal rights that you might
have under applicable national legislation governing the sale of consumer
goods in the country where you have bought your device. Your hearing care
professional may also have issued a warranty that goes beyond the clauses of
this international warranty. Please consult him/her for further information.
If you need service
Take your device to your hearing care professional, who may be able to sort
out minor problems and adjustments immediately. Your hearing care professional
can assist you in obtaining warranty service from the manufacturer. Your
hearing care professional may charge a fee for their services.
Warranty Certificate
Name of owner: ____ Hearing
care professional: Hearing
care professional’s address: ____ Hearing
care professional’s phone: ____ Purchase
date: ____ Warranty
period: Month: ____ Model:
___ Serial no.: __
Description of symbols and abbreviations used in this booklet
Warnings
Text marked with a warning symbol must be read before using the device.
Manufacturer
The device is produced by the manufacturer whose name and address are stated next to the symbol. Indicates the medical device manufacturer, as defined in EU Regulation 2017/745.
CE mark
The device complies with all required EU regulations and directives.
Electronic waste (WEEE)
Recycle hearing aids, accessories or batteries according to local regulations.
Hearing aid users can also return electronic waste to their hearing care professional for disposal. Electronic equipment covered by Directive 2012/19/EU on waste and electrical equipment (WEEE).
Regulatory Compliance Mark (RCM)
The device complies with electrical safety, EMC and radio spectrum requirements for devices supplied to the Australian or New Zealand markets.
Compliance statements
USA and Canada
This device complies with part 18 of the FCC Rules.
This device complies with FCC and Innovation, Science and Economic Development
Canada’s RF radiation exposure limits set forth for an uncontrolled
environment.
Use of other accessories not verified by the manufacturer may not ensure
compliance with FCC and Innovation, Science and Economic Development Canada’s
RF exposure guidelines.
User information
This equipment has been tested and found to comply with the limits pursuant to
part 18 of the FCC Rules. These limits are designed to provide reasonable
protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if
not installed and used in accordance with the instructions,
may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to other electronic equipment, which can be determined by turning the equipment OFF and ON, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the equipment
- Increase the separation between the equipment
Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.
This medical device complies with the Medical Device Regulation (EU) 2017/745.
Declaration of Conformity is available at the headquarters.
Oticon A/S Kongebakken 9
DK-2765 Smørum Denmark
www.oticon.global/doc